Equinoxe recallZ-1403-2024 · Exactech, Inc. · initiated Mar 6, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Exactech, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).
FDA root cause category: Nonconforming Material/Component
What the company is doing
Beginning 03/06/2024, Exactech issued an URGENT MEDICAL DEVICE RECALL notice and an URGENT DEAR HEALTHCARE PROFESSIONAL COMMUNICATION to its consignees via email. The recall notice explained the issue, clinical impact, and requested the following actions be taken: "" CAREFULLY REVIEW THIS RECALL NOTIFICATION to ensure that you fully understand the issue identified, the recall strategy, and all actions required. " IMMEDIATELY IDENTIFY AND QUARANTINE any of the subject devices in your inventory and/or customer s inventory listed Product Scope Listing. " EXTEND THE DESCRIPTION OF ISSUE AND CLINICAL IMPACT as described in the recall notification to your accounts that may have this product in their possession. " COMPLETE AND RETURN the attached Recall Inventory Response Form to Exactech via email at [email protected]. The communication to healthcare providers referenced the FDA Safety Communication issued on 01/16/2024, explained the issue, potential health risks, and suggested patient monitoring for those with implanted affected devices for potential device wear, failure, or bone loss and should consider performing X-rays to further evaluate the patient and their implanted device if you suspect a failed device. Contact your local Exactech Representative if you have any questions regarding this communication.
Recall details
- Recall number
- Z-1403-2024
- Status
- Open, Classified
- Initiated by the company
- Mar 6, 2024
- Posted by the FDA
- Mar 29, 2024
- Recalling firm
- Exactech, Inc.
- Quantity
- 2008 units
- Product code
- KWS
- FDA submissions
- K042021
- Distribution
- Nationwide and Australia, Austria, Belgium, Brazil, Chile, Canada , China, Colombia, France, Germany, Great Britain, Greece, Guatemala, India, Italy, Japan, Korea, Luxembourg, Malaysia, Netherlands, New Zealand, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand
Product description
Exactech Equinoxe GLENOID,KEELED, CEMENTED, mates with 47, 50, 53 head, beta curvature, Sizes: a) Small, Item Number 314-01-12, b) Medium, Item Number 314-01-13, c) Large, Item Number 314-01-14, Shoulder Arthroplasty
Code information (models, lots, serial numbers)
Show the full code information (18k characters)
a) Item Number 314-01-12, UDI/DI 10885862084439, Serial Numbers: 0744170, 0756603, 0744142, 0744169, 0744178, 0744141, 0744152, 0744182, 0756566, 0756592, 0756611, 0744134, 0744135, 0756591, 0744177, 0756567, 0756582, 0756606, 0756596, 0744179, 0744159, 0744145, 1117515, 1117547, 1117537, 1117517, 1117520, 1117548, 1117510, 1117543, 1117529, 1117502, 1117535, 1117499, 1117508, 1117546, 1117503, 1117528, 1117536, 1117506, 1117545, 1117521, 1117505, 1117512, 1117516, 1272400, 1361060, 1361065, 1272398, 1272401, 1272392, 1272403, 1272408, 0756602, 1117540, 1392234, 1350740, 1350736, 1361070, 1361071, 1392236, 1416355, 1428159, 1428162, 1428163, 1574777, 1574781, 1574788, 1574792, 1574778, 1574794, 1350744, 1574793, 1574780, 1574784, 1361073, 0744144, 1117523, 1117518, 1416362, 1574797, 1428168, 1428172, 1416358, 1392244, 1574795, 1428158, 1272409, 1272396, 1574799, 1574779, 1793114, 1633774, 1734408, 1703819, 