Mega Soft recallZ-2250-2024 · Megadyne Medical Products, Inc. · initiated Jun 17, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Megadyne Medical Products, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
To limit the use of the Mega Soft Pads to patients aged 12 or older to mitigate the potential risk of patient burns identified with use of Mega Soft pads.
FDA root cause category: Unknown/Undetermined by firm
What the company is doing
An URGENT MEDICAL DEVICE CORRECTION letter, dated 6/17/24, was mailed to consignees. Consignees are instructed to share the recall notification with all users of affected devices and confirm their understanding, post a copy of the communication for user awareness, and forward the recall notification to any facility to which product was shipped. The notification reads that Mega Soft and Mega 2000 devices are only to be used for patients 12 years old. The notification also reminds consignees to follow proper cleaning, placement, and setup steps. Consignees are asked to complete and return the provided Business Reply Form by fax to 844-441-2732 or by email to [email protected]. Customers with any questions about the communication can contact Ethicon at 1-877-384-4266 from Monday through Friday from 8:00 AM EST to 5:00 PM EST. UPDATE: Starting in September 2024, the firm communicated to new and existing consignees reiterating the use of affected devices in patients aged 12 years and older. Additionally, the firm's letter informs consignees that the eIFU of devices indicates that no additional materials should be placed between the patient and Mega Soft pad (example: no sheet).
Recall details
- Recall number
- Z-2250-2024
- Status
- Open, Classified
- Initiated by the company
- Jun 17, 2024
- Posted by the FDA
- Jul 3, 2024
- Recalling firm
- Megadyne Medical Products, Inc.
- Quantity
- 1463 units
- Product code
- GEI
- FDA submissions
- K031285
- Distribution
- Domestic: Nationwide distribution International: AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, CANADA, CHINA, GERMANY, GREECE, INDIA, ITALY, NETHERLANDS, NORWAY, JAPAN, SAUDI ARABIA, SINGAPORE, SOUTH KOREA, SPAIN, THAILAND, TURKEY, UNITED KINGDOM, UNITED STATES,
Product description
Brand Name: MEGADYNE" MEGA SOFT" Dual Reusable Patient Return Electrode Product Name: MEGADYNE" MEGA SOFT" Dual Reusable Patient Return Electrode Model/Catalog Number: 0835 Software Version: N/A Product Description: Reusable Patient Return Electrode, Mega Soft, Dual Cord, Adult, For Patients Weighing over 25lbs. (11.4kg), 117cm Long x 51cm Wide x 1.25cm Thick Component: No
Code information (models, lots, serial numbers)
Product Code: 0835; UDI-DI: 10614559101872; Lot Number: All distributed lot numbers. Product code has been discontinued from future production.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Mega Soft 0835Megadyne Medical Products, Inc. · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES2 recalls4 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 94837).
- Z-2248-2024 · Open, ClassifiedBrand Name: MEGADYNE" MEGA 2000" Patient Return Electrode Product Name: MEGADYNE" MEGA 2000" Patient Return Electrode Model/Catalog Number: 0800 Software Version: N/A Product Descr…
- Z-2249-2024 · Open, ClassifiedBrand Name: MEGADYNE" MEGA SOFT" Reusable Patient Return Electrode Product Name: MEGADYNE" MEGA SOFT" Reusable Patient Return Electrode Model/Catalog Number: 0830 Software Version:…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page