Equinoxe recallZ-1416-2024 · Exactech, Inc. · initiated Mar 6, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Exactech, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).
FDA root cause category: Nonconforming Material/Component
What the company is doing
Beginning 03/06/2024, Exactech issued an URGENT MEDICAL DEVICE RECALL notice and an URGENT DEAR HEALTHCARE PROFESSIONAL COMMUNICATION to its consignees via email. The recall notice explained the issue, clinical impact, and requested the following actions be taken: "" CAREFULLY REVIEW THIS RECALL NOTIFICATION to ensure that you fully understand the issue identified, the recall strategy, and all actions required. " IMMEDIATELY IDENTIFY AND QUARANTINE any of the subject devices in your inventory and/or customer s inventory listed Product Scope Listing. " EXTEND THE DESCRIPTION OF ISSUE AND CLINICAL IMPACT as described in the recall notification to your accounts that may have this product in their possession. " COMPLETE AND RETURN the attached Recall Inventory Response Form to Exactech via email at [email protected]. The communication to healthcare providers referenced the FDA Safety Communication issued on 01/16/2024, explained the issue, potential health risks, and suggested patient monitoring for those with implanted affected devices for potential device wear, failure, or bone loss and should consider performing X-rays to further evaluate the patient and their implanted device if you suspect a failed device. Contact your local Exactech Representative if you have any questions regarding this communication.
Recall details
- Recall number
- Z-1416-2024
- Status
- Open, Classified
- Initiated by the company
- Mar 6, 2024
- Posted by the FDA
- Mar 29, 2024
- Recalling firm
- Exactech, Inc.
- Quantity
- 11863 units
- Product code
- PHX
- FDA submissions
- K093275
- Distribution
- Nationwide and Australia, Austria, Belgium, Brazil, Chile, Canada , China, Colombia, France, Germany, Great Britain, Greece, Guatemala, India, Italy, Japan, Korea, Luxembourg, Malaysia, Netherlands, New Zealand, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand
Product description
Exactech Equinoxe REVERSE SHOULDER,36mm Humeral Liner: a) +0mm, Item Number 320-36-00, b) +2.5mm, Item Number 320-36-03; Shoulder Arthroplasty
Code information (models, lots, serial numbers)
Show the full code information (107k characters)
a) Item Number 320-36-00, UDI/DI 10885862086617, Serial Numbers: 1954421, 1954426, 1954423, 1954422, 1954424, 2084846, 2102178, 2095168, 2085991, 2142641, 2200474, 2200475, 2142642, 2223309, 2276145, 2399592, 2418172, 2418175, 2418174, 2399591, 2223310, 2223311, 2276143, 2276140, 1954425, 2276142, 2418173, 2276144, 2399593, 2399590, 2662625, 2528861, 2528862, 2528864, 3637099, 3673852, 3637100, 3637097, 2662624, 2788819, 2662626, 2788821, 3573735, 3573734, 3616397, 2565880, 2565882, 3616398, 3616399, 3616396, 3637096, 3694392, 2662627, 2823683, 3740075, 3754460, 3758537, 2200473, 2399594, 2276141, 3694390, 3740074, 3694391, 2565878, 3740073, 2823681, 2823684, 3918249, 3918248, 3989825, 3989840, 3827079, 3870844, 3918250, 3989838, 3870840, 3918260, 3870838, 3989830, 3989841, 3918261, 3827096, 3870834, 3870843, 3827084, 4028552, 3827081, 3827091, 3827095, 3827076, 3827094, 3827082, 3827078, 3827087, 3827088, 3827089, 3827097, 3827083, 3827090, 3827093, 3827080, 3918243, 3918247, 3918251, 3918254, 3918255, 3918242, 3918245, 3918252, 3918253, 3918258, 3918244, 3918246, 3918256, 3918257, 3918262, 3918263, 3989821, 3989822, 3989828, 3989833, 3989834, 3989842, 3989843, 3989846, 3870835, 3989835, 3989844, 3989824, 4028569, 4028574, 4028564, 4028565, 4028568, 4028551, 4028554, 4028557, 4028559, 4028560, 4028561, 4028562, 4028566, 4028567, 4028570, 4028571, 4028572, 4028573, 4028575, 4028576, 4028577, 4064490, 4064491, 4064492, 4064495, 3827077, 3827085, 3827086, 3827092, 3870837, 3870839, 3870841, 3870842, 4064499, 4064500, 4064511, 4064512, 4064489, 4064501, 4064503, 4064504, 4064510, 4064498, 4064502, 4064509, 4064513, 2528865, 2528863, 2565879, 3637098, 2823682, 2565881, 4371060, 4371047, 4316275, 4371061, 4469128, 4469126, 4469123, 4264327, 4371054, 4316262, 4264319, 3989832, 4064506, 4028558, 3989823, 3989837, 4254134, 4254132, 4371062, 4371042, 4316270, 4469109, 4469122, 4316281, 4469124, 4254125, 4469107, 4028555, 3989827, 4113866, 4316277, 4264314, 4469106, 4316265, 4469103, 4469125, 4028553, 4316272, 4028563, 4316285, 4254130, 3870836, 3989831, 4113862, 3989839, 4254129, 4316266, 4371051, 4316264, 4371053, 4371038, 4371055, 4316274, 4316280, 4316269, 4469116, 4469121, 4254131, 4264313, 4371046, 4371056, 4469118, 4113846, 4113850, 4254126, 4316278, 3918259, 4064493, 4064494, 4064497, 4064505, 4064507, 4064508, 4113852, 4113853, 4113859, 4064496, 4254136, 4264301, 4264302, 4264306, 4264308, 4264311, 4264312, 4264316, 4264317, 4264323, 4264330, 4264303, 4264304, 4264305, 4264307, 4264309, 4264310, 4264315, 4264318, 4264321, 4264322, 4264324, 4264325, 4264326, 4264328, 4264329, 4316250, 4316251, 4316252, 4316253, 4316254, 4316255, 4316257, 4316282, 4316283, 4316284, 4316260, 4316261, 4316271, 4316279, 4371044, 4371048, 4371050, 4371052, 4371057, 4371040, 4371041, 4371043, 4371058, 4371059, 4371064, 4469104, 4469111, 4469119, 4113851, 4113854, 4113857, 4113861, 4113863, 4113864, 4113865, 4113867, 4113860, 4113868, 4113847, 4113848, 4113849, 4113855, 4113856, 4113858, 4113869, 4113870, 4254122, 4254123, 4254128, 4254135, 4316256, 4316258, 4316259, 4316263, 4316267, 4316268, 4316273, 4371045, 4469113, 4469129, 4469110, 4469115, 4696477, 4696490, 4696468, 4696483, 4696471, 4696489, 4696492, 4696476, 4696478, 4696480, 4696481, 4696482, 4696484, 4696486, 4696491, 4696493, 3989829, 4264320, 4316276, 4469117, 4254124, 4028556, 4113845, 3989820, 4371063, 4371039, 4254133, 4938336, 4810176, 4959320, 4938335, 4993535, 5094629, 5134757, 5150137, 4696474, 4696467, 4771672, 4716450, 4810172, 4810151, 4837763, 4903692, 4897011, 4810166, 4696473, 4959324, 5013770, 4756201, 4707127, 4734395, 4756195, 4837776, 4771695, 5057125, 4837779, 5013786, 4959333, 4771677, 4881879, 5040533, 4959328, 4810149, 4993539, 4881894, 4881893, 4993560, 4903695, 4707125, 4707126, 4707128, 4810152, 4955976, 4756198, 4707131, 4716448, 4734403, 4734399, 4734406, 4837774, 4810147, 4993554, 4734404, 4810157, 4959318, 4756194, 4771679, 4756192, 4734394, 4707132, 4771685, 4938331, 4896998, 4707112, 4716445, 4707134, 4707113, 4756202, 4771683, 4707123, 4810161, 4707129, 4707108, 4837759, 4707110, 4810153, 4837772, 4837764, 4881888, 4938344, 4881890, 4956000, 4938329, 4896993, 4959309, 4938333, 5040540, 4993548, 5092587, 5040529, 4734417, 5092591, 5092584, 5092583, 5040535, 5065836, 4756212, 5134759, 5040538, 4771680, 4771686, 4756197, 4881877, 5013792, 4993543, 4707114, 4756190, 4771688, 4771684, 4938330, 4938353, 4810170, 5150136, 5057103, 4837773, 4696469, 4938338, 5013784, 4993533, 5040532, 4771676, 4993549, 4696470, 4734402, 4707117, 4756209, 4771674, 4771675, 4756185, 4959311, 4938343, 5040550, 4993536, 4959304, 4810167, 5092580, 4696479, 4734407, 4771687, 4771692, 4771690, 4896989, 4903711, 4903694, 4993542, 4837754, 5057102, 4881882, 5134761, 5057101, 4837767, 4881887, 4955983, 4955986, 4955992, 4716438, 4716446, 4716453, 4734409, 4734393, 4734397, 4734398, 4734405, 4734410, 4734411, 4734412, 4734413, 4771673, 4771681, 4771682, 4771693, 4771694, 4756186, 4756187, 4756203, 4756213, 4756214, 4810150, 4810155, 4810160, 4810164, 4810165, 4810171, 4810173, 4810174, 4810175, 4810179, 4837753, 4837762, 4837765, 4837766, 4837771, 4837778, 4881878, 4881884, 4897009, 4903708, 4903715, 4903721, 4959305, 4959306, 4959307, 4959308, 4959312, 4810162, 4810163, 4903717, 4959303, 4959310, 4959314, 4959317, 4959323, 4959325, 4959327, 4959332, 4993537, 4993541, 4993551, 4993552, 4993557, 4993561, 4993562, 4993553, 5013768, 5013771, 5013774, 5013777, 5013778, 5013782, 5013788, 5013791, 5040530, 5040534, 5040536, 5040537, 5040539, 5040546, 4707111, 4707121, 4696472, 4696487, 4696488, 4707116, 4707130, 4707133, 4707119, 4716449, 4707109, 4707122, 4716454, 4716459, 4716436, 4716437, 4716440, 4716442, 4716457, 4716460, 4716461, 4716439, 4716451, 4716452, 4716458, 4734400, 4734416, 4734418, 4756207, 4756208, 4716441, 4716444, 4716447, 4734396, 4734408, 4734414, 4734415, 4734419, 4756188, 4756196, 4756205, 4734401, 4756211, 4756189, 4756200, 4756204, 4756210, 4810158, 4810168, 4810169, 4881869, 4896983, 4896987, 4896991, 4897002, 4897003, 4897004, 4896997, 4897000, 4897005, 4897006, 4897007, 4881876, 4881886, 4896986, 4896995, 4897012, 4903690, 4903698, 4903701, 4903704, 4903716, 4903719, 4881889, 4903710, 4903712, 4938342, 4938355, 4771691, 4881875, 4903691, 4903693, 4938332, 4938354, 4955972, 4955974, 4955978, 4955980, 4955984, 4955985, 4955987, 4955988, 4955991, 4955993, 4955995, 4955998, 4955999, 4956001, 4881868, 4938328, 4938339, 4938340, 4938345, 4938347, 4938349, 4938350, 4955973, 4955977, 4955979, 4955981, 4955982, 4955989, 4955990, 4955994, 4955996, 4955997, 5013781, 5040547, 4938334, 4938341, 4938346, 4938351, 4938352, 4959326, 5013773, 5013776, 5040548, 5057100, 5057113, 5057123, 5057106, 5065837, 5065838, 5065840, 5065841, 5094615, 5094621, 5094623, 5094625, 5094631, 5057122, 5092597, 5040549, 4810159, 4881891, 4837769, 4903713, 4837758, 4897001, 4837777, 4896988, 5013764, 4993555, 5040541, 4707124, 4881881, 4881872, 5013775, 5013793, 5013767, 4810154, 4696485, 5057109, 5040545, 4881871, 4993538, 4881874, 4896999, 4896992, 4897010, 4903714, 4756199, 4903697, 4903699, 4903722, 3989836, 3989845, 3989826, 5188177, 5174850, 4837760, 5134764, 5199802, 5199797, 5174843, 5134760, 5199791, 5057104, 5199803, 5260626, 5150138, 4959331, 5013769, 5134755, 5057105, 4959319, 4881880, 5188182, 5188183, 5188172, 5260623, 5092600, 4959329, 5092593, 5174847, 5092594, 5092589, 5057108, 4771678, 5092602, 5174840, 5057114, 5174861, 5065835, 4771689, 5092586, 4881870, 5260619, 4896996, 5057120, 5057116, 5134756, 5188166, 5188176, 5057121, 5260606, 5260605, 5150130, 4993544, 5188180, 4993550, 4707115, 5199799, 5188164, 4993545, 5150110, 5188157, 5065834, 4938337, 5065839, 4881885, 4903709, 4756191, 5013790, 5092582, 4810156, 5040531, 4810177, 4959315, 5094630, 4896985, 4993547, 5150150, 5188163, 5188156, 5057127, 5150141, 5134754, 5092596, 5174837, 5150135, 5199796, 5260631, 5199785, 4897008, 4837755, 5174860, 4810178, 5150149, 4881883, 5150151, 5057111, 5199798, 5199805, 5013772, 4955975, 5013766, 4837775, 4837756, 5150157, 5150144, 5013789, 5092590, 5174849, 5092581, 5150153, 5134758, 5134765, 5094618, 5057110, 4881873, 5057118, 5199789, 5150102, 5150129, 5150146, 5134762, 5134763, 5150133, 5150155, 5155376, 5155378, 5174858, 5150124, 5188154, 5188159, 5188170, 5188179, 5188181, 5174853, 5260608, 5260612, 5260614, 5260618, 5260620, 5260622, 4993534, 4993546, 5040542, 5092595, 5188171, 5260613, 5150104, 5150108, 5150132, 5150145, 5150148, 5150121, 5150143, 5150147, 5174862, 5092578, 5092579, 5092585, 5150101, 5150105, 5150106, 5150109, 5150111, 5150134, 5155377, 5155380, 5174863, 5188162, 5199786, 5199790, 5199807, 5174842, 5174857, 5174866, 5188160, 5188165, 5188169, 5188174, 4993558, 5057117, 5150103, 5150125, 5150140, 5150142, 5174841, 5174846, 5188161, 5188173, 5188178, 5199792, 5199793, 5199806, 5057112, 5150099, 5188175, 5199804, 5260628, 4903703, 4959330, 5150126, 5174844, 5174845, 5174848, 5174865, 5188158, 5188168, 5260611, 5260616, 5260617, 5260624, 5260629, 5260632, 5260633, 5174851, 5092599, 4959321, 5150107, 5094614, 5174854, 5155379, 5092598, 4993540, 5094619, 5150152, 5092607, 5013783, 5092601, 5260604, 5150156, 5094611, 5174855, 5260607, 5094624, 5260615, 5174838, 5150098, 5174864, 5094616, 5150123, 5199795, 5150127, 5092604, 5150119, 4837757, 5040544, 5150122, 5260630, 5013779, 5094617, 5094628, 5094620, 5199801, 5092606, 5199800, 5174859, 5094626, 5094633, 5199788, 5260610, 5150112, 5094612, 5094613, 5188167, 4837770, 5174852, 5092603, 5150115, 5199794, 4993559, 5150139, 5092605, 5057119, 5174856, 5260609, 5260625, 5150118, 5260627, 5150114, 5094627, 5199787, 5094632, 5057124, 4903706, 5150117, 4881892, 4896994, 5150116, 5013785, 5150131, 4903702, 5094622, 5150113, 4938348, 4903718, 5150100, 4959316, 4959322, 5057129, 4716456, 4903720, 4938327, 5174839, 4938326, 4993556, 4903707, 5092592, 5150154, 4707120, 5057107, 5260621, 5188155, 5013780, 5013787, 5013765, 5150120, 5057115, 6027654, 5933476, 5869792, 5933436, 6169875, 5933418, 5869476, 6089982, 5911032, 5975002, 6027792, 6034421, 6190387, 5869842, 6100061, 6264910, 5911028, 6044205, 5870032, 5909991, 5910007, 5911112, 5933407, 5869834, 5869823, 5975006, 5869498, 6168355, 6064668, 6169994, 6027642, 6100038, 6170006, 6169740, 6089979, 6264928, 6169912, 6110379, 6246150, 6044196, 6207421, 6027663, 6027673, 5933483, 6034427, 5975113, 5974942, 6034430, 6168314, 6264929, 6207452, 6090002, 6095953, 5870040, 5933458, 6027680, 6090235, 5974970, 5869680, 5910070, 5974853, 6027690, 6064823, 6265003, 6190449, 6027661, 6168284, 6100057, 5869813, 6190405, 6027623, 5975223, 6169885, 6264915, 5974862, 6287259, 6190458, 6064649, 5933416, 5974984, 5974997, 6207450, 6207477, 6027681, 5909988, 5975221, 5933414, 5974962, 6264907, 6064821, 5869849, 5909996, 5909983, 5910084, 5909981, 5932750, 5933467, 5933421, 5975225, 5932680, 6169910, 5974842, 5869021, 5869844, 5911095, 5974944, 5974999, 6027624, 6027667, 6027814, 6105326, 6089977, 5911109, 6064835, 6064829, 6064837, 6100043, 6190396, 5974859, 6027811, 6168332, 5933484, 5974863, 5933465, 6064716, 5933442, 5975111, 5933444, 5933448, 5933457, 6095955, 5974993, 5974935, 6090243, 6064656, 5975226, 5975220, 6110372, 5911051, 6095950, 6095961, 6064684, 6044448, 6190432, 6176735, 5869821, 6027703, 6190402, 6246149, 6027638, 6027692, 6027693, 6027665, 5933487, 6207469, 6027672, 6168331, 6044203, 5933468, 5869822, 5974972, 5974854, 5974975, 5869833, 5974861, 5974967, 6168318, 5974846, 6027652, 5975008, 5975238, 5869807, 5869796, 5910001, 5975014, 5911084, 6027808, 5932658, 5974996, 5932687, 5932681, 5910000, 5975219, 6027637, 6027617, 5974947, 6027670, 6044222, 6190394, 6176719, 6109047, 6265000, 6168293, 6264921, 6064654, 6090003, 6090005, 6064840, 6297104, 6169745, 5974994, 5974974, 5974865, 5932670, 5869486, 5932685, 5974981, 5911087, 6034412, 5933461, 6064665, 5869814, 5974858, 