Centricity recallZ-3214-2024 · GE Healthcare · initiated Aug 5, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Healthcare or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
GE HealthCare has become aware of an issue where image storage is not possible when Centricity PACS reaches the maximum number of unique image identifiers (Ckeys). If this occurs, it can result in acquisition of a partial study.
FDA root cause category: Device Design
What the company is doing
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 08/05/2024 by letter delivered via traceable means. The notice explained the issue, potential risk to the patient, and requested the following: "You can continue to use your device if it has not reached the maximum number of Ckeys. In order to determine the remaining number of available Ckeys: " If GE HealthCare is remotely monitoring your Centricity PACS system as part of an existing service agreement, the number of available Ckeys is assessed routinely. If your device is close to maximum capacity, the GE HealthCare support team will contact you to discuss next steps. " If GE HealthCare is not remotely monitoring your system, please contact the GE HealthCare support team who will help you check your Ckey status. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions above." "GE HealthCare will provide an updated manual with CKey monitoring instructions at no cost to you. A GE HealthCare representative will contact you with further information." For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Recall details
- Recall number
- Z-3214-2024
- Status
- Open, Classified
- Initiated by the company
- Aug 5, 2024
- Posted by the FDA
- Sep 24, 2024
- Recalling firm
- GE Healthcare
- Quantity
- 2138 units
- Product code
- LLZ
- FDA submissions
- K110875, K043415
- Distribution
- Worldwide distribution.
Product description
GE Healthcare Centricty PACS, V2, V3, V4, V6, V7; Radiological imaging processing system.
Code information (models, lots, serial numbers)
Show the full code information (29k characters)
All versions of Centricity PACS: V2, UDI/DI Not Applicable; V3, UDI/DI Not Applicable; V4, UDI/DI 00840682124430; V6, UDI/DI 00840682104807; V7, UDI/DI 00840682145572 installed on the following systems: CARTI02312611, CARTI22214087T, 501268SSASW, 501268SSATSW, 928757VPASW, 928757VPBSW, 928757VPTSW, 602344PVSW2, 602344REPSITEB, MARICOPAC02308029, 602344TIMSVM, 480323PVSW, 661633PVSW, 805395PVSW, 805395PVTSW, V5SPORTS01804913, 310349PACSSWPV, UCLAMC2259090, 310349PACSSWBPV, UCLAMC2259063, 831624PVASW, UHSD02184496, 916781CPACSSRWT, PALMDALEM01810792, PALMDALEM16023955, UHSD16084343, PALMDALEM20508297, PALMDALEM21178954, MHL02186470, 909788HCAPACS1, 916781CPACSCKS, 916781CPACSBCKS, 650696PACSEESW, MPHS2259091, GSHCACPSW, GSH320037250, VSMY01820423, 209532SESW, SONORAREG19607823, RANCHO02255039, RANCHOPISU2, RANCHOPTESTISU, To be Provided , 805370HCALR, VMPACS69, LOSROBLES19666790, SOUTHWEST01815383, MCANPACSW, MCASPACSW, CENPACSW, 303329HCAPSWS, 303584HCAPVSW, 303584REPBSW, 303584HCAPVSWTESTIMS, SRMCPACSW, 719584VPSW, PKV2259020, 719584VPBSW, NSMCPACSW, 970476PVSW, VVMC15111181, 302674PACSSW, 302674PACSBSW, 302736SAEATSW, JFKSW, JFKR19980220B, INFINITYW01452121, 727462PVSW, 727462PVBSW, MPMH2259089T, 772461HCAPACSW, 772461HCAPACSW1, 772461HCAPACSWT, FWMCCPTSW, 352333PACSSW, NFRH19663978B, 904202BMCDSW, 904202BMCSSW, MJPACSW, MJPACTSW, 904202BMCDTSW, BIHT20915070-IMSTest2, BIHT20915071-IMSTEST4x, 407518COREPACS, HH1001398036, 407518CPREPB, 863680PVSW, KRMCOREPACSSW, KRMC19478073, 407888HCASWSS, 615344HCASW, PCMC420084442, 561694CAPACSSW, WFLHIV01, WESTFLORI19661467, 772335PACSSW, 772335PVSW, 941637PVSW, 941637TIMS90, 850875PVSW2, 904819PVSW, 904819PV2SW, 904819PVTSW, 727519PVBSW, 850325PACSCORE, CRMC820084433, 888663CPBSW, USFIMSSSASW, 727519PVSW, VSMAE01817090, NGDC01893569, 888663CPTSW, 888663CPB2SW, USFIMSSSABSW, USFIMSSSATSW1, USFIMSSSATSW2, 727519PVMSW, 727519PVTSW, VAMAF02345966, VSMI18957442DCC, 