Equinoxe recallZ-1420-2024 · Exactech, Inc. · initiated Mar 6, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Exactech, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

FDA root cause category: Nonconforming Material/Component

What the company is doing

Beginning 03/06/2024, Exactech issued an URGENT MEDICAL DEVICE RECALL notice and an URGENT DEAR HEALTHCARE PROFESSIONAL COMMUNICATION to its consignees via email. The recall notice explained the issue, clinical impact, and requested the following actions be taken: "" CAREFULLY REVIEW THIS RECALL NOTIFICATION to ensure that you fully understand the issue identified, the recall strategy, and all actions required. " IMMEDIATELY IDENTIFY AND QUARANTINE any of the subject devices in your inventory and/or customer s inventory listed Product Scope Listing. " EXTEND THE DESCRIPTION OF ISSUE AND CLINICAL IMPACT as described in the recall notification to your accounts that may have this product in their possession. " COMPLETE AND RETURN the attached Recall Inventory Response Form to Exactech via email at [email protected]. The communication to healthcare providers referenced the FDA Safety Communication issued on 01/16/2024, explained the issue, potential health risks, and suggested patient monitoring for those with implanted affected devices for potential device wear, failure, or bone loss and should consider performing X-rays to further evaluate the patient and their implanted device if you suspect a failed device. Contact your local Exactech Representative if you have any questions regarding this communication.

Recall details

Recall number
Z-1420-2024
Status
Open, Classified
Initiated by the company
Mar 6, 2024
Posted by the FDA
Mar 29, 2024
Recalling firm
Exactech, Inc.
Quantity
2806 units
Product code
PHX
FDA submissions
K182536, K180632
Distribution
Nationwide and Australia, Austria, Belgium, Brazil, Chile, Canada , China, Colombia, France, Germany, Great Britain, Greece, Guatemala, India, Italy, Japan, Korea, Luxembourg, Malaysia, Netherlands, New Zealand, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand

Product description

Exactech Equinoxe Humeral Liner: Item Numbers: a) 320-40-00, b) 320-40-03; Shoulder Arthroplasty

Code information (models, lots, serial numbers)

Show the full code information (25k characters)

a) 320-40-00, UDI/DI 10885862535030, Serial Numbers: 6173429, 5867573, 6007454, 6007478, 5927587, 6008225, 6007520, 5927562, 5927516, 5927575, 6197331, 6196344, 5867550, 5927586, 6196429, 5876444, 6196833, 5867568, 6174559, 5927640, 5927642, 6007467, 6173432, 5927563, 5876441, 6007452, 5876427, 5927631, 6196359, 5867564, 5927584, 5876450, 5927635, 6007456, 5927569, 6007567, 6196425, 6196773, 6173418, 5867565, 6196355, 6173443, 6196769, 6196776, 6196770, 6008226, 6007514, 6196351, 6196366, 5876433, 5876423, 6196505, 6174554, 5867558, 5927521, 6196424, 5927565, 5927591, 6173436, 6007475, 6007480, 6174553, 6008237, 6008243, 6196419, 6007548, 5927483, 6008351, 6174566, 5927485, 6196445, 6196433, 5867562, 5867567, 5876432, 5927498, 5876437, 5867556, 5927608, 5876430, 5876451, 5927561, 6008344, 6007543, 5927510, 6196364, 6196530, 6007568, 6196525, 6173428, 6196357, 5867557, 6196422, 6174552, 5876434, 5927543, 6173442, 6173434, 5927564, 6007465, 6196517, 5927530, 5927532, 6008346, 6196353, 6196349, 5927573, 5927492, 5927547, 6007571, 6196365, 6173445, 5876440, 5867563, 6007479, 6196764, 5867561, 6174557, 6196352, 6008348, 6196439, 