OLYMPUS Winter & Ibe GmbH · Video telescope "ENDOEYE HD II", 10 mm, 30°, autoclavable · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus ENDOEYE HD II video telescope is a laparoscope used in gynecologic surgery. The reports often describe poor image quality, a missing or flickering display, and breaks in the scope. FDA records list 6,045 reports, no deaths, 11 injuries and 6,034 malfunctions. There is 1 open recall, which began in November 2023 and concerns pink or green tinting of the image, which could mean swapping devices during surgery and possibly causing bleeding or tissue injury.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
6,135
Deaths reported
0
Injuries reported
12
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6,135 adverse event reports that list this exact device (DI 04042761074971), received between Sep 24, 2018 and Sep 30, 2026 — classified as 12 injuries, 6,123 malfunctions.

The most frequently reported specific device problem is “Poor Quality Image” (2,815 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 58 reportsDec 2023: 68 reportsJan 2024: 57 reportsFeb 2024: 41 reportsMar 2024: 53 reportsApr 2024: 66 reportsMay 2024: 135 reportsJun 2024: 215 reportsJul 2024: 223 reportsAug 2024: 122 reportsSep 2024: 199 reportsOct 2024: 207 reportsNov 2024: 95 reportsDec 2024: 121 reportsJan 2025: 283 reportsFeb 2025: 137 reportsMar 2025: 157 reportsApr 2025: 138 reportsMay 2025: 177 reportsJun 2025: 168 reportsJul 2025: 147 reportsAug 2025: 187 reportsSep 2025: 120 reportsOct 2025: 144 reportsNov 2025: 235 reportsDec 2025: 370 reportsJan 2026: 181 reportsFeb 2026: 198 reportsMar 2026: 295 reportsApr 2026: 304 reportsMay 2026: 203 reportsJun 2026: 98 reportsJul 2026: 63 reportsAug 2026: 129 reportsSep 2026: 150 reportsOct 2026: 0 reports202420252026370
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 6, 2023Open, Classified

    Reports of a pink or green coloration of the image. If the issue is encountered during a procedure, device exchange could potentially result in prolonged surgery; potential visual impacts mucosal injury or bleeding may occur.

    Z-0697-2024 · Root cause: Component design/selection · Recall details · FDA record ↗

Most reported problems

Device problems

  • Poor Quality Image2,815
  • Communication or Transmission Problem688
  • No Display/Image673
  • Break587
  • Optical Discoloration378
  • Erratic or Intermittent Display337
  • Fracture236
  • Dent in Material193

Patient problems

  • No Clinical Signs, Symptoms or Conditions6,112
  • Insufficient Information25
  • Arrhythmia1
  • Hemoptysis1
  • No Known Impact Or Consequence To Patient1
  • Respiratory Tract Infection1

Recent adverse event reports

The 25 most recent of 6,135 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Poor Quality Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50955058 ↗
  • MalfunctionSep 30, 2026Poor Quality Image; Output Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954687 ↗
  • MalfunctionSep 30, 2026Poor Quality Image; Connection Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50953656 ↗
  • MalfunctionSep 30, 2026Break; Poor Quality Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50953456 ↗
  • MalfunctionSep 30, 2026Poor Quality Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50952858 ↗
  • MalfunctionSep 30, 2026Poor Quality Image; Material Deformation; Optical Discoloration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50952205 ↗
  • MalfunctionSep 30, 2026Fracture; Optical Discoloration

    It was reported that during an unknown procedure with use of the gynecologic laparoscope, the viewable image developed a green tint/discoloration on the screen. The device was replaced with a back-up device and the procedure was completed. During the initial evaluation of the device, it was observed that the fibre bonding in the outer tube area (distal end) is damaged and the video cable is broken (which resulted in the green image.)No patient harm was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950835 ↗
  • MalfunctionSep 30, 2026Erratic or Intermittent Display; Fracture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950027 ↗
  • MalfunctionSep 30, 2026No Display/Image; Display or Visual Feedback Problem

    It was reported that during a diagnostic procedure, the videoscope's monitor displayed a black screen. The customer also reported that the image suddenly shifted entirely to green tones, and they had to replace the scope. There were no reports of any patient harm.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949673 ↗
  • MalfunctionSep 30, 2026Fracture; Poor Quality Image