1793343, 1793353, 1793350, 1635466, 1703814, 1734399, 1734402, 1793101, 1793095, 1793096, 1793344, 1793348, 1793367, 1793055, 1793355, 1793375, 1793369, 1793374, 1734422, 1633771, 1635464, 1633769, 1633775, 1633779, 1794653, 1793105, 1633777, 1392247, 1350735, 1416357, 1361066, 1361074, 1703820, 1633776, 1793341, 1793067, 1917888, 1917899, 1793100, 1917856, 1917895, 1734417, 1984866, 1703810, 1794659, 1793363, 1734413, 2174254, 1117526, 1384121, 1272405, 1416356, 1793365, 1793357, 1917892, 2017327, 1793043, 1793362, 1793366, 1794663, 1793046, 1794657, 1793098, 1984870, 2174258, 2174251, 2284394, 2284397, 2284392, 2174252, 2393054, 2404334, 2404335, 2404339, 1350731, 1416351, 1793051, 1793104, 1793099, 2536818, 2536808, 2536813, 2679975, 2679986, 2679991, 2679993, 2752696, 2752700, 1793097, 2017338, 1917865, 1793057, 1793059, 1734419, 2017326, 1917872, 2284399, 2469226, 2491265, 2508022, 2508031, 2536783, 2679981, 2752686, 2752684, 2980608, 2980607, 2980602, 2993861, 2980606, 2980609, 2993855, 2993850, 2993858, 3740570, 2679984, 2993867, 2017335, 2536782, 2536781, 2993859, 2993863, 2993869, 2993854, 3740565, 3740562, 3740568, 3740569, 2993865, 2993862, 2980603, 2980604, 2980605, 2993856, 2679977, 3950947, 3950892, 4092056, 2393056, 2752672, 4087712, 4087713, 4304826, 3950950, 4087706, 2752681, 2993857, 2993852, 4468485, 4912816, 4912811, 3740560, 4306258, 4304816, 4306265, 4468494, 4087725, 4087729, 4092059, 3950940, 4884357, 4912814, 4092061, 3950949, 3950963, 4304823, 4306261, 4468489, 4304820, 4304828, 4912806, 4971034, 4884356, 4971036, 4304818, 4468496, 4304808, 4304815, 4087726, 4971033, 4884355, 4884358, 4912808, 4912815, 4306263, 4912819, 6801011, 6800984, 6800988, 6800990, 6800998, 6801008, 6801009, 6801003, 6815714, 6801015, 6800999, 6815709, 6815706, 6801007, 6815703, 6815710, 6815711, 6815715, 6800995, 6800997, 6801017, 6801004, 6801016, 6815698, 6800987; b) Item Number 314-01-13, UDI/DI 10885862084446, Serial Numbers: 0742615, 0742397, 0742650, 0742412, 0742485, 0746496, 0742498, 0742608, 0742432, 0742423, 0742513, 0742616, 0742478, 0742488, 0742489, 0742495, 0742480, 0742494, 0742411, 0742407, 0742398, 0742399, 0742641, 0742516, 0742517, 0768435, 0742393, 0742405, 0742511, 0742637, 0742505, 0742627, 0742438, 0742619, 0742606, 0742609, 0742395, 0742621, 0742618, 0742476, 0742520, 0768458, 0742426, 0742428, 0768443, 0742508, 0742496, 0742429, 0742420, 0742439, 0742612, 0742418, 0768462, 0742493, 0742515, 0742408, 0742640, 0742409, 0742416, 0742396, 0742472, 0768441, 0742430, 0746540, 0742521, 0742503, 0768429, 0746542, 0746531, 0746507, 0742514, 0768440, 0742636, 0742431, 0742622, 0742506, 0768478, 0768430, 0768425, 0742605, 0742436, 0742486, 0742509, 0768448, 0768416, 0742406, 0746519, 0742504, 0746502, 0746522, 0746546, 0742646, 0746491, 0746534, 0746504, 0742647, 0746529, 0742404, 0746524, 0768419, 0746526, 0742624, 0746520, 0746543, 0768479, 0742638, 0746494, 0746518, 0746497, 0746523, 0768413, 0746513, 0746539, 1127270, 0742623, 0742497, 0746541, 0746544, 0746516, 0742649, 0768414, 0746500, 0742630, 0742392, 0746501, 0768453, 0742440, 0768446, 0742628, 0768417, 0746525, 0768468, 0746511, 0768463, 0746505, 0746545, 0768452, 0742419, 1127266, 1127249, 1127265, 1127221, 1127276, 1127282, 1127199, 1127242, 1127241, 1127222, 1127261, 1127259, 1127267, 1127239, 1127277, 1127288, 1127217, 1127260, 1127212, 