Skip to content

Truliant recallZ-0585-2024 · Exactech, Inc. · initiated Nov 21, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Exactech, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Exactech has received 3 complaints for 4 devices related to a loss of vacuum in the inner-most vacuum bag.

FDA root cause category: Packaging change control

What the company is doing

Exactech issued an URGENT MEDICAL DEVICE RECALL notice to its direct consignees on 11/21/2023 via email. The notice explained the issue, clinical impact, and requested the removal and return of all affected products. 11/22/2023, the firm initiated the recall OUS also by email.

Recall details

Recall number
Z-0585-2024
Status
Open, Classified
Initiated by the company
Nov 21, 2023
Posted by the FDA
Jan 18, 2024
Recalling firm
Exactech, Inc.
Product code
JWH
FDA submissions
K152170, K171045
Distribution
US NY, TX, IN, IL, MN, KY, VA, MD, TN, GA, NH, ME, SC, NV, FL, AR and Australia, Brazil, Chile, France, Germany, Guatemala, Italy, Japan, Korea, Slovenia, Spain, UK

Product description

Exactech Truliant Knee System, Catalog Numbers: a) 02-012-35-3509, b) 02-022-35-1509, c) 02-022-35-1513, d) 02-022-35-3010, e) 02-022-35-3509, f) 02-022-35-3510, g) 02-022-35-3511, h) 02-022-35-4511, i) 02-022-44-3512, j) 02-022-47-2509, k) 02-022-51-4519

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

a) 02-012-35-3509, GTIN 10885862304346, Serial Numbers: A730624, A730626, A730628, A730630, A730631, A730632, A730633; b) 02-022-35-1509, GTIN 10885862303974, Serial Numbers: A641578, A683756, A683757, A683758, A683759; c) 02-022-35-1513, GTIN 10885862304018, Serial Numbers: A662293; d) 02-022-35-3010, GTIN 10885862304261, Serial Numbers: A662100, A662102, A662104, A662105, A662106, A662107, A662109; e) 02-022-35-3509, GTIN 10885862304346, Serial Numbers: A652047, A652048, A652050, A652052, A652053, A652054, A730952, A730960, A730961; f) 02-022-35-3510, GTIN 10885862304353, Serial Numbers: A662121; g) 02-022-35-3511, GTIN 10885862304360, Serial Numbers: A662301, A662304, A662305; h) 02-022-35-4511, GTIN 10885862304544, Serial Numbers: A684499; i) 02-022-44-3512, GTIN 10885862305206, Serial Numbers: A641603, A641606, A641608, A641612; j) 02-022-47-2509, GTIN 10885862306326, Serial Numbers: A658478, A658486, A702263, A702265, A702267, A702268, A702269, A702271; k) 02-022-51-4519, GTIN 10885862307866, Serial Numbers: A737101, A737106, A737107;

Devices named in this recall

10 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

112 devices cleared under K152170, K171045 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 93507).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page