Bivona recallZ-2613-2024 · Smiths Medical ASD Inc. · initiated May 29, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smiths Medical ASD Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.
FDA root cause category: Under Investigation by firm
What the company is doing
Smiths Medical issued an URGENT MEDICAL DEVICE NOTIFICATION to its consignees on 05/29/2024 via letter. The notice explained the issue with the device, potential risk and requested the following: "Customer Required Actions: When using the device, all instructions, including warning and cautions contained in the Instructions for Use Documentation must be followed with heightened awareness. Please complete the following actions listed below: 1. Check all inventory locations within your institution for the affected catalog numbers listed in the notification and discontinue use. Discard all affected products following your institution s process for discarding. If discarding is not immediately possible at your facility, then the product should be quarantined until disposal. 2. Share this notification with all potential users of the devices to ensure they are aware of this notification and proposed mitigations. If the devices are used at another location, please ensure this communication is delivered there. 3. Complete and return the attached Customer Response Form to [email protected] within 10 days of receipt to acknowledge your understanding of this notification. Please contact your local representative for assistance with replacement product and/or credit. 3. DISTRIBUTORS: If you have distributed potentially affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to [email protected]" Questions: Global Complaint Management - 1-(866)-216-8806 or US: [email protected] Customer Service: [email protected] or 1-(800)-258-5361
Recall details
- Recall number
- Z-2613-2024
- Status
- Open, Classified
- Initiated by the company
- May 29, 2024
- Posted by the FDA
- Aug 23, 2024
- Recalling firm
- Smiths Medical ASD Inc.
- Product code
- BTO
- FDA submissions
- K912469
- Distribution
- Worldwide distribution.
Product description
PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) PEDIATRIC TRACHEOSTOMY TUBE 2.5mm TTS FLEXTEND V NECK, Product Code/List Number/Item Code 67PFS25; b) PEDIATRIC TRACHEOSTOMY TUBE 3.0mm TTS FLEXTEND V NECK , Product Code/List Number/Item Code 67PFS30; c) PEDIATRIC TRACHEOSTOMY TUBE 3.5mm TTS FLEXTEND V NECK , Product Code/List Number/Item Code 67PFS35; d) PEDIATRIC TRACHEOSTOMY TUBE 4.0mm TTS FLEXTEND V NECK , Product Code/List Number/Item Code 67PFS40; e) PEDIATRIC TRACHEOSTOMY TUBE 4.5mm TTS FLEXTEND V NECK , Product Code/List Number/Item Code 67PFS45; f) PEDIATRIC TRACHEOSTOMY TUBE 5.0mm TTS FLEXTEND V NECK , Product Code/List Number/Item Code 67PFS50; g) PEDIATRIC TRACHEOSTOMY TUBE 5.5mm TTS FLEXTEND V NECK , Product Code/List Number/Item Code 67PFS55
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
a) UDI/DI 10351688518590, Product Code/List Number/Item Code 67PFS25, Lot Numbers: 3961256, 4119091, 4213406, 4374410, 4395754; b) UDI/DI 10351688518606, Product Code/List Number/Item Code 67PFS30, Lot Numbers: 3853403, 3874324, 3884244, 3907973, 3930052, 3959311, 3961752, 4060198, 4083906, 4091413, 4105249, 4107582, 4116088, 4138998, 4145680, 4180479, 4213407, 4328215, 4334909, 4348419, 4380074, 4388119, 4395755, 4400635; c) UDI/DI 10351688518613, Product Code/List Number/Item Code 67PFS35, Lot Numbers: 3853405, 3868268, 3871039, 3877670, 3887715, 3897877, 3914571, 3923745, 3936566, 3939652, 3942529, 3952412, 3959318, 3959319, 3981991, 3981992, 3997804, 4005592, 4015530, 4018503, 4026671, 4029496, 4033708, 4041961, 4048011, 4051671, 4053726, 4060196, 4077102, 4083926, 4089293, 4095927, 4110435, 4119103, 4125804, 4136125, 4145702, 4153579, 4161986, 4173727, 4180483, 4213408, 4217433, 4221336, 4237359, 4260057, 4279323, 4328216, 4334912, 4334915, 4352772, 4365786, 4369563, 4383606, 4388120, 4389875, 4393393, 4399138; d) UDI/DI 10351688518620, Product Code/List Number/Item Code 