407518SCPACSHC, 850325HCASWOPT, WFLHIVT01, PCMC42259036, 352333PACSTSW, VSMAM02004646, WELLINGTON01810730, UHSD02289750, WELLINGTON19678279T, WELLINGTON20036598, TRMPACSW, TRMNC20021605, 100115-8-Centricity PACS-21222171, TRMTESTPACS, Doctors Ho22222417, 407772CORESW, 404605PVBSW, 404851PVSW, 404851PVTSW, 404851PVBSW, 404605PVTSW, 407772COREBSW, 407772CORETSW, 706651HCACORE, 912466PVSW, SGHS2259039, 912276PVSW, 912276PVTSW, 706278PACSSWPV, 706278PACSSWBPV, 706278PACSSWTPV, 404501PACSSWPV, 404501PACSSWBPV, 404501PACSSWATPV, 404501PACSSWBTPV, 706754PACSSW, 404605PVSW, 478633PACSSWPV, MCG2259011, MCG20084728, VSMQL01987331, 404325PACSSWPV, 404325PACSSWBPV, 404325PACSTSW, 406238PVSW, VSMZ01812192, 641422PVSW, PROSPECT2259057, PROSPECT2259084, PROSPECT22128566, 319272SCPSWULP, 319272CPDRSW, 319272CPDRBSW, 319272SCPTSWULP, 208COREPVSW, EID19660790, 208CORETIMS, 208743PVSW, 208743PVTSW, 847NWCPVSW, 847NWCPVSW2, 773880PVSW, 795079PACSEESW1, 795079PACSEESW2, 312926PACSSWPV, 312996PVSW2, JSHPACSSW, 312996PVSW1, 312996PVSWT1, 312996PVSWT2, 795079PACST, JHSJHOCC02259030, 773880PVBSW, 773880PVTSW, JSHPACSTSW, 217876PVSW, 217876PVBSW, 815285PACUV100, 630963PVSW, ADVPACSTSW1, 847437PVSW, 847437DCPVSW, 847437SAPVTSW, 847990PVSW, S046PACSSWPV, S801PACSSWPV, S046PACSSWTPV, ADVPACSSW, ADVPACSBSW, ADVPACSTSW, 815398CPACSSW, 219325PVSW1, 219325PVBSW1, 219325PVTSW1, 765983PVSW, 765983PVSW2, RHHCS02253964T, 812232PACSPV, 812232PACSTPV, 812232PACS2PV, THRH20084364, FRNKHAPACSPV, FRNKHAPACSTPV, FRANKFORT19633100, RRMCPACSW, RGMCPACSTW, RRMC120080697, VTP01810617T, VHT02334235P, VTP19617532, 985867CORESW, 985230PVSW, NOMC2259032, CFO18227240, 985230PVSWS, 504903CPACSSW, 504988SSASW, 504903CPACSTSW, 504988SSATSW, MCOLANO20039031, HTMC20085961, 504903CPACSIMS, 617638SSASW, 617638SSBSW, 617638SSATSW, 781979PVSW, 781979PVBSW, 781979PVTSW, 413447BCPVSW, 413447HCPVSW, 413447PVSW, 413748PVSW, 413748PVTSW, 413748PVBSW, 301698PVSW, 301698PVTSW, NORTHE02315652, NORTHE02370307T, 207795PVSWC3, 207795PVSWC1, 207795PVSWC2, 517353PACSEESW, 517353PACSEETSW, 906387VPSW, 248858PVASW, 248858PVBSW, 248858PVTSW, 906358PACUV100, 507288PVSW, 507288PVSWT, OMCR19625746, 651855PVSW, 651855PVBSW, 612831PVTSW, 660885PVSW, 660885PVTSW, S011PACSSWPV, S011PACSBSW, S011PACSTSW, 417623PVSW, 417623PVTSW, 417623PVBSW, BARTONCOU01819621, S009PACSSWPV, S009PACSSWBPV, S009PACSSWTPV1, S009PACSSWTPV2, 601824PVSW, 601859PVSW, MCMC2259045T, 601200PACSEESW, 601200PACSSWPV, 601376PVSW, 601936PACSSWPV, SDJM20032054, 601883PVSW, 601883PVTSW, DBCSW, 406345SCPACSTSW, 828326PVSW, 704660PVSW, 704660PVTSW, 402484PACSSW, AMIM02031832, 603433TSW1, 603433SW, PRH119666811, 609652PACSSW, 609652PACSBSW, 609652PACSTSW, 609835PACSSWPV, 609835PACSTSW, 609835PACSSWBPV, VHGPVSW, VHGPVBSW, VHGPVTSW, 856342CPACSSW, 856342CPACSBSW, 856342CPACSTSW, 973490PVSW, 973490PVTSW2, 973490PVTSW1, UMDNJPVPSW, UMDNJPVPBSW, UMDNJPVPTSW, 609383UVPACS100, 908522PVSW2, ARTHRITIS01452981, 505609PVSW, SJRMIMST20655929, 702733CPACSSW, SHPACSW, SUNPACSW, VALLEYHEA14460287, SUNPACSREPB, 20910879CPACST, NORTHERNN20284710, NORTHERNN20838223, NORTHERNN20838225, BLH01442255TST, BLH01451970, STO01440342, STO02157029, STO02157030T2, 718206PVSW, 718206PVTSW, 2435-1-Centricity PACS-20608371, 718206REPBSW, 315798PVSW1, 315798PTISU, COLPVPSWIMSA, MSKPVSW, MTSPACSSW, NEWYORK01976546, VSMAT01986937, MSI02099022, COLPVPSWIMSB, COLPVPSWIMSAT, COLPVPSWIMSBT, MSKPVBSW, MSKPVTSW2, MSKPVTSW3, MTSPACSBSW, ERM16295285, MSI19614991, 718920PVSW, 718920PVTSW1, BROOKHAVEN15531196, OAR01373900, 516763PVSW, 516763PVBSW, 516763PVTSW1, 516763PVTSW2, STO18806599, 914357PVSW, GSHP01370055, 914357PVBSW, 914357PVTSW, 315470CPACSASW, 315470CPACSBSW, 315470CPACSTSW, CRH21306236, 914681PVSW, 914681PVTSW, 718920PVYSW, 718920PVTSW2, 330543PACSSWPV, CHPA2259061, CHPA2259088T, 513747PVSW, 614234PVSW, 614722PVSW, MARKHZAN01465406, 614722PVBSW, SAH52259083, SAH52259110T, 614722PVTSW, 740348PACSSW, 740348PACSBSW, 740348PACSTSW, 419383PVSW, MUOO2259017, 