Olympus ENDOEYE HD II Rigid Videoscope recallZ-0697-2024 · Olympus Corporation of the Americas · initiated Nov 6, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Reports of a pink or green coloration of the image. If the issue is encountered during a procedure, device exchange could potentially result in prolonged surgery; potential visual impacts mucosal injury or bleeding may occur.
FDA root cause category: Component design/selection
What the company is doing
UPDATE: On 2/9/24 Olympus issued Correction to original letter dated November 6, 2023, regarding the IFU references in #2. The purpose of this Field Corrective Action is to reinforce the requirement in the IFU that users inspect the image prior to a clinical procedure (as described in Section 4.3.3 and 4.4), always have a spare laparoscope available (as described in Section 2.5 General Safety Notes) and to check the image for the spare equipment. Olympus issued Urgent Medical Device Correction letter on November 6, 2023 to US Operating Room Manager, Risk Management Department. Letter states reason for recall, health risk and action to take: This Field Corrective Action reminds customers to follow the steps in the instructions for use, especially to inspect the image prior to a clinical procedure, and always have a spare laparoscope available. Olympus is currently investigating technical solutions to address this issue. Carefully read the content of this notification. 2. Please make sure that all medical personnel are completely knowledgeable and thoroughly trained in the ENDOEYE HD II Instructions For Use (IFU). The purpose of this Field Corrective Action is to reinforce the requirement in the IFU that users inspect the image prior to a clinical procedure (as described in Section 7.4 Testing), always have a spare laparoscope available (as described in Section 2.5 General Dangers, Warnings and Cautions) and to check the image for the spare equipment. 3. If you have further distributed this product, identify your customers and forward them this Customer Notification. Please appropriately document your notification process and let us know the end-customer feedback accordingly. 4. Complete the enclosed response form and return it to our third-party partner, Sedgwick, via e-mail at [email protected] or by fax 866-551-2707. For any questions about the acknowledgement form, please call the Sedgwick team at 866-875-0524. If you require additional information o
Recall details
- Recall number
- Z-0697-2024
- Status
- Open, Classified
- Initiated by the company
- Nov 6, 2023
- Posted by the FDA
- Jan 10, 2024
- Recalling firm
- Olympus Corporation of the Americas
- Quantity
- WA50040A : 561 units; WA50042A: 857 units
- Product code
- HET
- FDA submissions
- K190744
- Distribution
- Worldwide - US Nationwide distribution.
Product description
Olympus ENDOEYE HD II Rigid Videoscope Catalog Numbers: WA50040A, WA50042A -used with a video system center, light source, documentation equipment, monitor, hand instruments, electrosurgical unit, and other ancillary equipment for endoscopy and endoscopic surgery within the thoracic and abdominal cavities including the female reproductive organs.
Code information (models, lots, serial numbers)
UDI-DI: WA50040A: 04042761074964 WA50042A: 04042761074971 All Serial Numbers
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Olympus WA50042AOLYMPUS Winter & Ibe GmbH · Laparoscope, Gynecologic (And Accessories)1 recall6,135 reports
- Olympus WA50040AOLYMPUS Winter & Ibe GmbH · Laparoscope, Gynecologic (And Accessories)1 recall954 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.