5927639, 5927632, 6173430, 6008347, 6008343, 6008240, 6173435, 5927522, 6196427, 5927515, 5927592, 5927589, 6196760, 5927488, 6174573, 6173426, 6008341, 6007519, 5867553, 5927524, 6196348, 5876436, 5876428, 5927518, 6007466, 5876446, 5927497, 6174565, 5927538, 5927576, 5927596, 5876435, 5876439, 5876445, 5876448, 5927512, 5927519, 5927619, 5927622, 5927623, 5927630, 5927494, 5927537, 6196461, 6196463, 6173433, 6174549, 6196428, 6196431, 6196443, 6196521, 6196522, 6196768, 6196772, 6196774, 6402708, 6497639, 6543125, 5927508, 6372322, 6498876, 6008219, 6402668, 6008233, 6372400, 6007539, 6543293, 5867575, 6197327, 5927590, 6372391, 6402685, 6402721, 6543417, 6197328, 6543410, 6196788, 6543121, 6196460, 6197330, 6372325, 6543142, 6542053, 6543124, 6543405, 6542583, 6498882, 6007556, 6196791, 6196794, 5927523, 6497905, 6497649, 6497680, 6196782, 6196766, 6372393, 6372403, 6402687, 5927617, 6498717, 6372404, 6498694, 6542056, 6543132, 6008231, 6008232, 6196346, 6196821, 6196775, 6372339, 5927493, 6498695, 6497871, 6372349, 5927486, 5927638, 6008239, 6008244, 6497655, 6542073, 6196804, 6196360, 6543115, 6008229, 6542080, 6196824, 5927579, 6372413, 6196779, 6174568, 6542569, 6007464, 6543402, 6402691, 6007545, 5927615, 6196771, 6543143, 6542069, 6372411, 6543122, 6543284, 6402675, 6543281, 5927513, 5927633, 5867577, 6008230, 6196781, 6196514, 5876442, 6007542, 6196785, 6543120, 6542572, 5876426, 6543418, 5927628, 6196354, 6543131, 6372407, 6543399, 6497860, 5927626, 6197336, 6007531, 6498718, 6196457, 6196797, 6497876, 6174569, 6196825, 6197317, 6497857, 6196778, 6497853, 6372334, 6497660, 6196468, 6196786, 6402656, 6196805, 6498879, 6542596, 6497604, 6542576, 5867574, 6007537, 6196507, 5927602, 6196831, 6543406, 6007453, 6007524, 6008234, 6497672, 6543413, 5927566, 6007515, 6196818, 6007481, 6498710, 6543302, 6498701, 6174572, 6372346, 5927624, 6497640, 6542575, 6497865, 6007552, 5927581, 5927542, 5927634, 6196465, 5927578, 5927556, 6498715, 5927550, 6497862, 6543139, 6008352, 6173441, 6402711, 6402673, 6543287, 5927517, 5927507, 6007529, 5927549, 6007532, 5927588, 5927570, 6372331, 6542070, 5927528, 6372330, 6498700, 6543420, 6543422, 6543299, 6007530, 5867571, 6196796, 6543396, 6196446, 5876431, 6543397, 5927577, 5927583, 6007558, 6196808, 6542064, 5927597, 6173424, 6402705, 6196510, 6196529, 6007513, 6372395, 5927501, 6372401, 5927545, 6402666, 6402719, 6498708, 6196528, 6498870, 6497863, 6543130, 6542568, 6173416, 5867570, 6196830, 6196438, 6498875, 5867566, 6497634, 6543126, 6497670, 5867559, 5867578, 6372345, 6196765, 6543307, 5867552, 6543411, 6196454, 5927625, 5867554, 6196811, 6196453, 6543117, 6497614, 6542588, 6196450, 6543129, 6543398, 6007458, 5927621, 6497906, 6196823, 6372412, 6372402, 6372343, 6372350, 6543280, 6196435, 6008238, 6173420, 6174576, 6197326, 6197339, 6372327, 6372394, 6372351, 6372405, 6174562, 6372333, 6542578, 6372347, 6543403, 6543295, 6497627, 6196783, 6498699, 6372341, 5927535, 6007476, 5876429, 6008349, 6008342, 6402733, 6497866, 6542594, 6542573, 6543289, 6542077, 6543395, 6543400, 6497677, 5927600, 6196839, 6197342, 6007473, 6372398, 6196809, 6542580, 6542589, 6174574, 5927534, 6542577, 6543123, 6542567, 6543138, 6542074, 6543415, 6542071, 6196835, 6196347, 6196449, 6196426, 6196800, 6372390, 6372326, 6372389, 6196448, 6196807, 6402718, 6007554, 6402669, 6402720, 6372332, 6497632, 6196456, 6497643, 6542581, 6497615, 6543116, 6543407, 6542574, 6542584, 6543134, 6402657, 