419383PVSWB, STMARYS-TISU, STMARYS-PISU, S040PACSSWBPV, S040PACSSWAPV, S040PACSSWTPV, 503257PACSSWPV, UHS101383446, PAMC19604625T, TMGSSASW, 215913CPBSW, 484526PVSW1, 484526PVTSW, 484526CPREPB, 215710PVSW, SMMC14898807, QUANTUMIM14640640A, QUANTUMIM14776702B, QUANTUMIM14776726T, 570523PVSW, EVAN02302786T, EVCH22176253, EVANVTPACSIMS, 724728SSASW, 724728SSABSW, 724728SSATSW2, STMARYME01476581, 412575CPACSSW, 412623CPBSW, UHS100555969, UHS101404909, 412623CPTSW, 215453PACSSWPV, GVH102398920T, GVH121323962, 401444PVBSW, 401444PVSW, 401444PVTSW, AIMC01810707, AIMC19632995, 864716PACSSWPV, 864716PACSSWT, 864716PACSSWBPV, TRMT2PACSW, 803461PACSSWPV, SCOA2259023, 843347PVSW, 843347PVTSW, CNOS02186078, 605353SSASW, 605333PVSW, 605882PVSW, 423493PACSW, 423553VPSW, 423855PACSW, 423493REPB, WTN00104692T, WTN2259070, WTN2259097, WTN21176121, 423837HCASWSU, 423431PVSW, 865977SSASW, 865977SSTSW, 806354CPACS, 806354CPACSB, NORTHWEST21314320, 806354CPACST7, 361694VPSW, DRISCOLLC02257225TIMS, 214820EEPVSW1, BSWNPACIMSDB1TR, SCW2259068, FORTDUNCA01810773, FORTDUNCA20854738T, FORTDUNCA21337445, DSMCCOREPACSSW, DSMCCOREPACSDRSW, DSMC-22886533, 214820PVSW, 972906GSPACSSW, HCA-NORT02332691, MEDIC01812756, 281274MSLPVSW, 713441MHPVSWA, 713704MHMSWA, 713704MHMSWB, 713704MHMSWTA, 713704MHMSWTB, 713741MHHPVSWB, VSMHE01893552, DHOL01810756, DHOL15291375T, DHOL20631410, DHOL23253841, 713770CVPACS, 210670PACSSWPV1, 254918PACSUV100, 214820EEPVSW2, BSWNPACIMSDB2T1, 254724PVSW, SCW2259095, VASOY02315718, 361572SSASW, CMDC2259015, HH1001979502, 801DASSW, 801DSWSP1, 801377PVSW1, 801SOFTWARE, 801SOFTWAREOPT, SMH820071869, 801924HCADASSW, MNTSW, 434654PVSW, APICPAX1CS07, 540433PVSW, CFHCAPACSW, RHC123290092, CMCPACSW, 352867PACSSW, HDHPACSW, CMCPACSTW, FHS219676614, CH721013826, CH721013832T, LGMSW, LGHI20084372, LGMSWOPT, RESTRAPVSW, RHC120038618, STSP21736781, STSP22125863, 802775SSASW, 802775SSATSW, 802775SSATSW4X, KVCH01462777, 206292PACSSWPV, 206292PACSSWTPV, 206364PVSW, 206545EESW, 206364PVBSW, UWTDB101, UWH20021873REPB, UWTDB110, 414465PVSW, 414465PVTASW, 414465PVTBSW, 414465PVBSW, PCHT01721604, 608745PVSW, COMMUNITY01973590, 304487PVSW, PCH12259035-TISU, 304536UVSW, DZ1906PP02, DZ1381PP01, DZ27??PP03, DZ2112CC01, To be Provided , To be Provided , To be Provided , HCIT4691855CPACS, HCIT1555339PACS, HCIT4825134PACS, HCIT1524287PACS, HCIT4691855PACS, HCIT1596840PACS, HCIT4430774CPACS, HCIT5184728CPACS, HCIT4859945PACS, 0910620047, 0910620047, 0910620047, 0910552308, 0910552310, 0910553037, 0910553037, ANZVICCAUL, ANZVICSAND, AH0104CS01, AH5045CS01, BH1038PC01, 22303605, To be Provided , 130041ST02, 030151ST09, 010120ST03, 010120ST52, 300013ST04, 300013ST09, 300013ST11, 300013ST21, 300013ST24, 441236VPSW, VSMAI01812924, HCIT4691866CPACS, HCIT4759526PACS, HCIT7102800PACS, HCIT7125379CPACS, To be Provided , HCIT5245250CPACS, HCIT5245250CPACS, HCIT4860936PACS, HCIT4432783PACS, HCIT5317135CPACS, HCIT4290686PACS, HCIT5245275CPACS, 323782-Centricity PACS-21288445, HCIT4790060PACS, HCIT4769017PACS, HCIT7203852PACS, HCIT4478566PACS, HCIT7134190CPACS, HCIT4568867PACS, HCIT7104315PACS, HCIT7103264IMS, 656687-Centricity PACS-21288460, HCIT5273606CPACS, HCIT5245248CPACS, HCIT5245248TESTCPACS, HCIT7229698PACS, 604520REPSITEA, 604520GEPACS, 604520TESTIMS, 604586REPSITEB, 709737PACS, 709752CORDMODA, 709752IMSAsw, 709752IMSBsw, 709752ITS1A, 709752ITS2B, 709752REPSRVRAsw, 709752GEPACS, 709752DCSDAS02sw, 709752DCSDAS04sw, 709752TESTIMSsw, 709752DASTST01sw, 709737DASEHOUTsw, 709752REPSERVERsw, 709752DCSDAS01sw, 709752DCSDAS03sw, 709752TDASNL01, 709752TDASNL02, 709752DASNLA01, 709752DASNLA02, 709752DASNLA03, 709752DASNLA04, 709752DASNLA05, 709752DASNLA06, 709752DASNLA07, 709752DASNLA08, 709752DASNLA09, 709752DASNLA10, 709752DASNLA11, 709752DASNLA12, 709752DASNLA13, 709752DASNLA14, 709752DASNLA15, 709752DASNLA16, 709752DASNLB01, 709752DASNLB02, 709752DASNLB03, 709752DASNLB04, 709752DASNLB05, 709752DASNLB06, 709752DASNLB07, 709752DASNLB07, 709752DASNLB08, 