Olympus recallZ-1700-2024 · Olympus Corporation of the Americas · initiated Mar 7, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Upon initial activation of the seal function, the POWERSEAL non-conformance results in an immediate Incomplete Seal Cycle tone with the accompanying message on the generator screen, and no energy would be delivered to the device resulting in delay in surgery.
FDA root cause category: Process control
What the company is doing
Olympus issued Urgent Medical Device Removal Letter on April 2, 2024 via email/visit. Letter states reason for recall, health risk and action to take: 1. Inspect your inventory and identify any POWERSEAL products having the lot number identified above. Please check all areas of the hospital to determine if any of these devices remain in inventory. Quarantine and cease use of the affected model/lots. 2. Please contact Customer Service at at 1-800-848-9024, option 2, with the quantity, model and lot number of the affected device(s). Olympus will issue a Return Material Authorization to return any affected product at no charge. Olympus will issue a credit to your facility upon return of the affected product. 3. Access the Olympus recall portal to indicate that you have received this notification. a. Go to https://olympusamerica.com/recall b. Enter the recall number "0444 . c. Complete the form as instructed and include your account ID number. 4. If you have distributed these devices outside your facility, please provide a copy of this letter to those facilities immediately. If you require additional information, please do not hesitate to contact me directly by phone at (647) 999-3203 or by e-mail at [email protected]
Recall details
- Recall number
- Z-1700-2024
- Status
- Open, Classified
- Initiated by the company
- Mar 7, 2024
- Posted by the FDA
- Apr 30, 2024
- Recalling firm
- Olympus Corporation of the Americas
- Quantity
- 850 units: (400 units US);(450 units OUS)
- Product code
- GEI
- FDA submissions
- K212643
- Distribution
- Worldwide distribution - US Nationwide and the countries of AU, CA, HK, KR.
Product description
POWERSEAL 5 mm, 37 cm, Curved Jaw Sealer and Divider, Double Action is indicated for use in general surgery and such surgical specialties as urologic, colorectal, bariatric, vascular, thoracic, and gynecologic. Model/Catalog Number: PS-0537CJDA
Code information (models, lots, serial numbers)
UDI-DI: 00821925044555 Lot numbers: CA182543, CA182544, CA182546, CA182548, CA191230
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Olympus PS-0537CJDAGyrus ACMI, LLC · Electrosurgical, cutting & coagulation & accessories2 recalls184 reports
Product lines
Other devices under the same FDA submission
2 devices cleared under K212643 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Olympus PS-0523CJDAGyrus ACMI, LLC · Electrosurgical, cutting & coagulation & accessories3 recalls90 reports
- Olympus PS-0544CJDAGyrus ACMI, LLC · Electrosurgical, cutting & coagulation & accessories2 recalls10 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 94375).
- Z-1699-2024 · Open, ClassifiedPOWERSEAL 5 mm, 23 cm, Curved Jaw Sealer and Divider, Double Action is indicated for use in general surgery and such surgical specialties as urologic, colorectal, bariatric, vascul…
- Z-1701-2024 · Open, ClassifiedPOWERSEAL 5 mm, 44 cm, Curved Jaw Sealer and Divider, Double Action is indicated for use in general surgery and such surgical specialties as urologic, colorectal, bariatric, vascul…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page