6196836, 6173417, 6196838, 6497864, 6542585, 6497658, 6497662, 6542081, 6402663, 5927603, 6402676, 6173425, 6007562, 6196434, 6196362, 6196816, 6402722, 6542593, 5927490, 6372344, 5927496, 6402712, 6196758, 6196356, 6007463, 6007474, 6174575, 6197334, 6007468, 6173444, 6402659, 6196799, 6402664, 6372396, 6372408, 6543424, 6497661, 6196519, 6372384, 6174551, 6174571, 6196452, 6543423, 6196469, 6197320, 6196829, 6196462, 6196790, 6372337, 5927526, 6542075, 6196432, 6402730, 6007471, 6174577, 6372335, 6007544, 6543303, 6173419, 6372406, 6402729, 6542054, 6008241, 5927541, 5927607, 5927644, 6197344, 6173427, 6173440, 6007569, 6498706, 6543296, 6007565, 6497878, 6008242, 6498712, 6497877, 6174555, 5927620, 6542057, 6498713, 6543128, 6372324, 6007540, 6196814, 6008235, 6007560, 6372386, 5927595, 6196761, 6007534, 6498720, 5876449, 6196789, 6497626, 6174556, 6497879, 6196350, 6372329, 6372338, 6402717, 6372323, 6196837, 6402662, 6402670, 6402679, 6402686, 6402725, 6402731, 6497654, 6498868, 6498877, 6497646, 6542052, 6542060, 6542065, 6543290, 6543291, 6543297, 6543301, 6543114, 6543133, 6543135, 6196793, 6196795, 6196806, 6197321, 6197324, 6497875, 6497887, 6497888, 6497893, 6497894, 6497895, 6542570, 6542579, 6572869, 6572866, 6561759, 6572874, 6552376, 6572823, 6563069, 6552897, 6572829, 6572737, 6561758, 6561777, 6552365, 6561773, 6563061, 6683102, 6572730, 6561766, 6572842, 6572846, 6561764, 6563068, 6561761, 6572884, 6572848, 6552380, 6552381, 6552911, 6563073, 6552868, 6572718, 6572728, 6572868, 6572700, 6668458, 6563049, 6572809, 6552894, 6572860, 6572808, 6572873, 6563077, 6622459, 6572807, 6552874, 6572804, 6572854, 6552372, 6552900, 6552862, 6552882, 6561772, 6572872, 6572841, 6572891, 6563067, 6572839, 6552870, 6563056, 6552902, 6671714, 6563064, 6552385, 6561757, 6572815, 6552892, 6552899, 6552371, 6552383, 6552360, 6561775, 6563065, 6572858, 6671727, 6572852, 6572799, 6552913, 6563062, 6563076, 6572887, 6572865, 6552378, 6561776, 6552384, 6572821, 6572707, 6552889, 6572870, 6572709, 6572727, 6563060, 6552869, 6552366, 6563075, 6572690, 6572813, 6572835, 6552896, 6572840, 6552883, 6563059, 6572719, 6563078, 6572685, 6572800, 6552860, 6552867, 6552872, 6552880, 6552886, 6552362, 6552370, 6552387, 6552876, 6561753, 6561768, 6563074, 6196368, 6497635, 6402682, 5867569, 6196762, 6196509, 6007551, 6543283, 6498698, 6497620, 6497673, 6497666, 6497676, 6497869, 6498696, 6196441, 6196447, 6402655, 6196451, 6402643, 6196784, 6497873, 6543306, 6497890, 6497896, 6497642, 6197340, 6173421, 6402648, 6402690, 6402665, 6196471, 6197332, 6402714, 6402713, 6372340, 6543421, 5927487, 6197333, 6497669, 6498869, 5876447, 6498707, 6007561, 6497872, 6543294, 6497665, 6497603, 6497900, 6497901, 6402642, 6497611, 6402680, 6497874, 6498871, 6372342, 6196508, 5927504, 6542062, 5927531, 5927514, 5927618, 6007570, 6007553, 5927554, 6542591, 5927560, 6497679, 5867560, 5927629, 5867555, 6197319, 6402632, 6497621, 6007535, 6543119, 6402726, 6543304, 6543401, 6402728, 6174548, 6196459, 6497648, 5867551, 6196810, 5927641, 6402674, 6498704, 6497855, 6196523, 6497859, 6498716, 6498703, 6543308, 6197341, 6497681, 6197322, 6007563, 6007555, 6402716, 6402651, 5927580, 6497898, 6196430, 5867576, 6196464, 6197338, 6542063, 5876424, 6196813, 6196506, 6497664, 6497880, 6402635, 6174560, 6008220, 6372348, 6543140, 6543409, 6008222, 6497616, 6196436, 6007459, 6497885, 6196798, 6497903, 6543419, 6402677, 6007516, 6007522, 6173431, 