Skip to content

Volar Bearing Plates VBEAL-3-7NRecalls & reported problems

Trimed, INC. · Volar Bearing Plate, 3-Hole, Lt, Narrow · Class 2 · Implant

All models of Volar Bearing Plates (16)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00842188124907), received between Nov 28, 2023 and Nov 28, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 28, 2023Class IIOpen, Classified

    Plates, part of a wrist fixation system, are affixed during surgery with bearings that may pop out from distal peg holes.

    Z-0932-2024 · Root cause: Labeling design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Off-Label Use1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 28, 2023Off-Label Use; Detachment of Device or Device Component; Off-Label Use; Detachment of Device or Device Component

    TRIMED'S MEDICAL ADVISOR REPORTED ON 05NOV2023 THAT A SURGEON INFORMED HIM THAT ON ONE CASE, HE WAS USING A WRIST FIXATION SYSTEM 3 VOLAR BEARING PLATE AND USED A STANDARD BONE SCREW TO PULL THE PLATE UP TO THE BONE. THIS IS NOT ON THE TRIMED SURGICAL TECHNIQUE MANUAL, THEREFORE THIS WAS OFF LABEL USE. WHEN THE SURGEON WENT TO PUT THE PEG IN, HE THINKS THAT THE BEARING MAY HAVE COME OUT OF THE BOTTOM OF THE PLATE, BUT FELT THAT FIXATION WAS EXCELLENT, WAS PLEASED WITH THE RESULT, AND DIDN'T FEEL REMOVAL WAS NEEDED OR NECESSARY. IT HAS BEEN REPORTED THAT THE PATIENT HAS DONE GREAT AND THIS DID NOT CAUSE ADDITIONAL HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18223020 ↗

Other Plate, fixation, bone devices

All Plate, fixation, bone devices →

Sources and updates