Skip to content

Ventlab VN4100MCFB-3Recalls & reported problems

Ventlab, LLC · SM CHILD BAG,3 MASKS W/CORRUGATED TUBING RESERVOIR, FILTER, MANO& NO BLOWBY · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Oct 11, 2023Open, Classified

    A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.

    Z-0367-2024 · Root cause: Device Design · FDA record ↗

Other Ventilator, Emergency, Manual (Resuscitator) devices