1127214, 1127245, 1127223, 1127235, 1127274, 1127283, 1252227, 1252140, 1252149, 1252154, 1252152, 1252141, 1252148, 1252195, 1252139, 1252168, 1252178, 1252172, 1252188, 1252133, 1252165, 1252218, 1252175, 1252179, 1252187, 1252189, 1252197, 1252211, 1252216, 1252224, 1252228, 1252236, 1252203, 0742434, 0742607, 0746515, 0746533, 0742610, 0768431, 0746537, 0768472, 0768421, 0768466, 0768432, 0768476, 0768461, 0742512, 0768451, 1127197, 1127269, 1127224, 1127198, 1127263, 1127268, 1127209, 1127231, 1127240, 1127201, 1428189, 1428213, 1252184, 1252166, 1252151, 1252225, 1372221, 1252234, 1428090, 1428089, 1372234, 1428208, 1252198, 1252231, 1252193, 1252150, 1372210, 1252206, 1252162, 1552833, 1252173, 1428092, 1428096, 1372206, 1252159, 1428093, 1252177, 1428202, 1252157, 1428095, 1428201, 1252180, 1372205, 1428211, 1252207, 1428209, 1252146, 1252238, 1428180, 1428198, 1372211, 1552827, 1552828, 1552830, 1552838, 1552839, 1552845, 1552847, 1574855, 1574859, 1574861, 1574869, 1574873, 1574874, 1552829, 1574862, 1574868, 1428204, 1428186, 1574856, 1252230, 1252232, 1372233, 1574871, 1552824, 1252160, 0768437, 0768444, 0742519, 0768459, 0742614, 1127250, 1252155, 1428206, 1428094, 1252229, 1428185, 1252156, 1252167, 1252164, 1428215, 1552834, 1428200, 1574858, 1428091, 1252220, 1428199, 1252143, 1574866, 1552848, 1552841, 1574863, 1252181, 1372219, 1252171, 1574857, 1428212, 1372231, 1552843, 1252223, 1428218, 1428217, 1252137, 1372208, 1428219, 1630175, 1715062, 1630158, 1715068, 1831880, 1630176, 1630169, 1673732, 1757619, 1630161, 1673728, 1630167, 1757642, 1757618, 1926220, 1673707, 1630168, 1834458, 1757628, 1757620, 1673710, 1807916, 1673712, 1807909, 1673720, 1630170, 1630171, 1718017, 1807903, 1673706, 1673718, 1757625, 1757637, 1630166, 1757638, 1630173, 1757611, 1757658, 1780276, 1780281, 1780291, 1780292, 1780294, 1757613, 1807910, 1807920, 1807924, 1807925, 1757644, 1915707, 1915709, 1915716, 1926222, 1673724, 1673727, 1673725, 1715054, 1715058, 1780288, 1718016, 1780284, 1807907, 1807911, 1807919, 1834444, 1834449, 1834450, 1780280, 1757632, 1757646, 1780277, 1780286, 1715056, 1757631, 1757648, 1715051, 1757622, 1630172, 1757617, 1127289, 1127256, 1127281, 1127286, 1127275, 1127219, 1127271, 1252221, 1372223, 1372218, 1252200, 1372209, 1252192, 1552837, 1372215, 1252136, 1252190, 1252153, 1428203, 1372227, 1428210, 1252134, 1252163, 1252169, 1252196, 1252202, 1252204, 1252215, 1252222, 1372224, 1552836, 1428190, 1630165, 1757653, 2043635, 2043620, 2043640, 1715069, 1956074, 1963710, 1757624, 1757615, 1831873, 1956075, 1831882, 1673730, 1673723, 1956069, 1956080, 1956084, 2043617, 1956064, 1963709, 2043618, 1834454, 1757635, 2043651, 1673715, 1926226, 1831877, 1926221, 1807926, 2043648, 1673731, 1915715, 1757649, 1757610, 1956073, 1834459, 1757623, 1715053, 1757630, 1956067, 2043645, 1956056, 1956082, 1715061, 1915706, 1915710, 1715066, 1926227, 1834447, 1956072, 1956085, 1807915, 1834442, 1926224, 1956061, 1956071, 1956078, 2043649, 2043629, 1757651, 1673709, 2043654, 1834439, 1807927, 1673729, 1807904, 1780293, 1715067, 1831878, 1963715, 1915712, 1757633, 2120786, 2120796, 2120814, 2116508, 2120800, 2120783, 2174521, 2174522, 2120804, 1127251, 1428188, 1552846, 1428192, 1372214, 1757657, 2043614, 1715059, 1715052, 2043652, 1956083, 2043631, 1831875, 1957587, 2043655, 1715065, 1834455, 1915705, 1673716, 1757641, 