6008228, 6543118, 6498693, 5927599, 6402644, 6008345, 6402641, 6497883, 6497884, 6196442, 6497624, 6497607, 5927548, 6196802, 6498697, 6497908, 6497902, 6007460, 6402660, 6497892, 5876425, 5927598, 5927558, 5927500, 5927601, 6542066, 6402715, 6542055, 6197329, 6543412, 6174570, 6497856, 6007559, 6402709, 6197323, 6542061, 6007546, 6196819, 6542058, 6372336, 6196513, 6402667, 5927520, 6196820, 6372328, 6542078, 6196343, 6498874, 6007525, 6497623, 6007527, 6008227, 6196444, 5927616, 6372388, 6196341, 6497691, 6497647, 6498709, 5927502, 6497867, 6196817, 5927567, 6497638, 6543282, 6497671, 6196472, 5927636, 5927525, 6497629, 5927572, 6497641, 6497686, 5927511, 6372399, 6197335, 6542590, 6497619, 6007557, 6543288, 6497650, 6497622, 6542586, 6196466, 6196787, 6497682, 5867572, 6196470, 5927571, 6196803, 6196512, 6008217, 6196832, 6542587, 6542571, 6572817, 6552903, 6671743, 6979068, 7035358, 6668464, 6672810, 6970784, 7047507, 6572705, 6646045, 6572687, 6572850, 6646090, 6646063, 6671717, 6668495, 6683106, 6672814, 6775151, 7021994, 7021831, 7047513, 6552917, 6646097, 6775117, 6668494, 6646094, 6646100, 6671702, 6671737, 6775138, 6668440, 6552890, 6552916, 6668441, 6672807, 6970779, 6572830, 6572723, 6668437, 6668467, 6683088, 6671746, 6646096, 6668488, 6671720, 6671750, 6668443, 6683087, 6672801, 6753954, 6737324, 6775105, 6552863, 6775109, 6775116, 6979057, 6775119, 7021978, 6646034, 6668451, 6572819, 6646076, 6668452, 6646069, 7021824, 6672800, 6646053, 6646032, 6622471, 6671729, 6671747, 6668492, 6672795, 6668489, 6753952, 6646066, 6683078, 6737317, 6552888, 6683099, 6979086, 6672816, 6737326, 6737333, 6979063, 7047518, 6671697, 6737347, 6572881, 6572716, 6646055, 6671734, 6646065, 7047472, 6552358, 6561767, 6572696, 6646087, 6672806, 6671744, 7021953, 6979085, 6979077, 7035328, 7021986, 6775123, 6671751, 7021961, 6668459, 6979067, 6552909, 6683110, 6775145, 6552359, 6646104, 6561771, 6668453, 6672802, 6683085, 7021947, 6775153, 6683105, 6737348, 6737325, 6979094, 6979040, 6979088, 6979069, 6561754, 6672812, 6737345, 6775129, 6970788, 7035367, 6672782, 6672780, 6622482, 6563053, 6668487, 6672811, 6561760, 6737342, 6646079, 6683117, 6671695, 7021809, 7047511, 6775108, 7021801, 6572713, 6668498, 6775148, 7021969, 6572689, 6672797, 6552906, 6561769, 6552912, 6646103, 6671713, 6671725, 6737332, 6775133, 6970787, 6979087, 6572833, 7047492, 6672788, 7047490, 6753971, 7021799, 6552865, 6668463, 6753953, 6563057, 6622461, 6646060, 6646101, 7047485, 6979049, 7047506, 6668445, 6668497, 7021984, 7047499, 7035371, 6552364, 6979082, 6668456, 6572867, 6572725, 6683112, 7035357, 6572856, 6572831, 6572698, 6646102, 6646083, 6668483, 6561774, 6671699, 6552908, 6683100, 6668490, 6753963, 6737351, 6753955, 6775140, 6753956, 6671706, 7035351, 7047476, 6646071, 6552879, 7021948, 6552895, 6753970, 7021798, 7021796, 7047482, 6561755, 6563072, 7047475, 6572828, 6572688, 6753966, 6552363, 6572734, 6775155, 6775106, 6979072, 6672781, 6979046, 7021819, 6572681, 7021807, 6622480, 6622472, 6572735, 6979074, 6979092, 7021940, 7035362, 6979050, 6572715, 6671754, 6646044, 6572853, 6622467, 6561750, 6622468, 6672792, 6672799, 6668478, 6646051, 6668444, 6668465, 6671716, 7021941, 7035350, 6737322, 6552905, 6552861, 6572805, 6775110, 6775137, 6552907, 6668472, 6646031, 6671740, 6683092, 6775112, 6552373, 6563066, 6683094, 6753962, 6668479, 6668455, 6683079, 6753961, 6753964, 7047500, 7047504, 6572803, 6622481, 6978981, 6775125, 6970793, 