Vital Signs recallZ-0365-2024 · SunMed Holdings, LLC · initiated Oct 11, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from SunMed Holdings, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
FDA root cause category: Device Design
What the company is doing
Customer notification was emailed October 11, 2023 advising customers that monitoring of tidal carbon dioxide in critically ill patients is recommended because use of accessories (e.g. PEEP valve) on the exhalation port of manual resuscitators may increase backward leak in the integrated manometer due to increased expiratory resistance. This affects all lots with duckbill-style patient valve with integrated manometer manufactured since 09/21/2019. Customers are to disseminate this information to all users and complete and return the response for. The notice also notes manual resuscitator IFUs will be updated to include this information as a warning statement.
Recall details
- Recall number
- Z-0365-2024
- Status
- Open, Classified
- Initiated by the company
- Oct 11, 2023
- Posted by the FDA
- Nov 21, 2023
- Recalling firm
- SunMed Holdings, LLC
- Product code
- BTM
- FDA submissions
- K012842
- Distribution
- US Nationwide distribution including in the states of AR, AZ, CA¿CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MN, NC, ND, NH, NJ, NM, NV, NY, OH, SC, TN, TX, UT, VA, & WV.
Product description
Ventlab, LLC SafeSpot manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. SafeSpot Infant Resuscitator REF #s: SS3200OBPC-T, SS3204MB-D12, SS3200MB-2, SS3200MB-D1, SS3200OBP, SS3200MB-MMC, SS3200MBP-M00, SS3200MB-PW, SS3200MBP-2I, SS3200MBF-2, SS3204MBF-2, SS3200MB- PWS10, SS3206MCP-S1, SS3204MB, SS3204MC-P2, SS3200MB, SS3200MBP-2, SS3200MBF, SS3200MB-I, SS3200MBP, SS3200OB, SS3200MC, SS3200MC-P2; 2. SafeSpot Safe Spot Bag, D1 Mask with Inflation Bag Reservoir, Manometer, PEEP Valve, and 40cm Pop-Off REF # PRO-1927.
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
UDI-DI: 10889483104107, (Case UDI: 30889483104105, Item # PRO-1927); UDI-DI: 10889483102954, (Case UDI: 30889483102958, Item # SS3200MB); UDI-DI: 10889483092590, (Case UDI: 30889483092594, Item # SS3200MB-2); UDI-DI: 10889483104330, (Case UDI: 30889483104334, Item #SS3200MB-D1); UDI-DI: 10889483172322, (Case UDI: 30889483172326, Item # SS3200MBF); UDI-DI: 10889483104354, (Case UDI: 30889483104358, Item # SS3200MBF-2); UDI-DI: 10889483104545, (Case UDI: 30889483104549, Item # SS3200MB-I); UDI-DI: 10889483105771, (Case UDI: 30889483105775, Item # SS3200MB-MMC); UDI-DI: 10889483086919, (Case UDI: 30889483086913, Item # SS3200MBP); UDI-DI: 10889483104903, (Case UDI: 30889483104907, Item # SS3200MBP-2); UDI-DI: 10889483107324, (Case UDI: 30889483107328, Item # SS3200MBP-2I); UDI-DI: 10889483103777, (Case UDI: 30889483103771, Item # SS3200MBP-M00); UDI-DI: 10889483103111, (Case UDI: 30889483103115, Item # SS3200MB-PW); UDI-DI: 10889483086902, (Case UDI: 30889483086906, Item # SS3200MB- PWS10); UDI-DI: 10889483103784, (Case UDI: 30889483103788, Item # SS3200MC); UDI-DI: 10889483103128, (Case UDI: 30889483103122, Item # SS3200MC-P2); UDI-DI: 10889483107904, (Case UDI: 30889483107908, Item # SS3200OB); UDI-DI: 10889483086926, (Case UDI: 30889483086920, Item # SS3200OBP); UDI-DI: 10889483231142, (Case UDI: 30889483231146, Item # SS3200OBPC-T); UDI-DI: 10889483107331, (Case UDI: 30889483107335, Item # SS3204MB); UDI-DI: 10889483158913, (Case UDI: 30889483158917, Item # SS3204MB-D12); UDI-DI: 10889483104354, (Case UDI: 30889483104358, Item # SS3204MBF-2); UDI-DI: 10889483160497, (Case UDI: 30889483160491, Item # SS3204MC-P2); UDI-DI: 10889483111734, (Case UDI: 30889483111738, Item # SS3206MCP-S1. All lots manufactured from 9/21/2018 to present. Latest expiration Date 10/11/2028.
Devices named in this recall
18 devices in the FDA device registry whose identifier appears in the recall's code information.
- Ventlab SS3200MB-MMCVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall1 report
- Ventlab SS3200MBPVentlab, LLC · Ventilator, emergency, manual (resuscitator)1 recall0 reports
- Ventlab SS3200OBPVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab SS3200MB-2Ventlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab SS3200MBVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab 10889483103128Ventlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab SS3200MBP-M00Ventlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab SS3200MCVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab SS3200MB-D1Ventlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab SS3200MB-IVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab SS3200MBP-2Ventlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab SS3200MBP-2IVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab SS3204MBVentlab, LLC · Ventilator, emergency, manual (resuscitator)1 recall0 reports
- Ventlab SS3200OBVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Vetnlab SS3206MCP-S1Ventlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab SS3204MC-P2Ventlab, LLC · Ventilator, emergency, manual (resuscitator)1 recall0 reports
- Ventlab SS3200MBFVentlab, LLC · Ventilator, emergency, manual (resuscitator)1 recall0 reports
- Vital Signs™ 9362Sunmed, LLC · Filter, Bacterial, Breathing-Circuit1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 93237).
- Z-0361-2024 · Open, ClassifiedVentlab, LLC AirFlow manual resuscitator devices with integrated manometer, multiple accessory configurations to include: 1. AirFlow Adult Resuscitator REF #s: AF1140MB-M5, AF114…
- Z-0362-2024 · Open, ClassifiedVentlab, LLC BreathTech manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. BreathTech Adult Resuscitator REF #s: BT5107F & BT5104P; …
- Z-0363-2024 · Open, ClassifiedVentlab, LLC Horizon manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Horizon Adult Resuscitator REF #s: 9122HZ, HZI9000, HZI9108, …
- Z-0366-2024 · Open, ClassifiedVentlab, LLC STAT-Check manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. STAT-Check Adult Resuscitator REF #: SC9101B; 2. STAT-Ch…
- Z-0367-2024 · Open, ClassifiedVentlab, LLC V-Care manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. V-Care Child Resuscitator REF #s: VN4100MB, VN4100MCFB-3, VN41…
- Z-0368-2024 · Open, ClassifiedCuraplex manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Curaplex Adult Premium BVM with Bag Reservoir, Manometer, Filter, Peep V…
- Z-0369-2024 · Open, ClassifiedMedline manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Adult Resuscitator REF #s: CPRM1116FPM, CPRM1116FPM4, CPRM1116M, CPRM1116…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page