1780282, 2043636, 1956070, 1956076, 1834457, 2043643, 1956058, 1718023, 1780283, 1715050, 1926229, 2120807, 2116510, 2174520, 2120799, 2120810, 2251358, 2116512, 2251357, 2174519, 2174526, 2120780, 2120790, 2120809, 2174523, 2251371, 2251379, 2120795, 2251375, 2251376, 2299861, 2343436, 2343448, 2343433, 2469241.2469246, 2469234, 2536838, 2343441, 2299871, 2251374, 2120794, 2299862, 2116517, 2120801, 2120808, 2299875, 2343449, 2299860, 2469250, 2116518, 2116519, 2251360, 1574867, 2043621, 1715064, 1915714, 1915717, 1926225, 1807922, 1834441, 1718018, 1757640, 1757643, 1926223, 1956077, 2043642, 1673733, 2116514, 2116516, 2120803, 2469245, 2505875, 2343438, 2505885, 2505858, 2536840, 2469238, 2536843, 2536846, 2536851, 2536837, 2116509, 2505859, 2505856, 2343431, 2299873, 2343432, 2343435, 2536824, 2251373, 2505860, 2299877, 2299856, 2343440, 2536847, 2536829, 2629660, 2629669, 2648798, 2648807, 2648808, 2648792, 2648795, 2629700, 2629672, 2648801, 2648805, 2648796, 2648797, 2758586, 2758587, 2758589, 2648803, 2629702, 2648799, 2648806, 2648809, 2648793, 2648802, 2648804, 2758583, 2629693, 1831879, 1915711, 1831871, 1715049, 1834438, 2043641, 1718021, 1957589, 1757639, 1956068, 2536825, 2536848, 2469236, 2536839, 2120815, 2343450, 2536844, 2536823, 2536830, 2469259, 2536850, 2960184, 2960188, 2960189, 2978378, 2978374, 2629674, 2629675, 2629677, 2629687, 2629681, 2629664, 2629678, 2629685, 2629698, 2648791, 2978370, 2978380, 2960176, 2960181, 2960182, 2629665, 2978368, 2978376, 2629659, 3587048, 3587053, 3587058, 3587054, 2978379, 3587040, 3587056, 3587078, 3587044, 3587055, 3587069, 2978366, 2978372, 2978373, 2978367, 2960180, 2629699, 2629654, 2758588, 2960173, 2758590, 2758591, 2758584, 2629689, 2629701, 1956066, 2043622, 1956060, 2536842, 2536841, 2120781, 2505879, 2120787, 2251362, 2174517, 2505862, 2251367, 3587065, 2629667, 2853384, 2721361, 3587047, 2960174, 2960178, 3587060, 3587074, 3587051, 3587059, 3587063, 3587072, 3587079, 3587064, 3587073, 3587041, 3587043, 3587066, 3587075, 3587077, 2978377, 2629691, 2960183, 2758593, 2629682, 2978371, 2629670, 3587067, 3915901, 3915906, 3915911, 3963332, 3963334, 3963337, 3963338, 3963347, 3963336, 3963342, 3963351, 3963352, 3915899, 3915900, 3915907, 2299864, 2505882, 2467476, 2299869, 2505866, 2536836, 2343447, 2536831, 3587046, 2853379, 2853395, 2629657, 2721369, 2853391, 2629684, 3587068, 3587042, 2648794, 2629692, 2648790, 2629676, 3963312, 4368744, 4087674, 4368747, 3963329, 4368734, 4400436, 4400433, 4400393, 4400399, 4087687, 3915908, 3963348, 4368732, 3915905, 3963333, 2467477, 2629658, 2629673, 3587070, 2978369, 3587049, 2960172, 2960175, 3587045, 4087678, 4400434, 3963311, 4087693, 4087654, 4400427, 4400416, 3963353, 3963355, 4087680, 4400401, 4368733, 4903234, 4903226, 4903223, 4903231, 3587082, 3963331, 4087677, 4368746, 4400413, 4368735, 4087671, 4087672, 4087668, 4087700, 4400432, 3963315, 4368738, 4400391, 3915898, 4400398, 4087702, 3963308, 3963317, 4087695, 4087657, 4400408, 4087681, 4400429, 3963343, 4931752, 4903238, 4875887, 4903233, 4400424, 4087673, 4400409, 4368741, 4087655, 4087682, 4087660, 3963322, 4400428, 4368730, 4368731, 4931762, 4931755, 4931770, 4931754, 4903239, 4931760, 4931767, 4875873, 4875878, 4875872, 4875886, 4931756, 4931768, 6134622, 6134646, 4087662, 4087670, 4400422, 4875880, 4931769, 4875877, 4875871, 4875875, 6134717, 