5975011, 6027658, 5869811, 5869817, 6264922, 5933427, 5869028, 6027659.5974980, 5870236, 5975242, 6044455, 6176708, 6027794, 5869801, 6169987, 6169761, 6168347, 6287180, 6110378, 6027813, 6168272, 5909992, 6027803, 6027619, 6169905, 5910003, 6027632, 6034413, 6090004, 5975234, 6246158, 6170016, 6176727, 6170019, 6190376, 6176723, 6170012, 5933417, 6190427, 5975114, 6190385, 5910069, 6090226, 5869828, 6064722, 6264905, 6090236, 5974995, 6027804, 6064685, 5869832, 5911027, 6100044, 6105318, 6109037, 6168345, 6207449, 5975217, 5933481, 5869666, 5870038, 5869691, 5909994, 5911111, 5911088, 5911100, 5911108, 5869481, 6027700, 6095966, 6110374, 5933471, 6095967, 5975235, 5933411, 5933469, 6027634, 6169981, 6168317, 6064842, 6207343, 6207338, 6027676, 6168290, 6246145, 6044439, 6064713, 6190380, 5932679, 6259703, 5933486, 6064655, 5870036, 5933474, 5911035, 6034422, 5933455, 5975237, 6027657, 6044218, 6064834, 6090246, 6176717, 6169894, 5869827, 5869826, 6207334, 6264988, 5911104, 6027806, 6190406, 6105315, 5975118, 5975013, 6168316, 5975125, 5974990, 5909989, 6044201, 5910006, 5975230, 6064846, 5974983, 5975233, 5974867, 6027699, 6027646, 6044438, 6044447, 6027694, 6100034, 6044442, 6207340, 6044443, 6207333, 6207424, 6168311, 6064663, 5869023, 6169759, 5869011, 5932770, 5974940, 5975108, 6090242, 6169769, 5911034, 6100059, 5932766, 6064686, 6169915, 5933482, 5911099, 5910087, 5933409, 6064672, 6027702, 6027688, 6027655, 5869013, 5910079, 6259717, 6169977, 6246146, 5974848, 5974869, 6027678, 5909987, 5974971, 5933412, 5910073, 6095942, 5975019, 5933413, 5911097, 5910005, 5974849, 5975000, 6064700, 5974843, 5975127, 5933428, 6027645, 6044217, 6027630, 6168350, 5932758, 5932686, 6034425, 6064824, 6169870, 6064653, 6100040, 6027797, 5869846, 5911052, 5932739, 6095962, 6207454, 5974931, 6044441, 6090229, 5932762, 5975120, 5910002, 6089980, 5869690, 5870037, 5870035, 5975010, 5975018, 6095944, 5932767, 5974855, 5975003, 6095945, 6027802, 5870056, 6170005, 6170017, 6027653, 5869678, 5910098, 6095968, 6027651, 6027674, 6064727, 6246148, 6190492, 6170010, 6095964, 5869681, 5870052, 5909999, 5910093, 5910096, 5909995, 5910097, 5869841, 5869837, 5910077, 5911090, 5911107, 5911033, 6044209, 5974845, 5932667, 5932666, 5933480, 5974991, 5911094, 5975241, 6027626, 6027622, 6089985, 5911044, 6027631, 6027796, 6089996, 6089992, 6044204, 6044459, 6044212, 6168302, 6064657, 6044436, 6064664, 6044445, 6207451, 6044202, 6207465, 6207453, 6090239, 6207475, 6090250, 6105303, 6105316, 6044220, 6207335, 6264906, 6100039, 6169882, 6264918, 6100049, 6287273, 6064712, 6064711, 5910086, 6027666, 6100036, 6168365, 6168326, 5933432, 6168327, 6246132, 6190398, 5933488, 5909982, 5911029, 6027648, 6168306, 6190455, 6100052, 6089998, 6027696, 6259705, 6027684, 6105309, 6027644, 6027787, 6064701, 6190428, 5933466, 5932663, 6264913, 5932768, 6095943, 6168280, 6169877, 6264983, 5910080, 6264981, 5910095, 5933478, 5932671, 6207471, 5975107, 6168308, 5975012, 6027805, 5975004, 6207467, 6169748, 6064839, 5974948, 6190463, 6027706, 5975236, 6105311, 5975240, 6190413, 6169763, 5869836, 5870057, 6027789, 5975244, 5911092, 5911093, 5932753, 5870034, 5911091, 5909990, 5933475, 6276309, 5974952, 6170011, 5911086, 6064662, 5975134, 5911042, 6090252, 5869668, 5910081, 5869692, 5933459, 5974964, 5974870, 6190404, 6190409, 6089981, 6190377, 6090245, 6064833, 6246147, 6044452, 6169766, 6264987, 5932769, 6110389, 6246121, 6190474, 5974998, 6089987, 6044433, 6190374, 6207344, 5910088, 6027682, 6064690, 5975112, 5975129, 5975117, 6095963, 6027800, 6044208, 6190454, 5932661, 5975116, 5909993, 5975109, 5910090, 5975245, 5869845, 5869017, 5870042, 5870046, 5932777, 5932682, 5870031, 5911031, 5933470, 5909997, 6100041, 6088942, 6169902, 5870058, 5911053, 5932678, 6064666, 6259719, 5870030, 6276315, 6064658, 6190425, 6276289, 6064661, 5869025, 5869027, 5869671, 5869672, 5869686, 5869688, 5869795, 5870232, 5910074, 5910083, 5932733, 5932746, 5932747, 6176706, 5932718, 5932721, 5932723, 5932728, 5932729, 5932730, 5932732, 5932734, 5932735, 5869663, 5869664, 5869665, 5869667, 5869669, 5869673, 5869675, 5869685, 5869687, 5869794, 5975005, 5975007, 5975020, 5975131, 6027636, 6027809, 6034424, 6034435, 6044199, 6044206, 6044207, 6044211, 6044215, 6044219, 6064641, 6064644, 6064645, 6064647, 6168307, 6169753, 6169754, 6169881, 6246110, 6246130, 5869001, 5869007, 5869010, 5869012, 5869016, 5869477, 5869480, 5869483, 5869484, 5869487, 5869488, 5869490, 5869670, 5869674, 5869676, 5869684, 5869804, 5869805, 5869806, 5870226, 5870228, 5870044, 5911030, 5911038, 5911039, 5911040, 5911043, 5911048, 5911050, 5911082, 5911085, 5911096, 5911098, 5911101, 5911102, 5911103, 5911105, 5975110, 5975115, 5975123, 5975128, 5975130, 5975132, 5974938, 5974954, 5974955, 5932752, 5932754, 5932756, 5932763, 5932765, 5932771, 5932773, 5932775, 5932660, 5932668, 5932669, 5932673, 5932674, 5932675, 5933420, 5933423, 5933425, 5933435, 5933437, 5933438, 5933439, 5933440, 5933441, 5933443, 5933447, 5933449, 5933450, 5933451, 5933452, 5933453, 5933454, 5933462, 6088924, 6088925, 6088926, 6088927, 6088928, 6088929, 6088930, 6088931, 6088932, 6088933, 6088934, 6088935, 6088937, 6088938, 6088939, 6088940, 6088941, 6088943, 6088945, 6088946, 6088948, 6088949, 6088950, 6088951, 6088952, 6088953, 5932720, 5974992, 5975001, 5975009, 5975124, 5975126, 5975216, 5974933, 5974934, 5974939, 5974941, 5974945, 5974946, 5974949, 5974951, 5974957, 5974958, 5974959, 5974960, 5974976, 5974988, 6034408, 6034409, 6034410, 6034414, 6034415, 6034416, 6034423, 6034426, 6034428, 6034429, 6034431, 6064673, 6064674, 6064678, 6064681, 6064682, 6064683, 6064687, 6064689, 6064691, 6064692, 6064693, 6064694, 6064695, 6064696, 6064819, 6064820, 6064826, 6064827, 6095947, 6095949, 6095952, 6095954, 6105304, 6044210, 6044213, 6044221, 6044225, 6064650, 6064670, 6064702, 6064704, 6064705, 6064706, 6064709, 6064714, 6064715, 6064728, 6090244, 6090248, 6105297, 6105298, 6105299, 6105300, 6109033, 6109034, 6109035, 6109036, 6109038, 6109039, 6109040, 6109041, 6109043, 6109044, 6109045, 6109046, 6110373, 6110375, 6110381, 6110382, 6110383, 6110385, 6110388, 6110390, 6110391, 6168319, 6090230, 6090233, 6090238, 6168288, 6168294, 6169739, 6169741, 6169746, 6169747, 6169750, 6169751, 6169755, 6169756, 6169770, 6169771, 6169772, 6169773, 6169774, 6169775, 6169871, 6169873, 6169876, 6169879, 6169883, 6169886, 6169892, 6169895, 6169896, 6169898, 6169901, 6169907, 6169916, 6169919, 6169986, 6169991, 6169992, 6169993, 6190457, 6190465, 6190467, 6190477, 6190479, 6190493, 6207460, 6207462, 6105322, 6105324, 6105325, 6168281, 6168291, 6168292, 6168304, 6190461, 6190476, 6190478, 6207458, 6246134, 6246137, 6246139, 6246140, 6246157, 6246159, 6259691, 6259697, 6259698, 6259700, 6259702, 6259714, 6259715, 6259718, 6264989, 6264992, 6264998, 6264999, 6265002, 6265004, 6265006, 6265007, 6265008, 6265009, 6319876, 6319877, 6319878, 6319879, 6319880, 6319265, 6319266, 6319267, 6319268, 6319269, 6319270, 6319271, 6319272, 6319273, 6319274, 6319275, 6319276, 6319277, 6319278, 6319279, 5869790, 5869791, 5869793, 5869798, 5869799, 5869800, 5869802, 5869809, 5869812, 5869819, 5869839, 5974982, 6190433, 5869482, 5909985, 5869478, 5932774, 6190412, 5869503, 5911089, 5933408, 6297118, 5870237, 5933419, 6095951, 6027641, 5869500, 5869689, 5974868, 5869679, 5869496, 5870234, 5911055, 6027660, 5933415, 5869830, 5975224, 6027671, 5869682, 5910085, 5869847, 5910072, 5932749, 6044460, 6044454, 5910094, 5974936, 5870055, 6264982, 6064646, 6064667, 6190437, 5869825, 6027695, 6044224, 6044444, 5869831, 6105323, 5932776, 5932772, 5932683, 5933424, 5975232, 6170020, 5933430, 5910075, 5869495, 6064832, 6064836, 5910078, 5869491, 5870227, 5869024, 5869485, 5869492, 5911036, 5932755, 5932751, 5974968, 5974961, 6168298, 6027669, 6168282, 5932677, 6064830, 6190440, 6064642, 5932742, 5869497, 5870059, 5932759, 5870041, 5911045, 6207468, 5911057, 6176714, 5932764, 6190419, 5932676, 6027668, 6027650, 6168300, 5870229, 5869019, 5869840, 5870045, 5932761, 6169906, 5869008, 5932665, 5869504, 5870235, 5869006, 5933426, 5869018, 5910071, 5869499, 5870231, 5911041, 5932757, 6190418, 6190426, 5911083, 5869501, 5869502, 5932684, 6027656, 5870033, 5933410, 5933422, 5869015, 5869505, 5910076, 5933433, 6168287, 5975243, 6297095, 5870053, 5869808, 5911054, 5975222, 6044437, 5870049, 5869848, 5911047, 5911037, 6044462, 6190431, 5933477, 5932664, 6264990, 5932736, 5869005, 5932672, 5869029, 5933472, 6027691, 5869022, 6190373, 6259707, 5975016, 6190443, 6246133, 6259696, 5910091, 5869493, 5869829, 6064703, 6064723, 5869000, 5869020, 6190399, 5910092, 5869004, 6168305, 5975122, 6259701, 4371049, 4254127, 4810148, 4707118, 4756193, 4837761, 4959313, 4696475, 5057128, 5040543, 6109031, 6319396, 6090231, 6176720, 6510544, 6207470, 6207430, 6506154, 6512533, 6100046, 6319428, 6319300, 6411945, 6411967, 6506566, 6510608, 6506140, 6246154, 6034436, 6089988, 6506537, 6510560, 6297066, 6506193, 6512513, 6411954, 6543363, 6064652, 6287271, 5869824, 6352074, 6489440, 6342215, 6342216, 6523634, 6510647, 6259730, 6319290, 6264997, 6170018, 6523623, 6512549, 6207289, 6543319, 6110377, 6169760, 6190370, 6190475, 6168324, 6297107, 6100045, 6523421, 6488895, 6523417, 6489407, 6523515, 6510587, 6207438, 6510595, 6510620, 6523495, 6506163, 6523670, 6264901, 6190386, 6259736, 6394525, 6264993, 6264900, 6297061, 6264919, 6207357, 6360339, 6352038, 6352022, 6523338, 6543348, 6489425, 6297063, 6352052, 6264914, 6044440, 6506161, 6190422, 6169878, 6034420, 6169985, 6259737, 6523415, 6319352, 6190393, 6190392, 5932719, 6489426, 6394627, 6169762, 6190462, 6168283, 6064699, 6319351, 6264903, 6190488, 6169880, 6169890, 6297098, 6064717, 6319340, 6506137, 6207323, 6170013, 6297054, 6523339, 6510607, 6319286, 6342253, 6089984, 6246102, 6169983, 6246115, 6297114, 6394635, 6319303, 6027801, 6352090, 6506559, 6246111, 6523464, 6510552, 6512534, 6523662, 6512523, 6506555, 6523674, 6523490, 6506169, 6168328, 6168329, 5869818, 6168353, 6190416, 6190414, 6319349, 6394523, 6512528, 6523465, 6523493, 6523431, 6523416, 6264917, 6090001, 6176729, 6488937, 6064838, 6170004, 6110392, 6100058, 6287174, 5975106, 6488952, 6319344, 6411944, 6489395, 6506153, 6489397, 6489439, 6543350, 6512510, 6489411, 5974844, 6297090, 6319366, 6352078, 6110384, 6297100, 6110380, 6297077, 6345417, 6319365, 6342214, 6352101, 5909984, 6207331, 6352089, 6297071, 6064659, 6523642, 6297080, 5909986, 6319325, 6488943, 6044435, 5975121, 6064721, 6264884, 6352039, 6543370, 6489441, 6100042, 6315772, 6352044, 6297058, 6352018, 6352042, 6510553, 6510577, 6342202, 6360338, 6319413, 6169908, 6510601, 6523678, 6089989, 6297105, 6064643, 6287255, 6342236, 6394640, 5974956, 6523468, 6512490, 6512517, 6523484, 6510581, 6543393, 6394536, 6523435, 6109030, 6342220, 6169996, 6190400, 6169888, 6089983, 6246122, 6543377, 6027697, 6319399, 6523630, 6352077, 6064698, 6523491, 6394572, 6411951, 6207325, 6287173, 6168360, 6190379, 6064719, 6064679, 6044456, 6342250, 6523412, 6287265, 6489424, 6510562, 6319388, 6543358, 6523500, 6319379, 5933456, 6510618, 6510603, 6319355, 6506138, 6543374, 6352045, 6319425, 6168301, 6319367, 6287257, 6489413, 5869014, 6027799, 6315776, 6027675, 6506170, 6489423, 5974965, 6510645, 6506533, 6207298, 6506187, 6543317, 6100037, 6506172, 6489416, 6510564, 6510578, 6207346, 5869677, 6510646, 5870051, 5911056, 6170007, 6512505, 6512491, 6523502, 6207472, 6506570, 6506518, 6287258, 6169988, 6506527, 6506136, 6510572, 6170024, 6506141, 6169903, 6394639, 6027705, 6523428, 6176726, 6105312, 6489400, 6523679, 6064677, 6259690, 6089995, 6089991, 6168334, 6264898, 6207337, 6297128, 6315771, 6207354, 6345408, 6319389, 6506544, 6488912, 5933473, 6246116, 6523433, 6319356, 6319381, 6089990, 6319338, 6064844, 6207435, 6319375, 6345405, 6394518, 6512516, 6297059, 6169897, 6169764, 6190460, 6190473, 6259733, 6297117, 6319411, 6168320, 6506551, 6489404, 6319333, 6510584, 6510546, 6190395, 6342212, 6259723, 6027807, 6027620, 6342238, 6168312, 6207295, 6246127, 6176730, 6360330, 6352036, 6352072, 6352017, 6506524, 6523469, 6352067, 6207358, 6190487, 6523626, 6523475, 6523503, 6510592, 6506158, 6319402, 6523489, 6510574, 6190383, 6523643, 6510616, 6169742, 6110376, 6190459, 5870050, 6259699, 6315770, 6297119, 6190480, 6342232, 6342240, 6352086, 6352102, 6489438, 6506554, 6352066, 6510633, 6510657, 6506548, 6523641, 6523644, 6297103, 6523455, 6352085, 6105320, 6506567, 6190435, 6027816, 6259709, 6319408, 6169984, 6342221, 6488951, 5869810, 6319323, 6489422, 6360323, 6489443, 6319410, 6315774, 5869479, 6027701, 6394647, 6259713, 6489449, 6489409, 6512519, 6168286, 6523472, 5974969, 6109048, 6105302, 6100050, 6207350, 6319348, 6190382, 6297101, 6342245, 6319345, 6352093, 6207322, 6352024, 6506195, 6489419, 6352075, 6489401, 5975239, 6169918, 6246125, 6319423, 6287179, 6095959, 6246155, 6246107, 6064651, 6523636, 6319302, 6169893, 6168367, 6190372, 6345403, 6034437, 6488961, 6034419, 6523653, 6543326, 6027647, 6489412, 6360325, 6207336, 6360336, 6264881, 6034433, 6319394, 6352032, 6319320, 6034432, 6510540, 6523335, 6506174, 6168351, 6287270, 6168357, 6342248, 6506148, 6523429, 6105308, 6273245, 6352046, 6319400, 6207463, 6168358, 6360313, 6510625, 6489398, 6543380, 6523422, 5869843, 6176712, 6264926, 6190466, 6207347, 6523519, 6506541, 6510563, 6489420, 6488905, 5975229, 5974852, 6287277, 6168340, 6027686, 6089999, 6342243, 6090232, 6489428, 6064688, 6352033, 6523483, 6489444, 6345414, 6489415, 5909979, 6543386, 6488911, 6506147, 6523482, 6489418, 6105310, 6319407, 6352100, 6394513, 6168349, 6360327, 6168342, 6488910, 6489394, 6523656, 6319301, 5910004, 6510648, 6345430, 6207423, 5869803, 6169982, 6246103, 6319295, 6315769, 6169990, 6169744, 6176728, 6207365, 6064680, 6319383, 6342227, 6169997, 6170001, 6264911, 6319285, 6207464, 6345419, 6352060, 6394545, 6352051, 6506149, 6489406, 6512535, 6297074, 6510550, 6523473, 6543376, 6523498, 6523470, 6543355, 6523486, 6523639, 6523336, 6543310, 5974943, 6190491, 5974989, 6506547, 6034434, 6411964, 6044449, 6345409, 6168335, 6319360, 5975133, 6352095, 6319382, 6506143, 6510613, 6207459, 6512536, 6190369, 6512548, 6090240, 6523676, 6523669, 6523671, 6044446, 6207426, 6543323, 6176725, 6276300, 6168354, 6089997, 6190472, 6259722, 6259744, 6190417, 6360311, 6297072, 6319405, 