VITROS recallZ-1230-2024 · Ortho-Clinical Diagnostics, Inc. · initiated Jan 26, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ortho-Clinical Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard rules, if selected, are not flagged or reported with the applicable QC results if the statistical criteria is met. This may result in acceptance of QC results that do not meet acceptance criteria, which may cause erroneous patient results to be reported.
FDA root cause category: Software Design Change
What the company is doing
On January 26, 2024, the firm issued a customer letter to all customers who have V3.8.0 or V3.8.1 installed on their VITROS Analyzer. Customers were informed of the issue and instructed to follow the instructions when using the VITROS System onboard QC package to use the configurable Westgard Rules. The software will be updated to mitigate this issue within the VITROS Software QC feature. The estimated availability is 2Q2024. If you have further questions, please contact Global Services Organization at 1-800-421-3311.
Recall details
- Recall number
- Z-1230-2024
- Status
- Open, Classified
- Initiated by the company
- Jan 26, 2024
- Posted by the FDA
- Mar 1, 2024
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Quantity
- 15 US; 43 OUS
- Product code
- JJE
- FDA submissions
- K031924
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
Product description
VITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished)
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
VITROS System Software Versions 3.8.0 or 3.8.1 Regular UDI-DI 10758750012343 Refurbished UD-DI 10758750033201 Serial Numbers: US: J46000208 J46000209 J46000240 J46000497 J46000864 J46000865 J46000898 J46001030 J46001064 J46001125 J46001132 J46001293 J46001462 J46001703 J46001704 OUS: J46000150 J46000933 J46000162 J46001029 J46000190 J46001041 J46000247 J46001042 J46000296 J46001087 J46000393 J46001119 J46000410 J46001186 J46000478 J46001187 J46000499 J46001210 J46000531 J46001217 J46000564 J46001289 J46000617 J46001541 J46000772 J46001565 J46000788 J46001571 J46000800 J46001609 J46000801 J46001629 J46000812 J46001639 J46000821 J46001645 J46000855 J46001680 J46000859 J46000948 J46000861 J46000983 J46000910
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- VITROS 6802445Ortho-Clinical Diagnostics, Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE9 recalls18 reports
- VITROS 6900440Ortho-Clinical Diagnostics, Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE5 recalls0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 93927).
- Z-1229-2024 · Open, ClassifiedVITROS 3600 Immunodiagnostic System Catalog Nos. 6802783 (regular) and 6802914 (refurbished)
- Z-1231-2024 · Open, ClassifiedVITROS 5600 Integrated System Catalog Nos. 6802413 (regular) and 6802915 (refurbished)
- Z-1232-2024 · Open, ClassifiedVITROS XT 3400 Chemistry System, Catalog No. 6844458
- Z-1233-2024 · Open, ClassifiedVITROS XT 7600 Integrated System, Catalog No. 6844461
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page