VITROS recallZ-1232-2024 · Ortho-Clinical Diagnostics, Inc. · initiated Jan 26, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ortho-Clinical Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard rules, if selected, are not flagged or reported with the applicable QC results if the statistical criteria is met. This may result in acceptance of QC results that do not meet acceptance criteria, which may cause erroneous patient results to be reported.

FDA root cause category: Software Design Change

What the company is doing

On January 26, 2024, the firm issued a customer letter to all customers who have V3.8.0 or V3.8.1 installed on their VITROS Analyzer. Customers were informed of the issue and instructed to follow the instructions when using the VITROS System onboard QC package to use the configurable Westgard Rules. The software will be updated to mitigate this issue within the VITROS Software QC feature. The estimated availability is 2Q2024. If you have further questions, please contact Global Services Organization at 1-800-421-3311.

Recall details

Recall number
Z-1232-2024
Status
Open, Classified
Initiated by the company
Jan 26, 2024
Posted by the FDA
Mar 1, 2024
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Quantity
25 US; 26 OUS
Product code
JJE
FDA submissions
K182063
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.

Product description

VITROS XT 3400 Chemistry System, Catalog No. 6844458

Code information (models, lots, serial numbers)

VITROS System Software Versions 3.8.0 or 3.8.1 UDI-DI 10758750031986 Serial Numbers: US: J34500140 J34500180 J34500257 J34500289 J34500319 J34500386 J34500422 J34500447 J34500473 J34500474 J34500510 J34500539 J34500547 J34500579 J34500588 J34500622 J34500656 J34500777 J34500818 J34500861 J34500887 J34500893 J34500896 J34500911 J34500931 OUS: J34500123 J34500128 J34500129 J34500130 J34500190 J34500261 J34500269 J34500271 J34500292 J34500301 J34500337 J34500356 J34500363 J34500515 J34500520 J34500536 J34500582 J34500609 J34500628 J34500658 J34500711 J34500793 J34500802 J34500940 J34500960 J34500962

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 93927).