Mega 2000 recallZ-2248-2024 · Megadyne Medical Products, Inc. · initiated Jun 17, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Megadyne Medical Products, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

To limit the use of the Mega Soft Pads to patients aged 12 or older to mitigate the potential risk of patient burns identified with use of Mega Soft pads.

FDA root cause category: Unknown/Undetermined by firm

What the company is doing

An URGENT MEDICAL DEVICE CORRECTION letter, dated 6/17/24, was mailed to consignees. Consignees are instructed to share the recall notification with all users of affected devices and confirm their understanding, post a copy of the communication for user awareness, and forward the recall notification to any facility to which product was shipped. The notification reads that Mega Soft and Mega 2000 devices are only to be used for patients 12 years old. The notification also reminds consignees to follow proper cleaning, placement, and setup steps. Consignees are asked to complete and return the provided Business Reply Form by fax to 844-441-2732 or by email to [email protected]. Customers with any questions about the communication can contact Ethicon at 1-877-384-4266 from Monday through Friday from 8:00 AM EST to 5:00 PM EST. UPDATE: Starting in September 2024, the firm communicated to new and existing consignees reiterating the use of affected devices in patients aged 12 years and older. Additionally, the firm's letter informs consignees that the eIFU of devices indicates that no additional materials should be placed between the patient and Mega Soft pad (example: no sheet).

Recall details

Recall number
Z-2248-2024
Status
Open, Classified
Initiated by the company
Jun 17, 2024
Posted by the FDA
Jul 3, 2024
Recalling firm
Megadyne Medical Products, Inc.
Quantity
259 units
Product code
GEI
FDA submissions
K080741
Distribution
Domestic: Nationwide distribution International: AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, CANADA, CHINA, GERMANY, GREECE, INDIA, ITALY, NETHERLANDS, NORWAY, JAPAN, SAUDI ARABIA, SINGAPORE, SOUTH KOREA, SPAIN, THAILAND, TURKEY, UNITED KINGDOM, UNITED STATES,

Product description

Brand Name: MEGADYNE" MEGA 2000" Patient Return Electrode Product Name: MEGADYNE" MEGA 2000" Patient Return Electrode Model/Catalog Number: 0800 Software Version: N/A Product Description: Reusable Patient Return Electrode, Mega 2000, Adult, For Patients Weighing over 25lbs. (11.4Kg), 91cm Long x 51cm Wide Component: No

Code information (models, lots, serial numbers)

Product Code: 0800; UDI-DI: 10614559100936; Lot Number: All distributed lot numbers. Product code has been discontinued from future production.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

5 devices cleared under K080741 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event