6360321, 6259721, 6169900, 6169999, 5974963, 6345429, 6506563, 6297082, 6506549, 6510571, 6523488, 6510586, 6319398, 6523487, 6352025, 6488925, 6523511, 6207342, 6352026, 6360331, 6512544, 6169979, 6342205, 6489431, 6265005, 5974841, 6489432, 6510551, 6264920, 6207428, 5974857, 5974864, 6394555, 6170015, 6319387, 6506188, 6169872, 6352098, 6064671, 6064847, 6168303, 6510594, 6543325, 5975218, 6190489, 6319359, 6352028, 6523492, 6169891, 6394650, 6259741, 6506145, 5974978, 6190438, 6064720, 5933485, 6319376, 6488913, 6287170, 6512515, 6523514, 6489442, 6287276, 6207319, 6297123, 6506536, 5909980, 6352054, 6259708, 6190397, 6207302, 6170002, 6360318, 6319326, 6297085, 5909998, 6345416, 6360328, 6169874, 6207370, 5974851, 6510585, 6510659, 6088947, 6064710, 6506535, 5869683, 6259710, 6287252, 6100060, 6190483, 6319404, 6207422, 6276314, 6190484, 6169914, 6342230, 6510542, 6510630, 6246126, 6207339, 6246104, 6488899, 6506175, 6506534, 6506194, 6523633, 6523327, 6246151, 6176713, 6297068, 5974986, 6523425, 6523666, 6064660, 6109032, 6360320, 5911046, 6543351, 5870039, 6246143, 6090251, 6246124, 6259711, 6168279, 6246109, 6510559, 6523655, 6168274, 6345411, 6394661, 6169913, 6190381, 6276294, 6276305, 6394642, 6276297, 6352092, 6345424, 6319385, 6488953, 6190468, 6512496, 6523635, 6510556, 6523680, 6489446, 6510548, 6352076, 6169909, 6506142, 6169749, 6512507, 6287256, 6034411, 5974866, 6064648, 6207293, 6342211, 5869816, 5910008, 6352103, 6411971, 6207287, 6394655, 6246160, 6360334, 6523438, 6506523, 6510602, 6510596, 6512525, 6488897, 6523660, 6297127, 6027643, 6319343, 6411952, 6297121, 6543366, 6264984, 6352105, 6064726, 6319307, 6510558, 6319292, 6489417, 6287263, 6319291, 6512495, 5870047, 6345423, 6176715, 6168295, 6027793, 6523505, 6543388, 6523646, 6089976, 5932743, 6523413, 6190450, 6488927, 6259704, 6169980, 6394529, 6523499, 6489429, 6523664, 6169767, 6342254, 6044453, 6207447, 6319358, 6394562, 6506552, 6543342, 6027629, 6543373, 6523640, 6190390, 6176716, 6246156, 6100056, 6168269, 5975119, 5974966, 6207300, 6207373, 6207372, 6319282, 6259731, 6352083, 6207376, 6297124, 6259724, 6352037, 6259732, 6259725, 6506139, 6523427, 6523516, 6489453, 6506528, 6512503, 6190486, 6543312, 6543327, 6543324, 6207324, 5974985, 6064707, 6276317, 6064669, 6207437, 6352031, 6352069, 6510583, 6287264, 6488903, 6319372, 6523508, 6512532, 6190384, 6319386, 6523638, 6287250, 5870048, 5932726, 6319322, 6319339, 6512546, 6190423, 6543357, 5932738, 6342237, 6319327, 6207456, 6100047, 6259706, 5932722, 6319395, 6411968, 6319369, 6488926, 6190424, 6512493, 5974973, 6100051, 6027687, 6489447, 6264996, 6319424, 6352080, 6168366, 6027683, 6207318, 5933434, 5974850, 6207429, 6176722, 6510575, 6090228, 6064697, 6169976, 6352084, 6506155, 5974950, 6489448, 6506538, 6543368, 6506167, 6190447, 6319283, 6319401, 6170008, 6523481, 6510555, 6510579, 6523478, 6345421, 6176710, 6523430, 5974860, 6168362, 6411949, 6100035, 6319412, 6345428, 6523485, 6352065, 5974977, 6352050, 6488906, 6297108, 6207478, 6510610, 6488949, 6109042, 6352088, 6264924, 6352048, 6246136, 6352104, 6169758, 6095939, 6264923, 6105319, 6207363, 6176721, 6100054, 6044223, 6168343, 6105314, 6287266, 6207366, 6264897, 6090000, 6089986, 6264902, 6352097, 6352087, 6264890, 6319280, 6287165, 6319281, 6345425, 6352049, 6352079, 6352019, 6352094, 6489434, 6512521, 6523672, 6510588, 6510557, 6489427, 6523333, 6506560, 6297120, 6207329, 6044200, 6044216, 6064843, 6168344, 5975228, 6360324, 6169911, 6352041, 6489405, 6207362, 6489445, 6523509, 6190388, 6523659, 6506545, 6352035, 6207361, 6168322, 6170014, 6319354, 6207368, 6168339, 6207284, 6044197, 6190469, 6543383, 6523518, 6512542, 6512538, 6488958, 6523314, 6027790, 6027788, 6207313, 5975135, 5870043, 6489399, 6207297, 6168337, 6105321, 6489436, 6489403, 6190407, 6319368, 6345404, 6506152, 6510554, 6207352, 6246114, 6027625, 6170000, 6510599, 6207364, 6510636, 6168341, 6394659, 6190391, 6506532, 6523675, 6352082, 6170025, 6044214, 6168364, 5932727, 6543340, 6169917, 6512518, 6360316, 6489451, 6095960, 6510604, 6190490, 6264880, 6506557, 6523629, 6190389, 6190456, 6394657, 6394554, 6411961, 6176724, 6342207, 6488904, 6506151, 6512540, 6506146, 6090223, 6543318, 6027810, 6319416, 5933446, 6352081, 6089994, 6394557, 6510600, 6523663, 6190446, 6411953, 6297093, 6523628, 6523507, 6064828, 6190430, 6510617, 6342256, 6523637, 6352030, 6319418, 6394586, 6512509, 6543337, 6190371, 6207457, 6169978, 6264994, 6319409, 6169899, 6090234, 6506150, 6064825, 6510640, 6345406, 6044198, 6044434, 6176734, 6207348, 6319313, 6190421, 6259729, 6510612, 6512522, 6190485, 6246113, 6259712, 6319324, 6360317, 6512537, 6287164, 6510590, 6506521, 6168321, 6319421, 6287274, 6489396, 6512547, 6169904, 6169765, 6259740, 6287261, 6027662, 6523658, 5933479, 6510632, 6345418, 6523510, 6411943, 6506562, 6506191, 6170021, 6506539, 6207476, 6034417, 6190420, 6523497, 6506159, 6510656, 6510541, 6265001, 6044457, 6411958, 6488942, 6523661, 6168336, 6176709, 6352020, 6489408, 6168277, 6034418, 6394559, 6543338, 6287262, 6190375, 6506519, 6287162, 6319403, 6394534, 6394654, 6345422, 6512508, 6543382, 5933460, 5975017, 6264995, 6176731, 6168309, 6027815, 6207292, 5910089, 6315778, 6095956, 6246120, 6342218, 6345410, 6170023, 6207431, 6488898, 6489430, 6489410, 6523411, 6512497, 6360335, 6510654, 6543375, 6523477, 5974987, 6510626, 6297111, 6297116, 6259726, 6506179, 6510614, 6360329, 6523681, 6287163, 6512531, 6170003, 6287172, 6342217, 6169998, 6190481, 6506171, 6506189, 6506192, 6512492, 6512499, 6512504, 6512512, 6512520, 6512527, 6512530, 6512545, 6411948, 6411965, 6246118, 6246119, 6264882, 6264883, 6264885, 6264886, 6264887, 6264888, 6264889, 6264891, 6264892, 6264894, 6264895, 6264896, 6264904, 6207290, 6207291, 6207299, 6207301, 6207326, 6207327, 6207332, 6207419, 6207433, 6207436, 6207341, 6207349, 6207439, 6246105, 6246106, 6246108, 6246117, 6246128, 6246129, 6259727, 6319370, 6319371, 6319373, 6319374, 6319377, 6319378, 6319380, 6319384, 6319390, 6319393, 6394509, 6394510, 6394511, 6394514, 6394520, 6394521, 6394522, 6394530, 6394531, 6259739, 6259742, 6259743, 6259745, 6259746, 6259747, 6259748, 6259749, 6273241, 6273242, 6273246, 6273247, 6273248, 6273252, 6273253, 6273255, 6273256, 6273257, 6273258, 6273259, 6273260, 6273261, 6273262, 6273264, 6273265, 6273266, 6319346, 6319347, 6319362, 6319363, 6207286, 6207288, 6207304, 6207305, 6207306, 6207307, 6207308, 6207309, 6207310, 6207311, 6207314, 6207315, 6207317, 6207328, 6207330, 6207351, 6207367, 6207369, 6207374, 6207375, 6207380, 6207381, 6207382, 6207383, 6207420, 6207425, 6207427, 6207434, 6207440, 6207441, 6207442, 6207444, 6207445, 6207446, 6207448, 6342241, 6342244, 6342247, 6342249, 6342251, 6342252, 6342255, 6287248, 6287251, 6287253, 6287254, 6287260, 6287267, 6287272, 6287166, 6287167, 6287175, 6287181, 6287182, 6287183, 6287184, 6287186, 6287187, 6287188, 6287189, 6287191, 6297086, 6297087, 6297088, 6297089, 6319287, 6319299, 6319310, 6319311, 6319312, 6319314, 6319315, 6319316, 6319317, 6319318, 6319319, 6319328, 6319329, 6319330, 6319331, 6319332, 6319334, 6319335, 6319336, 6319337, 6319342, 6360312, 6360314, 6360319, 6360322, 6360332, 6360337, 6394608, 6394609, 6394610, 6394611, 6394612, 6394613, 6394614, 6394615, 6394616, 6394617, 6394618, 6394619, 6394620, 6394621, 6394622, 6394623, 6394624, 6394625, 6394626, 6394628, 6394629, 6394630, 6394631, 6394632, 6394634, 6394636, 6394637, 6394638, 6394641, 6394644, 6394512, 6394528, 6352053, 6352057, 6352059, 6352061, 6352062, 6352063, 6352068, 6352070, 6352073, 6394508, 6394516, 6394527, 6394558, 6394563, 6394564, 6394565, 6394568, 6394573, 6394574, 6394575, 6394576, 6394577, 6394578, 6394579, 6394580, 6394581, 6394582, 6394583, 6394584, 6394588, 6394590, 6394591, 6394594, 6394598, 6394599, 6394600, 6394601, 6394602, 6394603, 6394604, 6394605, 6394606, 6394607, 6510605, 6510611, 6510622, 6510627, 6510629, 6510641, 6510644, 6510649, 6510650, 6510652, 6510653, 6523316, 6523317, 6523318, 6523319, 6523320, 6523321, 6523322, 6523323, 6523328, 6523329, 6523330, 6523331, 6523332, 6523334, 6523337, 6523448, 6523454, 6523325, 6523408, 6523418, 6523419, 6523420, 6523434, 6523436, 6523437, 6523440, 6523441, 6523442, 6523443, 6523444, 6523450, 6523451, 6523452, 6523456, 6523459, 6523460, 6523461, 6523462, 6523463, 6523466, 6523467, 6523521, 6523522, 6523524, 6523525, 6543341, 6543343, 6543344, 6543345, 6543346, 6543347, 6543349, 6543360, 6543361, 6543362, 6543364, 6543365, 6543367, 6543371, 6543372, 6543378, 6543381, 6543384, 6543390, 6543392, 6523650, 6489421, 6489437, 6319406, 6297096, 5869835, 6297094, 6488950, 6488928, 6394653, 6488929, 6523426, 6506190, 6488944, 6207296, 6506561, 6264927, 6394656, 6523424, 6273263, 6523652, 6297079, 6190429, 6488955, 6027635, 6506569, 6510543, 6190415, 6506517, 6273240, 6273250, 6506165, 6352021, 6394549, 6510566, 6264991, 6297076, 6394548, 6488920, 6168310, 6297062, 6523631, 5933431, 6394524, 6064841, 6523501, 6207360, 6273254, 6297069, 6394538, 6168289, 6273244, 6169752, 6190411, 6246144, 6297065, 6090237, 6506573, 6027685, 6345415, 6394544, 6506568, 6411963, 6543336, 6297081, 6176711, 6297064, 6319350, 6190482, 6523414, 6523457, 6394539, 6510569, 6510568, 6543335, 6543359, 6190378, 6095940, 6506183, 6190434, 6095948, 6169887, 6342222, 6510642, 6027621, 6064822, 6273249, 6297055, 6273238, 6394533, 6506565, 6506168, 6506550, 6523627, 6352029, 6190441, 6190464, 6246153, 6488940, 6411946, 6088936, 6506572, 6543321, 6190448, 6523645, 6273251, 6411957, 6352056, 6168275, 6297091, 6090224, 6207353, 6297073, 6506516, 6523409, 6510621, 6168273, 6319364, 6319353, 6510567, 6394542, 6394546, 6411962, 6394535, 6488909, 6342229, 6352027, 6510582, 6027618, 6411966, 6394526, 6523410, 6543339, 6523458, 6543394, 6176732, 6027689, 6523648, 6394541, 6090247, 6297125, 6190452, 6273239, 6297067, 6342219, 6411970, 6297056, 6394540, 6506522, 6488957, 6510615, 6506542, 6506543, 6488908, 6523624, 6273243, 6190410, 6394551, 6506571, 6488960, 6168297, 6273237, 6394547, 6488907, 6297097, 6342213, 6342209, 6095946, 6246138, 6523625, 6543322, 6264986, 6394543, 5932748, 6207359, 6168276, 6207356, 6297060, 6342224, 6190439, 6342208, 6523632, 6523657, 6523665, 6169743, 6207294, 6297070, 6510597, 6523649, 5975231, 6319361, 6506160, 6411969, 6510598, 6523667, 6506558, 6264985, 6394550, 6342206, 6190408, 6297057, 6342231, 6246141, 6488900, 6523512, 6352016, 6095941, 6523517, 6169884, 6089978, 6342226, 6523647, 6506162, 6543320, 6523668, 6168313, 6168270, 6342203, 6506173, 6543316, 6287169, 5910082, 6190436, 6207379, 6207377, 6297112, 6297099, 6394569, 5870230, 6259735, 6523439, 5869838, 6287185, 6297109, 6488914, 6246131, 6394658, 6506166, 6510624, 6259694, 6506180, 5870233, 6264925, 6319422, 6287268, 6523453, 5974979, 6319391, 5869003, 6394633, 6264909, 6027795, 6394648, 6512502, 6287275, 6259734, 6297126, 5869026, 6523445, 6488921, 6319341, 5932659, 6394643, 6207371, 5869009, 6287176, 6394660, 6394646, 6506178, 6297110, 6394593, 6287190, 6064725, 6394649, 6506184, 5911049, 6297122, 6297102, 6319397, 6647462, 6647319, 6675974, 6674372, 6646375, 6476164, 6634528, 6438946, 6468644, 6555477, 6655708, 6476163, 6466088, 6646385, 6649206, 6555524, 6447690, 6466962, 6476158, 6466722, 6466699, 6646388, 6447698, 6476148, 6447704, 6634478, 6655714, 6447715, 6438965, 6468941, 6555146, 6635462, 6635114, 6648816, 6635027, 6647469, 6634541, 6647332, 6647513, 6649247, 6476169, 6675231, 6466156, 6476175, 6476168, 6555498, 6635002, 6646395, 6438924, 6634461, 6655705, 6468640, 6647311, 6438928, 6468980, 6635012, 6648999, 6646485, 6646376, 6647471, 6439002, 6646363, 6466081, 6468651, 6476248, 6555269, 6722632, 6468649, 6647480, 6646389, 6468947, 6675179, 6476150, 6647539, 6466087, 6466719, 6647484, 6466163, 6447745, 6555201, 6438939, 6466101, 6476234, 6466710, 6555497, 6647527, 6555520, 6555175, 6447697, 6476237, 6476171, 6476159, 6635003, 6635490, 6634453, 6647317, 6649239, 6635136, 6555512, 6674369, 6447746, 6555246, 6468986, 6466096, 6447693, 6648842, 6555513, 6655662, 6555142, 6476222, 6476240, 6555261, 6634511, 6635001, 6646383, 6648838, 6438908, 6466704, 6634523, 6646355, 6647473, 6646447, 6649202, 6438919, 6438902, 6468972, 6655687, 6635005, 6647453, 6647340, 6647500, 6648991, 6438896, 6439011, 6466147, 6466143, 6466160, 6555226, 6555468, 6555178, 6555519, 6634517, 6555231, 6646463, 6647323, 6634547, 6466968, 6635457, 6647338, 6447717, 6647315, 6634534, 6466150, 6466963, 6646456, 6655703, 6635004, 6466083, 6647316, 6634493, 6648824, 6447696, 6555494, 6476170, 6655669, 6466158, 6634470, 6555488, 6634505, 6468973, 6646484, 6447705, 6468635, 6466965, 6649003, 6635148, 6555189, 6476155, 6647540, 6468653, 6648825, 6634490, 6438944, 6555253, 6675210, 6648856, 6675885, 6634510, 6555187, 6555275, 6635478, 6646367, 6647305, 6648830, 6648870, 6466981, 6555518, 6646470, 6647312, 6555509, 6469000, 6674379, 6675854, 6438984, 6635127, 6648823, 6555264, 6468646, 6468974, 6635141, 6555493, 6555237, 6555145, 6647344, 6555191, 6649227, 6447707, 6635142, 6646412, 6466960, 6466097, 6438979, 6466149, 6476245, 6438994, 6438971, 6476249, 6555278, 6635118, 6466707, 6635468, 6649224, 6646465, 6438920, 6438923, 6468994, 6476167, 6438933, 6476233, 6555196, 6555169, 6635466, 6555280, 6646405, 6646381, 6646364, 6646491, 6647455, 6675937, 6647346, 6675998, 6647302, 6675195, 6655697, 6675939, 6722673, 6555143, 6647476, 6649006, 6555265, 6647516, 6647520, 6647491, 6647303, 6648988, 6438952, 6634546, 6674375, 6722689, 6447699, 6555268, 6634515, 6646368, 6634500, 6438980, 6555485, 6466151, 6439008, 6466094, 6466964, 6466139, 6476238, 6675869, 6655704, 6649002, 6438907, 6466708, 6447706, 6468652, 6447689, 6466716, 6555267, 6447702, 6647535, 6647487, 6649248, 6468996, 6438899, 6438911, 6635014, 6468632, 6468971, 6634539, 6647529, 6555199, 6649197, 6468642, 6646404, 6447700, 6438894, 6447701, 6468988, 6555475, 6555170, 6635131, 6647548, 6555184, 6468633, 6466110, 6555168, 6635476, 6476232, 6466084, 6468634, 6646400, 6649194, 6635443, 6634456, 6468987, 6468998, 6646445, 6466141, 6647470, 6646473, 6468981, 6555190, 6635498, 6646393, 6635160, 6476228, 6555200, 6438927, 6438956, 6466961, 6635125, 6675868, 6476152, 6468978, 6466957, 6634998, 6555491, 6646466, 6555511, 6647538, 6647504, 6647509, 6649024, 6655667, 6649251, 6675891, 6675870, 