7047502, 6563058, 6646062, 6646108, 6672793, 6671738, 6683084, 7047508, 7021830, 6572814, 6572810, 6671723, 6683115, 7021805, 7035332, 6572802, 6970776, 6622484, 6552885, 6572682, 6572704, 6552910, 6572693, 6646040, 6646085, 6668448, 6572827, 6671741, 7047509, 7035349, 6572683, 7047470, 7021985, 6572686, 6561756, 6572849, 6646056, 6668496, 6671718, 6572717, 6646043, 6572680, 6671726, 6672796, 6622483, 7021950, 6572729, 6979090, 7035346, 7035331, 6979091, 7021959, 6646086, 6622473, 6979045, 6646092, 6671700, 6572824, 6671722, 6668481, 6646091, 6646088, 6672786, 6672804, 6737336, 6970778, 6737346, 6979098, 7035334, 6572736, 7035369, 6572726, 6552878, 6572722, 6672785, 6979070, 6668477, 6672783, 7021987, 6552891, 6572847, 6668480, 7021943, 6572801, 6646033, 6646074, 6646054, 6668493, 6753957, 6970774, 6970791, 7021956, 7021989, 7035376, 7035342, 6737320, 7035378, 6646030, 6572845, 6572883, 6572890, 6646041, 6646057, 6646064, 6646050, 6668469, 6646082, 6683097, 6683109, 6668462, 7035339, 7047484, 7047487, 7035377, 6970785, 6775128, 6671705, 6671709, 6775134, 7021954, 7021802, 7035356, 6572703, 6572844, 6622485, 6646081, 6668449, 6775141, 7021826, 6775143, 6668485, 6775150, 7047479, 7047510, 7047516, 6672815, 6671728, 6668473, 6672805, 6552866, 6572836, 6646089, 6646084, 6683111, 6552873, 6671749, 6775144, 6572878, 7021995, 7021983, 6775121, 6775118, 6563051, 6672813, 6561762, 6683091, 7021964, 6561763, 7021952, 6671731, 6563063, 6572885, 6552915, 6646106, 6979039, 6622469, 6671752, 6572826, 6552914, 6646078, 6737327, 6646038, 7047467, 6572820, 6671748, 6775115, 7021958, 6563055, 6572859, 6668470, 6671711, 6683093, 6978984, 6572863, 7021823, 6668502, 6552893, 6671735, 6683095, 7021835, 6622476, 6671721, 7021936, 6552379, 6552875, 7021825, 7047493, 6683101, 6668468, 6671730, 6572880, 6572694, 6775149, 7021820, 6646072, 6552859, 6572702, 6622470, 6622465, 6683090, 7021962, 6572889, 6572724, 6572811, 6572695, 6668500, 6563070, 7021797, 6622474, 6668446, 6672787, 6683104, 6683098, 6572822, 6668442, 6979078, 7047503, 7021821, 6671701, 6672809, 7035364, 7047477, 7021808, 6646042, 6646048, 6646049, 6646058, 6646059, 6646098, 6671708, 6671715, 6671742, 6737319, 6737334, 6737337, 6737331, 6737341, 6737343, 6737350, 6775124, 6775132, 7047488, 6572851, 7035338, 6737335, 6672791, 6572825, 6572855, 6672790, 6668454, 6970775, 6978996, 6668484, 6572684, 6646067, 6672794, 6672798, 6497645, 5927509, 6173438, 6007477, 6543137, 5927553, 6196516, 6402724, 6497653, 6008350, 6497685, 6007533, 5927503, 6402647, 6402640, 5927643, 6174567, 6007523, 6498691, 5927527, 5927540, 6196520, 6196342, 5927594, 5927484, 6498873, 6497909, 5927609, 6402633, 6497889, 6497613, 6372409, 6497683, 6497678, 6402678, 6497633, 6402654, 6497663, 6543309, 6543414, 6196504, 6498872, 6402683, 6498867, 6498880, 6497652, 6007538, 6196767, 6007536, 6196780, 6402638, 6497625, 6542072, 6007528, 6196420, 6196458, 5927495, 6007457, 6543127, 6196345, 5927482, 5927533, 6542076, 6497657, 5927551, 5927559, 6402672, 6497897, 6497602, 6497609, 6007461, 6542079, 6008236, 6497631, 6196828, 6497610, 6497637, 6497881, 6542068, 6497904, 6497858, 6402639, 6543416, 6498702, 6196440, 6543141, 6542595, 6402658, 6372387, 6196763, 5927539, 6007526, 6497899, 6372392, 6174550, 6196524, 6497861, 6498692, 6196815, 6497617, 6197318, 6196827, 5867579, 6007521, 6197325, 6542582, 6497687, 6622477, 7021937, 6552887, 7035370, 7021817, 6572843, 6970789, 6572708, 7035372, 6753958, 7021942, 