67PFS40, Lot Numbers: 3835787, 3853406, 3864263, 3864264, 3877654, 3884237, 3894877, 3901287, 3918018, 3920960, 3930015, 3931145, 3936583, 3939681, 3942528, 3952397, 3952398, 3959325, 3962585, 3981993, 3997805, 4008201, 4015531, 4029497, 4033709, 4035707, 4043938, 4051619, 4056251, 4060221, 4080783, 4086001, 4100895, 4107581, 4119124, 4136126, 4142427, 4145681, 4151010, 4156606, 4161963, 4171578, 4180476, 4213409, 4221337, 4237360, 4240130, 4254221, 4276451, 4283222, 4328217, 4331573, 4334920, 4344978, 4357248, 4362850, 4369572, 4388122, 4389876, 4395758, 4399139; e) UDI/DI 10351688518637, Product Code/List Number/Item Code 67PFS45, Lot Numbers: 3861137, 3868253, 3880679, 3887722, 3887723, 3904686, 3923731, 3936584, 3942523, 3952394, 3962590, 3981994, 3991132, 3997806, 4005593, 4018504, 4029498, 4041962, 4051665, 4060195, 4063170, 4077101, 4083932, 4089305, 4092641, 4097828, 4100890, 4119096, 4122440, 4130981, 4142451, 4151051, 4161982, 4171579, 4179342, 4217434, 4237361, 4240131, 4260058, 4276452, 4328218, 4334921, 4344979, 4355426, 4380075, 4383607, 4388123, 4389877, 4395761, 4399140; f) UDI/DI 10351688518644, Product Code/List Number/Item Code 67PFS50, Lot Numbers: 3868275, 3868276, 3874297, 3880681, 3894878, 3897867, 3918019, 3923726, 3930016, 3939676, 3942518, 3942519, 3952392, 3959323, 3962589, 3981995, 3987958, 4005595, 4011364, 4018505, 4029499, 4035708, 4053704, 4063171, 4077073, 4089307, 4097830, 4110436, 4116105, 4128042, 4136132, 4145688, 4153624, 4159962, 4179343, 4180474, 4213410, 4221338, 4283223, 4286795, 4322659, 4329838, 4334220, 4362864, 4388124, 4388125, 4388126, 4393394; g) UDI/DI 10351688518651, Product Code/List Number/Item Code 67PFS55, Lot Numbers: 3853408, 3864260, 3871050, 3877655, 3884263, 3891530, 3904671, 3904672, 3920942, 3930012, 3939670, 3939671, 3952349, 3965908, 3981996, 4008202, 4015532, 4018506, 4026690, 4035709, 4048025, 4053734, 4063172, 4077089, 4083905, 4096833, 4107580, 4110437, 4128016, 4138999, 4142428, 4151012, 4159955, 4163853, 4173728, 4179344, 4185410, 4196262, 4213411, 4279324, 4286796, 4331580, 4334931, 4352773, 4374411, 4380999, 4388127, 4389878, 4395762, 4399141
Devices named in this recall
7 devices in the FDA device registry whose identifier appears in the recall's code information.
- Bivona 67PFS50Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall35 reports
- Bivona 67PFS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall31 reports
- Bivona 67PFS45Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall23 reports
- Bivona 67PFS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall19 reports
- Bivona 67PFS55Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 reports
- Bivona 67PFS30Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall9 reports
- Bivona 67PFS25Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall0 reports
Product lines
Other devices under the same FDA submission
26 devices cleared under K912469 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Bivona 60P045Icu Medical, Inc. · Tube tracheostomy and tube cuff2 recalls12 reports
- Bivona 60SP040Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall20 reports
- Bivona 60SP035Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall3 deaths19 reports
- Bivona 60P055Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall16 reports
- Bivona 60SP050Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall13 reports
- Bivona 60P040Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall12 reports
- Bivona 60SP045Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall8 reports
- Bivona 60P050Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall7 reports
- Bivona 60SN040Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall5 reports
- Bivona 60SP055Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall5 reports
- Bivona 60SN035Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall4 reports
- Bivona 60SP030Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall4 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 94852).