6134732, 6134715, 6134708, 6134721, 6134726, 6134734, 6134725, 6134674, 6134624, 6134672, 6134654, 4400418, 4368736, 4400423, 4368745, 4875882, 4931757, 4875874, 4875881, 6134712, 6134665, 6134647, 6134639, 6134633, 6134648, 6134650, 6134676, 6134720, 6883308, 6883305, 6883316, 6883303, 7056640, 6883314, 6883295, 6883325, 6883298, 6883299, 6883318, 4875884, 4931761, 6134658, 6134640, 6134636, 6134644, 6134657, 6134662, 6134669, 6883311, 6883328, 6883321, 6134722, 6883307, 6883304, 6883327, 7056652, 7056641; c) Item Number 314-01-14, UDI/DI 10885862084453, Serial Numbers: 0742666, 0742660, 0742675, 0742662, 0742696, 0742663, 0742686, 0743043, 0742659, 0742992, 0742678, 0742682, 0742999, 0743006, 0743041, 0742657, 0742658, 0742661, 0742697, 0742668, 0743005, 0743003, 0743012, 0903127, 0743039, 0742993, 0743018, 0742688, 0742990, 0742672, 0743020, 0743035, 0742680, 0743011, 0743026, 0742991, 0903170, 0903164, 0903150, 0743023, 0903135, 0743037, 0743008, 0903159, 0742654, 0742669, 0742997, 0742687, 0742698, 0743022, 0743038, 0742989, 0743033, 0903167, 0903117, 0903118, 0903157, 0903145, 0903169, 0903147, 0903149, 0903163, 0903121, 0743015, 0903154, 0742674, 0742653, 0903137, 0903122, 0903140, 0903144, 0743025, 0903162, 0903146, 0742651, 0743028, 0743010, 0743007, 0903165, 0903124, 0742699, 0742681, 0903155, 0903138, 0903161, 0903168, 0903160, 0742694, 0903119, 0743030, 0903156, 0903148, 0903151, 1131991, 0983702, 0983699, 0983700, 0983729, 0983720, 0983728, 0983734, 0983696, 0983718, 0983691, 0983694, 0983735, 0983725, 0983732, 0983739, 0983710, 0903123, 0742691, 0742995, 0903166, 0742670, 0742656, 0742685, 0742684, 0903120, 0743001, 0903136, 0743017, 0742679, 0903133, 0903126, 0903142, 0743036, 0743019, 0742998, 1131996, 1132027, 1132083, 1132026, 1132088, 1132046, 1132069, 1132029, 1132080, 1132044, 1132085, 1132018, 1132072, 1132042, 1132082, 1132073, 1132086, 1132009, 1132070, 1132062, 1132016, 1132032, 1131998, 1132043, 1132000, 1132008, 1132060, 1132067, 1132040, 1132054, 1132065, 1132019, 1132005, 1132047, 1132050, 0983713, 0983692, 0983714, 0983693, 0983706, 0983741, 0983717, 0983708, 0983715, 0983737, 0983703, 0983730, 0983705, 1267905, 1267970, 1267869, 1267908, 1267965, 1267906, 1267897, 1267951, 1267968, 1267875, 1267895, 1267917, 1267918, 1267933, 1267934, 1267947, 1267954, 1267962, 1267963, 0903158, 0743029, 0903131, 0903130, 0743002, 0743040, 0742690, 1132041, 1132052, 1132055, 1132059, 1131997, 1132068, 1132013, 1132038, 1132030, 1132087, 1132066, 1132075, 1132048, 1132077, 0983740, 0983742, 0983704, 1392648, 1428099, 1267929, 1392649, 1267967, 1267925, 1392651, 1428113, 1428111, 1392657, 1267876, 1428102, 1267871, 1267939, 1267912, 1267879, 1267969, 1555103, 1428105, 1267956, 1267916, 1428116, 1416370, 1267928, 1428103, 1428107, 1267915, 1416375, 1428115, 1392642, 1267904, 1416368, 1267935, 1267885, 1416376, 1267884, 1267890, 1267936, 1267941, 1416371, 1555102, 1555120, 1416385, 1555106, 1555116, 1555118, 0903134, 0743021, 0742692, 1132089, 1132025, 1132074, 1131999, 1132078, 1132056, 1131992, 1132053, 1132024, 1132063, 1267887, 1267882, 1392653, 1267950, 1416373, 1555105, 1267919, 1267896, 1548306, 1392659, 1267886, 1392652, 1392656, 1267921, 1267907, 1548302, 1555109, 1555117, 1555115, 1555111, 1267880, 1267911, 1267937, 1548298, 1267883, 1630118, 1765768, 1713906, 1630123, 1765756, 1713901, 1713909, 1912923, 