6647501, 6722672, 6722665, 6722705, 6447712, 6634492, 6555148, 6438900, 6635049, 6439010, 6447767, 6466697, 6555227, 6555254, 6634479, 6438969, 6634459, 6555279, 6635011, 6634486, 6634513, 6649027, 6646451, 6634474, 6646472, 6675218, 6648827, 6648871, 6649018, 6655712, 6647331, 6468648, 6647325, 6555495, 6646411, 6634996, 6635047, 6675897, 6447749, 6466959, 6476243, 6555474, 6476230, 6555252, 6634532, 6634475, 6634508, 6655677, 6675229, 6447692, 6438910, 6646455, 6468984, 6466701, 6555521, 6438970, 6646486, 6647435, 6635038, 6468639, 6635103, 6647519, 6634483, 6438931, 6555165, 6647518, 6476149, 6438990, 6555258, 6675958, 6438967, 6438926, 6476235, 6647334, 6634467, 6555181, 6634544, 6634533, 6635158, 6555195, 6675873, 6648843, 6676017, 6648983, 6655688, 6635035, 6635048, 6634506, 6635000, 6635113, 6635042, 6647443, 6447708, 6647482, 6466973, 6476239, 6555198, 6555180, 6647449, 6466700, 6438909, 6466100, 6466095, 6649195, 6655694, 6466966, 6439006, 6438841, 6476151, 6555274, 6635029, 6635105, 6622455, 6634994, 6635152, 6466154, 6555523, 6647445, 6674371, 6647537, 6466144, 6555203, 6635032, 6438848, 6466714, 6555149, 6555160, 6635034, 6647488, 6439009, 6468645, 6555205, 6555185, 6635491, 6438959, 6634465, 6634466, 6634999, 6635016, 6635482, 6646471, 6622431, 6647328, 6647335, 6647336, 6555229, 6555147, 6635455, 6649015, 6447755, 6438968, 6646373, 6466146, 6647355, 6466713, 6466705, 6466098, 6476225, 6635018, 6675895, 6649026, 6476251, 6447765, 6466724, 6466138, 6647478, 6634514, 6647347, 6655701, 6555284, 6438954, 6466977, 6722693, 6466978, 6634502, 6634469, 6438922, 6555489, 6635442, 6675881, 6655656, 6634494, 6468979, 6634489, 6635140, 6647327, 6447743, 6635039, 6438961, 6468990, 6438916, 6466969, 6466140, 6468638, 6468982, 6555192, 6438917, 6635440, 6647349, 6648857, 6634526, 6647512, 6646356, 6439005, 6476246, 6468643, 6555238, 6466979, 6555197, 6635460, 6634997, 6635009, 6466955, 6635444, 6635104, 6675189, 6634516, 6476161, 6438977, 6555144, 6468641, 6468985, 6635447, 6466715, 6649241, 6438993, 6466161, 6555259, 6447716, 6466971, 6647356, 6634462, 6646357, 6555239, 6555486, 6555499, 6675860, 6634542, 6468993, 6647456, 6466090, 6438995, 6466086, 6555503, 6555517, 6635022, 6647541, 6647452, 6647475, 6647431, 6438925, 6476224, 6438957, 6466135, 6466975, 6447694, 6438903, 6447695, 6466720, 6466711, 6438996, 6476231, 6466956, 6466703, 6476229, 6476227, 6555466, 6634463, 6635488, 6635107, 6635151, 6635021, 6466972, 6476226, 6466698, 6635138, 6647474, 6635010, 6647465, 6447714, 6647460, 6647318, 6647306, 6648975, 6675975, 6647479, 6655661, 6675234, 6675996, 6635007, 6438974, 6438972, 6466157, 6468999, 6634484, 6635452, 6648980, 6722634, 6476242, 6555282, 6555154, 6635159, 6634455, 6438981, 6466137, 6468656, 6555228, 6646413, 6555155, 6476244, 6555159, 6634538, 6476160, 6635139, 6675971, 6674345, 6647547, 6647534, 6555163, 6476172, 6466093, 6555515, 6634485, 6439004, 6649200, 6447764, 6466970, 6555151, 6439007, 6649221, 6648989, 6447748, 6649236, 6635143, 6555481, 6466967, 6476154, 6466159, 6655692, 6438932, 6555179, 6635030, 6635112, 6647333, 6447742, 6647522, 6622434, 6648841, 6466152, 6476165, 6646490, 6646402, 6647523, 6647472, 6646460, 6647342, 6649240, 6647546, 6646365, 6466155, 6634481, 6722683, 6655679, 6466089, 6476223, 6635040, 6634522, 6466142, 6646476, 6646495, 6646384, 6675227, 6555262, 6635037, 6555241, 6438898, 6555285, 6655702, 6648995, 6466103, 6647454, 6647432, 6646496, 6646361, 6635144, 6647343, 6649034, 6649249, 6466980, 6634509, 6646369, 6655699, 6634477, 6635013, 6675851, 6555476, 6555230, 6635153, 6646391, 6675930, 6555232, 6468945, 6555277, 6646392, 6634548, 6646477, 6647545, 6634487, 6655695, 6555525, 6646371, 6649021, 6466712, 6555249, 6647503, 6555492, 6555193, 6646382, 6466974, 6555240, 6646474, 6555260, 6438983, 6438912, 6634549, 6438929, 6466145, 6438958, 6476250, 6476176, 6555244, 6675981, 6647544, 6649230, 6675982, 6655671, 6634521, 6646452, 6648815, 6648837, 6675209, 6634501, 6634488, 6634995, 6635017, 6635020, 6438960, 6646379, 6646380, 6646390, 6648813, 6648818, 6646399, 6647459, 6647483, 6647486, 6647489, 6647448, 6647457, 6647458, 6647463, 6647464, 6647466, 6647467, 6647468, 6675865, 6675876, 6675880, 6675888, 6675889, 6675893, 6675896, 6675899, 6438849, 6438850, 6438855, 6438856, 6438857, 6438860, 6438861, 6438864, 6438865, 6438870, 6438871, 6438872, 6438876, 6438877, 6438878, 6438879, 6438880, 6438881, 6438882, 6438884, 6438885, 6447739, 6447740, 6447744, 6447750, 6447751, 6447753, 6447757, 6447758, 6468942, 6468943, 6468944, 6468946, 6468948, 6468949, 6468950, 6468951, 6468952, 6468953, 6468954, 6468955, 6468956, 6468957, 6468960, 6468961, 6468963, 6468965, 6468968, 6555243, 6555255, 6555257, 6622430, 6622432, 6622437, 6622439, 6622440, 6622441, 6622442, 6622445, 6622446, 6622447, 6622448, 6622449, 6622451, 6622452, 6622454, 6622457, 6635006, 6635019, 6635023, 6635025, 6635026, 6635108, 6635109, 6635110, 6635111, 6635128, 6635129, 6635132, 6635133, 6635134, 6635445, 6635449, 6635450, 6635459, 6635461, 6635465, 6635467, 6635469, 6635471, 6635474, 6635475, 6635479, 6635480, 6635483, 6635486, 6635492, 6635494, 6635495, 6635496, 6655658, 6655666, 6655668, 6655674, 6655675, 6655682, 6655684, 6674321, 6674322, 6674323, 6674324, 6674325, 6674326, 6674327, 6674328, 6674329, 6674330, 6674331, 6674332, 6674333, 6674334, 6674335, 6674336, 6674337, 6674338, 6674339, 6674340, 6674341, 6674342, 6674343, 6674344, 6674346, 6674347, 6674348, 6674349, 6674350, 6674351, 6674352, 6674353, 6674354, 6674355, 6674356, 6674357, 6674358, 6674359, 6674360, 6674361, 6674363, 6674364, 6674365, 6674367, 6674368, 6647436, 6648831, 6648832, 6648833, 6648834, 6648835, 6648836, 6648844, 6648852, 6648853, 6648858, 6648859, 6648863, 6648865, 6648867, 6648868, 6648869, 6722602, 6722603, 6722604, 6722605, 6722606, 6722608, 6722609, 6722610, 6722613, 6722614, 6722615, 6722616, 6722617, 6722618, 6722619, 6722620, 6722621, 6722622, 6722623, 6722624, 6722626, 6722628, 6722629, 6722633, 6722635, 6722639, 6722641, 6722642, 6722643, 6722644, 6722645, 6722646, 6722647, 6722648, 6722649, 6722650, 6722651, 6722652, 6722654, 6722655, 6722656, 6722657, 6722659, 6438904, 6648828, 6438895, 6438976, 6555272, 6634460, 6555263, 6468983, 6635106, 6438854, 6438883, 6722694, 6646478, 6438888, 6635477, 6555507, 6438867, 6634520, 6722703, 6722968, 6555508, 6635497, 6438862, 6438889, 6635122, 6675882, 6466721, 6438863, 6438858, 6466717, 6635481, 6635120, 6634497, 6438975, 6646387, 6438999, 6476157, 6438887, 6466085, 6646409, 6466706, 6635472, 6723007, 6438866, 6646366, 6646374, 6646449, 6646370, 6635123, 6646386, 6555504, 6675847, 6635031, 6647307, 6722962, 6438987, 6468995, 6647530, 6466726, 6438892, 6438988, 6466104, 6438891, 6722988, 6466725, 6476177, 6438875, 6555505, 6635116, 6646410, 6635489, 6635101, 6722996, 6438997, 6675883, 6675904, 6646492, 6646479, 6649209, 6438874, 6646408, 6634498, 6438890, 6722999, 6438906, 6438897, 6647531, 6647350, 6722992, 6647536, 6438842, 6466091, 6466723, 6476174, 6646487, 6438886, 6438868, 6438869, 6646489, 6646396, 6646378, 6646360, 6635130, 6438873, 6466718, 6635119, 6646462, 6635024, 6438982, 6438936, 6622450, 6466106, 6646504, 6635454, 6622443, 6438973, 6466099, 6447703, 6635458, 6468975, 6466164, 6622444, 5057126, 6352091, 6510638, 5932662, 6100053, 5975015, 6394589, 6488901, 6264893, 6394571, 6319392, 6523324, 6523449, 6100033, 6523446, 6411960, 6512500, 6488935, 6169995.6488896, 6352096, 6523476, 5974937, 6394515, 6506144, 6489450, 6246161, 6176707, 6488938, 6064676, 6543314, 6510635, 5932760, 6488956, 6246123, 6319427, 5933464, 6027704, 6488936, 6345412, 6488918, 6168368, 6168271, 6489433, 6510573, 6342204, 6360315, 6523651, 6287269, 6190495, 6506526, 6105317, 6394561, 6394567, 6297075, 6089993, 6506181, 6543330, 6207461, 6027640, 6190451, 6207474, 6523504, 6488946, 6207378, 6506529, 6488954, 6510634, 6190444, 6090227, 6510589, 6352064, 6352071, 6394553, 6543379, 6342228, 6488902, 6394560, 6510593, 6510609, 6352047, 6246142, 6169889, 6105313, 5911106, 6488922, 6319321, 6488959, 6170022, 6543354, 6543331, 6523506, 6190403, 6345432, 6044458, 6510561, 6512526, 6168285, 5974856, 6523654, 6523496, 6345427, 6345426, 6506525, 6523677, 6506515, 6523471, 6264899, 6488923, 6394592, 6510637, 6394552, 6352058, 6207345, 6100055, 6190453, 6027628, 6207316, 5933445, 6190445, 6523479, 6190470, 6170009, 6510631, 6342210, 6168299, 6394645, 6506176, 6411942, 6027627, 6543391, 6190442, 6207466, 5911110, 6488947, 6488948, 6064724, 6510565, 5869489, 5932725, 6510547, 6543329, 6319288, 6523494, 6342223, 6394570, 6543385, 6506182, 6523526, 6510570, 6319306, 6319308, 6207443, 6027649, 6523432, 6394519, 6276302, 6394587, 6394556, 6044451, 6411950, 6512524, 6168296, 6168333, 6394517, 6512529, 6523527, 6342225, 5932741, 6169768, 6168352, 6168325, 6168323, 6411959, 6488945, 6489435, 6095957, 6190471, 6110386, 6506520, 6506185, 6506540, 6207355, 6523480, 6523523, 6027698, 6523673, 6510549, 6287168, 6352099, 6489414, 6264980, 6168356, 6168338, 6512543, 6512511, 6090241, 6315775, 6176718, 6297115, 6512541, 6352043, 6168315, 6523622, 6319297, 6027639, 6510651, 6510576, 6064675, 6027664, 6488919, 6297083, 6488933, 6342235, 6342239, 6394585, 6394597, 6394595, 6264912, 6489402, 6510658, 6246112, 6345431, 6090225, 6523423, 6207432, 6168278, 6345407, 6297084, 6510628, 6488924, 6506157, 6488916, 6543356, 6506156, 6044450, 6168361, 6064831, 6319289, 6169757, 6319294, 6207285, 6105307, 5974847, 6543328, 6319304, 5932737, 6095958, 5932724, 6319298, 5869002, 6319309, 6105301, 6319296, 6360310, 6168330, 6488932, 6543353, 6207320, 6489452, 6345420, 6510580, 6510643, 6510606, 6352034, 6543313, 6394651, 6510545, 6506164, 6264916, 6965010, 6649029, 6808460, 6898660, 6899073, 6827331, 6898612, 6898751, 6622435, 6555153, 7025374, 7017776, 6866236, 6959819, 6447761, 7050478, 6674377, 6991064, 7046843, 6964867, 6655707, 6774867, 6438949, 6808456, 6813654, 6827547, 6813667, 6834047, 6901112, 6959934, 6959827, 7005046, 6964873, 7005054, 7005042, 6964909, 7050427, 7024996, 7023790, 7047012, 7043298, 7047014, 6898757, 6812438, 6870678, 6898756, 6648978, 6634512, 6796871, 6964888, 6786597, 6870673, 6991014, 7046831, 7044514, 6796878, 6722964, 6675237, 6555202, 6870126, 6870117, 7044108, 7044037, 6965147, 6649217, 6827315, 6647494, 6866233, 6634503, 6960197, 6960198, 6960207, 7044048, 6898594, 6964893, 6675863, 6647313, 6808459, 6646359, 6812454, 6827521, 6866245, 6901086, 6901110, 6923314, 6648864, 6966787, 6991013, 6808457, 7031853, 7044078, 6647481, 7045089, 7005079, 7046811, 6812307, 6722667, 7043223, 6649243, 6786603, 6655659, 6634525, 6722691, 6648998, 6796898, 6870682, 6898762, 6898839, 6959860, 6655710, 6870669, 6960287, 6827280, 6960290, 7044534, 7044543, 7044005, 7046817, 6655673, 6808438, 6786619, 6923318, 7004965, 6990983, 6723006, 6722984, 6722736, 6870672, 6870147, 6466107, 6786580, 6796867, 6812469, 6834055, 6808411, 6898595, 6959918, 6960196, 6898873, 7004934, 6866251, 6438989, 6812428, 6991018, 6813656, 6964872, 7046997, 7012427, 6468964, 6655678, 6813659, 6898601, 6812354, 6813663, 6960235, 6959887, 6959926, 7043218, 7050377, 6655713, 6898596, 6466082, 6970976, 6870131, 6965134, 6838515, 6649001, 6966783, 6991067, 6898581, 6965013, 6979158, 6898657, 7023754, 6675957, 7046823, 6991055, 6870150, 6649228, 6649014, 6634454, 6964859, 7044105, 7046972, 6898602, 6808447, 6898753, 7043734, 6990705, 7046825, 6634504, 7050376, 6722997, 6722709, 6722979, 6901084, 6476162, 6898821, 6834106, 6898585, 6990696, 6674362, 6649025, 6722678, 6827512, 6813644, 6813665, 6827335, 6647326, 6898724, 6901089, 6923331, 6923330, 6898738, 6870674, 6959832, 6555158, 6990975, 6990965, 7031840, 7043282, 7043294, 6964912, 7047913, 6899029, 7043260, 6898583, 6634480, 6468647, 6555469, 6812303, 6827549, 6827277, 6898754, 6870665, 6960240, 6722708, 6960289, 6964936, 7050432, 7046784, 7047037, 6555510, 7045093, 6655686, 7044115, 6966793, 6970984, 6898860, 6634499, 6786620, 6555283, 6675230, 6635102, 6438851, 6675901, 6870102, 6649212, 6812352, 6898622, 6899040, 6898614, 6870108, 6899067, 6901087, 6898749, 6959859, 6960044, 6923324, 6966797, 6635155, 6959848, 6964978, 7031828, 7005061, 6827320, 7047008, 7046847, 7047002, 7047025, 6796872, 6649033, 6468655, 7024990, 6898714, 6812485, 6649233, 6555473, 6827332, 6447713, 6647434, 6899024, 6796897, 6964900, 6964988, 6960098, 7023785, 7045110, 7017764, 7050477, 7044516, 7043211, 7046790, 7043276, 7045098, 6796859, 6812479, 6898579, 6675942, 6899025, 6964982, 6964997, 6964904, 6808450, 6960283, 7057547, 6555186, 6722757, 6786599, 6990689, 6634536, 6647498, 6786584, 6812317, 6870114, 6960193, 6899079, 6870109, 6979136, 7023770, 7045121, 7046959, 6898865, 6722756, 6722727, 6959866, 6964968, 6635124, 6555233, 6447688, 6870629, 6899059, 6899069, 6960191, 6960192, 6959837, 6960245, 6991034, 6991068, 6634518, 6722722, 6649207, 6648994, 6812323, 6834056, 6834079, 6870092, 6898565, 6870143, 6898844, 6965005, 7044017, 6990704, 6827291, 6898682, 6722972, 6959874, 6723005, 7024963, 7043721, 6635149, 6788667, 6647497, 6649232, 6786631, 6648992, 6899070, 6898603, 6965028, 6960212, 7023783, 7050422, 7043293, 7047930, 6675952, 6647324, 6648987, 6555522, 6774834, 6813642, 6722713, 6808442, 6812432, 6827265, 6870142, 6866232, 6870159, 6923326, 6870137, 6866241, 6965009, 6965021, 6964996, 6959876, 6960246, 6959829, 7005023, 7008793, 6979115, 6996155, 7023771, 7043752, 7043259, 7044540, 6991061, 7043267, 7046866, 6979109, 7046854, 7043252, 6965022, 6966829, 6870112, 7008777, 6827536, 6722985, 6774885, 6468650, 6796855, 6796884, 6649005, 6812443, 6812326, 6786611, 6824494, 6827307, 6722653, 6898872, 6899060, 6964865, 7044087, 7024973, 7044009, 6923336, 6722735, 6946763, 6899064, 6898897, 6898740, 7046742, 6648849, 6438963, 6870640, 7046859, 6649208, 6870152, 6866256, 6675906, 6834065, 6899038, 6675977, 6901129, 6649032, 6647550, 6774869, 6812496, 6991048, 7050448, 7047027, 6964887, 7025382, 6960215, 7043227, 6675964, 6675236, 6647438, 6675965, 6675226, 6812328, 6946760, 6965011, 6960256, 6898604, 7024942, 6634491, 7044537, 6901111, 7044542, 7050468, 7044106, 6649235, 7024978, 6959854, 6555482, 6812308, 6990998, 6646501, 6812441, 6675928, 6786613, 6824500, 6827259, 6866253, 6923351, 6648854, 6979147, 7043288, 6866230, 7043248, 6990974, 7057538, 6979153, 7050386, 6898727, 