709752DASNLB09, 709752DASNLB10, 709752DASNLB11, 709752DASNLB12, 709752DASNLB13, 709752DASNLB14, 709752DASNLB15, 709752DASNLB16, 709737UTILITYsw, 709752ITS2sw, 709752ITS1sw, 709752CORDMODsw, 709752DASTST02sw, 709637DAS01sw, 709637DAS02sw, 709637DAS03sw, 709637DAS04sw, 709637DAS05sw, 709637DAS06sw, 709637DAS07sw, 709637DAS09sw, 709637DASWH10sw, 709637DASWH11sw, 709637DASWH12sw, 709637DAS10sw, 709637DAS8sw, 709637DAS11sw, 709637DAS12sw, 709637DAS13sw, 709637UTILITYsw, 709637CDPSsw, 709651DASGA01, 709651DASGA02, 709651DASGA03, 709651DAS01sw, 709651DAS02sw, 709651DAS03sw, 709651UTILITYsw, 709292DAS01sw, 709292DAS02sw, 709292DAS03sw, 709292DAS04sw, 709292UTILITYsw, 709292PACS, 709454PACS, 519843IMS, 519843TESTIMS, 519843GEPACS, 905522GEPACS, 905521GEPACS, 905521IMSA, 905522IMSB, 905521TESTIMS, 613548IMSA, 613548IMSB, 613548TESTIMS, 613548GEPACS, 519685REFERENCEIMS, 519685REFERENCEDAS, 905575IMS, 905576REPSITEA, 905576REPSITEB, 905576GEPACS, 905576TESTIMS, 613732GEPACS, 613732IMS, 613732TESTIMS, 519426IMS, 519426TESTIMS, 519426GEPACS, 416813REPSITEB, 416813REPSITEA, 416813TESTIMS, 416813GEPACS, 416326IMS1, 416326IMS2, 416326GEPACS, 416756GEPACS, 416756TESTIMS, 416756IMS, 905372GEPACS, 905372IMSA, 905372IMSB, 905372TESTIMS, HCIT4602438PACS, HCIT4358052PACS, HCIT4949579PACS, HCIT4920063PACS, HCIT4096494PACS, HCIT4567313PACS, HCIT4601968PACS, HCIT5026141PACS, HCIT5183890CPACS, HCIT4888543PACS, To be Provided , To be Provided , To be Provided , To be Provided , 082499110172, 082499310155, 082455210055, 082455150052, 082455150052, 082455160221, 082455160221, 082455160134, 082499180109, 082499180102, 082455160255, 082427190180, 082445110005, 082403010015, 082455190146, 082499130790, 082499130125, 082499130790, 082499020074, 082455040339, 082407040684, 082407040684, 082455040204, 82455040334, 082499040230, 082455040337, 082455040339, 082455040147, 082455040274, 082455040352, 082445040028, 082445040028, 082455040317, SiteID-Centricity PACS-20608297, 082407040684, 082499110183, 082455190147, 82499220115, 082499100382, 082455150074, 082455150068, 082455150039, To be Provided , 082455180040.082499180080, 4807212, 082455170088, 082499010124, 082499110172, 082499060054, 082499100386, 082499100386, 082499100386, 082499100386, 082455100286, 082499250135, 082499250042, 082455070037, 082455120548, 082455120548, 082455100352, 082455100179, 082455100179, 082455160241, 082455200034, 082433120225, 082499120374, 082455120507, 082499250135, 082455180041, 082455120545, 082455120552, 082455120510, 082499100188, 082499100188, SiteID-Centricity PACS-20595026, To be Provided , 082455130065, 082455130335, 82455290014, 082499130780, 082455130366, 082455130394, 082455290015, 082499180085, 082445030013, 082455160035, 082499180106, 082499130792, 082455120226, 082455080025, 082455050079, 082455050076, 082455050069, 082455120371, 082455130415, 082427150109, 082455060004, 082455120544, 13767625-Centricity PACS-21273767, 082499130816, 082499130826, 082407120583, 082455100357, 082407100616, 082499100385, 082455100353, 082499100415, 082499230063, 082499260011, 082499140323, HCIT4825460CPACS, 2939580-Centricity PACS-21294025, HCIT5231090CPACS, HCIT4668793PACS, HCIT50000107PACS, HCIT1516613IW, HCIT1543321PACS, HCIT4946501PACS, 40549-1CP01, 40549-1CP01, DK9022PAC1, DK9154CA01, DK9066IMS1, DK9066TESTPAC1, HCIT4277575PACS, HCIT4692615PACS, HCIT4796839PACS, EG1526PP02, EG1017PP02, EG1055PC01, HCIT5162696CPACS, ZA2879PP01, M4477025, X129763408, A4167665, 22106615, A4498636, M61421909, M100323902, M2030921, M2753440, M4171826, M4476437, M2001113, X59655539, To be Provided , M4206540, M5172536, M5172537, M5172568, P4172826, P4172827, M178185410, X387005305, MQ1234502, M386578303, A27411, M4144112, M4045536, M4046544, M4047061, M40480163, M4188439, B4191728, C4191734, D4191718, E4191720, M54807112, NC1028, To be Provided , To be Provided , M4491536, PF1000PC02, M143153109, M4143924, M4143984, M220995201, 02398902, M4825009 / M4825010 / M4825011, M4204530, H4858984, D4858905, D4858906, D4858918, M838835602, M9164747, M4209925, M4209926, M4209927, Utuora, HNO, M4505229, HC0399PA01, HC2641CS01, tobecreated, 530019CS01, PC0075CS01, tobecreated, 744459PA02, HC0220CS01, 518086CS01, HC2720CS02, HC4017CS01, HC0074CS01, HC1627CS01, HC4438CS01, HC0881CS01, UC2847CS01, UESS01CS01, 126292CS01, UJJF01CS01, 416005CS01, HC4912CS01, 530366PC03, UC1220CS01, HC0468CS02, 556162CS01, UHAM02CS02, HC0727CS01, tobecreated, UHAN01CS01, 362520SV01, HHAN01SV01, HHEI01CS01, HHEI02CS01, UC1989CS01, HC7048CS02, 410014SV02, HC1992CS01, HC1282CS01, HC2554CS01, tobecreated, HC0847CS01, HC3019SV03, 624751CS01, HC0014CS01, To be Provided , HC1467PA01, 560496CS01, 145774CS01, HCNE01CS01, HC0094CS01, 553160CS01, 434487CS01, HCSC01CS01, HC3076CS01, UC4372SV02, HC0863CS01, 424881PA01, GH1069PP01, GH1022PAC1, HCIT4494109PACS, HCIT4494109UV, HCIT523983PACS, 828550053, 0828550074, 0828550085, 0828550070, 828550079, 0828550105, 0828550072, 828550052IB01, 828550064, 828550066, 828550052IB01, 828550095, 0828550073, 828550057, 083055283270418, 083055247345516, MAX-DC-PACS, 083055820666618, MCCT-CPACS, 083055863242715, 083055868428111, SCTIMST-IN-CPACS, RCC - CPCS 7.0.0.4.8, 083055201766451, 083055204252715, Wockhardt-CPACS7.0.0.4.8, INHS-ASWINI-CPACS, 083055207081618, AIIMS-NM-PACS, 083055003435816, SIRHNHOSPITALTRUST-CPACS, 083055211206517, QUALITYCARE01, 083055803467916, 083055801005510, 083055806995816, 083055801058716, JIPMER01820558, 083055808240917, NARUVI -CPACS, 083055840003710, CHL-CPACS-6.0.10.3, 083055604732915_CPACS, RSUDMadaniPalu-CPACS, PTUROGEN01, MayapadaBandung-CPACS, TEASIAHEALTHCARE-CPACS, PTGLOBALMP-CPACS, PTMKMas02, PTSentosa1, SiteID-Centricity PACS-20535377, SANTOSA-CPACS, MKCikarang-CPACS, MKBekasiBarat-CPACS, MKBEKASITIMUR_CPACS, MITRADEPOK001, RSUDTarakan-CPACS, RSUDSlawi-CPACS, RSUDTarakanKaltara-CPACS, PTMULYA-PACS, RSReginaMaris-CPACS, RSUDKlungkung-CPACS, RSPBINTARO2, RSMKG-SERPONG01, MITRAKELUARGATEGAL-CPACS, RSUDKabupatenB-CPACS, RSUDMukomuko-CPACS, RSUDCIAWI-CPACS, RSUDPadangan-CPACS, RSPOLRIB-CPACS, RSMITRAKD-CPACS, PTUNISIAMEDIKAFARMA_CPACS, RSMK-KelapaGading01, RSUPFATMAWATI-CPACS, RSKELUARGAKALIDERES-CPACS, RSHARAPANKITA-CPACS, RSPJATINNEGARA1, PREBO001, SULIANTI001, MitraKeluargaSlawi-CPACS, RSMKKEMAYORAN-CPACS, RSMK-Cibubur01, RSUDPandanArang-CPACS, RSMardiRahayu-CPACS, RSALDr.Ramelan-CPACS, RSUDKKM-CPACS, RSUDPasirian-CPACS, RSUDProvinsiNTB-CPACS, RSHajiSurabaya-CPACS, RSUDBHAKTIDHARMAHUSADA-CPACS, RSPHCSurabaya-CPACS, RSJDSurakarta-CPACS, RSUDSOREANG-CPACS, RSUDDRHKB-CPACS, RSUDHMB-CPACS, RSDDRSJ-CPACS, RSUPSOERADJI-CPACS, RSUDDRMAP-CPACS, Dr.MOHAMMADZYN_CPACS, RSUDAMDS-CPACS, MitraKeluargaDeltamas-CPACS, RSUDBEKASI-CPACS, RSUPM.DJAMIL-CPACS, RSUDPalembang_CPACS, RSMITRAKK-CPACS, RSMITRAKELUARGABINTARO-CPACS, RSUDSIJUNJUNG-CPACS, RSUDBERKAHP-CPACS, RSUDPANDEGA-CPACS, RSUDMThDjamanS-CPACS, RSUPKARIADI-CPACS, RSUDKRMTW-CPACS, Sidoarjo-SysID, RSUDSidoarjo-CPACS, RSMITRAKW-CPACS, RSUDTuban-CPACS, SYAMSUDIN001, RSPSURABAYA1, RSUPDr.Wahidin-CPACS, HoesinPALEMBANG-CPACS, PTSEJANU01, BETHSAIDA-CPACS, MitraKeluargaPamulang-CPACS, RSUDLeimina-CPACS, RSUDWALED-CPACS, RSMITRA-CPACS, PTHNU02, KeluargaGrandWisata-CPACS, RSMKSS-CPACS, MitraKeluargaJatiasih-CPACS, RSBinaHusadaCibinong-CPACS, IQ1906PC03, IQ1727PC03, 81100DAS05, IL1018IT01, IL1059IT01, To be Provided , IL1484IT01, IL1469IT08, IL1098IT10, IL1098IT18, To be Provided , TBA, D5104305, B5104381, A51043144, A5802803, C5368613, A5368605, A5801809, B5368603, A5630621, A5121809, TBA, tbc, A5353013, TBA, A5660044, 02172341, A5768722, A5643504, A6085807, TBA, A5105147, A670060230, A5156018, A5119431, A5719805, C5603910, A5684351, to be created, 02286983, B5105141, A5145726, A5113125, HCIT4293918PACS, JD1102, YP1016-Centricity PACS-20584816, YP5025-Centricity PACS-02170316, JA0842, JL1419, JL1421, JD1260, JD1259, JD1258, JD1256, JD1257, JL1703, JA0896, JD1113, JA0738, YP4000-Centricity