Ventlab, LLC V-Care manual resuscitators with integrated… recallZ-0367-2024 · SunMed Holdings, LLC · initiated Oct 11, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from SunMed Holdings, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
FDA root cause category: Device Design
What the company is doing
Customer notification was emailed October 11, 2023 advising customers that monitoring of tidal carbon dioxide in critically ill patients is recommended because use of accessories (e.g. PEEP valve) on the exhalation port of manual resuscitators may increase backward leak in the integrated manometer due to increased expiratory resistance. This affects all lots with duckbill-style patient valve with integrated manometer manufactured since 09/21/2019. Customers are to disseminate this information to all users and complete and return the response for. The notice also notes manual resuscitator IFUs will be updated to include this information as a warning statement.
Recall details
- Recall number
- Z-0367-2024
- Status
- Open, Classified
- Initiated by the company
- Oct 11, 2023
- Posted by the FDA
- Nov 21, 2023
- Recalling firm
- SunMed Holdings, LLC
- Product code
- BTM
- FDA submissions
- K012842
- Distribution
- US Nationwide distribution including in the states of AR, AZ, CA¿CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MN, NC, ND, NH, NJ, NM, NV, NY, OH, SC, TN, TX, UT, VA, & WV.
Product description
Ventlab, LLC V-Care manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. V-Care Child Resuscitator REF #s: VN4100MB, VN4100MCFB-3, VN4100MCP; 2. V-Care Infant Resuscitator REF #s: VN3100OB, VN3100MB-2, VN3100MB-M01, VNW3100MB-2, VN3100MBP-3M, VN3100MB, VN3100MB-D1, VN3100MB-M00-0, VN3100MB-M1, VN3100OBP, VN3100MB-B, VN3100MB-I, VN3100MBP-000, VN3100MBP-2, VN3100MB-PW2, VN3100MBP, VN3100MBP-D1, VN3100MBPW-D2, VN3100MBP-F, VN3100MXK-2, VN3100MX-F, VN3100MC-2, VN3100MC, VN3100MC-B, VN3100MC-D1, VN3100MC-2U, VN3106MC-M1-2V, VN3104MCFB- M01, VN3106MC-D2V, VN3106MC-M1V, VN3100MCP-2, VN3100MCP, VN3106MCPF- M1V, VN3100MBF; 3. V-Care Pediatric Resuscitator REF #s: VN2100MB, VN2102MB, VN2102MBF, VN2102OBP, VN2100MBF, VN2100MBP, VN2100MB-PW, VN2102MX, VN2102MX-F, VN2100MX-M123, VN2100MC, VN2102MC, VN2102MC-B, VN2102MCP, VN2100MCP; 4. V-Care Small Adult Resuscitator REF #s: VN5000MB, VN5002MB, VN5100MB-F, VN5000MBPF, VN5000MBPF-I, VN5000MX, VN5100MX-F, VN5000MX-N, VN5000MXP, VN5000MXP-K, VN5000MC, VN5102MC, VN5100MCP, VN5000MC-N.
Code information (models, lots, serial numbers)
Show the full code information (5k characters)
ITEM # VN2100MB, UDI-DI: 10889483095379, (Case UDI: 30889483095373); ITEM # VN2100MBF, UDI-DI: 10889483092774, (Case UDI: 30889483092778); ITEM # VN2100MBP, UDI-DI: 10889483108260, (Case UDI: 30889483108264); ITEM # VN2100MB-PW, UDI-DI: 10889483108253, (Case UDI: 30889483108257); ITEM # VN2100MC, UDI-DI: 10889483108277, (Case UDI: 30889483108271); ITEM # VN2100MCP, UDI-DI: 10889483094525, (Case UDI: 30889483094529); ITEM # VN2100MX-M123, UDI-DI: 10889483098936, (Case UDI: 30889483098930); ITEM # VN2102MB, UDI-DI: 10889483105542, (Case UDI: 30889483105546); ITEM # VN2102MBF, UDI-DI: 10889483110966, (Case UDI: 30889483110960); ITEM # VN2102MC, UDI-DI: 10889483111994, (Case UDI: 30889483111998); ITEM # VN2102MC-B, UDI-DI: 10889483107393, (Case UDI: 30889483107397); ITEM # VN2102MCP, UDI-DI: 10889483088470, (Case UDI: 30889483088474); ITEM # VN2102MX, UDI-DI: 10889483095751, (Case UDI: 30889483095755); ITEM # VN2102MX-F, UDI-DI: 10889483105863, (Case UDI: 30889483105867); ITEM # VN2102OBP, UDI-DI: 10889483108895, (Case UDI: 30889483108899); ITEM # VN3100MB, UDI-DI: 10889483096994, (Case UDI: 30889483096998); ITEM # VN3100MB-2, UDI-DI: 10889483105900, (Case UDI: 30889483105904); ITEM # VN3100MB-B, UDI-DI: 10889483113585, (Case UDI: 30889483113589); ITEM # VN3100MB-D1, UDI-DI: 10889483105566, (Case UDI: 30889483105560); ITEM # VN3100MBF, UDI-DI: 10889483100622, (Case UDI: 30889483100626); ITEM # VN3100MB-I, UDI-DI: 10889483109380, (Case UDI: 30889483109384); ITEM # VN3100MB-M00- 0, UDI-DI: 10889483095980, (Case UDI: 30889483095984); ITEM # VN3100MB-M01, UDI-DI: 10889483110386, (Case UDI: 30889483110380); ITEM # VN3100MB-M1, UDI-DI: 