6722674, 6647352, 6827319, 6655709, 6866254, 6635147, 6901079, 6960046, 7046961, 6870649, 6812329, 6675867, 6647358, 6646475, 6870679, 6898687, 6966788, 6774904, 6813632, 6648845, 6946728, 6965008, 6960297, 6647299, 6786638, 6634519, 6796860, 6813639, 6898722, 6898737, 6946729, 6870685, 7024975, 7046787, 7047932, 7044527, 6946730, 6979124, 6774833, 6675219, 6774893, 6555273, 6468997, 6812397, 6834098, 6870663, 6870664, 6965139, 6964884, 6966809, 6994824, 7023803, 6964999, 6979114, 6812338, 7046760, 7012423, 7046958, 6898725, 7045106, 6622456, 7005063, 6649245, 6979113, 7017783, 6870670, 6899057, 6901092, 6647439, 6959897, 6966822, 7046743, 6649246, 6649252, 6648872, 6438935, 6555480, 6901103, 6898787, 6796896, 6901128, 6946727, 6946726, 6870632, 6966828, 6996154, 6946757, 6959937, 6827529, 6635015, 7024958, 7023772, 7031876, 7024943, 7023763, 6808478, 7044509, 7043296, 7044088, 7043240, 7043239, 7044066, 7008780, 7050368, 7044071, 6898870, 7043213, 6870643, 6827533, 6866229, 7023769, 6555250, 6827257, 6723000, 6827302, 7023789, 7057553, 6555141, 6647511, 6898684, 6827510, 6646407, 7023798, 6655706, 6635117, 6827314, 6812336, 7023806, 7023797, 6960059, 6647433, 6634545, 6647493, 6827321, 6898890, 6899046, 6675856, 6923335, 7044014, 6979118, 7044507, 7045086, 6796882, 6786136, 6649199, 6870630, 6898690, 6870654, 6898600, 6959908, 6966806, 7047948, 6898588, 7047020, 6898564, 7043281, 7043307, 6676023, 6898811, 6870657, 6812453, 6898771, 6960250, 7044020, 6675905, 6866249, 6866221, 6866248, 7050378, 6966786, 6675902, 6898814, 6901091, 6959905, 6966820, 7031839, 7008787, 7045117, 6649210, 6476153, 6675938, 6646467, 6635121, 6722710, 6827300, 6808482, 6812372, 6555236, 6555235, 6960291, 6723013, 6959869, 6970993, 6979126, 7024985, 7046818, 6960292, 6722612, 6468636, 6468966, 6866224, 6648840, 6655672, 6648839, 6827301, 6960085, 7023807, 7047903, 6959896, 6960063, 6647337, 6655685, 6468629, 6827267, 6866259, 7046978, 7050451, 6675228, 6468958, 6647330, 6899044, 6555204, 6622438, 6898597, 6870125, 6898744, 7047904, 7047918, 7050465, 7057555, 7047949, 6946710, 7043277, 6898784, 7045111, 7024960, 6635485, 6675936, 6901117, 6812415, 6722714, 6675213, 6634537, 6649022, 6649019, 6786614, 6788657, 6555251, 6635051, 6808476, 6808433, 6808469, 6635441, 6812500, 6827540, 6447759, 6447763, 6870681, 6898773, 6870648, 6808437, 6964984, 6960279, 6979135, 6870158, 6960282, 7025371, 6964981, 7024995, 6635053, 7031945, 7031864, 7031871, 7031877, 7043241, 7046974, 6647321, 6960083, 6959907, 6898855, 6722696, 6960288, 6722684, 6774895, 6796902, 6788651, 6808434, 6812384, 6722715, 6722660, 6812467, 6870684, 6834099, 6898878, 6898781, 6898750, 6834078, 6898823, 6923337, 6965020, 6965031, 6959895, 6965012, 6964919, 6960302, 7044109, 6960253, 7045126, 7031813, 7046749, 6838500, 6827309, 6898679, 6964958, 6796853, 6990972, 6959883, 6635150, 6722752, 6722726, 6838502, 6555140, 6965030, 6990962, 7023794, 6649223, 7046927, 6675972, 6870151, 6870118, 6722762, 6898589, 6870087, 6438921, 6960276, 6964902, 6960227, 6964905, 7012420, 7046785, 6898710, 7047912, 7046973, 7043242, 7008792, 6964925, 6970989, 6466702, 6960295, 6898741, 7043311, 6447756, 6447747, 6898820, 6722677, 6870116, 6898716, 6438943, 6827276, 7045108, 6723014, 6722720, 6722712, 6634529, 6649204, 6649205, 6655693, 6812447, 6723011, 6774848, 6812434, 6827288, 6827329, 6774863, 6960041, 6901090, 6901119, 6827271, 7024956, 7044079, 6966821, 6979145, 7044508, 7046841, 7005056, 6466148, 6786130, 6808415, 6812499, 6827260, 6635473, 6786117, 6959838, 7005027, 7023768, 6648814, 7046773, 6813660, 6813635, 6827283, 6901126, 6808485, 6959912, 6991052, 6959932, 7008786, 7046813, 7046786, 7050438, 7047936, 7050430, 6898681, 6675866, 6655676, 6788644, 6647341, 6786126, 6898664, 6898827, 6923343, 6960239, 6898686, 6898606, 7024962, 6901116, 7031865, 6964881, 7044545, 7044038, 7005080, 6438901, 6438964, 6923349, 6722700, 6555467, 6812414, 6722740, 6813653, 6786593, 6827554, 6722969, 6722990, 6827286, 6898889, 6634507, 6946764, 6990976, 6965023, 7024976, 7044539, 7044067, 7057552, 7045138, 7023788, 7047925, 7046792, 6722631, 6649238, 6870103, 6965016, 6646377, 6635135, 7047894, 6649211, 6959903, 7046988, 7046795, 6555502, 6647304, 7023799, 6649020, 6960233, 6674378, 6901102, 6960281, 7005057, 7024974, 6870656, 7047935, 6979154, 6646454, 6555157, 6788646, 6812344, 6786128, 6898809, 7005035, 7017769, 6959904, 6555167, 7023774, 6555176, 6898843, 7005021, 6722970, 7017774, 6923321, 6898735, 6923346, 6966785, 7043233, 7046824, 6555172, 6438938, 6788640, 6447738, 6774847, 6786630, 6774872, 6808444, 6808418, 6675969, 6827293, 6635146, 6438893, 6870634, 6796903, 6870148, 6870638, 6870633, 6438945, 7005031, 6966803, 7024999, 6964965, 6898706, 6898620, 6898702, 6964889, 6960040, 6994845, 7043301, 6438840, 6647300, 6722749, 6647521, 6898691, 6899065, 6898838, 6923317, 6555500, 6970972, 6966795, 6990966, 7044523, 7045128, 6991041, 6808458, 6827317, 6813643, 6646406, 6786581, 6898615, 6866244, 6959879, 6964879, 6959861, 7043284, 7046842, 7046791, 7005066, 7047941, 6447709, 6655696, 7046993, 6649013, 6722611, 6468991, 6898708, 6827261, 6812304, 6946755, 6646448, 6970987, 6555150, 6898902, 7005037, 7043747, 6648985, 6555483, 6812315, 6812368, 6796858, 7043235, 6812382, 6675973, 6555276, 6812411, 6675959, 6870083, 6812458, 6827517, 6774880, 6923339, 6965001, 6901075, 6970983, 7004944, 6990964, 6898901, 6812457, 7023761, 7050463, 7005038, 7045084, 6899034, 7024950, 6990699, 6923323, 6970985, 6970958, 6722760, 6774857, 6635470, 6808487, 6965002, 6812389, 6468992, 7047915, 6722738, 6555171, 6774881, 6635033, 7017771, 7047900, 7044118, 6675202, 6646498, 6655689, 6649214, 6946737, 6439018, 6990702, 6966811, 6923344, 6813657, 6796890, 6646398, 6647308, 6774889, 6786134, 6796854, 6796866, 6796865, 6786591, 6786135, 6812403, 6827524, 6786127, 6675933, 6870652, 6827552, 6898763, 6923340, 6808440, 6923350, 6898605, 6959917, 6960047, 6959929, 6960261, 6959886, 6966835, 7005024, 7005029, 7005045, 6964875, 6476241, 6991053, 6966805, 7050400, 6827296, 6438914, 6786600, 6827526, 6675970, 6991000, 7043302, 7046846, 7045132, 6898582, 6827508, 6991056, 6646459, 6655691, 6827282, 6646488, 6646401, 6965157, 7050460, 6960081, 7023808, 6812486, 6959841, 6960266, 6960264, 6996176, 7044095, 6722730, 6796891, 6898846, 6964894, 7043255, 7057545, 6722640, 7047003, 7031805, 7047899, 7012414, 7045101, 7045090, 7046776, 7047905, 7050428, 6722982, 6722671, 6722686, 6722717, 6622433, 6722658, 6788664, 6722995, 6786608, 6646358, 6812337, 6796887, 6812460, 6827535, 6796881, 6898669, 6622453, 6901104, 6827560, 6824504, 6898587, 6824492, 6901108, 6898729, 6870146, 6901088, 6898621, 6964923, 6979148, 6964932, 6898786, 6964976, 6979132, 6966807, 6964969, 6959930, 7012416, 6966830, 7012413, 7005051, 6991005, 7012428, 6991002, 7043274, 7005039, 7012425, 7043221, 7043273, 7050423, 6965143, 6476236, 7043746, 6647447, 6647440, 6813662, 6991065, 6646403, 6964987, 6964989, 6970988, 6898826, 6970961, 6979128, 6722991, 6647492, 6722662, 6649196, 6722978, 6786594, 6824491, 6675874, 6808470, 6786590, 6808468, 6827285, 6866225, 6812320, 6812342, 6634482, 6635137, 6898850, 6898854, 6901080, 6898574, 6898609, 6898867, 6901071, 6959847, 6964967, 6964973, 6991021, 6964933, 6964917, 6991007, 6991017, 6990971, 7017773, 7023809, 6960226, 6964948, 7046763, 6959850, 7043314, 7044513, 7050476, 7043246, 7031953, 7046971, 7043251, 6796879, 6723017, 6959919, 6964979, 7044064, 7046856, 6647345, 6827290, 6996185, 6796857, 6722744, 7024968, 6898819, 6964911, 6964980, 6960223, 6812316, 6675940, 6960055, 6960076, 6649237, 6647505, 6946736, 7044016, 6648846, 6555245, 6646482, 7046788, 6675861, 6960091, 6965133, 6898720, 6647514, 6808430, 6827322, 6675968, 6960241, 6813633, 6960086, 6774865, 6796900, 6722987, 6946712, 6827528, 6960057, 6675946, 6959899, 6960201, 6960262, 6959863, 7046767, 6649203, 6898790, 6788653, 6555471, 6834054, 6970956, 6964935, 6979157, 7025375, 7025385, 6812375, 6827334, 6899028, 6647477, 6959935, 6960093, 7046867, 6808484, 6827299, 6870127, 6634531, 6965006, 6970977, 6970982, 6996153, 7047005, 6946735, 6964961, 6964960, 7005026, 7031856, 6786610, 6827308, 6898766, 6960307, 7044083, 6990960, 6647301, 6827327, 6647320, 6675894, 6827273, 6722741, 6994816, 6996152, 7024947, 7023778, 7047942, 6635050, 6634476, 6634550, 6647450, 6898836, 6870687, 6870671, 6898748, 6899030, 6722742, 6991044, 6964946, 6964942, 6946742, 6964941, 7031847, 7008784, 7045129, 6722980, 6675962, 6655680, 6722989, 6646414, 6774837, 6786583, 6827272, 6796880, 6870122, 6898715, 6866239, 6647329, 6898849, 6901109, 6655700, 6899078, 6923334, 6774835, 6964876, 6964857, 6964858, 6960255, 6959909, 6960072, 6979134, 7004964, 7005049, 7005069, 7017779, 7031809, 7023773, 6990697, 7044044, 7043733, 6964928, 7044063, 7050433, 7046748, 7050426, 7044018, 7050401, 7046925, 7046857, 6676020, 6786628, 6649219, 6901114, 7012424, 6647437, 6834051, 6648826, 6812480, 6899076, 6723009, 6635463, 6722751, 6813668, 6899071, 6796863, 6439013, 6812385, 6812318, 6870124, 6647446, 6870123, 6898875, 6898730, 6898745, 6960056, 6438940, 6959889, 6959920, 6960254, 6990989, 6990995, 6990963, 7024949, 7023779, 7043719, 7005028, 7044006, 7043720, 7047953, 7044090, 7044510, 6675220, 7047036, 7047035, 7046780, 7047929, 7043243, 6991001, 7046832, 7046960, 7047931, 7047946, 6964940, 6635036, 6774840, 6796852, 6866243, 6447752, 6647348, 6898810, 6898599, 6898607, 6898861, 6898598, 6965135, 6970995, 6870085, 7031806, 6722679, 6788659, 6649250, 6786121, 6796862, 6786125, 6813649, 6812392, 6796845, 6898734, 6827511, 6813666, 6959885, 6990978, 6812437, 7017790, 7024989, 6965141, 6438918, 6964892, 7046828, 6649010, 6438852, 6796851, 6834103, 6870651, 6722734, 6990993, 7044076, 7044102, 6438950, 6675980, 6722728, 6866238, 6898721, 6960075, 6838509, 6959902, 6960217, 7005041, 7043269, 6960219, 7050414, 7050435, 7046768, 7044057, 7046844, 7047921, 7045105, 6675864, 6722750, 6723012, 6648819, 6466162, 6788643, 6788650, 6796899, 6870090, 6808488, 6898728, 6923347, 6923329, 6960248, 6964871, 7023804, 6979105, 7043254, 7046837, 7012419, 7046829, 6439003, 6648984, 6788652, 6466982, 6970962, 6994851, 6991015, 7017800, 7023765, 7046783, 7046798, 6674366, 6827262, 6959891, 6970957, 7047030, 6674374, 6674376, 6674373, 6675877, 6774898, 6796849, 6786120, 6796883, 6808427, 6824493, 6808445, 6824499, 6812473, 6812325, 6812470, 6723015, 6812359, 6866227, 6870161, 6870160, 6899035, 6899050, 6866220, 6898592, 6898822, 6870141, 6960278, 6960294, 6959880, 6965138, 7046840, 6959839, 6959844, 7045080, 6959833, 7046796, 7043250, 6722748, 6555162, 6946700, 6960243, 6959831, 7017768, 6870144, 7043731, 7024980, 6675963, 6788655, 6788645, 6808436, 6813634, 6808443, 6870680, 6901124, 6866228, 6964883, 6901113, 6555194, 6647542, 6964878, 6796876, 6722994, 6796868, 6648866, 6827311, 6898851, 6979149, 7031956, 6964998, 7043247, 6901125, 6438846, 6808451, 6555152, 6796894, 6959916, 6966794, 7004971, 6991020, 6991033, 6676015, 7043289, 7046870, 7043299, 6648821, 6899041, 6796869, 6964890, 6996171, 6722695, 6808464, 6964949, 6991012, 7023800, 6438966, 6901099, 7046836, 6722974, 6786606, 6827268, 6722721, 6722753, 6722704, 6960272, 6722706, 6722755, 6898662, 6870086, 7023781, 6722998, 6722719, 6866255, 6870104, 7031832, 6959835, 6966789, 6960048, 6965018, 6965035, 6965032, 7025000, 6649222, 6774884, 6438934, 6898772, 6964945, 6966796, 7057540, 7044536, 6438951, 6646502, 6655715, 6439014, 6555164, 6555188, 6964975, 6796848, 6466109, 7017780, 6960293, 7005040, 6979150, 6964974, 6979121, 6647490, 6812313, 6870135, 6965137, 6964957, 6990706, 6960285, 6722993, 6834053, 6901121, 6898760, 6959924, 6965153, 6964921, 6870128, 7005030, 6991070, 6722743, 6648993, 6812348, 6898805, 6964926, 7005055, 7023784, 6438986, 6675966, 6439016, 6722680, 6796877, 6788654, 6827274, 6812474, 6827304, 6834093, 6813638, 6866237, 6870121, 6898739, 6827502, 6870660, 6965152, 6960213, 6808446, 6959834, 7004945, 6959865, 7012412, 6959842, 7024951, 7044007, 7043316, 7046833, 7046830, 7005064, 7046947, 7057554, 7046861, 6675225, 6870098, 7005032, 7047038, 7044089, 7046834, 7050472, 7047022, 7050475, 6438845, 6649225, 6647322, 6870156, 6959872, 6994830, 7023760, 7023777, 6901096, 6655690, 6786123, 6774888, 6786118, 6812399, 6870133, 6812456, 6870157, 6796885, 6808479, 6898783, 6898802, 6898788, 6923315, 6946754, 6960208, 6965168, 6960221, 6960077, 6966791, 6960200, 6960051, 6994819, 7024952, 7043328, 7023792, 7044515, 7043292, 7047944, 7050365, 6555271, 6466092, 6786607, 6675976, 6827295, 6827264, 6899033, 6923342, 6866222, 6634527, 6964950, 6960079, 6960244, 7005044, 7043283, 7046814, 6655665, 6722668, 6788669, 6960088, 7024957, 6965167, 7024941, 7024966, 7024954, 7043231, 6996162, 6960053, 6960224, 6946738, 7008788, 6870082, 6834095, 6898789, 7043326, 6827278, 6898800, 6774836, 7031826, 6827281, 6468967, 6438915, 6994847, 7023780, 6870107, 6675223, 6722711, 6722731, 6635487, 7043285, 6960080, 6447754, 6965136, 6964910, 7045123, 7057539, 6675221, 6899058, 6812393, 6646457, 6959820, 7023796, 7050474, 7046848, 6990692, 6812390, 6555234, 6438913, 6959870, 6959906, 6827313, 6649242, 6960214, 6959821, 7031852, 6635041, 7046793, 7047919, 6723018, 6964952, 6675931, 6655660, 6898894, 6634457, 6898882, 7045109, 6647533, 6655664, 6723016, 6796870, 6898831, 6827530, 6964956, 6964963, 6923345, 6959818, 7025380, 6979111, 7050429, 7047934, 6635464, 6439015, 6796846, 6675967, 6827506, 6901100, 6827543, 6870631, 6923316, 6960052, 7024983, 7024955, 7012431, 6555506, 6468657, 6959875, 6555248, 6808475, 6923338, 6959823, 6979141, 6960095, 6723004, 6647339, 7023767, 6438941, 6996170, 6991019, 7046754, 6634471, 6964964, 7047956, 6996168, 6447711, 6722685, 6808473, 6675871, 6827305, 6898608, 6812371, 6901085, 7043278, 6675862, 6812387, 6786634, 6964898, 6990999, 7023782, 7005052, 7031858, 7057551, 6647532, 6722737, 6808467, 6635451, 6898891, 6649023, 6898853, 7023795, 6946759, 6959890, 6898880, 6960251, 6812306, 6834071, 6899055, 6959882, 6959922, 6960299, 6964903, 6966798, 6827298, 7043741, 6675989, 7005070, 6965170, 6649244, 6675991, 6647485, 6466958, 6722718, 6774892, 6786622, 6788658, 6723002, 6555478, 6786604, 6722732, 6786639, 6675850, 6774899, 6813664, 6796886, 6647442, 6812498, 6808466, 6796892, 6786623, 6647524, 6812374, 6812310, 6813645, 6827541, 