6622466, 6753967, 6672808, 6572862, 6683089, 6572838, 6978982, 6668503, 7047498, 6775126, 6979071, 6646047, 6572871, 6572711, 7035337, 6753969, 6979064, 6622458, 7047481, 7021814, 6979047, 6683114, 7021990, 6671739, 6668476, 6572691, 6683116, 6671696, 6552904, 6672789, 6683103, 6775114, 7047478, 6775107, 7047497, 6683083, 6552374, 6561751, 6737318, 6646105, 6979084, 6572731, 6552369, 6572710, 6552361, 7035366, 6572876, 6775156, 6970790, 7047469, 6668486, 6561748, 7047496, 7047474, 7035336, 6646052, 6572837, 6672803, 6979043, 6775139, 7021966, 6563050, 6668460, 7035359, 7047480, 6978987, 6563052, 6552382, 6979061, 7047501, 7021965, 7021992, 6979056, 7047486, 6572857, 6683107, 7035335, 6561752, 6672777, 6671733, 6572692, 6646070, 6683108, 7021970, 6775147, 6753965, 6572714, 6775130, 7035344, 7047512, 6572818, 6552858, 7021810, 6552864, 7035343, 7021945, 7021975, 6424098, 7021949, 7047514, 6978985, 7021832, 7047468, 6775127, 6646073, 6775113, 7047491, 6646095, 7021981, 7047471, 7035379, 6978993, 6753960, 6775111, 6737339, 7021828, 6572816, 6622464, 7047489, 7047495, 6646093, 6646080, 6572864, 7021803, 7035345, 7035329, 6646107, 7021800, 6646046, 7035374, 6683082, 6775104, 7021991, 6683096, 6671704, 6671719, 6572834, 7021977, 6671736, 6978980, 6552386, 7035333, 6970782, 6978988, 6646068, 6737349, 6970781, 6572679, 6775135, 7021993, 6970786, 6970780, 6978989, 6646036, 6572882, 6402688, 6497854, 6543292, 6498719, 5927593, 6402636, 5927544, 6007472, 6196812, 6497606, 6007541, 6402653, 5927546, 6402732, 6498878, 6543298, 6402671, 5927574, 6007564, 6008245, 6196437, 6543136, 6196777, 6196423, 6498714, 6196826, 6197343, 5927585, 5927499, 6497618, 6497907, 6196511, 6008223, 6402637, 5876443, 5927605, 6174561, 6197337, 6372397, 6196515, 6196503, 6174564, 5927582, 6196361, 6196367, 6008221, 6196792, 6402646, 6372410, 5927637, 6007518, 6543286, 6497870, 5927604, 5927568, 6497690, 6196822, 6498711, 6498705, 6646075, 6552871, 7021972, 6737316, 7021971, 6978983, 7021957, 6646037, 7035348, 7035340, 7021988, 6978986, 7035368, 7021946, 7021818, 7035341, 6668475, 7021827, 6978998, 6572699, 7047473, 6552375, 6671707, 6668457, 7035375, 6979079, 6775122, 6671698, 6563071, 6646099, 6979062, 7035330, 6978992, 6775157, 7035355, 7047517, 6572879, 6622479, 6775142, 6561765, 6671724, 7021806, 6552898, 7047515, 6683081, 6775136, 6668466, 7021955, 6572886, 6552368, 7021815, 6668501, 7047466, 7021794, 6979075, 7021960, 7021792, 6552377, 6572701, 6978995, 7021829, 6561749, 6979093, 6979065, 6671753, 6572712, 6970792, 7021793, 6979042, 7021980, 6646039, 7021813, 7021816, 7021833, 7021804, 6668439, 6979053, 6672779, 6978990, 7035360, 6671745, 6668471, 6978991, 7047483, 6561770, 6671710, 6542592, 6007549, 6173422, 5927529, 6196421, 6543404, 7035373, 6979081, 6979066, 7021976, 6572877, 6979048, 7021939, 6671703; b) 320-40-03, UDI/DI 10885862535047, Serial Numbers: 5910013, 5910544, 6176692, 6168221, 5933375, 5932710, 5910540, 6176689, 6176705, 5932700, 5910546, 6170253, 6168239, 6168242, 6168231, 5910030, 6176686, 5910541, 5933351, 5910011, 5932804, 6176684, 5932796, 5932716, 5933391, 5932701, 5932782, 5933401, 6176688, 5932793, 5932645, 6168245, 5910015, 6168249, 6168251, 5933381, 5910021, 5932717, 5910023, 6170225, 6168230, 6168402, 5910033, 5910018, 5933356, 5933360, 6168228, 6168243, 6168261, 6168267, 6170227, 6176666, 6168380, 6168409, 6168411, 6413497, 6397898, 5910551, 5932789, 6168224, 6168226, 6397830, 6527360, 6176676, 6535182, 