    The customer reported purple stripes in the image during inspection for use. During the initial evaluation of the device, it was observed that the fiber bonding in the outer tube area (distal end) was damaged. involving an Olympus gynecologic laparoscope. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949262 ↗
  • MalfunctionSep 30, 2026Poor Quality Image; Material Integrity Problem

    The customer reported that the image appears green involving an Olympus Laparoscope, Gynecologic. Additionally, during the initial evaluation of the device, damaged fibre bonding in the outer tube area was observed. There was no patient injury reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948591 ↗
  • MalfunctionSep 30, 2026Failure to Clean Adequately

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947890 ↗
  • MalfunctionSep 30, 2026Degraded; Fracture

    The customer returned the Olympus Laparoscope Gynecologic for a separate issue. During the initial evaluation of the returned device, protector of the video cable section is overstressed and the fiber bonding in the outer tube area (distal end) is damaged observed. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947849 ↗
  • MalfunctionSep 30, 2026Communication or Transmission Problem

    It was reported that during an unspecified therapeutic procedure, the gynecologic laparoscope experienced a B-31 error code. The reported event resulted in a procedural prolongation of less than 30 minutes. There were no reports of patient harm.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947557 ↗
  • MalfunctionSep 30, 2026No Display/Image; Poor Quality Image

    It was reported that at the start of an unknown procedure using the gynecologic laparoscope, green and purple lines were seen at the top of the screen. The customer reported that they restarted the device after a short time, the lines reappeared then the screen went black. Another restart was performed without image problems, and the procedure was completed. There was no report of patient harm or adverse patient event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947207 ↗
  • MalfunctionSep 30, 2026Crack; Poor Quality Image

    The customer reported poor image quality involving an Olympus Laparoscope, Gynecologic. Additionally, during the initial evaluation of the device, a green image and a cracked cover glass in the Insertion section were observed. There was no patient injury reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50944764 ↗
  • MalfunctionSep 29, 2026Break; Dent in Material; Material Integrity Problem; Optical Problem; Sharp Edges

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50938659 ↗
  • MalfunctionSep 29, 2026Break; Erratic or Intermittent Display; No Display/Image; Fracture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50934233 ↗
  • MalfunctionSep 29, 2026Break; Optical Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50934012 ↗
  • MalfunctionSep 29, 2026Crack; Poor Quality Image

    The customer reported that the image looked green involving an Olympus Laparoscope. During initial evaluation, cracked coverglass of the insertion section was observed. There were no reports of patient harm.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50933050 ↗
  • MalfunctionSep 28, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50924219 ↗
  • MalfunctionSep 28, 2026Break; Degraded; Erratic or Intermittent Display

    The customer reported image discoloration during an unspecified procedure involving an Olympus Laparoscope, Gynecologic, which was completed with the same device. Additionally, during the initial evaluation of the device, a broken Charge Coupled Device unit, distal end of the insertion section with contour sharpness, and a green image were observed. There was no patient injury reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50920327 ↗
  • MalfunctionSep 28, 2026Break; Erratic or Intermittent Display; Electrical /Electronic Property Problem; Fracture; Communication or Transmission Problem

    The customer reported the Olympus Gynecologic Laparoscope had an Error E226 during an unspecified procedure. During the initial evaluation of the device it was observed, the fiber bonding in the outer tube area is damaged, the Charged Coupled Device is broken, the dielectric strength is inadequate and stripes in image occur were observed. No patient harm was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50917926 ↗
  • MalfunctionSep 28, 2026Leak/Splash; Failure to Clean Adequately

    The customer returned the Gynecologic Laparoscope¿to the service center for evaluation related to a separate issue. During testing, the service repair technician identified dirt on the grip and water leakage in the device. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50917237 ↗
  • MalfunctionSep 28, 2026No Display/Image; Communication or Transmission Problem

    It was reported that during an unspecified therapeutic procedure, the Rigid Video Scope displayed the error "Fog Free function has been disabled" after being placed in a scope warmer, during which the scope image went black. The fog-free function subsequently remained nonfunctional. The procedure was completed using a backup device with no reports of further patient harm.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50916598 ↗

Other Laparoscope, Gynecologic (And Accessories) devices

All Laparoscope, Gynecologic (And Accessories) devices →