6786627, 6827556, 6812487, 6439000, 6827258, 6870145, 6898893, 6901095, 6827548, 6675872, 6898845, 6898613, 6898832, 6722666, 6898616, 6827516, 6834058, 6812416, 6898858, 6827550, 6722729, 6898899, 6964868, 6964874, 6960286, 6960263, 6964880, 6966824, 6970974, 6966832, 6964907, 7017770, 6964906, 6979131, 7005077, 6964886, 7017772, 7005036, 7043217, 7050412, 7047897, 6964916, 7050391, 7047922, 6965155, 6964862, 7050420, 6646397, 6866235, 6675953, 6964901, 6964915, 6646500, 6634530, 6438962, 6635499, 6722976, 6468630, 6796847, 6812445, 6812302, 6898879, 6964990, 6722724, 6960249, 6898863, 6649016, 6964991, 6838516, 6647354, 6646450, 6796888, 6813641, 6812351, 6675932, 6555156, 6476173, 6964861, 6964914, 7046849, 6959878, 6959933, 6898812, 6634543, 6647441, 6722701, 6796901, 6796875, 6634458, 6675961, 6675943, 6812494, 6870644, 6634464, 6649226, 6870677, 7012430, 6808431, 6946740, 7031860, 6827561, 7047910, 6959851, 6827303, 6555177, 6990986, 6960094, 7046746, 7050446, 7023762, 6786129, 6796874, 6647351, 6827330, 6965169, 6648817, 6786605, 6964971, 6722697, 6901077, 6966819, 7057544, 6964972, 6722698, 6812400, 6898677, 6786609, 6812418, 6901107, 6901072, 6959923, 6555247, 6722761, 6722630, 6870659, 6964955, 7008778, 6898777, 6960232, 6866231, 6898829, 6813658, 6870132, 6898567, 6813655, 6964938, 6979177, 6946762, 6960231, 6991008, 6966790, 7046868, 7005053, 7044070, 6960090, 6675954, 6646468, 6812471, 6964934, 6965154, 6990961, 7045127, 6648862, 6468989, 6649216, 6786624, 6675960, 6901123, 6439019, 6834073, 6959936, 6959888, 6990693, 6959915, 6960058, 6991045, 7031844, 7045113, 7050456, 6990988, 6647309, 6870099, 6870093, 6648820, 6866242, 6870140, 6870688, 6901115, 6960092, 6964920, 6964962, 6964931, 7023764, 6991038, 7043225, 7031870, 6866246, 6898733, 6898573, 6960070, 6960071, 6960049, 6960067, 6960189, 7004983, 6964929, 6774890, 6827270, 6812346, 6866223, 6898617, 6898791, 6870676, 6899053, 6979122, 7031859, 6898584, 6722965, 6649009, 6466983, 6796861, 6812451, 6812305, 6870635, 6898887, 6786629, 6898896, 6898898, 6923322, 6870637, 6960074, 6965004, 6965000, 6959893, 6960045, 6959910, 6960238, 6960257, 6675205, 6774886, 6675217, 6808463, 6808435, 6722747, 6722607, 6812319, 6899027, 6648996, 6960068, 6960242, 7005047, 6966825, 6647444, 6870666, 6870100, 6898874, 6898591, 6964953, 6991062, 6647517, 6675858, 6808429, 6870110, 6964951, 6994818, 6964922, 7044012, 7023756, 7046758, 7047034, 7047024, 6898755, 6901073, 6722971, 6834082, 6870153, 6866234, 6899023, 6946744, 6438947, 6899037, 7046826, 6960258, 6827324, 6866247, 7005062, 6964937, 6774882, 6827266, 6827316, 6827325, 6870149, 6991011, 7024986, 7044093, 6786621, 6991063, 6774825, 6812327, 6901078, 7023757, 7047004, 6647451, 6722973, 6827269, 6827312, 6808449, 6827294, 6923333, 7023793, 7047011, 7047015, 7046838, 7044051, 7046991, 7045124, 6899036, 6994837, 6649231, 6964908, 6649229, 6648847, 6649198, 6923319, 6870655, 6946756, 6964897, 6555183, 6959862, 6960084, 7047907, 6959853, 6960096, 7050355, 6655681, 6468654, 6960273, 7044065, 6675926, 6990968, 6965142, 6812396, 6901082, 6959881, 6827323, 6946781, 6965159, 7023766, 7023811, 6898866, 6808454, 6646394, 6655657, 6813650, 6827310, 6964930, 7012422, 6960199, 7031850, 7031836, 6898864, 6960268, 6964927, 6622436, 6649220, 6812309, 6675929, 6646446, 6723001, 6866240, 6722981, 6899082, 6901074, 6898881, 6946779, 6990981, 7043275, 6898847, 6635126, 6675900, 6655683, 6722670, 6788656, 6476166, 6898837, 6901127, 6898856, 6898782, 6964866, 6979116, 7008779, 6964947, 7008791, 7043716, 7025369, 6648855, 6774849, 6812324, 7046852, 6979171, 6648990, 6648861, 6970959, 6966823, 7046987, 7046778, 6812497, 6635156, 6647461, 6827279, 6827333, 6901098, 6898886, 6898877, 6834094, 6960271, 6870120, 6899080, 6898593, 7005058, 6827534, 7008781, 6966827, 6990990, 6990984, 7031810, 6796844, 6655663, 6786131, 6812448, 6966834, 7031803, 6674380, 6960050, 6964994, 6964993, 6996159, 7005075, 7047031, 7046751, 7046941, 6722975, 6675945, 6723010, 6722739, 6722961, 6675983, 6812409, 6796895, 6870096, 6870101, 6870115, 6899066, 6898580, 6899031, 6808477, 6959836, 6960089, 6966801, 6796893, 6959855, 7005025, 6960073, 6996174, 7023775, 6964995, 6960065, 7043324, 7047026, 7044022, 7044086, 7047000, 6834077, 6870658, 6870139, 6898732, 6898778, 6960042, 6991077, 7023791, 6960230, 7044027, 7044085, 7044074, 7046850, 6870155, 6901076, 6901101, 6991009, 6991039, 6994825, 6994834, 6994848, 6994850, 7023787, 7023801, 7023802, 7023805, 6774875, 6774876, 6774877, 6774883, 6774891, 6774894, 6774897, 6774900, 6774902, 6774903, 6996158, 6996160, 7050374, 7050383, 7050413, 7050452, 6796850, 6796856, 6796873, 6796889, 6555490, 6647515, 6813640, 6813646, 6834066, 6834068, 6898785, 6898792, 6898796, 6898797, 6898801, 6898806, 6898813, 6898834, 6898661, 6898667, 6898675, 6898683, 6898685, 6898688, 6898693, 6898695, 6898704, 6898709, 6898711, 6898712, 6960274, 6960275, 6991076, 7004936, 7004941, 7004947, 7004948, 7004949, 7004950, 7004952, 7004953, 7004956, 7004978, 7005068, 7012417, 7012418, 7012421, 7012426, 7012429, 7031867, 7031868, 7031872, 7031873, 7031875, 7043723, 7043725, 7043736, 7043754, 7044002, 7044004, 7044013, 7044036, 7044031, 7044531, 7044533, 7044030, 7044041, 7044049, 7044053, 7044059, 7044062, 7044082, 7045083, 7045092, 7045095, 7045097, 7045102, 7045115, 7045122, 7045130, 7046762, 7046777, 7031829, 6649213, 6722627, 7031815, 6946697, 7031811, 6808461, 6812332, 6813671, 6870097, 6812380, 7024993, 6870089, 6994841, 7044098, 6812343, 6812383, 6898610, 6960209, 6991004, 7044096, 7025372, 7025384, 6675232, 6813669, 6675235, 6555270, 6898736, 6866252, 6646469, 6870134, 6813670, 6979175, 6675941, 6648851, 6899054, 6812391, 6812388, 6901105, 6812476, 7046998, 6786132, 6722966, 6722688, 6991024, 6722637, 6960259, 7046865, 6996140, 6946751, 7050467, 6675925, 6722967, 6898869, 6898779, 6979103, 7031804, 6898566, 7044120, 6898888, 7004954, 6675193, 6923332, 6946770, 6649218, 6996139, 6960305, 6824497, 7004955, 7044077, 7047914, 6812501, 6722746, 6946753, 7044101, 7046770, 7044541, 7043730, 7043265, 6812417, 6898577, 6959864, 7025383, 7031938, 7047018, 6786585, 6812477, 6959868, 6646464, 6675200, 6648997, 6959845, 6812322, 6675201, 6812495, 7031952, 6722723, 6722977, 6774841, 6898563, 6959913, 7004943, 6970992, 7004979, 6812345, 6946747, 6946775, 7046741, 6946768, 7044113, 7044524, 7043256, 7043750, 6812475, 6774839, 6870661, 6898892, 6996145, 6970964, 6898885, 6827306, 6959840, 6970963, 6898769, 7004942, 6946750, 7025378, 7046862, 7031951, 6970990, 6991073, 6898848, 6838503, 6959871, 7031842, 7031940, 6722963, 6898774, 6722960, 6722601, 6834081, 6946707, 7044058, 7050375, 6649234, 6970960, 6898572, 6959914, 6996143, 6898578, 6812350, 6979127, 6960218, 6824507, 6722636, 6675955, 6774858, 6555484, 6774831, 7004935, 7047924, 6649215, 6870639, 6970966, 6675848, 6812440, 6649007, 6946772, 7043262, 6812483, 6812331, 6970980, 6946698, 7031942, 7043219, 6946777, 7004939, 6946752, 7044119, 6979108, 6960267, 6649201, 6675951, 6438942, 6722661, 6960265, 6898571, 6870667, 6960300, 6946696, 6812492, 6946699, 6946703, 6959822, 7044052, 7050415, 6675878, 6834074, 6834088, 7031946, 6970967, 7044100, 6946771, 7044116, 7044103, 7031957, 7044114, 7044525, 6722638, 6646461, 6812353, 7031869, 7044073, 7046821, 6824496, 6722681, 6555470, 6923320, 6675956, 6946749, 6649193, 6959877, 6555501, 6722763, 6786616, 6812468, 6899063, 6675206, 6675903, 6722702, 6838505, 7025386, 7031947, 7050447, 6870653, 6970965, 6634496, 6970986, 6898895, 6960260, 6898775, 7025377, 7031939, 7043236, 6648979, 6898767, 6647496, 6870113, 6722716, 6438948, 6786637, 6786587, 6824509, 6870084, 6722675, 6834086, 6898569, 6838506, 6827263, 6959873, 6996142, 6898859, 6824502, 6812330, 7044047, 7031941, 7044522, 7047917, 6812478, 6812430, 6812466, 6870647, 6996150, 6898852, 7046992, 6722745, 6870641, 6812412, 6898765, 6722725, 6786586, 6946773, 7046815, 6675947, 6675187, 6808428, 6722676, 6834096, 6996129, 6991025, 6812333, 6824510, 6870668, 6466108, 6812341, 6870119, 6786588, 6812356, 6959828, 7031955, 7031950, 6870675, 6646481, 7045096, 7004977, 7047916, 7044042, 6898776, 7050410, 6675853, 6870642, 7017791, 6834067, 6990994, 6959901, 6898576, 7025368, 7050462, 6898758, 6960277, 7025379, 7050409, 6870683, 6990985, 6870645, 6722754, 6898871, 6898768, 6555182, 7044050, 6635115, 6722733, 6675875, 6970971, 6960284, 6675950, 6834090, 7031944, 7043728, 6812484, 6959928, 7046753, 6555281, 6634535, 6555256, 6649030, 6675948, 6722682, 6898570, 6996144, 7004940, 6648848, 6675978, 6812334, 6946748, 6946704, 6824508, 7045104, 6722959, 6898761, 6946778, 6959825, 7044081, 6898752, 6774854, 6812424, 6812335, 6946776, 6898884, 6959858, 7025376, 7031825, 6722699, 6970969, 7045119, 6649031, 6675949, 6722758, 6786625, 6834101, 6870106, 6959900, 6959921, 6979133, 7025381, 7046761, 6722986, 6812423, 6959830, 6824506, 6898764, 6675887, 6722663, 6946769, 6898868, 6970994, 6898759, 6812339, 7046864, 6786133, 6786592, 6898575, 7004933, 7043263, 6959846, 6959826, 7025370, 7044092, 7031954, 6946774, 6646453, 6675944, 6722669, 6824503, 6870094, 6870646, 6923313, 6824498, 7044075, 7044056, 6466105, 6634540, 6646483, 6898747, 6722687, 6979172, 6676012, 6960304, 6979174, 6774829, 6676003, 6774842, 6813647, 6970970, 6675997, 7005071, 6675222, 7031874, 7050416, 7031819, 7031833, 7045087, 7031822, 7046772, 7045085, 7031821, 7031824, 6960061, 7044117, 7044112, 7047927, 7050362, 6899075, 6898794, 6870136, 6970973, 6722664, 6812321, 6870088, 6866250, 6774826, 6960270, 6649004, 6866257, 6960210, 6812311, 7024994, 7046990, 6675208, 6774878, 6870138, 7005022, 7043226, 6990996, 6812340, 6675999, 6555266, 6676004, 6812377, 6675992, 6866258, 6946716, 6979176, 7031857, 7025373, 6990980, 6808486, 6675990, 6812312, 6808471, 6898742, 6866226, 6827287, 6960298, 7004951, 6870129, 7005076, 6813637, 6813652, 6970968, 6970979, 6813651, 6824505, 7004981, 6808439, 7031866, 6635456, 6646458, 6646493, 6966784, 7050443, 6786598, 6813661, 6870095, 6812386, 7005078, 6979178, 7043729, 7044099, 4771671, 6100048, 6319420, 6488939, 6411947, 6319429, 5974932, 6207473, 5932740, 6488917, 6488941, 6488930, 6105305, 6523474, 5870054, 6207321, 6259738, 5932731, 6259695, 6095965, 6109029, 6411955, 6394537, 6523326, 6297113, 6512514, 6506546, 6510591, 6506556, 6064718, 6352040, 6488915, 6264908, 6168346, 6345413, 6027677, 6100032, 5974953, 5932745, 6168348, 6110387, 6510623, 6319419, 6319305, 6027798, 6543369, 6352023, 6207312, 6543352, 6964959, 7046745, 6960296, 7046963, 7023810, 7046755, 6635154, 7050481, 6960280, 7044033, 6898862, 7044511, 7045125, 7044519, 7046851, 7050440, 6959856, 6964918, 6898883, 6965140, 7044054, 7046989, 7046968, 6923341, 6722983, 6959898, 6468976, 7044068, 7047006, 7046986, 6901120, 6834100, 6964895, 6960236, 6555516, 6960064, 6466709, 6960211, 6964954, 6646372, 6439012, 6647526, 6979120, 7047898, 6899042, 6827326, 6964977, 7025387, 6991003, 6990959, 6898726, 7031949, 7047023, 7044546, 7004967, 7050424, 6964944, 6959867, 7012415, 6555173, 7045131, 6965025, 6965015, 6991036, 7017809, 7046750, 7008776, 7046771, 6964882, 6827289, 7047952, 6635446, 6447762, 7044003, 6655698, 7057542, 7044021, 6996181, 6979139, 7047939, 7031817, 7043229, 7044043, 6646503, 7044055, 7044026, 7047010, 7046775, 7047009, 7024959, 6965007, 6898743, 6788636, 6438953, 7017788, 6476156, 6960066, 7050395, 6994821, 6898665, 7046946, 7050394, 7045081, 7046764, 7043253, 6898770, 7047896, 6827328, 6946722, 6964924, 6946723, 7046853, 7031807, 7050466, 6898833, 7050431, 7047957, 6990982, 7047926, 6959824, 6990973, 6990691, 7047920, 7047940, 6774901, 6870154, 7045114, 7050359, 7024998, 6870105, 6964869, 6786635, 7047945, 6649028, 6964970, 7044040, 6722759, 6555166, 7043300, 6812314, 7050388, 7044532, 6994820, 7047007, 6964860, 7046820, 6979144, 7031814, 7047902, 7005050, 6901093, 6923328, 7046985, 7046816, 7043748, 6960220, 6827284, 7017778, 7050417, 7050445, 7004961, 6808489, 6635028, 7050385, 7044529, 6827297, 6898723, 6834105, 6438859, 6555479, 6946780, 6965029, 7017806, 7023755, 7031846, 7046937, 6964896, 7045099, 6812449, 7046966, 7044046, 7045116, 6946739, 7045118, 7043245, 7043745, 6675181, 6960269, 6796864, 7050425, 7046774, 7045091, 6946761, 6996147, 6991016, 7044010, 7057536, 7047016, 6994817, 7043222, 6946765, 7050390, 7057543, 6964864, 6898568, 7047895, 7046999, 7023752, 7047909, 6439017, 6966804, 6959925, 7043295, 7044015, 7047955, 7044094, 6468977, 7043743, 7044512, 6959849, 6923348, 7046740, 7047028, 6964863, 6964885, 6964877, 7046739, 7050441, 7043305, 7004986, 6808480, 6959927, 7046981, 6964939, 6964986, 7046756, 7050366, 6959911, 7050464, 6990969, 7043737, 7050360, 7047001, 6990991, 6870686, 7046779, 7050471, 7045120, 6959884, 6959843, 7050450, 6970975, 7043234, 6838507, 7046995, 7050406, 6959892, 6648986, 7031808, 6923325, 6964870, 6996164, 7047933, 7045133, 7047033, 6991010, 6647525, 7047928, 7046863, 7044039, 7031943, 7023786, 7024944, 6438992, 7046845, 6959852, 6960097, 7045082, 7050393, 6960054, 7045100, 7050369, 6996175, 7046994, 7047911, 7043717, 6965014, 6438853, 6647508, 6870662, 6827318, 6808441, 6959894, 7057546, 7045134, 7046759, 6946685, 6964992, 7005072, 7046766, 6898663, 7004970, 6812493, 7047029, 6979100, 7045107, 6965163, 7017781, 6960060, 6964985, 7046789, 7043735, 7043261, 7044028, 7044023, 6786582, 7046860, 6786595, 7047013, 7045136, 6898900, 7044029, 6901122, 7017784, 7047017, 6898658, 6898825, 6898824, 6898876, 7024992, 7050418, 6990992, 6898659, 6827275, 7057541, 7047032, 7024971, 7046797, 7043268, 6923312, 6965003, 7044097, 7044061, 7050439, 7017782, 7046752, 6827557, 6827292, 6960247, 7044528, 7045139, 6722625, 6901097, 6898731, 7050379, 7050473, 7047019, 7043257, 7043317, 7047943, 7046747, 6996163, 7008789, 7050407, 7046977, 6965161, 7024988, 6946758, 7046794, 6923327, 6964966, 6898817, 6647314, 7031861, 7044069, 6946724, 6996156, 6996157, 7031820, 7045088, 7043220, 7047923, 7005065, 6813636, 7005074, 7044110, 7046984, 6812347, 6824501, 7045094, 7050364, 6994846, 6970978, 7043230, 7043718, 7031838, 7031835, 6994814, 7057537, 7050382, 7050444, 7050402, 7046769, 7031848, 7050454, 7043738, 7050389, 7050380, 7047908, 7050421, 7031837, 7043744, 7050411, 7043739, 7043751, 7031841, 7043266, 7057550, 7044019, 7031843, 7044526, 7043727, 7043726, 7044025, 7050449, 7031948, 7057548, 