Captus 4000e Thyroid Uptake Systems- Intended to be used by… recallZ-1069-2024 · Mirion Technologies (Capintec), Inc. · initiated Nov 27, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Mirion Technologies (Capintec), Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Spring arm failure with Captus 4000e Thyroid Uptake Systems, can cause the collimator to fall downward to its lowest point of travel, causing potential injury with a patient or operator
FDA root cause category: Device Design
What the company is doing
Capintec issued Product Urgent Safety Alert Preventative Action letter on 11/27/23 via email. Letter states reason for recall, health risk, and action to take: 1. Review this Alert and ensure that all affected personnel, including all operators of Captus 4000e Thyroid Uptake Systems, are aware of the contents. 2. Instruct affected personnel and operators to comply with the following steps in the interest of patient and operator safety: a. Prior to moving the arm, ensure that the positioning locks are released. b. Use caution when moving the spring arm vertically. Ensure that when the spring arm is moved vertically, the operator and patient are not below the spring arm or in its travel path. Move the spring arm using an outstretched arm, ensuring no body parts are in the spring arm s travel path. c. Once in position, secure the arm by tightening the locking handle prior to performing Thyroid Uptake and Bioassay Procedures. d. Perform Thyroid Uptake and Bioassay Procedures with patient (or employee, for Bioassay Procedures) in a seated position, rather than a supine position on a table. e. Store the arm in an upright position. This places the least amount of stress on the internal components. Engage the locking handle when the spring arm is not being raised or lowered. If you have any questions about this Safety Alert, the Preventive Action steps outlined above, or your Captus 4000e Thyroid Uptake System, please contact Capintec Customer Support at the phone number or email listed below. If you notice any unusual change in the performance or functional response of the arm in your system (e.g. arm is making an unusual noise or arm does not move smoothly at any articulation joint), please contact Capintec Customer Support: Capintec Customer Support: 1-800-631-3826 (201) 825-9500 [email protected]
Recall details
- Recall number
- Z-1069-2024
- Status
- Open, Classified
- Initiated by the company
- Nov 27, 2023
- Posted by the FDA
- Feb 9, 2024
- Recalling firm
- Mirion Technologies (Capintec), Inc.
- Quantity
- 467 units
- Product code
- IZD
- Distribution
- Nationwide Foreign: AUSTRIA BANGLADESH BRASIL Brazil CANADA CHILE Costa Rica Ecuador France Hanoi Vietnam Hong Kong INDIA ISRAEL ITALY Malaysia PAKISTAN Philippines POLAND Saudi Arabia SOUTH KOREA Spain SWEDEN TAIWAN THAILAND TURKEY UNITED ARAB EMIATES UNITED KINGDOM VIETNAM
Product description
Captus 4000e Thyroid Uptake Systems- Intended to be used by trained Nuclear Medicine Technologists or Nuclear Medicine Physicians to perform Thyroid Uptake Procedures. Models: 5430-30151, 5430-30152, and 5430-30154
Code information (models, lots, serial numbers)
UDI: 0859942006096 and 0859942006102 Serial Numbers: 940000 through 940535
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Captus 4000e Thyroid Uptake SystemMirion Technologies (capintec), Inc. · Probe, Uptake, Nuclear3 recalls5 reports
- Captus 4000e Thyroid Uptake System with Well Captus 4000e Thyroid System with WellMirion Technologies (capintec), Inc. · Probe, Uptake, Nuclear3 recalls4 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.