Venclose System EVSRF Catheter recallZ-0371-2025 · Bard Peripheral Vascular Inc · initiated Sep 4, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Bard Peripheral Vascular Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The Venclose" RF Ablation Catheters may have been manufactured with internal wiring error. The red and yellow signal wires were soldered on swapped ports of the catheter PCB. This error in wire configuration would lead to power being applied through the proximal 3/4th of the heating coil, bypassing the thermocouple, which will then cause the generator to overdrive the catheter in an attempt to reach desired temperature set point. The wire attachment error causes the device to reach temperatures exceeding the levels intended for therapeutic efficacy, while the console is falsely indicating that temperature has not yet been reached, thus increasing the risk of harm due to excessive temperature.
FDA root cause category: Employee error
What the company is doing
On 10/16/2024, the firm sent an "URGENT: Medical Device Product Removal" letter via FedEx or email to customers informing them that the Venclose RF Ablation Catheters may have been manufactured with internal wiring errors that could cause inaccurate temperature detection resulting in devices reaching temperatures higher than intended, while the user receives incorrect warnings that the catheter has not reached temperature. Customers are instructed to: 1. Please cease use of product listed in this letter, and immediately quarantine all affected product remaining in their possession. 2. The Venclose" RF Generator, which is used in conjunction with the Venclose" RF Ablation Catheters, contains software that requires an upgrade to be performed prior to using the impacted Venclose" RF Ablation Catheters. Venclose" RF Ablation Catheters that are not listed as impacted as per attachment 1, can continue to be used. 3. The affected devices should not be used until after the Venclose" RF Generator software has been updated by a BD representative. 4. A BD Representative will contact your facility in order to arrange the upgrade to your facility s Venclose" RF Generator software. The upgrade will be performed by a BD Representative upon completion of the Customer Response Form. 5. After the confirmed Venclose" RF Generator software has been updated, the products listed in this letter can be used as normal. 6. Share this notification with all users within your facility network of the product to ensure they are also aware of this field action. If the affected products have been distributed further, please ensure that these entities are contacted and informed of the recall. 7. Complete and return the attached Customer Response Form and return it to the BD contact noted on the form whether or not you have any of the impacted product so that BD may acknowledge your receipt of this notification. Indicate on the response form that the impacted product has been was quarantined by
Recall details
- Recall number
- Z-0371-2025
- Status
- Open, Classified
- Initiated by the company
- Sep 4, 2024
- Posted by the FDA
- Nov 13, 2024
- Recalling firm
- Bard Peripheral Vascular Inc
- Quantity
- 222,373 catheters
- Product code
- GEI
- FDA submissions
- K160754
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI and WV. The countries of Australia, Singapore, Austria, Belgium, Czechia, Denmark, France, Germany, Greece, Ireland, Italy, Poland, Slovenia Spain, Sweden, Switzerland, United Kingdom, United Arab Emirates, and South Africa.
Product description
The EVSRF catheter is a sterile, single-use disposable medical device for endovenous radiofrequency ablation, a procedure in which a refluxing vein is thermally coagulated to permanently eliminate the vein from blood circulation. The catheter has a long shaft for insertion into a vein, a handle with start/stop button, and a connector cable. The catheter shaft has a heating element that is energized by the digiRF radiofrequency generator, which is a multi-voltage energy delivery system with touchscreen control that automatically sets the non-adjustable treatment parameters for the catheter. A button on the catheter (or a foot pedal attached to the generator) begins an automated treatment cycle of 20 seconds long at a set temperature of 120 C, and the treatment stops automatically when complete. The catheter is used within a sterile operative field, while the generator remains outside the sterile operative field at all times.
