Volar Bearing Plates recallZ-0932-2024 · TriMed Inc. · initiated Nov 28, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from TriMed Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Plates, part of a wrist fixation system, are affixed during surgery with bearings that may pop out from distal peg holes.

FDA root cause category: Labeling design

What the company is doing

On 11/28/23, recall notices were emailed to customers who were asked to do the following: 1) Stop further use of the device and return any unused products to the recalling firm. 2) Provide a copy of this recall notice to the hospital/surgeon/physician that is using the device. 3) Complete and return the response form to [email protected] Contact [email protected] for any additional questions.

Recall details

Recall number
Z-0932-2024
Status
Open, Classified
Initiated by the company
Nov 28, 2023
Posted by the FDA
Jan 31, 2024
Recalling firm
TriMed Inc.
Quantity
270
Product code
HRS
FDA submissions
K222637
Distribution
US: FL, MI, AL, NY, CA

Product description

Volar Bearing Plate. REF: VBEAL-3-7N, VBEAL-5-7N, VBEAL-3-7S, VBEAL-5-7S, VBEAL-7-7S, VBEAL-3-7W, VBEAL-5-7W, VBEAR-3-7N, VBEAR-5-7-N, VBEAR-3-7S, VBEAR-5-7S, VBEAR-7-7S, VBEAR-3-7W, VBEAR-5-7W. Intended to be used as an aid to the treatment of certain types of fractures, non-unions or osteotomies.

Code information (models, lots, serial numbers)

REF/UDI-DI/Lots: VBEAL-3-7N/00842188124907/ME23071, ME23072; VBEAL-5-7N/00842188124914/ME23073; VBEAL-3-7S/00842188124938/ME23031, ME23067, ME23082, ME23097; VBEAL-5-7S/00842188124945/ME23068; VBEAL-7-7S/00842188124952/ME23004, ME23086; VBEAL-3-7W/00842188124969/ME23033; VBEAL-5-7W/00842188124976/ME23036; VBEAR-3-7N/00842188124990/ME23069, ME23084, ME23085; VBEAR-5-7-N/00842188125003/ME23074; VBEAR-3-7S/00842188125027/ME23029, ME23087; VBEAR-5-7S/00842188125034/ME23019; VBEAR-7-7S/00842188125041/ME23022; VBEAR-3-7W/00842188125058/ME23034; VBEAR-5-7W/00842188125065/ME23041, ME23037

Devices named in this recall

14 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

1 device cleared under K222637 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page