7050419, 7057549, 7031851, 7045137, 6990979, 6675934, 6898619, 6648976, 6990701, 7050372, 7044045, 7045135, 7050363, 6960303, 6979117, 7043228, 6994835, 6994842, 7045112, 7047937, 6813648, 6870091, 6675993, 6960301, 7044121, 6970991, 6979110, 7004938, 7050358, 6722690, 6994833, 6675979, 6466102, 7045103, 7005033, 7047938, 7005073, 7047958, 6488931, 6207303, 6394596, 6319357, 6064708, 6506186, 6168359, 6176733, 6506564, 6027679, 6506531, 6523520, 6523513, 6510639, 6512494.6646362, 7046757, 7046869, 7024961, 7044111, 7044080, 6991047, 6979142, 7043286, 6960087, 7017797, 6991051, 7043732, 6466136, 6990967, 7005059, 6898808, 7044060, 6898707, 6722692, 6808432, 7046982, 6675855, 6834089, 6991040, 7046782, 7047021, 6960252, 7023776, 6468628, 6960099, 6965148, 7046954, 6447760, 6648977, 6965149, 6964913, 7024987, 6634495, 7023759, 6991072, 6635448, 7046781, 6655670, 6964899, 6959857, 6996178, 7043742, 7017777, 6838511, 7043749, 6960069, 6960062, 7017765, 6438955, 7050437, 6994839, 6635043, 6634468, 6635453, 7031816, 6648850, 6648822, 6438998, 6648829, 6634473, 6648860, 6774830, 6169989, 6510619, 7017803, 6991028, 6979173, 6898828, 6438847, 6990977, 6634524; b) Item Number 320-36-03, UDI/DI 10885862086624, Serial Numbers: 1954405, 2219058, 2219059, 1954403, 2478133, 2219057, 2635737, 2878165, 3748189, 3752637, 3758376, 1954406, 1954408, 3591103, 3795157, 3795158, 2878162, 2878164, 2878166, 2878171, 2878163, 2878168, 2878169, 2878170, 2878167, 3918304, 3866343, 3866337, 3866339, 3866338, 3866340, 3918311, 3918314, 3918302, 3918303, 3918305, 3918315, 3918316, 3918320, 3918307, 3918308, 3918309, 3918312, 3918313, 3918317, 3918319, 3918321, 3918300, 3918318, 4127477, 4127478, 3866341, 3866342, 3866344, 3866345, 3866347, 3998795, 2478131, 2878172, 2635738, 3741451, 4264287, 4316293, 4264290, 4264294, 4254141, 3918301, 4254137, 4264288, 3997773, 4254143, 4254138, 4316296, 3997775, 3918306, 3998802, 4254140, 3998799, 4127472, 4127473, 4127476, 4254142, 4316294, 4316295, 3997776, 3998798, 3997772, 3998796, 4127474, 4316286, 2635739, 3758378, 4316289, 4264286, 4264289, 4316288, 4264291, 4264298, 4264300, 4254139, 3918310, 4316299, 4316298, 4264296, 3998797, 4827471, 4965937, 4827476, 4861982, 4861980, 5092480, 4965931, 4888404, 5032829, 4827475, 4965946, 4965928, 4965934, 4965938, 4965941, 5032824, 5032822, 4827477, 4827478, 4827484, 4861974, 4861975, 4861976, 4861977, 4861983, 4888407, 4888413, 4888415, 4888417, 4888416, 4965940, 4965927, 5152273, 5152281, 5152287, 5032821, 5032839, 5032835, 5032831, 5032836, 5092473, 5092469, 5130478, 4965932, 4965947, 4861981, 5032827, 3758377, 3998801, 4316297, 5152270, 4965935, 5152277, 5092476, 5152276, 5130479, 5247514, 4827485, 4965929, 4965945, 5247510, 5152283, 5092478, 5092471, 5092483, 4965944, 4827480, 5032834, 5247521, 5247515, 4827481, 4888410, 5247499, 5165793, 5032838, 5152292, 5152286, 5165796, 5165797, 5165807, 5247497, 5247518, 5152294, 5152278, 5152279, 5152288, 5152290, 5152291, 5152293, 5165812, 5152285, 5165799, 5165800, 5165802, 5165804, 5165813, 5247513, 5247519, 5247502, 5165801, 5247509, 5247517, 5247501, 5247504, 4888414, 5092481, 5032837, 5032820, 5032830, 4888408, 5092477, 4888409, 5247508, 4888405, 4888412, 4888418, 5032825, 5092474, 5152289, 4965949, 4965936, 4316291, 4264297, 3997774, 4264292, 4316290, 4861978, 4965948, 5152271, 5247505, 5165792, 4965942, 5092475, 5092482, 5247523, 5152269, 5247498, 5032841, 5092479, 5247500, 5032826, 5165809, 5247516, 5032832, 5152272, 5247503, 5152282, 5152280, 4965943, 4888411, 5165814, 5130477, 5165795, 5032828, 5247512, 6007497, 6190366, 6023115, 5868133, 6170143, 6009014, 6009042, 6023114, 5869617, 5869603, 6190236, 6023920, 5868475, 5868115, 5870079, 6170132, 6169792, 6007489, 5868174, 5868152, 5868104, 5868469, 6007482, 6007652, 5870077, 6189357, 5868109, 5869610, 6008321, 6190277, 6007635, 6190256, 5868116, 6189328, 6167533, 5869611, 6169793, 6007646, 5868113, 5870081, 5868127, 5868153, 5869623, 6007501, 5868156, 5869622, 6169789, 5868474, 6008320, 5868162, 6023915, 5868120, 5870084, 5868488, 5868480, 5870070, 6190309, 5868478, 5868483, 5868176, 6008317, 6190264, 6023925, 5868163, 6008326, 5868130, 6007639, 6189331, 5868472, 5869614, 5870060, 6023951, 6023907, 5868105, 5868167, 6007643, 6007498, 6007507, 6008328, 5868117, 6190345, 6009020, 6007650, 6169784, 6009038, 6023953, 5868151, 5868150, 5868148, 5869605, 5869602, 6007655, 6007510, 6009025, 6007659, 6009016, 6190237, 6190355, 5868135, 6169783, 6189360, 6189362, 6189367, 6189368, 6189373, 5869609, 5869620, 5868159, 5868121, 5868126, 5868145, 5868171, 5868175, 5868178, 5868179, 5868487, 6169778, 6169779, 6189344, 6189349, 6189352, 6189355, 6023934, 6190299, 6190305, 6190328, 6190331, 6190289, 6190322, 6190323, 6190324, 6190245, 6190254, 6190271, 6190282, 5868098, 5868106, 5868494, 5868099, 5868119, 4316287, 5092472, 4827470, 4827474, 4827483, 5032819, 5032840, 5165805, 4827479, 5247506, 4827482, 5165811, 6189389, 6323615, 6189384, 6543436, 6538312, 6189375, 6190333, 6189396, 6189336, 6543442, 6354343, 6190330, 6190249, 6023928, 5868090, 5870074, 5868134, 6190244, 6190269, 6023952, 6190266, 6323610, 6543472, 6190346, 6189358, 6543218, 6167520, 6009034, 6008314, 6009041, 6394741, 6323620, 6543427, 6190248, 6484606, 6007504, 6023921, 6543484, 5868144, 6167513, 6190338, 6170136, 6190303, 6543253, 6489158, 6543431, 6008338, 6167524, 6023123, 6394758, 5868157, 6169782, 5868493, 6354348, 6189386, 6489194, 5868484, 5868173, 6354347, 6190235, 6543469, 6323619, 6190319, 5870086, 6009026, 6023119, 6489175, 6169799, 6394721, 6189316, 6023101, 6170142, 6489148, 6190300, 6190279, 6354374, 6190265, 6189356, 6489168, 6543269, 6190261, 6394735, 6190358, 6489155, 6543225, 6190359, 5869612, 6323618, 6007662, 6354334, 5870068, 6543217, 6543219, 6008313, 6489151, 6190312, 6323614, 6543254, 6543473, 6323607, 6190267, 5868123, 6170146, 6007642, 5868102, 6009017, 6170123, 6169790, 6189324, 6323602, 6023950, 6354361, 6543276, 6354356, 5868165, 6394834, 6484598, 6489169, 5868142, 6543439, 6323611, 6543456, 6543428, 5868096, 6008331, 6454872, 6454898, 6354372, 6190294, 6538291, 6543443, 6190353, 6023113, 6190318, 6008311, 6008340, 6190368, 6007483, 6323621, 6354365, 6489159, 6543220, 6394725, 6394823, 6190258, 6190292, 6394754, 6354369, 6023912, 6538292, 6170130, 6169797, 6538303, 6023922, 6007500, 6023905, 6538317, 6543265, 6543453, 5868491, 6170129, 6007494, 6007503, 6190351, 6543261, 5868476, 6009037, 5868158, 6323598, 6190278, 6323608, 6189330, 6323599, 6484641, 6484640, 6489185, 6543247, 6543460, 6489154, 6543461, 6323594, 6007488, 6189369, 6354346, 6167515, 5868100, 6167534, 5868112, 5869601, 6489153, 6489160, 6189338, 6489192, 6394738, 6543249, 6543457, 6009028, 6007487, 6023126, 6190297, 6323613, 6538315, 6394728, 5870089, 6189314, 6169787, 6543229, 6023929, 6190298, 6189378, 6170144, 6170145, 6008336, 6023118, 5868489, 6023110, 5868111, 6169794, 5868122, 6543267, 6543430, 6543478, 6543275, 6023111, 6023121, 6189391, 6023943, 6023924, 6190262, 6008329, 6394806, 6489187, 6189388, 6167531, 6023109, 6394836, 6189323, 5868490, 6489174, 5868492, 5869619, 6007491, 5869626, 6008333, 6007640, 6489180, 6543270, 6007660, 6023937, 6190342, 6008323, 6167522, 6354376, 6190243, 6190336, 6543479, 6189322, 6008325, 6169786, 5869624, 5868155, 6007641, 6023129, 6167525, 6023107, 6489157, 6023911, 6394760, 6023919, 6394807, 6008334, 6189380, 6009040, 6167532, 6009018, 6503950, 6190320, 5868154, 5868473, 6167519, 6190356, 6023935, 6189321, 6323593, 6323606, 6190329, 6454894, 6489156, 6394759, 6354353, 6543273, 6190341, 6190273, 6394751, 6484599, 6484607, 6023944, 6023125, 6543462, 6189353, 6189327, 6394731, 6189359, 6354337, 6009036, 6454886, 6170128, 6543244, 6023909, 6190304, 6489161, 6323612, 5870063, 6489173, 6489162, 6543483, 6007658, 6543259, 6323597, 6543425, 6190263, 6484625, 6190234, 6023910, 6394832, 6484608, 6543222, 6189372, 6323595, 6189382, 6543272, 6543224, 5870066, 6189381, 6009023, 6190274, 6394750, 6007502, 6167517, 6189394, 6007656, 6167516, 6008315, 6354340, 6169788, 5870083, 6454887, 6354344, 6394740, 6489172, 6007649, 5870073, 5868132, 6489179, 6007665, 6394844, 6167537, 6007657, 6543482, 6007490, 6543277, 6190238, 5868143, 6190281, 6354358, 6023947, 6543250, 6543477, 6484591, 6484617, 6543467, 6354360, 6354364, 6189387, 6190283, 6394718, 6394719, 6394722, 6394723, 6394729, 6394734, 6394736, 6394737, 6394744, 6394748, 6394755, 6394757, 6394762, 6394726, 6489167, 6489195, 6189335, 6454873, 6489149, 6489150, 6538313, 6543236, 6543252, 6543256, 6543257, 6543429, 6543445, 6543449, 6543464, 6543221, 6543223, 6543235, 6543242, 6543246, 6543248, 6543278, 5868114, 6190307, 6489181, 5868169, 6023127, 6489170, 6007493, 6394813, 6394838, 5870088, 5868466, 6007511, 6007496, 6007647, 6008318, 6189329, 5870072, 6190349, 6190268, 5868481, 6190240, 5870064, 5868118, 5868107, 6689905, 6554975, 6554098, 6554056, 6658240, 6455338, 6554989, 6689901, 6698797, 6609731, 6554105, 6455337, 6455348, 6554969, 6455331, 6554082, 6689908, 6698793, 6554052, 6554106, 6455346, 6689923, 6609732, 6554066, 6609726, 6698812, 6689879, 6609759, 6609750, 6698824, 6698799, 6554972, 6698819, 6554077, 6658246, 6554948, 6554961, 6698841, 6554950, 6554979, 6689933, 6698768, 6554050, 6609752, 6554058, 6554079, 6554094, 6554084, 6554076, 6689884, 6609745, 6609730, 6698780, 6689889, 6689891, 6698777, 6609756, 6698831, 6698825, 6555019, 6455352, 6455335, 6555018, 6554089, 6689921, 6554093, 6555017, 6554095, 6455344, 6554974, 6554047, 6609733, 6554986, 6455351, 6554069, 6554978, 6555006, 6554992, 6689922, 6554947, 6554967, 6554061, 6554955, 6554977, 6554996, 6554071, 6455332, 6698772, 6554097, 6609736, 6455339, 6455341, 6455342, 6455347, 6455357, 6554952, 6554962, 6554997, 6554998, 6555005, 6555011, 6609744, 6609746, 6609765, 6689896, 6689897, 6689898, 6689911, 6689914, 6689915, 6689918, 6689925, 6689927, 6689929, 6689931, 6698764, 6698782, 6698788, 6698789, 6698828, 6698769, 6698822, 6454513, 6454503, 6454530, 6454505, 6454517, 6454518, 6454522, 6454527, 6454536, 6454512, 4316292, 4827473, 4965939, 5152274, 4888406, 5092484, 5247520, 4965933, 5152275, 5165810, 4827486, 5247522, 5130482, 5165794, 5092468, 5165808, 5130475, 6543466, 6454881, 6394745, 6394821, 6394828, 6167523, 6190246, 6394812, 6170127, 6189354, 6189365, 6023914, 6023923, 6023116, 6394802, 6023938, 6454889, 6538300, 6008319, 6354368, 6023906, 6009033, 6169795, 6189393, 6394742, 5868177, 6190242, 6008332, 6323596, 6007485, 6489176, 6489186, 6354359, 6543237, 5870076, 6394840, 6190321, 6543240, 6484645, 6007506, 6190311, 6189348, 6189342, 6394841, 6189376, 6023100, 6023102, 5870078, 6394822, 6189351, 6323617, 6323616, 6538309, 6189395, 6190255, 6190288, 6190276, 6170126, 6484594, 6484611, 6394845, 6543243, 6190260, 6484643, 6543459, 6007654, 6484637, 6394724, 6354355, 6354366, 6538308, 6538295, 6538318, 6543251, 5868103, 5868136, 6484647, 6169798, 6538299, 6543458, 6538290, 6489183, 6354338, 6543465, 6543474, 5868097, 5868468, 6489188, 6189364, 6489182, 6484623, 6454879, 6394733, 6484610, 5868164, 6543234, 5868147, 6454875, 6484616, 6538305, 6023926, 6538306, 6543463, 6503946, 6454895, 6454892, 6167514, 6007495, 6007505, 6454874, 6394732, 6543216, 6190337, 6454896, 6543268, 6538307, 6538296, 6007637, 6538310, 6354367, 6543233, 6394743, 6484635, 6189315, 6009027, 6543452, 6170135, 6394730, 6538314, 6394818, 6007663, 6543279, 6394746, 6190250, 5870085, 6489165, 6543426, 5870071, 6167527, 5868095, 6167529, 6189347, 6394829, 6489163, 6023120, 6189319, 5869607, 6170140, 6543227, 6354352, 6189374, 6354342, 6394803, 6169800, 6454891, 6394817, 5868477, 6484633, 5868149, 6008322, 6394805, 6190347, 6190253, 6354345, 6190251, 5868486, 6543260, 6484638, 5868170, 6009031, 6190332, 6189370, 6190361, 6023917, 6484593, 6354375, 6354371, 6009015, 5869616, 6484622, 6189385, 6189334, 6454882, 6189390, 6170122, 6007486, 6394814, 6167521, 6543432, 6009022, 6543454, 6543274, 6454890, 6394752, 6323600, 6484600, 5868131, 5868138, 6484605, 5869615, 6190350, 5870062, 6543476, 6007661, 6167526, 6489190, 6489178, 5869613, 6484596, 6489164, 6394833, 6023103, 6543228, 6189346, 6543435, 6543231, 6538316, 6543255, 6189339, 6007492, 6454893, 6354336, 6190280, 6008316, 6190327, 6543444, 6007636, 6354349, 5868168, 6503951, 6543271, 6543232, 6394756, 5870080, 6190360, 6543262, 6489147, 5869618, 6503945, 6323605, 5868124, 6190295, 6394819, 6454883, 6394827, 6543447, 5868108, 6394835, 6394842, 6189377, 6023104, 5868172, 5869625, 6008324, 6394826, 6354341, 6489171, 6394739, 6007638, 6009030, 6007653, 5868485, 6394720, 6190310, 6190286, 6489184, 6190314, 6190285, 6484601, 6190340, 6008335, 6007648, 6007644, 6543263, 5868101, 7031181, 7031232, 6935275, 6689881, 6786315, 6698836, 6898633, 6899587, 6808747, 6970241, 6970239, 7016613, 6898643, 7031202, 7031212, 7031258, 7031167, 7049043, 7031259, 7016594, 7049042, 7049033, 6898649, 6689880, 7016582, 6455343, 6689874, 7024567, 6899096, 7031318, 6554100, 6924766, 6846891, 6924881, 7049068, 6808750, 6924851, 6689876, 6555003, 6898646, 6899098, 6924771, 7031290, 6812629, 6554987, 7031283, 7057207, 6689885, 6935260, 6898655, 6609757, 6609751, 6554954, 6924733, 6924878, 6898625, 7016573, 6924848, 6899596, 7016596, 6455334, 6924875, 6924764, 6935300, 6554102, 6689888, 6554968, 6899591, 6689875, 6924773, 6924737, 6924768, 6924736, 6924755, 6924871, 6935276, 6935245, 6973663, 6832171, 6924864, 6689902, 7031206, 6554985, 7016612, 6924892, 6935280, 7031234, 6554087, 6812620, 6698800, 7024576, 6846916, 7049027, 6555007, 6899581, 6689900, 6846906, 6899580, 6924734, 6924896, 7031211, 6846900, 6935350, 6554957, 6899100, 6924756, 6609725, 6554993, 6555001, 6609737, 6846884, 7031285, 6898627, 6935256, 6924752, 6935340, 7057183, 7057186, 7031265, 6973681, 6924731, 6554086, 6846912, 7024558, 7024584, 7031205, 6808744, 6609761, 6924866, 7031280, 7016574, 7031306, 6698785, 6846904, 7049051, 6698802, 6786311, 6808771, 6924757, 6455354, 6812619, 6924860, 6698813, 7049028, 6554988, 6609755, 6808768, 6899099, 6899604, 6899597, 6935291, 6935348, 6924867, 6970251, 6808759, 7057208, 7031287, 6924744, 7023956, 6899583, 7049040, 7049083, 7031309, 7031164, 6554062, 7049093, 6554088, 6899083, 6812648, 6846863, 6832177, 6698806, 6455329, 6846890, 6554965, 6812606, 7031245, 7049089, 6924749, 7016597, 6455356, 6689924, 6808742, 