Code information (models, lots, serial numbers)
Show the full code information (9k characters)
Catalog Number: VC-10A2.5-6F-100 UDI does: (01)00858254006015(17)241009(10)65497423 (01)00858254006015(17)241006(10)66030181 (01)00858254006015(17)241011(10)66599182 (01)00858254006015(17)241017(10)66634308 (01)00858254006015(17)241020(10)66784768 (01)00858254006015(17)241026(10)66793709 (01)00858254006015(17)241028(10)66793710 (01)00858254006015(17)241103(10)66897727 (01)00858254006015(17)241113(10)66947733 (01)00858254006015(17)241117(10)67054619 (01)00858254006015(17)241130(10)67308611 (01)00858254006015(17)250322(10)69548248 (01)00858254006015(17)250419(10)69548249 (01)00858254006015(17)250423(10)69548250 (01)00858254006015(17)250427(10)69567357 (01)00858254006015(17)250511(10)69567358 (01)00858254006015(17)250515(10)70322501 (01)00858254006015(17)250607(10)70523716 (01)00858254006015(17)250523(10)70597994 (01)00858254006015(17)250525(10)70751594 (01)00858254006015(17)250606(10)70791660 (01)00858254006015(17)250613(10)71054985 (01)00858254006015(17)250620(10)71197454 (01)00858254006015(17)250703(10)71197455 (01)00858254006015(17)250801(10)71197574 (01)00858254006015(17)250622(10)71319092 (01)00858254006015(17)250630(10)71361455 (01)00858254006015(17)250712(10)71589682 (01)00858254006015(17)250721(10)71780480 (01)00858254006015(17)250724(10)71780481 (01)00858254006015(17)250821(10)71974718 (01)00858254006015(17)250803(10)72167973 (01)00858254006015(17)250821(10)72216631 (01)00858254006015(17)250809(10)72216758 (01)00858254006015(17)250810(10)72282293 (01)00858254006015(17)250828(10)72440542 (01)00858254006015(17)250914(10)72491306 (01)00858254006015(17)250905(10)72609982 (01)00858254006015(17)250912(10)72735944 (01)00858254006015(17)250920(10)72845915 (01)00858254006015(17)250927(10)73024457 (01)00858254006015(17)250929(10)73184835 (01)00858254006015(17)251005(10)73218082 (01)00858254006015(17)251011(10)73319890 (01)00858254006015(17)251012(10)73424182 (01)00858254006015(17)251102(10)73441733 (01)00858254006015(17)251121(10)73808592 (01)00858254006015(17)251125(10)74103378 (01)00858254006015(17)251127(10)74171115 (01)00858254006015(17)251129(10)74219123 (01)00858254006015(17)260112(10)74355356 (01)00858254006015(17)241108(10)V67114001 (01)00858254006015(17)241108(10)V67114002 (01)00858254006015(17)241108(10)V67114003 Lot Numbers: 65497423 66030181 66599182 66634308 66784768 66793709 66793710 66897727 66947733 67054619 67308611 69548248 69548249 69548250 69567357 69567358 70322501 70523716 70597994 70751594 70791660 71054985 71197454 71197455 71197574 71319092 71361455 71589682 71780480 71780481 71974718 72167973 72216631 72216758 72282293 72440542 72491306 72609982 72735944 72845915 73024457 73184835 73218082 73319890 73424182 73441733 73808592 74103378 74171115 74219123 74355356 V67114001 V67114002 V67114003 Catalog Number: VC-10A2.5-6F-60 UDI codes: (01)00858254006008(17)241013(10)66186400 (01)00858254006008(17)241024(10)66793696 (01)00858254006008(17)241102(10)66793697 (01)00858254006008(17)241107(10)66793698 (01)00858254006008(17)241021(10)66802218 (01)00858254006008(17)241031(10)66863470 (01)00858254006008(17)241110(10)66981017 (01)00858254006008(17)241120(10)67017611 (01)00858254006008(17)241125(10)67105253 (01)00858254006008(17)241115(10)67190839 (01)00858254006008(17)241123(10)67241133 (01)00858254006008(17)241128(10)67429950 (01)00858254006008(17)241204(10)67429951 (01)00858254006008(17)241204(10)67429952 (01)00858254006008(17)250410(10)69548196 (01)00858254006008(17)250327(10)69548197 (01)00858254006008(17)250416(10)69548198 (01)00858254006008(17)250331(10)69659895 (01)00858254006008(17)250502(10)69659896 (01)00858254006008(17)250505(10)69748757 (01)00858254006008(17)250509(10)69928277 (01)00858254006008(17)250601(10)70414634 (01)00858254006008(17)250517(10)70487066 (01)00858254006008(17)250519(10)70634659 (01)00858254006008(17)250529(10)70673130 (01)00858254006008(17)250604(10)70868426 (01)00858254006008(17)250608(10)70933358 (01)00858254006008(17)250612(10)70994254 (01)00858254006008(17)250614(10)71075353 (01)00858254006008(17)250618(10)71197363 (01)00858254006008(17)250704(10)71197364 (01)00858254006008(17)250705(10)71197365 (01)00858254006008(17)250806(10)71197533 (01)00858254006008(17)250625(10)71278440 (01)00858254006008(17)250628(10)71278441 (01)00858254006008(17)250707(10)71609707 (01)00858254006008(17)250727(10)71629666 (01)00858254006008(17)250714(10)71650328 (01)00858254006008(17)250710(10)71650329 (01)00858254006008(17)250717(10)71818971 (01)00858254006008(17)250719(10)71818972 (01)00858254006008(17)250731(10)71913951 (01)00858254006008(17)250814(10)72151238 (01)00858254006008(17)250816(10)72216513 (01)00858254006008(17)250807(10)72248924 (01)00858254006008(17)250825(10)72405570 (01)00858254006008(17)250823(10)72475111 (01)00858254006008(17)250903(10)72524414 (01)00858254006008(17)250831(10)72576947 (01)00858254006008(17)250907(10)72640519 (01)00858254006008(17)250910(10)72768920 (01)00858254006008(17)250918(10)72812308 (01)00858254006008(17)250922(10)72978442 (01)00858254006008(17)250925(10)73048080 (01)00858254006008(17)250927(10)73141510 (01)00858254006008(17)251008(10)73184825 (01)00858254006008(17)251015(10)73441652 (01)00858254006008(17)251017(10)73493157 (01)00858254006008(17)251025(10)73542584 (01)00858254006008(17)251029(10)73656671 (01)00858254006008(17)251123(10)73656672 (01)00858254006008(17)251201(10)74137006 (01)00858254006008(17)251204(10)74137007 (01)00858254006008(17)251206(10)74288518 (01)00858254006008(17)260114(10)74355347 (01)00858254006008(17)260121(10)74989575 Lot Numbers: 66186400 66793696 66793697 66793698 66802218 66863470 66981017 67017611 67105253 67190839 67241133 67429950 67429951 67429952 69548196 69548197 69548198 69659895 69659896 69748757 69928277 70414634 70487066 70634659 70673130 70868426 70933358 70994254 71075353 71197363 71197364 71197365 71197533 71278440 71278441 71609707 71629666 71650328 71650329 71818971 71818972 71913951 72151238 72216513 72248924 72405570 72475111 72524414 72576947 72640519 72768920 72812308 72978442 73048080 73141510 73184825 73441652 73493157 73542584 73656671 73656672 74137006 74137007 74288518 74355347 74989575 Catalog Number: VC10A256F100 UDI codes: (01)00858254006015(17)260222(10)74959653 (01)00858254006015(17)260227(10)75563807 (01)00858254006015(17)260402(10)76413778 (01)00858254006015(17)260524(10)77536423 (01)00858254006015(17)260529(10)77750390 Lot Numbers: 74959653 75563807 76413778 77536423 77750390 Catalog Number: VC10A256F60 UDI codes: (01)00858254006008(17)260215(10)75087737 (01)00858254006008(17)260304(10)75439758 (01)00858254006008(17)260306(10)75784764 (01)00858254006008(17)260311(10)75828822 (01)00858254006008(17)260314(10)75940087 (01)00858254006008(17)260326(10)76272657 (01)00858254006008(17)260412(10)76413711 (01)00858254006008(17)260422(10)76960626 (01)00858254006008(17)260424(10)77048331 (01)00858254006008(17)260425(10)77093851 (01)00858254006008(17)260502(10)77115614 (01)00858254006008(17)260603(10)77272719 (01)00858254006008(17)260605(10)77948128 (01)00858254006008(17)260611(10)77984934 (01)00858254006008(17)260617(10)78094204 (01)00858254006008(17)260621(10)78220237 (01)00858254006008(17)260626(10)78295294 (01)00858254006008(17)260701(10)78387756 Lot Numbers: 75087737 75439758 75784764 75828822 75940087 76272657 76413711 76960626 77048331 77093851 77115614 77272719 77948128 77984934 78094204 78220237 78295294 78387756 Catalog Number: VCOUS6F100 UDI codes: (01)00858254006244(17)251005(10)73310842 (01)00858254006244(17)251009(10)73319925 (01)00858254006244(17)251214(10)73319926 (01)00858254006244(17)251023(10)73388088 (01)00858254006244(17)251109(10)73623196 (01)00858254006244(17)251112(10)73923359 (01)00858254006244(17)251114(10)73954408 (01)00858254006244(17)251116(10)73987507 (01)00858254006244(17)251214(10)74020834 (01)00858254006244(17)260107(10)74507761 (01)00858254006244(17)260110(10)74889457 (01)00858254006244(17)260116(10)74933204 (01)00858254006244(17)260124(10)75016007 (01)00858254006244(17)260130(10)75127795 (01)00858254006244(17)260204(10)75204432 (01)00858254006244(17)260206(10)75276852 (01)00858254006244(17)260214(10)75305699 (01)00858254006244(17)260404(10)76413837 (01)00858254006244(17)260406(10)76480490 (01)00858254006244(17)260506(10)76526332 (01)00858254006244(17)260507(10)77379776 (01)00858254006244(17)260507(10)77379777 Lot Numbers: 73310842 73319925 73319926 73388088 73623196 73923359 73954408 73987507 74020834 74507761 74889457 74933204 75016007 75127795 75204432 75276852 75305699 76413837 76480490 76526332 77379776 77379777 Catalog Number: VCOUS6F60 UDI codes: (01)00858254006251(17)251020(10)73319896 (01)00858254006251(17)251107(10)73573950 (01)00858254006251(17)251119(10)73890876 (01)00858254006251(17)251214(10)74070677 (01)00858254006251(17)260208(10)74507759 (01)00858254006251(17)260409(10)76413830 (01)00858254006251(17)260507(10)76593887 Lot Numbers: 73319896 73573950 73890876 74070677 74507759 76413830 76593887
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Venclose System EVSRF Catheter VC10A256F60Venclose Inc · Electrosurgical, Cutting & Coagulation & Accessories1 recall1 death68 reports
- Venclose System EVSRF Catheter VC10A256F100Venclose Inc · Electrosurgical, Cutting & Coagulation & Accessories1 recall52 reports
Product lines
Other devices under the same FDA submission
2 devices cleared under K160754 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Venclose System digiRF Generator VCRFG1Venclose Inc · Electrosurgical, Cutting & Coagulation & Accessories1 recall63 reports
- Venclose™ EVSRF™ Catheter VC-10A2.5-6F-100 [10X]Venclose Inc · Electrosurgical, Cutting & Coagulation & Accessories2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page