Skip to content

Olympus WA50040ARecalls & reported problems

OLYMPUS Winter & Ibe GmbH · Video telescope "ENDOEYE HD II", 10 mm, 0°, autoclavable · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus ENDOEYE HD II video laparoscope is a camera used during keyhole surgery. The 942 reports mostly describe poor image quality, transmission problems, a blank display and breakage. One report describes a full-thickness burn. The reports include no deaths and 4 injuries. There is 1 recall, open, started November 2023, about pink or green image tints, which could mean swapping devices mid-surgery and potential tissue injury or bleeding.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
954
Deaths reported
0
Injuries reported
4
Recalls naming it
1

Summary

The FDA's MAUDE database contains 954 adverse event reports that list this exact device (DI 04042761074964), received between Dec 9, 2022 and Sep 30, 2026 — classified as 4 injuries, 950 malfunctions.

The most frequently reported specific device problem is “Poor Quality Image” (406 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 6 reportsDec 2023: 9 reportsJan 2024: 7 reportsFeb 2024: 5 reportsMar 2024: 11 reportsApr 2024: 10 reportsMay 2024: 32 reportsJun 2024: 45 reportsJul 2024: 38 reportsAug 2024: 25 reportsSep 2024: 29 reportsOct 2024: 26 reportsNov 2024: 21 reportsDec 2024: 25 reportsJan 2025: 34 reportsFeb 2025: 31 reportsMar 2025: 25 reportsApr 2025: 21 reportsMay 2025: 24 reportsJun 2025: 16 reportsJul 2025: 32 reportsAug 2025: 27 reportsSep 2025: 25 reportsOct 2025: 18 reportsNov 2025: 24 reportsDec 2025: 69 reportsJan 2026: 20 reportsFeb 2026: 24 reportsMar 2026: 45 reportsApr 2026: 43 reportsMay 2026: 32 reportsJun 2026: 19 reportsJul 2026: 19 reportsAug 2026: 20 reportsSep 2026: 22 reportsOct 2026: 0 reports20242025202669
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 6, 2023Open, Classified

    Reports of a pink or green coloration of the image. If the issue is encountered during a procedure, device exchange could potentially result in prolonged surgery; potential visual impacts mucosal injury or bleeding may occur.

    Z-0697-2024 · Root cause: Component design/selection · Recall details · FDA record ↗

Most reported problems

Device problems

  • Poor Quality Image406
  • Communication or Transmission Problem131
  • No Display/Image105
  • Break103
  • Optical Discoloration42
  • Fracture36
  • Dent in Material35
  • Erratic or Intermittent Display34

Patient problems

  • No Clinical Signs, Symptoms or Conditions952
  • Full thickness (Third Degree) Burn1
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 954 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Poor Quality Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50953875 ↗
  • MalfunctionSep 30, 2026Break; Poor Quality Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50953229 ↗
  • MalfunctionSep 29, 2026Poor Quality Image

    The customer reported a green image involving an Olympus rigid video laparoscope. There was no patient injury reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50935097 ↗
  • MalfunctionSep 28, 2026Poor Quality Image

    The customer reported focus image issue during laparoscopic cholecystectomy therapeutic procedure that was completed with an Olympus back-up device involving an Olympus rigid video laparoscope. There were no reports of patient harm.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50914573 ↗
  • MalfunctionSep 24, 2026Poor Quality Image

    The customer reported the Olympus rigid video laparoscope had image interference/lines and image foggy/misty during installation. There was no patient involvement reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50889549 ↗
  • MalfunctionSep 24, 2026Material Split, Cut or Torn

    The customer returned the rigid video laparoscope for a separate issue. During the initial evaluation of the returned device, fiber bonding in the outer tube area (distal end) was damaged. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50882954 ↗
  • MalfunctionSep 23, 2026Break; Poor Quality Image

    The customer reported the Olympus rigid video scope image became green during inspection for use. During the initial evaluation of the device, a defective light guide bundle at the distal end was observed. There were no reports of patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50872930 ↗
  • MalfunctionSep 23, 2026No Display/Image; Fracture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50872918 ↗
  • MalfunctionSep 23, 2026Gas/Air Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50871482 ↗
  • MalfunctionSep 22, 2026Poor Quality Image

    The customer reported a noisy image during maintenance involving an Olympus Rigid Video Laparoscope. Additionally, during the initial evaluation of the device, multiple horizontal stripes of noise were observed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50857290 ↗
  • MalfunctionSep 22, 2026Crack; Poor Quality Image

    The customer returned the rigid video laparoscope to the service center for a separate issue. During the device analysis, the service repair technician indicated that a light guide lens with a crack and a noisy image were found. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50855614 ↗
  • MalfunctionSep 22, 2026Break; Erratic or Intermittent Display; Poor Quality Image

    The customer reported the rigid video laparoscope had an abnormal image and the screen became noised during reprocessing. During the initial evaluation of the device, it was observed the light guide bundle of insertion section was slightly interrupted and weakened; the universal cord malfunctioned and the insertion beam was slightly dimmed. There were no reports of patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50853055 ↗
  • MalfunctionSep 21, 2026Crack; Scratched Material

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50848716 ↗
  • MalfunctionSep 21, 2026Break; Poor Quality Image; Output Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50847716 ↗
  • MalfunctionSep 21, 2026Break; Display or Visual Feedback Problem

    The customer reported that the rigid video laparoscope displayed image noise during therapeutic laparoscopy procedure while the patient was under sedation. The intended procedure was completed using a backup device. During the initial evaluation of the device, the fiber bonding in the outer tube area at the distal end was damaged, and no image was displayed. No patient harm was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50846453 ↗
  • MalfunctionSep 21, 2026No Display/Image; Communication or Transmission Problem; Material Integrity Problem

    The customer reported the rigid video laparoscope no image during Inspection for use. During the initial evaluation of the device it was observed, E218 (scope error) and the fiber bonding in the outer tube area (distal end) was damaged. No patient involvement was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50838912 ↗
  • MalfunctionSep 14, 2026No Display/Image; Material Integrity Problem

    The customer returned the rigid video laparoscope to the service center for a separate issue. During the device analysis, the service repair technician indicated that no image and damaged fibre bonding in the outer tube area at the distal end were found. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50756325 ↗
  • MalfunctionSep 11, 2026Communication or Transmission Problem

    The customer reported E104 fog free function for an Olympus rigid video laparoscope. There were no reports of patient harm.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50734039 ↗
  • MalfunctionSep 11, 2026Break; Crack; Poor Quality Image; Dent in Material; Scratched Material

    The customer reported a broken lens during reprocessing involving an Olympus rigid video laparoscope. Additionally, during the initial evaluation of the device, a cracked objective lens, cracked light guide connector and cover glass, dents and scratches on the outer tube, and a cracked image were observed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50733894 ↗
  • MalfunctionSep 10, 2026Poor Quality Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50722085 ↗
  • MalfunctionSep 9, 2026Break; Erratic or Intermittent Display; Electrical /Electronic Property Problem; Residue After Decontamination; Communication or Transmission Problem; Material Deformation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50692069 ↗
  • MalfunctionSep 1, 2026Material Puncture/Hole; Optical Discoloration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50534596 ↗
  • MalfunctionAug 26, 2026Scratched Material

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50460823 ↗
  • MalfunctionAug 25, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50438731 ↗
  • MalfunctionAug 18, 2026Fogging

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50303649 ↗