6924750, 6970238, 6698779, 6812637, 6898630, 6924903, 6924882, 6973669, 6935308, 6698810, 6698781, 7024564, 7016583, 6924779, 6698794, 6812621, 6698787, 6554994, 6846883, 6554078, 6832170, 7023949, 7031201, 7049082, 6555015, 6924751, 6812603, 6935247, 6812655, 7031253, 6786318, 6689930, 6973670, 6935284, 6935312, 6555010, 6924884, 6846892, 6554973, 7049079, 6935344, 6554990, 6924741, 6812614, 6935309, 6846870, 6689892, 7057231, 7049057, 6812658, 6698783, 7016571, 6846885, 6808778, 7057201, 7031228, 7031269, 7016591, 6554060, 6554081, 6698801, 6812626, 6808762, 6846886, 6846860, 6846873, 6898650, 6846901, 6924758, 6924776, 6832178, 6924767, 6924858, 6935298, 6935316, 6924774, 7016593, 7016579, 7016602, 7016586, 7024581, 7016607, 6899599, 7049088, 7024577, 6658242, 6808770, 6812622, 7031238, 6555008, 7016615, 7016581, 7031171, 6899605, 6698838, 7023943, 6832172, 7031184, 7057196, 7057203, 7057220, 7057193, 7016592, 6812602, 6812633, 6846859, 7016617, 6935244, 6924845, 6808776, 6609753, 6812659, 6554092, 6935358, 6935326, 6935311, 6554045, 7031217, 7049086, 6846910, 6899103, 7049050, 6554959, 6455336, 6554099, 7031241, 6689894, 6698809, 6846867, 7016572, 7031295, 7057218, 6554101, 6698820, 6899112, 6812650, 6935287, 6924874, 7024582, 7016611, 7024583, 6924775, 7049066, 6808743, 6808752, 7031242, 6609734, 7057212, 6689919, 6924869, 6554999, 6808763, 6812605, 6924765, 6554073, 6973687, 6924763, 6924787, 7049078, 6554063, 6973664, 6609758, 6698835, 6812615, 6812623, 6924728, 6973667, 7049087, 7057202, 6689903, 7024578, 6698778, 6812604, 6935313, 6935269, 6609754, 7049029, 7057222, 6898652, 6935337, 6554980, 6554963, 6554951, 7031291, 7031190, 7031297, 6698840, 6808780, 6808754, 6899593, 6935294, 6970242, 6973657, 6924901, 7031284, 7057182, 7049063, 7057223, 7049024, 6899576, 6554049, 6609764, 6898624, 6898645, 6899588, 6609769, 6846895, 6698771, 6898647, 6899586, 6935274, 7031226, 6935346, 7031227, 7049032, 7049092, 7049062, 6970243, 6935321, 6689916, 6812639, 6973660, 6898632, 6698803, 7057237, 6924888, 7024571, 6698818, 6935264, 6924897, 6935259, 7016609, 7057228, 6808746, 6935293, 6808758, 6935336, 6609747, 6554059, 6554981, 6898648, 6689886, 6973654, 7016616, 7031220, 7031163, 7031246, 6455349, 6924748, 6812609, 6455355, 6924846, 6808753, 6554976, 6689904, 6808772, 6846915, 6935250, 7016603, 6808781, 6554064, 6698784, 6698837, 6808766, 6689907, 6924862, 6846882, 6808751, 6554960, 6924854, 7031277, 7057241, 6698798, 6808774, 7031188, 6698832, 6555004, 6555000, 6846913, 6658243, 6970252, 6935347, 7031302, 6973674, 6554067, 6786320, 6846879, 6846914, 6898623, 6935343, 6846903, 6924868, 7016577, 7031215, 7031176, 6935251, 7016585, 7031312, 6555012, 6924784, 6924781, 6935318, 6812635, 6689920, 6899582, 6924885, 7024594, 6555014, 6808765, 6846907, 6899601, 7031237, 7057187, 7023970, 6554964, 6935297, 6924856, 7031311, 6658241, 6924850, 6554057, 6935296, 6554048, 7031307, 7057226, 6899579, 6812654, 6689910, 6554090, 6808769, 7031236, 6698839, 6698834, 7049071, 6554091, 6554944, 6786307, 7023940, 6924745, 7024569, 7031174, 6935277, 6812607, 6832184, 6554070, 6846899, 6846902, 6973689, 6970254, 6899094, 6554949, 6846874, 6554051, 6935286, 7024595, 6924852, 6808745, 6812624, 6924777, 6935270, 7057206, 6786312, 6786322, 6924895, 6924899, 7023933, 7031178, 7023937, 7023938, 7031193, 7031207, 7031218, 7031299, 7031304, 7016600, 6554053, 6698786, 6689928, 6899102, 6832169, 6935342, 6689932, 6924902, 6832174, 6935278, 6689912, 6689917, 6698790, 7023952, 6898656, 6935310, 6555020, 6609748, 6899109, 6812651, 6698775, 7031315, 6698814, 6609760, 6935288, 6812640, 6454514, 6454529, 6454524, 6454538, 6454525, 6454523, 6454507, 6454537, 6454532, 6454508, 6454535, 6454521, 6454519, 6454528, 6454516, 5092470, 5032833, 6189398, 6484620, 5870069, 6009029, 6354339, 6170133, 5868479, 6009032, 6538294, 6538298, 6484595, 6538311, 6023916, 5868141, 6170124, 6394810, 6484636, 6189397, 6170134, 6454899, 6023106, 6543245, 6167528, 6394816, 6009043, 6543481, 6454871, 6190293, 6454897, 6167535, 6394811, 6189392, 6543448, 6023105, 6543450, 6189337, 6190317, 6484614, 6394843, 6023918, 6189333, 6484629, 6394747, 6008339, 6538304, 6503949, 6170139, 6167530, 6394801, 6354354, 6543480, 6394753, 6008312, 6167536, 6503952, 5869604, 6543470, 6323609, 5868470, 6484613, 6484597, 6023124, 6008330, 6543238, 6454888, 6354362, 6023936, 6484648, 5868110, 6454878, 6189332, 6484592, 6023122, 6394831, 6189313, 6543241, 6484628, 6008337, 5870087, 6190296, 6484602, 6170141, 5869606, 6023904, 6170137, 6543434, 6354370, 6190252, 6023930, 6489193, 6190259, 6543433, 6323592, 6454876, 6190364, 6484618, 6484630, 6543226, 5870065, 5870075, 6543471, 5868094, 6189345, 6190357, 6543455, 6023942, 6543266, 6189341, 6189350, 6484642, 7016619, 7031166, 6935248, 7057200, 7049054, 6970247, 6846897, 7016576, 6554982, 7049065, 6812632, 6846888, 7023957, 7049067, 7049039, 6924891, 6846908, 6970253, 7016587, 7031254, 6924738, 7031170, 7031300, 7031257, 6689909, 7024560, 6808764, 7016604, 6609743, 6935303, 7049052, 6935345, 7024580, 6689878, 7024566, 6899577, 7049085, 6698796, 7049041, 7024570, 7057197, 7031310, 6658245, 6935330, 6698792, 7031191, 6455358, 7049046, 7023953, 6698811, 6935262, 6924870, 7016618, 6698770, 6935257, 6698817, 6899608, 7049034, 6698807, 6812627, 6935325, 7016601, 7016605, 7016580, 7057221, 6935332, 6924883, 6935349, 6786317, 7031267, 6786305, 7049048, 7031270, 7031288, 6935271, 6554104, 7049091, 7031298, 6973676, 7049022, 7057195, 6786321, 6554970, 6924743, 6899589, 6924900, 7031260, 6973671, 6554096, 6554946, 6846889, 6924742, 7057224, 6554983, 7016569, 7016589, 7031272, 6973661, 6554953, 6786319, 6846893, 6935282, 6924877, 6935253, 7057205, 7024586, 6935335, 7031293, 6899114, 7031203, 6786310, 7024565, 6609768, 6924904, 7049055, 6698763, 6924894, 7024575, 7057229, 7031317, 7016595, 7023934, 7057213, 7016575, 6924759, 6698827, 7057227, 7024563, 6935305, 7057214, 7024574, 6786306, 7057204, 7049072, 6924747, 7023967, 7049053, 6898626, 7049037, 7049058, 7057219, 7031276, 7057230, 6808757, 6846896, 6808756, 6689882, 7031244, 7031316, 7016584, 7031271, 7049031, 7016570, 7031305, 6786308, 6554991, 6554083, 7023958, 6808775, 7031273, 6924761, 7057185, 6924739, 7057198, 6898639, 7031262, 6973665, 7024587, 6698804, 7031233, 6609729, 6898640, 7016588, 6899594, 6846909, 7057225, 7057188, 7031177, 7031303, 6935268, 7031197, 6698808, 6935328, 7016606, 6935306, 7057211, 7031301, 7057238, 6808773, 7031247, 6846894, 6554043, 7057209, 6455340, 7031210, 6924886, 6924879, 7024573, 6935331, 6698767, 7057235, 6924857, 6846917, 7057215, 6899578, 6924782, 6924873, 6899607, 7023950, 7049059, 6554074, 7023955, 7057216, 6935329, 6555013, 7023951, 7016621, 6554966, 6935281, 7031169, 6455353, 7031313, 7049047, 6609740, 7031172, 7031196, 6812643, 6455345, 7031264, 6970246, 7031289, 7031186, 7057232, 6935292, 6832181, 7016614, 7016622, 7016610, 6935327, 6935334, 6970244, 6935304, 6924753, 6898654, 7031278, 7049074, 6935324, 6832186, 6832176, 6812634, 6898651, 6698791, 6935333, 6832183, 6935353, 6832175, 6935320, 6609738, 6832179, 6832182, 6698765, 6935302, 6832173, 6935301, 6924754, 6970255, 6832180, 6924889, 6698833, 6832185, 6898631, 6812630, 7031175, 6554075, 6698815, 6846858, 6554046, 6935341, 7016578, 7057210, 6924880, 6689887, 6698826, 7024593, 7024589, 6454506, 6454531, 6454534, 6454515, 6454511, 6454526, 6169796, 6009024, 6007484, 6484615, 5870061, 6189343, 6189320, 6489177, 6543258, 5869621, 6538297, 6484624, 6169777, 6008327, 6454885, 6190275, 5870067, 6023112, 6189361, 5868465, 6394839, 6190308, 6394761, 6189371, 6484646, 6007509, 6484619, 6189340, 6484590, 6007651, 6538289, 6169780, 6190354, 6538302, 6484604, 6190313, 6394815, 6454877, 6354350, 6007499, 6167518, 6354373, 6023108, 6503953, 6007664, 6023931, 6489152, 6538293, 6489189, 6503948, 6190334, 6484603, 6484621, 6023941, 6484639, 6190343, 6543441, 6394749, 6812610, 6924855, 7031243, 6554085, 6554065, 6935323, 6924861, 6935317, 6935267, 7023932, 6899101, 6973655, 6898637, 7024562, 7023971, 6924780, 6455333, 7049026, 6846869, 6935241, 7023944, 7023946, 6898634, 6898641, 6812645, 7024579, 6898638, 6973659, 7057236, 7031198, 6970250, 7031308, 6698821, 6554072, 7031192, 6609735, 6554055, 7023966, 6973656, 7049045, 6808779, 6812613, 6812661, 7031268, 6698773, 7031281, 6973684, 6935258, 6973690, 6554995, 6924732, 6846878, 6935357, 6689899, 6898642, 7031183, 6658239, 7057190, 7031235, 6924893, 6808761, 7024588, 7031252, 7031266, 6555002, 6935273, 6970257, 6846887, 6698829, 7057194, 7031274, 7031229, 7057217, 6924890, 7031213, 7023941, 6812616, 6832188, 7049025, 6924876, 6808748, 7049080, 6455330, 6554080, 6698805, 7049075, 6554958, 6935252, 6812647, 6924785, 6554068, 6899598, 6924730, 7031194, 6832187, 7057199, 7016599, 7024585, 7031314, 7031185, 6924859, 7024597, 7031282, 7031168, 7049081, 7057184, 7016620, 7049056, 7049060, 7023954, 7031182, 6609762, 6698776, 7031160, 6846866, 7024591, 6924898, 6454502, 6454510, 6454884, 6609749, 7031195, 7049094, 7031249, 6973688, 6554103, 6698795, 6609767, 6846862
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Equinoxe 320-36-00Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented2 recalls27 reports
- Equinoxe 320-36-03Exactech, Inc. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented2 recalls6 reports
Product lines
Other devices under the same FDA submission
3 devices cleared under K093275 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Equinoxe 320-36-10Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented2 recalls1 report
- Equinoxe 320-36-13Exactech, Inc. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented2 recalls1 report
- Equinoxe 320-01-36Exactech, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented7 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 94092).
- Z-1402-2024 · Open, ClassifiedExactech Equinoxe GLENOID,KEELED, CEMENTED, Mates with 38, 41, 44 head, alpha curvature, Sizes: a), Small, Item Number 314-01-02; b) Medium, Item number 314-01-03; c) Large, Item …
- Z-1403-2024 · Open, ClassifiedExactech Equinoxe GLENOID,KEELED, CEMENTED, mates with 47, 50, 53 head, beta curvature, Sizes: a) Small, Item Number 314-01-12, b) Medium, Item Number 314-01-13, c) Large, Item Nu…
- Z-1404-2024 · Open, ClassifiedExactech Equinoxe GLENOID,KEELED, CEMENTED, mates with 47, 50, 53 head, beta curvature, Sizes: a) Small, Item Number 314-02-02, b) Medium, Item Number 314-02-03, c) Large, Item Nu…
- Z-1405-2024 · Open, ClassifiedExactech Equinoxe GLENOID,KEELED, CEMENTED,47, 50, 53 head, beta curvature, Sizes: a) Small, Item Number 314-02-12, b) Medium, Item Number 314-02-13, c) Large, Item Number 314-02…
- Z-1406-2024 · Open, ClassifiedExactech Equinoxe GLENOID, POSTERIOR AUGMENT,PEGGED, CEMENTED, 8o, LEFT, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-02-22, b) 41, 44, 47, 50 head, Medium, Item …
- Z-1407-2024 · Open, ClassifiedExactech Equinoxe GLENOID, POSTERIOR AUGMENT,PEGGED, CEMENTED, 8o, RIGHT, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-02-32, b) 41, 44, 47, 50 head, Medium, Item…
- Z-1408-2024 · Open, ClassifiedExactech Equinoxe GLENOID,POSTERIOR AUGMENT, PEGGED, 12o, LEFT, FOR CEMENTED USE ONLY, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-04-22, b) 41, 44, 47, 50 head, …
- Z-1409-2024 · Open, ClassifiedExactech Equinoxe GLENOID,POSTERIOR AUGMENT, PEGGED, 12o, RIGHT, FOR CEMENTED USE ONLY, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-04-32, b) 41, 44, 47, 50 head,…
- Z-1410-2024 · Open, ClassifiedExactech Equinoxe GLENOID,POSTERIOR AUGMENT, PEGGED, 16o, LEFT, FOR CEMENTED USE ONLY, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-06-22, b) 41, 44, 47, 50 head, …
- Z-1411-2024 · Open, ClassifiedExactech Equinoxe GLENOID,POSTERIOR AUGMENT, PEGGED, 16o, RIGHT, FOR CEMENTED USE ONLY, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-06-32, b) 41, 44, 47, 50 head,…
- Z-1412-2024 · Open, ClassifiedExactech Equinoxe CAGE GLENOID, CEMENTED, alpha curvature, Mates with: a) 38, 41, 44 head, Small, Item Number 314-13-02, b) 38, 41, 44 head, Medium, Item Number 314-13-03, c) 3…
- Z-1413-2024 · Open, ClassifiedExactech Equinoxe CAGE GLENOID, CEMENTED, beta curvature, Mates with: a) 47, 50, 53 head, Medium, Item Number 314-13-13, b) 47, 50, 53 head, Large, Item Number 314-13-14, c) 47…
- Z-1414-2024 · Open, ClassifiedExactech Equinoxe CAGE GLENOID,POSTERIOR AUGMENT, CEMENTED, Left, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-13-22, b) 41, 44, 47, 50, head, Medium, Item Number …
- Z-1415-2024 · Open, ClassifiedExactech Equinoxe CAGE GLENOID,POSTERIOR AUGMENT, CEMENTED, Right, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-13-32, b) 41, 44, 47, 50, head, Medium, Item Numbe…
- Z-1417-2024 · Open, ClassifiedExactech Equinoxe REVERSE SHOULDER,36mm Constrained Humeral Liner: a) +0mm, Item Number 320-36-10, b) +2.5mm, Item Number 320-36-13; Shoulder Arthroplasty
- Z-1418-2024 · Open, ClassifiedExactech Equinoxe REVERSE SHOULDER, 38mm Humeral Liner: a) +0mm, Item Number 320-38-00, b) +2.5mm, Item Number 320-38-03; Shoulder Arthroplasty
- Z-1419-2024 · Open, ClassifiedExactech Equinoxe REVERSE SHOULDER,38mm Constrained Humeral Liner: a) +0mm, Item Number 320-38-10, b) +2.5mm, Item Number 320-38-13, Shoulder Arthroplasty
- Z-1420-2024 · Open, ClassifiedExactech Equinoxe Humeral Liner: Item Numbers: a) 320-40-00, b) 320-40-03; Shoulder Arthroplasty
- Z-1421-2024 · Open, ClassifiedExactech Equinoxe Constrained Humeral Liner: Item Numbers: a) 320-40-10, b) 320-40-13, Shoulder Arthroplasty
- Z-1422-2024 · Open, ClassifiedExactech Equinoxe REVERSE SHOULDER,42mm Humeral Liner, a) +0mm, Item number 320-42-00, b) +2.5mm, Item Number 320-42-03; Shoulder Arthroplasty
- Z-1423-2024 · Open, ClassifiedExactech Equinoxe REVERSE SHOULDER,42mm Constrained Humeral Liner, a) +0mm, Item number 320-42-10, b) +2.5mm, Item Number 320-42-13; Shoulder Arthroplasty
- Z-1424-2024 · Open, ClassifiedExactech Equinoxe REVERSE SHOULDER,46mm Humeral Liner, a) +0mm, Item number 320-46-00, b) +2.5mm, Item Number 320-46-03; Shoulder Arthroplasty
- Z-1425-2024 · Open, ClassifiedExactech Equinoxe REVERSE SHOULDER,46mmConstrained Humeral Liner, a) +0mm, Item number 320-46-10, b) +2.5mm, Item Number 320-46-13; Shoulder Arthroplasty
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page