10889483102763, (Case UDI: 30889483102767); ITEM # VN3100MBP, UDI-DI: 10889483100622, (Case UDI: 30889483100626); ITEM # VN3100MBP-000, UDI-DI: 10889483110393, (Case UDI: 30889483110397); ITEM # VN3100MBP-2, UDI-DI: 10889483104262, (Case UDI: 30889483104266); ITEM # VN3100MBP-3M, UDI-DI: 10889483100639, (Case UDI: 30889483100633); ITEM # VN3100MBP-D1, UDI-DI: 10889483110751, (Case UDI: 30889483110755); ITEM # VN3100MBP-F, UDI-DI: 10889483102770, (Case UDI: 30889483102774); ITEM # VN3100MB-PW2, UDI-DI: 10889483110744, (Case UDI: 30889483110748); ITEM # VN3100MBPW-D2, UDI-DI: 10889483111192, (Case UDI: 30889483111196); ITEM # VN3100MC, UDI-DI: 10889483100646, (Case UDI: 30889483100640); ITEM # VN3100MC-2, UDI-DI: 10889483111208, (Case UDI: 30889483111202); ITEM # VN3100MC-2U, UDI-DI: 10889483112113, (Case UDI: 30889483112117); ITEM # VN3100MC-B, UDI-DI: 10889483095768, (Case UDI: 30889483095762); ITEM # VN3100MC-D1, UDI-DI: 10889483112021, (Case UDI: 30889483112025); ITEM # VN3100MCP, UDI-DI: 10889483110768, (Case UDI: 30889483110762); ITEM # VN3100MCP-2, UDI-DI: 10889483113622, (Case UDI: 30889483113626); ITEM # VN3100MX-F, UDI-DI: 10889483112038, (Case UDI: 30889483112032); ITEM # VN3100MXK-2, UDI-DI: 10889483111215, (Case UDI: 30889483111219); ITEM # VN3100OB, UDI-DI: 10889483099964, (Case UDI: 30889483099968); ITEM # VN3100OBP, UDI-DI: 10889483105924, (Case UDI: 30889483105928); ITEM # VN3104MCFB- M01, UDI-DI: 10889483110775, (Case UDI: 30889483110779); ITEM # VN3106MC-D2V, UDI-DI: 10889483107539, (Case UDI: 30889483107533); ITEM # VN3106MC-M1- 2V, UDI-DI: 10889483095997, (Case UDI: 30889483095991); ITEM # VN3106MC-M1V, UDI-DI: 10889483097991, (Case UDI: 30889483097995); ITEM # VN3106MCPF- M1V, UDI-DI: 10889483099001, (Case UDI: 30889483099005); ITEM # VN4100MB, UDI-DI: 10889483111338, (Case UDI: 30889483111332); ITEM # VN4100MCFB-3, UDI-DI: 10889483111239, (Case UDI: 30889483111233); ITEM # VN4100MCP, UDI-DI: 10889483098042, (Case UDI: 30889483098046); ITEM # VN5000MB, UDI-DI: 10889483112052, (Case UDI: 30889483112056); ITEM # VN5000MBPF, UDI-DI: 10889483112229, (Case UDI: 30889483112223); ITEM # VN5000MBPF-I, UDI-DI: 10889483113646, (Case UDI: 30889483113640); ITEM # VN5000MC, UDI-DI: 10889483095393, (Case UDI: 30889483095397); ITEM # VN5000MC-N, UDI-DI: 10889483108352, (Case UDI: 30889483108356); ITEM # VN5000MX, UDI-DI: 10889483107782, (Case UDI: 30889483107786); ITEM # VN5000MX-N, UDI-DI: 10889483107553, (Case UDI: 30889483107557); ITEM # VN5000MXP, UDI-DI: 10889483089828, (Case UDI: 30889483089822); ITEM # VN5000MXP-K, UDI-DI: 10889483110409, (Case UDI: 30889483110403); ITEM # VN5002MB, UDI-DI: 10889483100660, (Case UDI: 30889483100664); ITEM # VN5100MB-F, UDI-DI: 10889483107560, (Case UDI: 30889483107564); ITEM # VN5100MCP, UDI-DI: 10889483098554, (Case UDI: 30889483098558); ITEM # VN5100MX-F, UDI-DI: 10889483095782, (Case UDI: 30889483095786); ITEM # VN5102MC, UDI-DI: 10889483109403, (Case UDI: 30889483109407); ITEM # VNW3100MB-2, UDI-DI: 10889483100677, (Case UDI: 30889483100671). All lots manufactured from 9/21/2018 to present. Latest expiration Date 10/11/2028.
Devices named in this recall
59 devices in the FDA device registry whose identifier appears in the recall's code information.
- Ventlab VN3100OBVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall1 report
- Ventlab VN5000MXPVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN2100MBFVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN2100MCPVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN2100MBVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN5000MCVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN2102MXVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN3100MC-BVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN5100MX-FVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN3100MB-M00-0Ventlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN3106MC-M1-2VVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN3100MBVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN3106MC-M1VVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN4100MCPVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN5100MCPVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN3106MCPF-M1VVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN3100MBFVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN3100MBP-3MVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab 10889483100646Ventlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN5002MBVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VNW3100MB-2Ventlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN3100MB-M1Ventlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN3100MBP-2Ventlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN2102MBVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN3100MB-D1Ventlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN2102MX-FVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN3100MB-2Ventlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN3100OBPVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN2102MC-BVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN3106MC-D2VVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN5000MX-NVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN5100MB-FVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN5000MXVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN2100MBPVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN2100MCVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN5000MC-NVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN2102OBPVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN3100MB-IVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN5102MCVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN3100MB-M01Ventlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN3100MBP-000Ventlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN5000MXP-KVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN3100MB-PW2Ventlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN3100MBP-D1Ventlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN3100MCPVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN3104MCFB-M01Ventlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN2102MBFVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN3100MC-2Ventlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN3100MXK-2Ventlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN4100MCFB-3Ventlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN4100MBVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN3100MC-D1Ventlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN3100MX-FVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN5000MBVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN3100MC-2UVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN5000MBPFVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN3100MB-BVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN3100MCP-2Ventlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
- Ventlab VN5000MBPF-IVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93237).
- Z-0361-2024 · Open, ClassifiedVentlab, LLC AirFlow manual resuscitator devices with integrated manometer, multiple accessory configurations to include: 1. AirFlow Adult Resuscitator REF #s: AF1140MB-M5, AF114…
- Z-0362-2024 · Open, ClassifiedVentlab, LLC BreathTech manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. BreathTech Adult Resuscitator REF #s: BT5107F & BT5104P; …
- Z-0363-2024 · Open, ClassifiedVentlab, LLC Horizon manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Horizon Adult Resuscitator REF #s: 9122HZ, HZI9000, HZI9108, …
- Z-0365-2024 · Open, ClassifiedVentlab, LLC SafeSpot manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. SafeSpot Infant Resuscitator REF #s: SS3200OBPC-T, SS3204MB-…
- Z-0366-2024 · Open, ClassifiedVentlab, LLC STAT-Check manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. STAT-Check Adult Resuscitator REF #: SC9101B; 2. STAT-Ch…
- Z-0368-2024 · Open, ClassifiedCuraplex manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Curaplex Adult Premium BVM with Bag Reservoir, Manometer, Filter, Peep V…
- Z-0369-2024 · Open, ClassifiedMedline manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. Adult Resuscitator REF #s: CPRM1116FPM, CPRM1116FPM4, CPRM1116M, CPRM1116…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page