1765779, 1765757, 1805348, 1912944, 1713899, 1713898, 1716697, 1716701, 1630115, 1765769, 1713894, 1630109, 1630121, 1713902, 1713914, 1765766, 1765777, 1805342, 1805352, 1805357, 1805363, 1805364, 1805367, 1765759, 1765767, 1805344, 1630122, 1713913, 1765778, 1765782, 1630116, 1713888, 1713915, 1713910, 1132028, 1132010, 1132011, 1132014, 1132035, 0983723, 1416379, 1416387, 1392650, 1267966, 1428110, 1428112, 1555110, 1555101, 1267961, 1267959, 1267923, 1555123, 1392646, 1416372, 1416383, 1555114, 1267926, 1267867, 1267900, 1428100, 1267955, 1416377, 1267957, 1555108, 1267881, 1416382, 1267866, 1267878, 1267888, 1267893, 1267894, 1267932, 1267943, 1267946, 1267948, 1267949, 1267964, 1765776, 1912926, 1834392, 1713905, 1805351, 1765781, 1805355, 1713907, 1957582, 1716704, 1805361, 1975033, 1805360, 1912932, 1805354, 1912935, 1912936, 1630126, 1912934, 1985482, 1834401, 1985492, 1975038, 1985487, 1834383, 1630112, 1805358, 1834403, 2017374, 1834402, 1957578, 2017378, 2017384, 1765764, 1805350, 1834400, 1975036, 1805347, 1985484, 1912943, 1765765, 1765755, 1713892, 1713908, 1765770, 1805371, 2118249, 2118240, 2118243, 2118259, 2147129, 2147107, 2193365, 2193371, 2193375, 2118241, 2118244, 2118254, 2147128, 1132007, 1132039, 1267874, 1267909, 1548300, 1392644, 1267899, 1267940, 1428117, 1428101, 1975044, 1630114, 1912933, 1975039, 1912924, 1805356, 1912938, 1912929, 2017371, 1805362, 1985490, 1765760, 1713890, 1716700, 1912941, 1716703, 1985489, 1985479, 2147112, 2288245, 2147102, 2147120, 2288258, 2147113, 2147127, 2147098, 2193369, 2193352, 2193358, 2118248, 2288265, 2338111, 2366599, 2366602, 2366627, 2366634, 2147114, 2193346, 2288261, 2338102, 2338104, 2338115, 2366615, 2366616, 2366624, 2433032, 2433040, 2338109, 2193344, 2193366, 2433031, 2433033, 2433039, 2147123, 2118242, 2147115, 2147131, 2147108, 2147103, 2288255, 2338113, 2118247, 2288252, 1548307, 1267960, 1267931, 1985481, 1713891, 1912928, 1713889, 1834386, 2017380, 2433038, 2506110, 2193367, 2433021, 2366631, 2536855, 2506113, 2288259, 2433043, 2506087, 2366612, 2366614, 2366613, 2147106, 2147118, 2147111, 2338108, 2193364, 2629805, 2552358, 2552359, 2552377, 2552386, 2552361, 2552360, 2552387, 1555112, 1912921, 1834393, 1834387, 1805343, 2017377, 1765758, 2338106, 2288263, 2147100, 2433024, 2118245, 2193363, 2506105, 2536879, 2536874, 2506090, 2366628, 2366617, 2366600, 2366630, 2506093, 2629778, 2552368, 3617061, 2629791, 2552378, 2552381, 2552362, 2552385, 2552389, 2552391, 2629784, 2989554, 2552363, 2629774, 2967051, 2967052, 2967053, 2629782, 2629788, 1912922, 1957580, 2366608, 2536865, 2536863, 2506101, 2506099, 2433027, 2506114, 2147119, 2288264, 2629777, 3617070, 3591939, 3591941, 3591946, 3617069, 3617077, 3617078, 2552374, 3591933, 3591936, 3617058, 3617067, 3617068, 3617073, 2989553, 2552392, 2989546, 3963403, 3963398, 3963402, 3963416, 2193377, 2118239, 2433035, 2536862, 2629798, 2552379, 2989545, 2629776, 3591937, 4127780, 4221474, 4221476, 4221466, 4221448, 4127778, 2433018, 3963415, 4221460, 4221457, 4221467, 4221482, 4221473, 3963408, 4127766, 4045774, 4854386, 4906782, 4906769, 4906765, 4906772, 4906779, 4906759, 4906783, 3617057, 3591945, 2629779, 2989552, 4045770, 4221459, 4221475, 4221456, 4221458, 4127789, 4045761, 4127769, 4045762, 4221471, 4221447, 4221455, 4221451, 4127768, 4221465, 4045773, 4906780, 4906771, 4854393, 4931779, 4906764, 4931773, 4906785, 4854395, 4854390, 4906777, 4221472, 4127775, 4221441, 4127786, 3963400, 4127777, 3963404, 4854397, 6134698, 6134688, 6134781, 6134778, 6134684, 6134680, 6134707, 4931774, 4906763, 4906776, 4906766, 6134690, 6134692, 6134787, 6134686, 6134693, 6134699, 6134704, 6134705, 6134792, 6134772, 6879883, 6879887, 7056745, 6134687, 6134771, 6879896, 6879898, 7056741, 6879888, 6134775, 6134779, 6134790, 6134776, 6879901, 6879894
