Incisive CT recallZ-1871-2024 · Philips North America Llc · initiated Apr 15, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Metal mounting box on the rotating scanner on rotor (heat change box) located within the Incisive CT system, may become compromised, resulting in potential contact with other gantry components located within the Incisive CT system during rotation. Resulting in a fragment of a damaged component expelled at a low velocity
FDA root cause category: Under Investigation by firm
What the company is doing
Philips issued URGENT Medical Device Correction Letter on 4/15/24 to US consignees. [International global markets were notified of the issue and instructed to distribute the FSN to affected consignees in accordance with local requirements]. Letter states reason for recall, health risk and action to take: " Please continue to use your system in accordance with its intended use. If you need to remain in the room for the duration of a scan, Philips recommends using protective eyewear. " Circulate this Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue. Please retain this letter with your system(s) until a solution is installed on your system; ensure the letter is in a place likely to be seen/viewed. " Place this URGENT Medical Device Correction Letter with your system documentation. " Please complete and return the attached response form to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction Letter, understanding of the issue, and required actions to be taken . Philips will contact you to schedule a time for a Field Service Engineer (FSE) to visit your site and install a solution to address the issue (reference FCO72800819 and FCO72800820). If you need additional information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377).
Recall details
- Recall number
- Z-1871-2024
- Status
- Open, Classified
- Initiated by the company
- Apr 15, 2024
- Posted by the FDA
- May 20, 2024
- Recalling firm
- Philips North America Llc
- Quantity
- 18 units
- Product code
- JAK
- FDA submissions
- K180015
- Distribution
- Worldwide - US Nationwide distribution.
Product description
Incisive CT - Computed Tomography X-Ray System intended to produce cross-sectional images of the body. Model Number: 728143
Code information (models, lots, serial numbers)
UDI-DI: (01) 00884838085015 (21) + Serial number Serial Numbers- Applied to FCO72800819: 500046, 500064, 500074, 500102, 500110, 500111, 500145, 500168, 500178, 500211 500216, 500217, 500222, 500225, 500230, 500240, 500242. Serial Number-Applied to FCO72800820 : 500615
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Incisive CTPhilips Healthcare (Suzhou) Co., Ltd. · System, x-ray, tomography, computed13 recalls5 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 94536).
- Z-1872-2024 · Open, ClassifiedIncisive CT - Computed Tomography X-Ray System intended to produce cross-sectional images of the body. Model Number: 728144
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page