Skip to content

Centricity CPACS 7.0 (UDI 00840682145558)Recalls & reported problems

Ge Healthcare · Class 2

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
2

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840682145558), received between Nov 26, 2024 and Nov 26, 2024 — classified as 1 injury.

2 FDA recall records name this device's identifier, 2 still open.

9 other recalls cover devices cleared under the same FDA submission (K110875); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 15, 2022Open, Classified

    Inaccurate Distance and Area measurements with use of Global Stack viewport.

    Z-1566-2022 · Root cause: Process design · FDA record ↗

  2. Apr 28, 2022Open, Classified

    Potential for Distance and Area measurements to display inaccurate measurement values when performed on magnified images, and specific to ZFP, Distance and Area measurements can display inaccurate measurement values when performed on lossy images that are scaled down from their original resolution.

    Z-1233-2022 · Root cause: Software design · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ED PHYSICIAN REVIEWED CHEST X-RAY ON CENTRICITY PACS - RA1000 CLINICAL WORKSTATION WITH INCLUDE COMPARISON ENABLED. PHYSICIAN DID NOT RECOGNIZE CHEST X-RAY IMAGE WAS A COMPARISON AND TREATED PATIENT FOR A PNEUMOTHORAX FROM AN OLD COMPARISON CHEST X-RAY BY PLACING A CHEST TUBE INTO LUNG. PATIENT'S CURRENT CHEST X-RAY WAS NORMAL (NO PNEUMOTHORAX). PACS ADMIN REPORTS NO OTHER PATIENT IMPACT.

    Patient problems: Insufficient Information

    MAUDE report 20789000 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K110875. One submission often covers many models, so check the product and lot details in each record.

  • Z-3168-2026 · Jul 31, 2026 · Open, Classified
    Centricity PACS V2, V3, V4, V6, V7, software product is intended for the storage, reading, diagnostic review, analysis, annotation, distribution, printing, editing, and processing
  • Z-3169-2026 · Jul 31, 2026 · Open, Classified
    Centricity PACS Server V7, intended for managing digital medical images and associated data, All versions
  • Z-3167-2026 · Jul 31, 2026 · Open, Classified
    Enterprise Archive V8, software product for receiving, archiving, and sending electronic medical data, All versions
  • Z-3166-2026 · Jul 31, 2026 · Open, Classified
    Centricity Enterprise Archive V2, V3, and V4, software product for receiving, archiving, and sending electronic medical data, All Versions
  • Z-3170-2026 · Jul 31, 2026 · Open, Classified
    Workflow Core Services V8, intended to provide set of services for storage and retrieval of patient data, exams and performs system Administration, common configuration management
  • Z-3171-2026 · Jul 31, 2026 · Open, Classified
    Core Services Hub V8, a set of backend services including Cross Enterprise (XE) and Workflow Core (WFC) services that enables enterprise imaging workflows, All versions
  • Z-3214-2024 · Aug 5, 2024 · Open, Classified
    GE Healthcare Centricty PACS, V2, V3, V4, V6, V7; Radiological imaging processing system.
  • Z-1466-2022 · Jun 14, 2022 · Terminated
    Centricity PACS Software Version 7.0 SP0.0.4.7
  • Z-1911-2019 · May 16, 2019 · Terminated
    Centricity PACS Foundation Centricity PACS software product is intended for the storage, reading, diagnostic review, analysis, annotation, distribution, printing, editing, and pro

Other Laparoscope, general & plastic surgery devices

All Laparoscope, general & plastic surgery devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.