Captus 4000e Thyroid Uptake System with Well Captus 4000e Thyroid System with WellRecalls & reported problems

Mirion Technologies (capintec), Inc. · The Captus 4000e is a PC based Thyroid Uptake System, on a ergonomically designed mobile stand. It includes a 1024 channel MCA, automated Quality Assurance, Thyroid Uptake Module, Bioassay module, Wipe Test Module, and Lab Tests. · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00859942006102), received between Mar 2, 2019 and Jan 10, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Mechanical Problem” (4 reports).

3 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 20, 2026Open, Classified

    Potential for failure of the tension bolt which would allow the spring arm to unexpectedly fall to its lowest position, which is approximately 25 inches (63.5 cm) from the ground.

    Z-3193-2026 · Root cause: Device Design · Recall details · FDA record ↗

  2. Apr 22, 2024Open, Classified

    Complaints of unexpected detachment of the collimator have been reported.

    Z-1893-2024 · Root cause: Device Design · Recall details · FDA record ↗

  3. Nov 27, 2023Open, Classified

    Spring arm failure with Captus 4000e Thyroid Uptake Systems, can cause the collimator to fall downward to its lowest point of travel, causing potential injury with a patient or operator

    Z-1069-2024 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Mechanical Problem4
  • Detachment of Device or Device Component1

Patient problems

  • No Consequences Or Impact To Patient2
  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 10, 2024Mechanical Problem; Detachment of Device or Device Component; Mechanical Problem; Detachment of Device or Device Component

    CAPTUS 4000E THYROID UPTAKE SYSTEM S/N (B)(6) WAS SHIPPED TO (B)(6) MEDICAL CENTER (NOW (B)(6)HEALTH) ON FEBRUARY 5, 2021. THE NUCLEAR MEDICINE TECHNICIAN WAS PERFORMING QUALITY ASSURANCE. THE COLLIMATOR UNEXPECTED DETACHED FROM THE SYSTEM ARM. PRIOR TO THIS EVENT, THE SYSTEM HAD BEEN FUNCTIONING PROPERLY. NO INJURY WAS REPORTED BY THE OPERATOR. THE COLLIMATOR ASSEMBLY WEIGHS 28 POUNDS, AND THERE IS THE POTENTIAL FOR SERIOUS INJURY IF THE COLLIMATOR WERE TO FALL AND COME INTO CONTACT WITH A PATIENT OR OPERATOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18485762 ↗
  • MalfunctionJul 3, 2020Mechanical Problem; Mechanical Problem

    ON (B)(6) 2020, THE LEAD NUCLEAR MEDICINE TECHNOLOGIST FROM (B)(6) CONTACTED CAPINTEC TO REPORT THAT THE COLLIMATOR DETACHED FROM THE SPRING ARM OF THEIR CAPTUS 4000E SYSTEM, S/N (B)(4). THE COLLIMATOR ASSEMBLY WEIGHS 28 POUNDS, AND THERE IS THE POTENTIAL FOR INJURY IF THE COLLIMATOR WERE TO FALL AND COME INTO CONTACT WITH A PATIENT OR OPERATOR. THE PATIENT WAS SEATED ON A CHAIR AND THE TECHNOLOGIST WAS PREPARING TO MEASURE THE PATIENT'S THYROID. THE TECHNOLOGIST WAS ABLE TO HOLD THE COLLIMATOR TO PREVENT IT FROM FALLING. NO INJURY OCCURED TO EITHER PATIENT OR OPERATOR.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10228437 ↗
  • MalfunctionOct 8, 2019Mechanical Problem; Mechanical Problem

    ON 9/19/19 END USER CONTACTED CAPINTEC TO REPORT THAT THE COLLIMATOR FELL OFF A CAPTUS 4000E SYSTEM, S/N (B)(4). USER WAS PERFORMING QC ON THE SYSTEM, AND FOUND THE COLLIMATOR DIFFICULT TO TURN. CONTINUING TO MOVE THE COLLIMATOR CAUSED IT TO FALL OFF THE ARM. THE FALL WAS BROKEN BY THE CABLE. NO INJURY OCCURRED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9166032 ↗
  • MalfunctionMar 2, 2019Mechanical Problem; Mechanical Problem

    ON (B)(6) 2019 CAPINTEC WAS INFORMED THAT THE COLLIMATOR FELL OFF A CAPTUS 4000 THYROID UPDATE SYSTEM- SERIAL NUMBER (B)(4). TECHNOLOGIST JUST COMPLETED QC ON THE SYSTEM. SHE WAS PERFORMING CAPSULE COUNTS WHEN THE COLLIMATOR FELL OFF. NO INJURIES OCCURRED. SERVICE PERSONNEL REPLACED THE COLLIMATOR ASSEMBLY ON SITE. USER WAS INSTRUCTED TO RETURN FAILED PARTS FOR EVALUATION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8384799 ↗

Other Probe, Uptake, Nuclear devices