6538379, 6535206, 6535179, 6527446, 6397822, 6176690, 6168214, 6168399, 6170254, 6351454, 6527393, 6527375, 6397813, 6527438, 5932792, 6527404, 5933377, 5932632, 5932639, 5932784, 5932698, 6168257, 5933354, 6170237, 6538362, 6397836, 5933355, 6176682, 6170235, 6397828, 5910038, 5932703, 5932695, 5932648, 5933406, 5932714, 6351444, 5910558, 6527371, 6176693, 6176665, 6351446, 6351451, 6176696, 6335152, 6170246, 6397827, 6527410, 6527395, 6170209, 5910022, 6168218, 6397810, 5933364, 6168384, 6170255, 6351441, 5910553, 5933366, 6351453, 6168389, 5933626, 5932707, 6176674, 6397860, 6397858, 6168385, 6168412, 6397865, 5910559, 6535166, 6170219, 5932646, 5932633, 5910032, 6397863, 6527418, 5932791, 6168215, 6168404, 6535185, 6538384, 6535199, 6538372, 5910036, 6397842, 5933365, 6170232, 5910010, 5933353, 6170228, 6170233, 5933374, 6397857, 6527310, 6170216, 6538367, 5933392, 6527388, 5932688, 6535202, 6168417, 6170236, 6397840, 5932797, 6535187, 6168401, 6170256, 6176704, 6335157, 6397877, 6527295, 6535169, 6535198, 6397829, 5910026, 5932787, 6168392, 6168394, 6176673, 6527434, 5932641, 6168255, 6176664, 6170252, 6168253, 6397831, 6168415, 5932693, 6527396, 6397809, 6168391, 6168250, 6168373, 6351450, 5910017, 6397851, 5932647, 6535177, 6397835, 6170229, 5933362, 5910009, 5932715, 6538375, 6351435, 5933357, 5932798, 6538369, 6170248, 6170221, 6170251, 5910550, 6535186, 5933403, 6335155, 5932699, 6527442, 6397871, 6397881, 5932705, 6538385, 5932653, 6176695, 6535163, 6527421, 6168219, 5910019, 6397824, 6397837, 6397846.6397880, 6535167, 6535170, 6535178, 6527294, 6527298, 6527299, 6527317, 6527328, 6527331, 6671489, 6622528, 6674049, 6609123, 6575875, 6609111, 6575870, 6575872, 6575934, 6413491, 6527385, 6168238, 6527398, 6538361, 6527415, 5910549, 6535205, 6351459, 6527376, 6351452, 6170230, 6535181, 6170211, 6170247, 6413492, 6527431, 6527364, 6176677, 6351433, 6168377, 6527333, 6538382, 5932651, 6168263, 6397891, 6168241, 6535168, 5933394, 6538354, 6527401, 5933361, 6527403, 5932656, 6168386, 6397811, 6535172, 5932713, 6538383, 5933402, 6527414, 6176669, 6538386, 6527334, 6176681, 6527412, 5932788, 6527339, 6527343, 6527313, 6527314, 5932790, 5933376, 5932780, 6527296, 6527324, 6527308, 6168260, 6397890, 6535160, 6535165, 6168265, 6527392, 6538357, 6168244, 6535189, 6397843, 6397832, 6168407, 6168371, 6168233, 6170212, 6527318, 6538371, 6527348, 6397868, 6397845, 6170208, 6351439, 5910035, 6170240, 6168396, 6538353, 5933387, 5932708, 6535161, 6168237, 6397825, 6168388, 6397894, 5910560, 6413489, 6176685, 6527397, 5910020, 6535171, 5932692, 6538374, 5910557, 5932712, 6527353, 5910014, 5932794, 6176670, 6397897, 6397821, 6527386, 6538364, 6176702, 6397870, 5910556, 6168403, 5933382, 6397815, 6168397, 6535184, 6527406, 6527444, 5910024, 6527370, 6538366, 6527341, 6527411, 5932779, 6527354, 5932631, 5933393, 6170244, 6168413, 6397806, 6351447, 6527428, 5932655, 6527312, 6170224, 5932637, 6397869, 6535157, 6397817, 6535196, 5933399, 6527394, 6176697, 6176694, 6527380, 6527382, 5910016, 6527447, 6538390, 6527358, 6335158, 6535194, 5933404, 5933405, 6170222, 6527437, 6397862, 6351457, 6527435, 6575871, 6622515, 6575911, 6674000, 6575919, 6671486, 6674008, 6674009, 6730388, 6674045, 7023551, 6622540, 6575897, 6673994, 6673996, 7023565, 7023575, 7023531, 6575882, 6575929, 6979013, 6671481, 6674048, 6609127, 6622538, 6622514, 6671490, 6671477, 7023573, 6575928, 6674006, 6671492, 7023569, 7023564, 7023527, 7023529, 