VITROS recallZ-1229-2024 · Ortho-Clinical Diagnostics, Inc. · initiated Jan 26, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ortho-Clinical Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard rules, if selected, are not flagged or reported with the applicable QC results if the statistical criteria is met. This may result in acceptance of QC results that do not meet acceptance criteria, which may cause erroneous patient results to be reported.
FDA root cause category: Software Design Change
What the company is doing
On January 26, 2024, the firm issued a customer letter to all customers who have V3.8.0 or V3.8.1 installed on their VITROS Analyzer. Customers were informed of the issue and instructed to follow the instructions when using the VITROS System onboard QC package to use the configurable Westgard Rules. The software will be updated to mitigate this issue within the VITROS Software QC feature. The estimated availability is 2Q2024. If you have further questions, please contact Global Services Organization at 1-800-421-3311.
Recall details
- Recall number
- Z-1229-2024
- Status
- Open, Classified
- Initiated by the company
- Jan 26, 2024
- Posted by the FDA
- Mar 1, 2024
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Quantity
- 9 US; 66 OUS
- Product code
- JJE
- FDA submissions
- K083173
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
Product description
VITROS 3600 Immunodiagnostic System Catalog Nos. 6802783 (regular) and 6802914 (refurbished)
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
VITROS System Software Versions 3.8.0 or 3.8.1 Regular UDI-DI 10758750002979 Refurbished UD-DI 10758750007103 Serial Numbers: US: J36000109 J36000292 J36000489 J36000955 J36001167 J36001222 J36001283 J36001330 J36001348 OUS: J36000148 J36001000 J36001495 J36000340 J36001012 J36001513 J36000348 J36001024 J36001515 J36000406 J36001026 J36001519 J36000407 J36001027 J36001528 J36000427 J36001051 J36001530 J36000446 J36001088 J36001535 J36000511 J36001108 J36001558 J36000517 J36001111 J36001570 J36000580 J36001112 J36001573 J36000612 J36001117 J36001575 J36000626 J36001154 J36001579 J36000688 J36001229 J36001581 J36000753 J36001237 J36001582 J36000861 J36001245 J36001583 J36000877 J36001246 J36001595 J36000920 J36001401 J36001600 J36000923 J36001403 J36001602 J36000929 J36001427 J36001608 J36000933 J36001451 J36001618 J36000941 J36001488 J36001632 J36000942 J36001489 J36001640
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- VITROS 6802783Ortho-Clinical Diagnostics, Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE8 recalls8 reports
- VITROS 6802914Ortho-Clinical Diagnostics, Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE6 recalls0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 93927).
- Z-1230-2024 · Open, ClassifiedVITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished)
- Z-1231-2024 · Open, ClassifiedVITROS 5600 Integrated System Catalog Nos. 6802413 (regular) and 6802915 (refurbished)
- Z-1232-2024 · Open, ClassifiedVITROS XT 3400 Chemistry System, Catalog No. 6844458
- Z-1233-2024 · Open, ClassifiedVITROS XT 7600 Integrated System, Catalog No. 6844461
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page