Captus 4000e Thyroid Uptake SystemRecalls & reported problems
Mirion Technologies (capintec), Inc. · The Captus 4000e is a PC based Thyroid Uptake System, on a ergonomically designed mobile stand. It includes a 1024 channel MCA, automated Quality Assurance, Thyroid Uptake Module, and Bioassay module. · Class 1
3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 3
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00859942006096), received between Aug 14, 2023 and Nov 30, 2023 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (5 reports).
3 FDA recall records name this device's identifier, 3 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Jul 20, 2026Open, Classified
Potential for failure of the tension bolt which would allow the spring arm to unexpectedly fall to its lowest position, which is approximately 25 inches (63.5 cm) from the ground.
Z-3193-2026 · Root cause: Device Design · FDA record ↗
- Apr 22, 2024Open, Classified
Complaints of unexpected detachment of the collimator have been reported.
Z-1892-2024 · Root cause: Device Design · FDA record ↗
- Nov 27, 2023Open, Classified
Spring arm failure with Captus 4000e Thyroid Uptake Systems, can cause the collimator to fall downward to its lowest point of travel, causing potential injury with a patient or operator
Z-1069-2024 · Root cause: Device Design · FDA record ↗
Most reported problems
Device problems
- Detachment of Device or Device Component5
- Mechanical Problem5
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Skin Discoloration1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 30, 2023Mechanical Problem; Detachment of Device or Device Component; Mechanical Problem; Detachment of Device or Device Component
THYROID UPTAKE SYSTEM S/N (B)(6) WAS SHIPPED ON 2/23/2016. THE TECHNICIAN WAS PERFORMING A THYROID UPTAKE ON PATIENT WHEN THE ARM UNEXPECTEDLY FELL TO ITS LOWEST POSITION. REVIEW BY ENGINEERING INDICATED A SPRING ARM FAILURE. NO INJURIES OCCURRED. A REPLACEMENT ASSEMBLY WAS SHIPPED AND USER INSTRUCTED TO RETURN THE DEFECTIVE PART TO MANUFACTURER FOR ENGINEERING EVALUATION AND IDENTIFICATION OF ROOT CAUSE. THE COLLIMATOR AND ARM WEIGH AND SPRING ARM ASSEMBLY WEIGH 45 POUNDS, WHICH HAS THE POTENTIAL FOR SERIOUS INJURY TO EITHER PATIENT OR OPERATOR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18236365 ↗MalfunctionNov 22, 2023Mechanical Problem; Detachment of Device or Device Component; Mechanical Problem; Detachment of Device or Device Component
THYROID UPTAKE SYSTEM S/N (B)(6) WAS SHIPPED ON APRIL 19, 2018. THE TECHNICIAN WAS PERFORMING ROUTINE QC WHEN THE SYSTEM'S ARM UNEXPECTEDLY FELL TO ITS LOWEST POSITION. REVIEW BY ENGINEERING INDICATED A SPRING ARM FAILURE. NO INJURIES OCCURRED. A REPLACEMENT ASSEMBLY WAS SHIPPED AND USER INSTRUCTED TO RETURN THE DEFECTIVE PART TO MANUFACTURER FOR ENGINEERING EVALUATION AND IDENTIFICATION OF ROOT CAUSE. THE COLLIMATOR AND ARM WEIGH AND SPRING ARM ASSEMBLY WEIGH 45 POUNDS, WHICH HAS THE POTENTIAL FOR SERIOUS INJURY TO EITHER PATIENT OR OPERATOR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18190383 ↗MalfunctionNov 22, 2023Mechanical Problem; Detachment of Device or Device Component; Mechanical Problem; Detachment of Device or Device Component
