Total immunoglobulin M (IgM total) IVD, kit… recallZ-1003-2024 · Siemens Healthcare Diagnostics, Inc. · initiated Dec 15, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for a negative bias with quality control (QC) and patient sample results when using the Atellica CH Immunoglobulin M_2 (IgM_2) reagent.
FDA root cause category: Under Investigation by firm
What the company is doing
An Urgent Medical Device Correction notification letter dated 12/19/23 was sent to customers. Actions to be Taken by the Customer: Siemens Healthineers recommends batch testing samples for Atellica CH IgM_2 as follows: 1. Remove and discard any Atellica CH IgM_2 reagent packs onboard the analyzer. 2. Load a single fresh Atellica CH IgM_2 reagent pack onto the analyzer. 3. Perform a Lot calibration and process Quality Control (QC). 4. Immediately process a batch of patient samples and conclude with a repeat run of QC. Patient results should not be reported until the QC performed at the end of the batch run has been assessed. - If the QC results are within the established range, patient results can be reported. - If the QC results are not within the established range, do not report patient results and repeat steps 1 4 above. 5. Remove and discard the Atellica CH IgM_2 reagent pack at the end of the batch run. Siemens Healthineers does not recommend using the ADVIA IgM_2 reagent on the Atellica CH or Atellica CI Analyzers. Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. We apologize for the inconvenience this situation may cause. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
Recall details
- Recall number
- Z-1003-2024
- Status
- Open, Classified
- Initiated by the company
- Dec 15, 2023
- Posted by the FDA
- Feb 7, 2024
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 7,075 units (927 US, 6148 OUS)
- Product code
- CFN
- FDA submissions
- K151767, K992662
- Distribution
- Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, WI, and PR; The countries of Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, Colombia, Croatia, Curacao, Czech Republic, Denmark, Egypt, Estonia, Finland France, Germany, Greece, Hong Kong, Hungary, India, Iran, Ireland, Israel, Italy, Ivory Coast, Japan, Jordan, Kenya, Kuwait, Latvia, Macedonia, Mexico, Morocco, Netherlands, New Zealand, Norway, Oman, P.R. China, Pakistan, Paraguay, Peru, Poland, Portugal, Republic South Korea, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Turkey, U.A.E, United Kingdom, Uruguay, Vatican City, Vietnam, Zaire.
Product description
Total immunoglobulin M (IgM total) IVD, kit, nephelometry/turbidimetry
Code information (models, lots, serial numbers)
Siemens Material Number (SMN): 11097620; UDI DI: 00630414595627; Lot No. (Exp. Date): 221764 (10/01/2023), 221829 (02/01/2024), 231915 (06/01/2024), 232025 (11/01/2024).
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Atellica CH IgM_2 11097620Siemens Healthcare Diagnostics Inc. · METHOD, NEPHELOMETRIC, IMMUNOGLOBULINS (G, A, M)1 recall0 reports
Other devices under the same FDA submission
7 devices cleared under K151767, K992662 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Atellica CH 930 Analyzer 11067000Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE5 recalls679 reports
- Atellica CH Diluent 11099300Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall0 reports
- Atellica A-LYTE™ Integrated Multisensor Na+ / K+ / Cl- 11099315Siemens Healthcare Diagnostics Inc. · ELECTRODE, ION SPECIFIC, POTASSIUM25 reports
- Atellica Direct Load 11068008Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE3 reports
- ADVIA® Chemistry IGM_2 Reagents 10309032Siemens Healthcare Diagnostics Inc. · METHOD, NEPHELOMETRIC, IMMUNOGLOBULINS (G, A, M)2 reports
- Atellica CH Cleaner 11099303Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 report
- Atellica CH AlbP 11097530Siemens Healthcare Diagnostics Inc. · BROMCRESOL PURPLE DYE-BINDING, ALBUMIN1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.