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- Equinoxe 314-01-13Exactech, Inc. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall2 reports
- Equinoxe 314-01-12Exactech, Inc. · Prosthesis, Shoulder, Hemi-, Humeral, Metallic Uncemented1 recall0 reports
- Equinoxe 314-01-14Exactech, Inc. · Prosthesis, Shoulder, Hemi-, Humeral, Metallic Uncemented1 recall0 reports
Product lines
Other devices under the same FDA submission
69 devices cleared under K042021 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Equinoxe 300-01-11Exactech, Inc. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall44 reports
- Equinoxe 314-02-02Exactech, Inc. · Prosthesis, Shoulder, Hemi-, Humeral, Metallic Uncemented1 recall10 reports
- Equinoxe 314-02-14Exactech, Inc. · Prosthesis, Shoulder, Hemi-, Humeral, Metallic Uncemented1 recall8 reports
- Equinoxe 314-02-03Exactech, Inc. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall4 reports
- Equinoxe 314-02-13Exactech, Inc. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall4 reports
- Equinoxe 314-01-02Exactech, Inc. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall2 reports
- Equinoxe 314-02-04Exactech, Inc. · Prosthesis, Shoulder, Hemi-, Humeral, Metallic Uncemented1 recall1 report
- Equinoxe 314-01-03Exactech, Inc. · Prosthesis, Shoulder, Hemi-, Humeral, Metallic Uncemented1 recall0 reports
- Equinoxe 314-01-04Exactech, Inc. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall0 reports
- Equinoxe 314-02-12Exactech, Inc. · Prosthesis, Shoulder, Hemi-, Humeral, Metallic Uncemented1 recall0 reports
- Equinoxe 315-25-00Exactech, Inc. · PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED27 reports
- Equinoxe 300-01-09Exactech, Inc. · Prosthesis, Shoulder, Hemi-, Humeral, Metallic Uncemented26 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 94092).
- Z-1402-2024 · Open, ClassifiedExactech Equinoxe GLENOID,KEELED, CEMENTED, Mates with 38, 41, 44 head, alpha curvature, Sizes: a), Small, Item Number 314-01-02; b) Medium, Item number 314-01-03; c) Large, Item …
- Z-1404-2024 · Open, ClassifiedExactech Equinoxe GLENOID,KEELED, CEMENTED, mates with 47, 50, 53 head, beta curvature, Sizes: a) Small, Item Number 314-02-02, b) Medium, Item Number 314-02-03, c) Large, Item Nu…
- Z-1405-2024 · Open, ClassifiedExactech Equinoxe GLENOID,KEELED, CEMENTED,47, 50, 53 head, beta curvature, Sizes: a) Small, Item Number 314-02-12, b) Medium, Item Number 314-02-13, c) Large, Item Number 314-02…
- Z-1406-2024 · Open, ClassifiedExactech Equinoxe GLENOID, POSTERIOR AUGMENT,PEGGED, CEMENTED, 8o, LEFT, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-02-22, b) 41, 44, 47, 50 head, Medium, Item …
- Z-1407-2024 · Open, ClassifiedExactech Equinoxe GLENOID, POSTERIOR AUGMENT,PEGGED, CEMENTED, 8o, RIGHT, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-02-32, b) 41, 44, 47, 50 head, Medium, Item…