THYROID UPTAKE SYSTEM S/N (B)(6) WAS SHIPPED ON SEPTEMBER 2015. THE TECHNICIAN WAS PERFORMING AN INSPECTION AND NOTED DAMAGED CABLE AND BROKEN STOPPING PIN. REVIEW BY ENGINEERING INDICATED A SPRING ARM FAILURE. NO INJURIES OCCURRED. A REPLACEMENT ASSEMBLY WAS SHIPPED AND USER INSTRUCTED TO RETURN THE DEFECTIVE PART TO MANUFACTURER FOR ENGINEERING EVALUATION AND IDENTIFICATION OF ROOT CAUSE. THE COLLIMATOR AND ARM WEIGH AND SPRING ARM ASSEMBLY WEIGH 45 POUNDS, WHICH HAS THE POTENTIAL FOR SERIOUS INJURY TO EITHER PATIENT OR OPERATOR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18190382 ↗MalfunctionAug 14, 2023Mechanical Problem; Detachment of Device or Device Component; Mechanical Problem; Detachment of Device or Device Component
CAPTUS 4000E THYROID UPTAKE SYSTEM S/N (B)(6) WAS SHIPPED TO RAPP-ISO GMBH, GERMANY IN AUGUST 2020. ON JUNE 24, 2023 THE TECHNICIAN ATTEMPTED TO OPERATE THE CAPTUS 4000E TO MEASURE A PHANTOM. THE PROBE UNEXPECTEDLY DISLODGED FROM THE CARRIAGE ARM. THE ARM'S ARTICULATION HAD BECOME LOOSE AND EVENTUALLY DETACHED. PRIOR TO THIS EVENT, THE SYSTEM HAD BEEN FUNCTIONING PROPERLY. USER REPORTED NO INJURY. THE COLLIMATOR ASSEMBLY WEIGHS 28 POUNDS, AND THERE IS THE POTENTIAL FOR SERIOUS INJURY IF THE COLLIMATOR WERE TO FALL AND COME INTO CONTACT WITH A PATIENT OR OPERATOR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17539463 ↗MalfunctionAug 14, 2023Mechanical Problem; Detachment of Device or Device Component; Mechanical Problem; Detachment of Device or Device Component
THYROID UPTAKE SYSTEM S/N (B)(6) WAS SHIPPED ON APRIL 2016. THE TECHNICIAN WAS PERFORMING ROUTINE CALIBRATION WHEN THE SPRING ARM FELL AND HIT THE TECHNICIAN'S LEG, CAUSING A BRUISE. NO MEDICAL TREATMENT WAS REQUIRED. DISCUSSION WITH USER INDICATES THAT THE TECHNICIAN ROUTINELY MOVED THE ARM UP AND DOWN WITHOUT RELEASING THE LOCKING MECHANISM. A REPLACEMENT ASSEMBLY WAS SHIPPED AND USER INSTRUCTED TO RETURN THE DEFECTIVE PART TO MANUFACTURER FOR ENGINEERING EVALUATION AND IDENTIFICATION OF ROOT CAUSE. THE COLLIMATOR AND ARM WEIGH AND SPRING ARM ASSEMBLY WEIGH 45 POUNDS, WHICH HAS THE POTENTIAL FOR SERIOUS INJURY TO EITHER PATIENT OR OPERATOR.
Patient problems: Skin Discoloration
MAUDE report 17539457 ↗
Other Probe, Uptake, Nuclear devices
- Captus 4000e Thyroid Uptake System with Well Captus 4000e Thyroid System with WellMirion Technologies (capintec), Inc. · Probe, Uptake, Nuclear3 recalls4 reports
- Captus 4000e Thyroid Uptake System with 2 in well 5430-30154Mirion Technologies (capintec), Inc. · Probe, Uptake, Nuclear2 recalls0 reports
- neoprobe NPCU3Devicor Medical Products, Inc. · Probe, Uptake, Nuclear1 recall0 reports
- TruNode Surgical Gamma Probe 120-807605Hologic, Inc. · Probe, uptake, nuclear11 reports
- GAMMA FINDER® IIIW.O.M. World of Medicine GmbH · Probe, Uptake, Nuclear1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.