PACS-02321044, JL1701, JD1247, JD1246, JD1245, JD1248, JD1249, JA0894, JL1741, JD1276, JD1278, JD1279, JD1280, JL1718, JDS038, JDS039, JDS040, JDS041, JAS010, JL1387, JD0981, JD0982, JD0983, JD0984, JD0985, JA0734, YP5020-Centricity PACS-02097541, JDS003, YP5013-Centricity PACS-18045997, JL1702, JD1250, JD1252, JD1254, JD1253, JD1255, JA0895, JD1251, JDS022, JDS023, JDS024, JDS025, JL1160, JD1168, JL1728, JAS024, JDS087, JDS088, JDS089, JDS090, JDS091, JDS092, JDS093, JL1745, YP5037-Centricity PACS-15349241, JDS058, JDS059, JL1412, JL1408, JL1733, YP5022-Centricity PACS-02285739, JDS033, JDS034, JL1460, JA0750, JD1181, JD1182, JD1183, JD1184, JD1185, JD1186, YP5049-Centricity PACS-20978976, YP5030-Centricity PACS-02274229, JDS032, JL1750, YP4011-Centricity PACS-20584824, YP1017-Centricity PACS-02344060, JL1399, JL1732, JDS123, YP5009-JP-532345-Centricity PACS-20574025, YP4002-Centricity PACS-14539502, JL1168, JD0828, JAS022, JDS082, JDS083, JL1719, JL1144, JDS101, JDS102, YP5042-Centricity PACS-20101281, YP5031-Centricity PACS-02332510, JDS042, JL1396, JL1738, JL1290, JD1100, YP1012-Centricity PACS-20584811, JL1152, JL1721, JDS062, JDS063, JDS064, JAS017, JD1116, YP4001-JP-407999-Centricity PACS-19736997, JL1730, JDS111, JL1420, JD1081, JD1082, JD1095, JD1096, JD1097, JD1098, JA0830, JL1148, JDS001, JAS001, JDS002, YP5012-JP-100014-Centricity PACS-20574034, JL1407, JD1065, JD1069, JD1068, JD1066, JL1739, JD0959, JL1712, JA0900, JD1271, JD1272, JD1273, YP1014-Centricity PACS-01978327, JL1747, JD1099, JD1101, YP1013-Centricity PACS-20584812, JL1102, JL1413, JD1148, JD1138, JD1140, JD1146, JD1147, JD1139, JA0831, JD1150, JD1153, JD1159, YP5002-JP-341108-Centricity PACS-20568516, YP4006-Centricity PACS-20584822, JL1740, JL1748, JL1161, JD0767, JD0810, JD0811, JA0288, YP5048-Centricity PACS-20644650, JDS146, JL1736, YP5034-Centricity PACS-14194563, JDS047, YP5039-Centricity PACS-19629898, JAS016, JDS061, JL1417, JA0835, JD1187, JD1188, JD1189, JD1190, YP5027-Centricity PACS-02255460, JDS019, JDS020, YP5032-Centricity PACS-02360541, JDS006, YP5014-JP-280363-Centricity PACS-20574036, JL1752, JD1152, JD1151, YP1020-Centricity PACS-20584818, JL1753, YP5038-Centricity PACS-16450968, JA0834, JD1177, JD1176, JD1178, JD1179, YP5007-JP-280203-Centricity PACS-20574023, YP5004-JP-080362-Centricity PACS-20568519, JL1754, JA0832, JD1164, JD1165, JD1166, YP4012-JP-080218-Centricity PACS-20579603, JL1743, JD1091, JD1092, JD1093, JD1094, YP1011-JP-081007-Centricity PACS-20579601, JD1160, JD1167, YP5003-JP-080145-Centricity PACS-20568518, JL1751, JD1051, JD1052, JL1601-Centricity PACS-20584807, YP5024-Centricity PACS-02179873, JDS010, JDS011, JAS002, YP5010-JP-143756-Centricity PACS-20574031, JL1392, JD0999, YP5023-Centricity PACS-02256894, JAS006, JDS026, JDS027, JL1401, JD1013, JL1724, JAS020, JDS073, JDS074, JDS075, JDS076, JL1450, JD1143, JD1141, JD1145, JD1142, JD1144, JA0829, JL1746, JDS012, JDS013, JAS003, JDS031, YP5021-JP-240130-Centricity PACS-20574046, YP5001-JP-040967-Centricity PACS-20568514, JL1744, JDS021, JL1715, JAS007, JDS028, JDS029, JDS030, JL1726, JAS023, JDS084, JDS085, JD1169, JD1170, YP5006-JP-040008-Centricity PACS-20568522, JA0741, JD1161, JD1162, JD1163, JL1416, JL1720, JAS013, JDS052, JDS053, JDS054, JDS055, YP5035-Centricity PACS-15111123, JL1173, JL1390, JD0256, JD0257, JD0258, JD0259, JD0260, JD0261, JD0995, JD0993, JD0992, JD0991, JD0994, JA0301, JAS015, JL1734, JDS134, JDS135, JDS136, JDS137, JAS031, JL1737, JDS154, JA0899, JD1268, JD1269, JD1270, JL1705, JL1146, JD0774, JD0775, JD0962, JD1128, JD1125, JD1126, JD1127, JD0961, JL1710, JD1261, JD1262, JD1263, JA0901, YP5019-Centricity PACS-02042638, JL1749, JD1154, YP5000-Centricity PACS-02277426, YP5036-Centricity PACS-15261448, JDS056, JDS057, JD1130, JD1136, JD1137, JA0740, YP4004-Centricity PACS-20584821, YP5050-Centricity PACS-21180376, JL1742, JL1440, JD1121, JD1124, JDS105, JL1731, JL1159, JD0812, JL1723, JAS019, JDS069, JDS070, JDS071, JDS072, JD1149, JD1277, YP4005-JP-110249-Centricity