- Z-1408-2024 · Open, ClassifiedExactech Equinoxe GLENOID,POSTERIOR AUGMENT, PEGGED, 12o, LEFT, FOR CEMENTED USE ONLY, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-04-22, b) 41, 44, 47, 50 head, …
- Z-1409-2024 · Open, ClassifiedExactech Equinoxe GLENOID,POSTERIOR AUGMENT, PEGGED, 12o, RIGHT, FOR CEMENTED USE ONLY, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-04-32, b) 41, 44, 47, 50 head,…
- Z-1410-2024 · Open, ClassifiedExactech Equinoxe GLENOID,POSTERIOR AUGMENT, PEGGED, 16o, LEFT, FOR CEMENTED USE ONLY, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-06-22, b) 41, 44, 47, 50 head, …
- Z-1411-2024 · Open, ClassifiedExactech Equinoxe GLENOID,POSTERIOR AUGMENT, PEGGED, 16o, RIGHT, FOR CEMENTED USE ONLY, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-06-32, b) 41, 44, 47, 50 head,…
- Z-1412-2024 · Open, ClassifiedExactech Equinoxe CAGE GLENOID, CEMENTED, alpha curvature, Mates with: a) 38, 41, 44 head, Small, Item Number 314-13-02, b) 38, 41, 44 head, Medium, Item Number 314-13-03, c) 3…
- Z-1413-2024 · Open, ClassifiedExactech Equinoxe CAGE GLENOID, CEMENTED, beta curvature, Mates with: a) 47, 50, 53 head, Medium, Item Number 314-13-13, b) 47, 50, 53 head, Large, Item Number 314-13-14, c) 47…
- Z-1414-2024 · Open, ClassifiedExactech Equinoxe CAGE GLENOID,POSTERIOR AUGMENT, CEMENTED, Left, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-13-22, b) 41, 44, 47, 50, head, Medium, Item Number …
- Z-1415-2024 · Open, ClassifiedExactech Equinoxe CAGE GLENOID,POSTERIOR AUGMENT, CEMENTED, Right, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-13-32, b) 41, 44, 47, 50, head, Medium, Item Numbe…
- Z-1416-2024 · Open, ClassifiedExactech Equinoxe REVERSE SHOULDER,36mm Humeral Liner: a) +0mm, Item Number 320-36-00, b) +2.5mm, Item Number 320-36-03; Shoulder Arthroplasty
- Z-1417-2024 · Open, ClassifiedExactech Equinoxe REVERSE SHOULDER,36mm Constrained Humeral Liner: a) +0mm, Item Number 320-36-10, b) +2.5mm, Item Number 320-36-13; Shoulder Arthroplasty
- Z-1418-2024 · Open, ClassifiedExactech Equinoxe REVERSE SHOULDER, 38mm Humeral Liner: a) +0mm, Item Number 320-38-00, b) +2.5mm, Item Number 320-38-03; Shoulder Arthroplasty
- Z-1419-2024 · Open, ClassifiedExactech Equinoxe REVERSE SHOULDER,38mm Constrained Humeral Liner: a) +0mm, Item Number 320-38-10, b) +2.5mm, Item Number 320-38-13, Shoulder Arthroplasty
- Z-1420-2024 · Open, ClassifiedExactech Equinoxe Humeral Liner: Item Numbers: a) 320-40-00, b) 320-40-03; Shoulder Arthroplasty
- Z-1421-2024 · Open, ClassifiedExactech Equinoxe Constrained Humeral Liner: Item Numbers: a) 320-40-10, b) 320-40-13, Shoulder Arthroplasty
- Z-1422-2024 · Open, ClassifiedExactech Equinoxe REVERSE SHOULDER,42mm Humeral Liner, a) +0mm, Item number 320-42-00, b) +2.5mm, Item Number 320-42-03; Shoulder Arthroplasty
- Z-1423-2024 · Open, ClassifiedExactech Equinoxe REVERSE SHOULDER,42mm Constrained Humeral Liner, a) +0mm, Item number 320-42-10, b) +2.5mm, Item Number 320-42-13; Shoulder Arthroplasty
- Z-1424-2024 · Open, ClassifiedExactech Equinoxe REVERSE SHOULDER,46mm Humeral Liner, a) +0mm, Item number 320-46-00, b) +2.5mm, Item Number 320-46-03; Shoulder Arthroplasty
- Z-1425-2024 · Open, ClassifiedExactech Equinoxe REVERSE SHOULDER,46mmConstrained Humeral Liner, a) +0mm, Item number 320-46-10, b) +2.5mm, Item Number 320-46-13; Shoulder Arthroplasty
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page