Vantage recallZ-0587-2024 · Exactech, Inc. · initiated Nov 21, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Exactech, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Exactech has received 3 complaints for 4 devices related to a loss of vacuum in the inner-most vacuum bag.
FDA root cause category: Packaging change control
What the company is doing
Exactech issued an URGENT MEDICAL DEVICE RECALL notice to its direct consignees on 11/21/2023 via email. The notice explained the issue, clinical impact, and requested the removal and return of all affected products. 11/22/2023, the firm initiated the recall OUS also by email.
Recall details
- Recall number
- Z-0587-2024
- Status
- Open, Classified
- Initiated by the company
- Nov 21, 2023
- Posted by the FDA
- Jan 18, 2024
- Recalling firm
- Exactech, Inc.
- Product code
- HSN
- FDA submissions
- K152217
- Distribution
- US NY, TX, IN, IL, MN, KY, VA, MD, TN, GA, NH, ME, SC, NV, FL, AR and Australia, Brazil, Chile, France, Germany, Guatemala, Italy, Japan, Korea, Slovenia, Spain, UK
Product description
Exactech Vantage Ankle System, Catalog Numbers: a) 350-21-04, b) 350-22-03
Code information (models, lots, serial numbers)
a) 350-21-04, GTIN 10885862276247, Serial Numbers: A644020; b) 350-22-03, GTIN 10885862276353, Serial Numbers: A641653, A641655, A641659, A641660, A641662, A641663, A641666, A641667, A641668, A641669, A641672, A641654
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Vantage 350-22-03Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer2 recalls2 reports
- Vantage 350-21-04Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer2 recalls0 reports
Product lines
Other devices under the same FDA submission
58 devices cleared under K152217 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Vantage 350-22-02Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall5 reports
- Vantage 350-21-02Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall4 reports
- Vantage 350-22-01Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall3 reports
- Vantage 350-22-11Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall3 reports
- Vantage 350-21-01Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall2 reports
- Vantage 350-21-03Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall2 reports
- Vantage 350-21-13Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall2 reports
- Vantage 350-21-22Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall2 reports
- Vantage 350-21-23Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall2 reports
- Vantage 350-22-12Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall2 reports
- Vantage 350-22-23Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall2 reports
- Vantage 350-22-43Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93507).
- Z-0577-2024 · Open, ClassifiedAcuMatch Hip System, Catalog Number 142-32-27
- Z-0578-2024 · Open, ClassifiedExactech Alteon Hip XLE Liner, Catalog Numbers: a) 01-030-40-0636, b) 01-030-40-0640, c) 01-030-42-0536
- Z-0579-2024 · Open, ClassifiedExactech Equinoxe Reverse Shoulder System, Catalog Numbers: a) 320-36-00, b) 320-36-03, c) 320-36-04, d) 320-36-10, e) 320-36-13, f) 320-38-00, g) 320-38-13, h) 320-40-00,…
- Z-0580-2024 · Open, ClassifiedExactech Equinoxe Shoulder System, Catalog Numbers: a) 314-23-02, b) 314-23-03, c) 314-23-13, d) 314-23-14, e) 314-23-15, f) 314-24-22, g) 314-24-23, h) 314-24-24, i) 314…
- Z-0581-2024 · Open, ClassifiedExactech Novation Hip System, Catalog Numbers: 134-36-45
- Z-0582-2024 · Open, ClassifiedExactech Optetrak Knee System, Catalog Numbers: a) 200-02-26, b) 200-02-29, c) 200-02-32, d) 200-02-35, e) 200-02-38, f) 200-02-41, g) 200-03-32, h) 200-07-29, i) 200-07-…
- Z-0583-2024 · Open, ClassifiedExactech Optetrak Logic Knee System, Catalog Numbers: a) 02-012-47-3011, b) 02-012-47-5009, c) 02-012-47-5011, d) 02-012-65-3013
- Z-0584-2024 · Open, ClassifiedExactech Optetrak Logic PS Knee System, Catalog Numbers: 02-012-35-5009
- Z-0585-2024 · Open, ClassifiedExactech Truliant Knee System, Catalog Numbers: a) 02-012-35-3509, b) 02-022-35-1509, c) 02-022-35-1513, d) 02-022-35-3010, e) 02-022-35-3509, f) 02-022-35-3510, g) 02-022-…
- Z-0586-2024 · Open, ClassifiedExactech Truliant, Activit-E Knee Insert, Catalog Numbers: a) 02-023-02-0029, b) 02-023-02-0032, c) 02-023-02-0035, d) 02-023-02-0038, e) 02-024-35-1512, f) 02-024-35-1515, …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page