PACS-20535391, JL1405, JD1057, JL1141, YP5026-Centricity PACS-02253241, JDS017, JDS018, JAS005, YP5033-Centricity PACS-02399305, JAS011, JDS044, JDS045, JDS046, JD1119, JD1120, YP4003-Centricity PACS-02276197, JDS004, YP5015-JP-221077-Centricity PACS-20574038, JL1704, JAS004, JDS014, JDS015, JDS016, JD1275, YP5018-Centricity PACS-02013203, JL1722, JAS018, JDS065, JDS066, JDS067, JDS068, JL1158, JD0978, JD0971, JD0973, JD0974, JD0976, JD0977, JD0972, JD0975, JA0736, JL1713, JAS056, YP1005-Centricity PACS-17452182, JL1391, JD0990, JD1061, JD0989, YP1004-Centricity PACS-20584809, JL1153, JD0784, JL1397, JL1393, JD1041, JD1015, JD1016, JDS133, JL1163, JL1418, JL1706, YP5044-Centricity PACS-20484083, JAS027, YP5005-JP-134658-Centricity PACS-20568520, JL1394, JL1386, JL1411, JD1017, JD0967, JD0968, JD0969, JD1108, JD1107, JD1109, JD0966, JA0733, JDS060, YP5028-Centricity PACS-02248290, JL1174, JD0839, JD0858, JA0290, JL1725, JAS021, JDS077, JDS078, JDS079, JDS080, JDS081, YP5011-Centricity PACS-02178701, JDS124, YP5046-Centricity PACS-20578187, JL1729, JAS025, JDS094, JDS095, JDS096, JDS097, JL1180, JAS026, JDS099, JDS100, JA0897, JD1264, JD1265, YP5016-JP-130422-Centricity PACS-20574042, JL1169, JD1267, YP5017-JP-160102-Centricity PACS-20574044, JD1050, JD1062, JDS112, JDS113, YP5045-Centricity PACS-20509113, JL1430, JD1071, YP1006-Centricity PACS-02309861, JL1735, JL1717, JAS009, JDS035, JDS036, JDS037, JDS043, KE2701CP01, KE0802CP01, KE0402CP01, KE1402PAC01, KE1110CP01, KE2001CP01, KE2202CP01, KE0301CP01, KE0202CP01, KE4705CP01, KE1502CP01, KE0201CP01, KE1602CP01, KE1701CP01, KE0302CP01, KE1111CP01, KE1201CP01, KE4701CP01, KE3004CP01, KE4704CP01, KE4601CP01, KE4101CP01, KE3402CP01, KE0601CP01, KE0702CP01, 0850550196, 0850550301, 0850550115, 0850550306, 0850550300, 0850550322, 0850550321, 0850550296, To be Provided , 0850550139, 0850550238, 852004_RAD-Centricity PACS-20542654, KW1190PC03, KW1011PC04, KW1018PP01, KW1147WS01, KW1155PC09, KW1155PC02, KW1102PC02, KW116PC06, KW1005PC01, KW1012PC01, KW1003-CPACSv7, KW1007PP02, KW1004PC02, KW1019SAHP, KW1013PC04, KW1001PC06, KW1096PC01, KW1019SAHP, KW1093PC03, LB1160SV01, LB1439PC02, LB1462PC01, LB1000PC01, LB1279PC02, LB1485PC01, LB1001PC04, LB1497PC04, To be Provided , MMC2, AMPANG6, IPPT4, SERDANG1, UTAR-CPACS, HSB-CPACS, HUSM008, UMMC-CPACS7, HPU-CPACS, IIUM-CPACS, REGENCY-CPACS, PSH-CPACS7, UNIMAS-CPACS, SGB-CPACS, PKHS-CPACS, Lifecare-CPACS, MSU-MY-PACS, Hospital Sultan Abdul Halim, Sungai Petani, SERDANG1, LIFECARE6, BMALTA25, MT1014PC01, HCIT4579183PACS, HCIT4665855PACS, HCIT4376494PACS, To be Provided , HCIT4550532PACS, HCIT4825663CPACS, HCIT4825663DAS, To be Provided , HCIT4784198PACS, HCIT4753028CPACS, HCIT4429184PACS, HCIT4820102PACS-SPS, HCIT4820102PACS, HCIT4865124PACS-IMS, HCIT4753028CPACS, AMS3960, OM1109PC01, OM1181PC01, To be Provided , HCIT5109010PACS, HCIT4369313PACS, HCIT4433887PACS, HCIT4714397CPACS, HCIT5255446PACS, HCIT4274455PACS, HCIT1693543PACS, HCIT1693543TESTPACS, HCIT4537778PACS, HCIT4392798PACS, HCIT1584981PACS, QA1102PP01, QA1114PC05, RU6265ID05, RF1647ID05, tobecreated, RU3143ID01, RU3143ID01, RU2050ID01, RF1383ID01, To be Provided , RU7579ID05, To be Provided , RF0931ID01, RU3464ID01, RU3427ID01, SA2706PP01, SA2247PP02, SA1246PP02, SA1304PP06, SA2160PP04, SA1277PP06, SA2591PP01, SA2128, SA1075PP01, SA2135, SA1025CPACS, SA2410PP01, SA1233PP06, SA1013PP17, SA2385PP01, SA2433PP02, SA1039 - CPACS, SA2315PP12, SA2581PP02, SA3053PP02_Prod, SA3053PP02_Test, SA2039PP01, SA2192, SA2329PP01, SA1005PP01 - RAD, SA1060PP08, SA1176PP01, SA1208PP01, SA2000PP01, SA1006PP07, SA2275PP02, SA2468PP01 - DF, SA2235PP01, SA1001PP28_PROD, SA1008PP01_PACS_PROD, SA1011_PACS_PROD, SA1209PP09, SA1054_PROD, SA2110PP03, SA2191PP10, SA2424PP03, SA2281PP01, SA1001PP28_TEST, SA1054_TEST, SA2645PP01, SA1008_PACS_TEST, SA1058_CPACS_WH_PROD, SA1011_PACS_TEST, SA2393PP01, SA2006PP01, SA1064PP02, SA1175-Centricity PACS EE, SA2177PP01, SA1071 - CPACS, SA2268PP01, SA2958_PACS, SA2477PP01, SA1339_CPACS7, NNI-SG-CPACS, KTPH8, ASIAMEDIC02, RADLINK10, AH-CPACS, NNI-SG-CPACS7, KTPH-SG-CPACS7, 4547195 - 6966037 - 3227712-Centricity PACS-24283964, NTFGH3, NUH-PACS, TTSH-RAD-CPACS6, TTSH-RAD-PACS-TEST, WHC-CPACS7, TTSH-RAD-CPACS7, ZA1352PP05, ZA2241PAC3, ZA2895PP02, ZA2812PAC1, ZA2412PAC21, FPG10802, FPG10902, EPP05803, FEG85600, FEG85608, FEG34405, FEG34406, FEB92518, FEB92519, FEE07A22, FEE07A23, FEG85700, FEG85704, FE410911, FE410912, DE619840, DE619846, DEM88511, DEM88512, DEM87116, DEM87121, DEI12311, DEI12312, DEP00203, 20034126, MPX06502, MPX06503, MPX23600, MPX91100, MPX43176, MPX43177, MPX43178, MPX43179, MPX43180, MPX43181, MPX91302, MPX28301, NE465120, NE670325, NE673137, NE300609, NE219643, FPD10805, FPD10806, FPD10807, FE816139, FE816140, FE8161102, DPP03300, DPP03301, DPP03302, DPP03303, DPP03305, FEG51800, FEF70522, FEF70523, YE151130, YE162A34, YP600802, GPX45803, GPX45903, MEX70164, MEX67400, MEX25523, MEX67605, MEX61704, MEX689477, MEX67314, GEX18404, GEQ20315, GEX00300, GEC21836, GE448033, GE449501, DPN08303, DEN27229, DEL26465, DEL26479, DE993820, DE993828, DE993832, DE993833, DE492A43, DE492A50, DPN08101, DPN29611, DPN29612, DPN29613, DPN29619, DPN94102, DPN96701, DPB37408, DPN69000, DPN887737, DPN95906, DPM89207, DPN87756, DPL32994, 24277780, NPV73804, NE203141, NE740521, NE216225, NE487826, DPD01000, DE539281, DE539290, DPD33502, FPD36800, NPD28000, NE723605, NEV07027, NE748614, NEV13618, NEV06838, FPD07500, NE304907, NE223245, NE419140, NEV44001, NPV84521, NPV79338, NEV44017, NEV44018, NPV79367, NPV79368, NP715206, NEV44034, NE741437, NEV35826, NE688826, UP285A29, UP285A55, GEE34022, 529010CS02, 150010PA01, 529020PC03, 437918CS01, 437918SV03, 722050CS01, 529060CS01, 475290WS05, 730140CS01, tobecreated, 0853550429, 0853550209, 0853550142, 0853550675, 0853550199, CGMHKEL20541717, 0853550804, 0853551073, 0853550616, 0853551139, 0853550223, 0853551064, CGMHTPE20541713, 0853550312, 0853550218, CGMHTAO20541704, HCIT4783733PACS, P06127PAC2, 06002PAC37, 06172PAC01, 06162PAC03, 06680PAC05, 06051PAC01, 16003PAC02, 21022PAC01, 34296PAC02, 34001PAC02, To be Provided , To be Provided , 34340PAC02, 34455PAC02, 34433PAC01.2, 34449PAC01.2, 59025PAC01.2, 34696PAC01, To be Provided , 34361PAC04.2, 34361PAC04.3, 34361PAC04.4, 34361PAC04.5, 34361PAC04.6, Temp_34118PAC01, To be Provided , 41400PAC05, 06004PAC23, AE0002PP02, AE1802PP01, AE1564PP03, AE1761PAC2, 81564SUN01, RRJ0500038, 81001VIR04, RWJ0100069, RJN7100049, 80327ISU01, 80164ISU01, DDC01VM003, RFSDAVM014, RW60200064, 85235DBS01, RVV0100002, RVVKC00067, RPA0200134, RWF0300076, RWF0300084, RVV0900052, RPA1700007, RPA2900005, RRF0100004, RM30100088, RRF0200014, RWDDAVM011, PRDC0VM045, RNQ51VM014, 80257HOS01, To be Provided , To be Provided , To be Provided , To be Provided , PRDC0VM056, PRDC0VM084, PRDC0VM100, PRDC0VM116, PRDC0VM134, PRDC0VM148, RNS01HOS01, RNS01TEST1, 85252LOA01, RX1LBTEST1, RK5BCVM002, RWFTW00008, 80674IMS01, RBK0200069, BacLieuGH-CPACS, VINMECDN07, VINMECHL07, THANHNHAN01, VINMECHAIPHONG07, VINMECPQ7, VinUni-CPACS, CHORAY01, BINHCHANHGH-CPACS, VINMECTANCANG-CPACS, VINMECNHATRANG07, NINHBINH001, VINMECINT-CPACS.
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- Centricity CPACS 6.0Ge Healthcare · System, image processing, radiological3 recalls0 reports
- Centricity CPACS 7.0Ge Healthcare · System, image processing, radiological3 recalls0 reports
- Centricity CPACS 4.0Ge Healthcare · System, image processing, radiological2 recalls0 reports
Product lines
Other devices under the same FDA submission
3 devices cleared under K110875, K043415 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Centricity CPACS 7.0Ge Healthcare · Laparoscope, general & plastic surgery2 recalls1 report
- Centricity CPACS RA1000 6.0Ge Healthcare · Laparoscope, general & plastic surgery1 recall0 reports
- Centricity CPACS RA1000 4.0Ge Healthcare · System, image processing, radiological1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page