- Z-2588-2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) TTS TRACHEOSTOMY TUBE 7.0MM TIGHT TO SHAFT CUFF, Product Code/List Number/Item Code 670170; b) TTS TRACHEOSTOMY TUBE 8.0…
- Z-2589-2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) AIRE-CUF TRACHY TUBE 6.0MM MID-RANGE , Product Code/List Number/Item Code 750160; b) AIRE-CUF TRACHY TUBE 7.0MM MID-RAN…
- Z-2590-2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) NEONATAL TRACHEOSTOMY TUBE , Product Code/List Number/Item Code 60N025
- Z-2591-2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) NEONATAL TRACH TUBE , Product Code/List Number/Item Code 60N030; b) NEONATAL TRACH TUBE , Product Code/List Number/Item…
- Z-2592-2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) NEONATAL TRACHEOSTOMY TUBE 2.5FLEXTEND PLUS , Product Code/List Number/Item Code 60NFP25; b) NEONATAL TRACHEOSTOMY TUB…
- Z-2593-2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: 2.5MM UNCUFFED NEONATAL FLEXTEND PLUS TRACHEOSTOMY TUBE W/NEO STRAIGHT STRAP, Product Code/List Number/Item Code 60NFPS25
- Z-2594-2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) 3.0MM FLEXTEND NEO STRAIGHT TRACH TUBE , Product Code/List Number/Item Code 60NFPS30; b) 3.5MM FLEXTEND NEO STRAIGHT T…
- Z-2595-2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) PAED. TRACHEOSTOMY TUBE 2.5MM UNCUFFED , Product Code/List Number/Item Code 60P025; b) PAED. TRACHEOSTOMY TUBE 3.0MM UNC…
- Z-2596-2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) PAED. TRACHEOSTOMY TUBE 4.0MM FLEXTEND PLUS, Product Code/List Number/Item Code 60PFP40; b) PAED. TRACHEOSTOMY TUBE 4.0M…
- Z-2597-2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: 6.0MM FLEXTEND PEDIATRIC PLUS VEE TRACH TUBE , Product Code/List Number/Item Code 60PFP60
- Z-2598-2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) 4.0MM FLEXTEND PEDIATRIC PLUS STRAIGHT TRACH TUBE , Product Code/List Number/Item Code 60PFPS40; b) 4.5MM FLEXTEND PEDI…
- Z-2599-2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) PAED. TRACHEOSTOMY TUBE 2.5MM STANDARD FLEXTEND PLUS , Product Code/List Number/Item Code 60PFS25; b) PAED. TRACHEOSTOMY…
- Z-2600-2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: 2.5MM PEDIATRIC UNCUFFED FLEXTEND PLUS TRACHEOSTOMY TUBE WITH NEONATE STRAIGHT NECKSTRAP, Product Code/List Number/Item Code…
- Z-2601-2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) 3.0MM FLEXTEND PEDIATRIC STANDARD STRAIGHT TRACH TUBE , Product Code/List Number/Item Code 60PFSS30; b) 3.5MM FLEXTEND …
- Z-2602-2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) 2.5 UNCUFFED NEONATAL TRACHEOSTOMY TUBE , Product Code/List Number/Item Code 60SN025; b) 3.0 UNCUFFED NEONATAL TRACHEOST…
- Z-2603-2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) NEONATAL TRACHEOSTOMY TUBE , Product Code/List Number/Item Code 65N025; b) NEONATAL TRACHEOSTOMY TUBE , Product Code/Lis…
- Z-2604-2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) PAED. TRACHEOSTOMY TUBE 2.5MM AIRE-CUF , Product Code/List Number/Item Code 65P025; b) PAED. TRACHEOSTOMY TUBE 3.0MM AIR…
- Z-2605-2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: 2.5MM AIRE-CUF NEONATAL TRACHEOSTOMY TUBE , Product Code/List Number/Item Code 65SN025; 3.0MM AIRE-CUF NEONATAL TRACHEOSTOM…
- Z-2606-2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) TTS NEONATAL TRACH TUBE , Product Code/List Number/Item Code 67N025; b) TTS NEONATAL TRACH TUBE , Product Code/List N…
- Z-2607-2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: TTS NEONATAL TRACHY TUBE 4.0MMTIGHT TO SHAFT CUFF , Product Code/List Number/Item Code 67N040
- Z-2608-2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) NEONATAL TRACHEOSTOMY TUBE 2.5mm TTS FLEXTEND V FLANGE , Product Code/List Number/Item Code 67NFP25; b) NEONATAL TRACHEO…
- Z-2609-2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) NEONATAL TRACHEOSTOMY TUBE 2.5mm TTS FLEXTEND STRAIGHT FLANGE , Product Code/List Number/Item Code 67NFPS25; b) NEONATA…
- Z-2610-2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) TTS PEDIATRIC TRACH TUBE , Product Code/List Number/Item Code 67P025; b) TTS PEDIATRIC TRACH TUBE , Product Code/List Nu…
- Z-2611-2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) PEDIATRIC TRACHEOSTOMY TUBE 4.0mm TTS FLEXTEND PLUS V FLANGE , Product Code/List Number/Item Code 67PFP40; b) PEDIATRIC…
- Z-2612-2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) PEDIATRIC TRACHEOSTOMY TUBE 4.0mm TTS FLEXTEND PLUS STRAIGHT FLANGE , Product Code/List Number/Item Code 67PFPS40; b) PE…
- Z-2614-2024 · Open, Classifieda) PEDIATRIC TRACHEOSTOMY TUBE 2.5mm TTS FLEXTEND STRAIGHT NECK , Product Code/List Number/Item Code 67PFSS25; b) PEDIATRIC TRACHEOSTOMY TUBE 3.0mm TTS FLEXTEND STRAIGHT NECK , Pr…
- Z-2615-2024 · Open, ClassifiedPORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) 2.5mm TTS CUFFED NEONATAL TRACHEOSTOMY TUBE , Product Code/List Number/Item Code 67SN025; b) 3.0mm TTS CUFFED NEONATAL T…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page