6674016, 6575885, 6575896, 6730384, 6575932, 6575881, 6674043, 7023524, 6674015, 6622517, 6674051, 6674031, 7023567, 7023535, 6609114, 6575891, 6730401, 6674010, 6674003, 6609112, 6671482, 7023561, 6575878, 6575933, 6622530, 6674004, 6674044, 6979012, 7023571, 6674021, 7023520, 6730381, 7023549, 6674022, 7023543, 6622532, 6609115, 6609128, 6674034, 6575869, 6671488, 6730376, 6972844, 6674001, 6575913, 7023534, 6674017, 6609124, 6575888, 6673998, 6555225, 6575899, 6674011, 6609110, 6622516, 6730371, 6413487, 5910547, 6535175, 6170207, 6176698, 6168379, 6168387, 5910545, 6527389, 6538380, 6176701, 6176678, 6351456, 6527346, 5932802, 6527326, 6527300, 6538376, 6176703, 6535162, 6168266, 5910542, 6397812, 6527408, 6527436, 6527443, 6168246, 6527372, 6168259, 6168258, 5932799, 6176672, 5932630, 6535190, 6535174, 6168252, 6527344, 6397849, 6397853, 6397895, 5932704, 6168382, 6170231, 6527356, 6176700, 6527338, 6527387, 5933383, 5933396, 6527325, 6168223, 6397872, 6397847, 6168225, 6170215, 6538377, 5933625, 6527365, 5932694, 6535164, 6538359, 6168405, 5932689, 6527448, 5910539, 5910548, 6170249, 5932803, 6527384, 6527306, 5910561, 5933349, 5933359, 6538373, 6527391, 5932628, 5932640, 6527305, 5933373, 6168418, 6168234, 5932635, 6527332, 6527297, 5932786, 6413485, 6168374, 6168414, 6397867, 6168240, 6535195, 6538360, 6535200, 6176675, 6397838, 6168375, 6176687, 5933379, 5932642, 6538388, 6535197, 6168264, 6527433, 6527335, 6168410, 6168217, 6351448, 7023580, 7023539, 6555211, 7023518, 6972809, 6575908, 7023550, 6673992, 6575876, 6609118, 7023574, 6972838, 6671479, 7023576, 6622518, 7023554, 6972834, 6972814, 6972817, 7023523, 6555219, 7023542, 6622524, 6575900, 6555208, 6555207, 6671478, 7023547, 6674042, 7023521, 6555212, 7023560, 7023517, 6671491, 7023558, 6674007, 6674026, 6730373, 6575936, 7023557, 6671480, 6575930, 6575903, 6609117, 6972845, 6979014, 6674033, 6674028, 6575901, 6575914, 6622533, 6972840, 7023546, 7023519, 7023528, 6972843, 6674020, 6555222, 6972835, 6575909, 6674047, 6730368, 7023533, 6575884, 6575907, 7023530, 6555210, 6730403, 7023563, 6671495, 6575910, 6527407, 6413496, 5933400, 5933624, 6168227, 6527452, 5933384, 6170241, 6170250, 6527322, 6535191, 5933358, 6397896, 6527400, 5933370, 6538368, 6170245, 6170226, 6527430, 6397833, 6335159, 6535180, 6527352, 5932795, 6527329, 6397852, 5932690, 6170217, 6527368, 6527405, 6397876, 6527373, 6397823, 6535176, 6527302, 6397841, 5933390, 6527303, 6168236, 6351436, 5910029, 6176699, 5910552, 5932706, 5933623, 6538352, 6538381, 6527307, 6527340, 5932691, 5932711, 6527451, 6527369, 5932629, 6176671, 6397889, 6168235, 6527379, 6397834, 6413498, 6170243, 6168406, 5932806, 6168247, 6535201, 6176683, 6527426, 5933367, 6538358, 5932650, 6170214, 6538378, 6527425, 5932783, 6397854, 6538355, 6527445, 6176668, 6351432, 6335153, 6351442, 6527419, 6527423, 6351449, 6535158, 6674012, 6972846, 6622521, 7023579, 6972812, 6972822, 7023559, 6972813, 6730404, 7023568, 6972832, 6730385, 6575915, 7023556, 6972837, 6979005, 7023562, 6730369, 6730400, 6575905, 6575883, 6609109, 7023541, 6972811, 6979010, 6972821, 6622539, 7023536, 7023532, 6673997, 6671484, 7023555, 6622534, 6609125, 6575935, 6622531, 6972836, 6972825, 6674035, 7023537, 6575874, 6972816, 6555221, 6575912, 6575922, 6972829, 6622525, 6674013, 6671494, 6575931, 6972826, 6972842, 6609121, 6673999, 6397878, 6413490, 6527357, 6351437, 6168369, 6170239, 6979002, 6972818, 6575924

Devices named in this recall