Exactech Optetrak Logic Knee System, Catalog Numbers recallZ-0583-2024 · Exactech, Inc. · initiated Nov 21, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Exactech, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Exactech has received 3 complaints for 4 devices related to a loss of vacuum in the inner-most vacuum bag.
FDA root cause category: Packaging change control
What the company is doing
Exactech issued an URGENT MEDICAL DEVICE RECALL notice to its direct consignees on 11/21/2023 via email. The notice explained the issue, clinical impact, and requested the removal and return of all affected products. 11/22/2023, the firm initiated the recall OUS also by email.
Recall details
- Recall number
- Z-0583-2024
- Status
- Open, Classified
- Initiated by the company
- Nov 21, 2023
- Posted by the FDA
- Jan 18, 2024
- Recalling firm
- Exactech, Inc.
- Product code
- JWH
- FDA submissions
- K150890, K111400
- Distribution
- US NY, TX, IN, IL, MN, KY, VA, MD, TN, GA, NH, ME, SC, NV, FL, AR and Australia, Brazil, Chile, France, Germany, Guatemala, Italy, Japan, Korea, Slovenia, Spain, UK
Product description
Exactech Optetrak Logic Knee System, Catalog Numbers: a) 02-012-47-3011, b) 02-012-47-5009, c) 02-012-47-5011, d) 02-012-65-3013
Code information (models, lots, serial numbers)
a) 02-012-47-3011, GTIN 10885862174352, Serial Numbers: A676497; b) 02-012-47-5009, GTIN 10885862174543, Serial Numbers: A641353; c) 02-012-47-5011, GTIN 10885862174550, Serial Numbers: A641341; d) 02-012-65-3013, GTIN 10885862613400, Serial Numbers: A655561
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- Optetrak Logic 02-012-47-5009Exactech, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer2 recalls23 reports
- Optetrak Logic 02-012-47-3011Exactech, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer2 recalls9 reports
- Optetrak Logic 02-012-47-5011Exactech, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer2 recalls6 reports
- Truliant CC 02-012-65-3013Exactech, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall1 report
Product lines
Other devices under the same FDA submission
164 devices cleared under K150890, K111400 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Optetrak Logic 02-012-47-4009Exactech, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer1 recall29 reports
- Optetrak Logic 02-012-49-4011Exactech, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer1 recall23 reports
- Optetrak Logic 02-012-47-2509Exactech, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer1 recall21 reports
- Optetrak Logic 02-012-49-4009Exactech, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer1 recall20 reports
- Optetrak Logic 02-012-47-3009Exactech, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer1 recall19 reports
- Optetrak Logic 02-012-49-2009Exactech, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer1 recall16 reports
- Optetrak Logic 02-012-48-4009Exactech, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer1 recall16 reports
- Optetrak Logic 02-012-49-3509Exactech, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer1 recall15 reports
- Optetrak Logic 02-012-48-2009Exactech, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer1 recall14 reports
- Optetrak Logic 02-012-49-3011Exactech, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer1 recall14 reports
- Optetrak Logic 02-012-49-2011Exactech, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer1 recall13 reports
- Optetrak Logic 02-012-48-3509Exactech, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer1 recall12 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93507).
- Z-0577-2024 · Open, ClassifiedAcuMatch Hip System, Catalog Number 142-32-27
- Z-0578-2024 · Open, ClassifiedExactech Alteon Hip XLE Liner, Catalog Numbers: a) 01-030-40-0636, b) 01-030-40-0640, c) 01-030-42-0536
- Z-0579-2024 · Open, ClassifiedExactech Equinoxe Reverse Shoulder System, Catalog Numbers: a) 320-36-00, b) 320-36-03, c) 320-36-04, d) 320-36-10, e) 320-36-13, f) 320-38-00, g) 320-38-13, h) 320-40-00,…
- Z-0580-2024 · Open, ClassifiedExactech Equinoxe Shoulder System, Catalog Numbers: a) 314-23-02, b) 314-23-03, c) 314-23-13, d) 314-23-14, e) 314-23-15, f) 314-24-22, g) 314-24-23, h) 314-24-24, i) 314…
- Z-0581-2024 · Open, ClassifiedExactech Novation Hip System, Catalog Numbers: 134-36-45
- Z-0582-2024 · Open, ClassifiedExactech Optetrak Knee System, Catalog Numbers: a) 200-02-26, b) 200-02-29, c) 200-02-32, d) 200-02-35, e) 200-02-38, f) 200-02-41, g) 200-03-32, h) 200-07-29, i) 200-07-…
- Z-0584-2024 · Open, ClassifiedExactech Optetrak Logic PS Knee System, Catalog Numbers: 02-012-35-5009
- Z-0585-2024 · Open, ClassifiedExactech Truliant Knee System, Catalog Numbers: a) 02-012-35-3509, b) 02-022-35-1509, c) 02-022-35-1513, d) 02-022-35-3010, e) 02-022-35-3509, f) 02-022-35-3510, g) 02-022-…
- Z-0586-2024 · Open, ClassifiedExactech Truliant, Activit-E Knee Insert, Catalog Numbers: a) 02-023-02-0029, b) 02-023-02-0032, c) 02-023-02-0035, d) 02-023-02-0038, e) 02-024-35-1512, f) 02-024-35-1515, …
- Z-0587-2024 · Open, ClassifiedExactech Vantage Ankle System, Catalog Numbers: a) 350-21-04, b) 350-22-03
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page