INVOS recallZ-1731-2024 · Covidien · initiated Jan 9, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Covidien or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
manufacturing defects to sensor cable connectors may result in error codes or non-functioning channel
FDA root cause category: Process design
What the company is doing
On January 9, 2024 Medtronic (parent company of Covidien) issued a Urgent Medical Device Recall notification via UPS. Medtronic asked consignees to take the following action: 1. Quarantine and stop using all product from the affected lots of INVOS" Reusable Sensor Cables PMAC71RSC and PMAC71RIC. " See attachment A for guidance on how to identify affected devices via cable box packaging or cable hang tag. 2. Complete and return Customer Confirmation Form. Complete and return the enclosed Customer Confirmation Form even if you do not have any affected lots remaining. 3. Reach out to Medtronic to obtain RGA # /Returned Goods Authorization number. " Your Medtronic Representative / Medtronic Customer Service at [email protected] can assist in initiating product return as necessary. 4. Return all product from the affected lots in your possession to Medtronic as described on the Customer Confirmation Form. 5. Pass on this notice to all those who need to be aware within your organization or to any organization where the potentially affected product from the specified lots has been transferred or distributed.
Recall details
- Recall number
- Z-1731-2024
- Status
- Open, Classified
- Initiated by the company
- Jan 9, 2024
- Posted by the FDA
- Apr 30, 2024
- Recalling firm
- Covidien
- Quantity
- 1122 units
- Product code
- MUD
- FDA submissions
- K211561
- Distribution
- US: AK AZ CA CT FL GA IA IN LA MA MD MI MN MO NC NY OH OK PA SC TN TX VA WA WI OUS: Australia Austria Belgium Canary Islands Chile Czech Republic Denmark Finland France Germany Guatemala Indonesia Ireland Italy Japan Korea, Republic Of Kuwait Lithuania Mayotte New Zealand Nicaragua Poland Portugal Romania Slovakia Spain Sweden Switzerland Taiwan United Kingdom
Product description
Medtronic Invos, Reusable Infant Sensor Adapter Cable for PM7100, REF: PMAC71RIC
Code information (models, lots, serial numbers)
Lot # 20221220 and 20230721; UDI: 10884521806436
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- INVOS PMAC71RICCovidien LP · Oximeter, tissue saturation1 recall4 reports
Product lines
Other devices under the same FDA submission
9 devices cleared under K211561 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- INVOS PM7100Covidien LP · Oximeter, tissue saturation21 reports
- INVOS PM7100Covidien LP · Oximeter, tissue saturation18 reports
- INVOS PMPAMP71Covidien LP · Oximeter, tissue saturation11 reports
- INVOS PM7100Covidien LP · Oximeter, tissue saturation9 reports
- INVOS PMPAMP71Covidien LP · Oximeter, tissue saturation7 reports
- INVOS ISCovidien LP · Oximeter, tissue saturation1 death5 reports
- INVOS PMPAMP71Covidien LP · Oximeter, tissue saturation5 reports
- INVOS ISCovidien LP · Oximeter, tissue saturation3 reports
- INVOS PMSENS71-PCovidien LP · Oximeter, tissue saturation1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 93883).
- Z-1732-2024 · Open, ClassifiedMedtronic Invos, Reusable Sensor Cable for PM7100, REF: PMAC71RSC
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page