Venclose System EVSRF Catheter VC10A256F60Recalls & reported problems
Venclose Inc · Endovenous Radiofrequency Ablation (RFA) Catheter, 60cm · Class 2
All models of Venclose System EVSRF Catheter (2)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Venclose vein ablation catheter is used for radiofrequency treatment of veins. Reports mostly describe heating problems with the catheter running too hot or not heating enough, generator errors and connection failures, and mention burns, pain and blood clots. The 62 reports mention 1 death and 25 injuries. There is 1 open recall from September 2024 about internal wires swapped during manufacturing, and reports fell to 4 in the last 12 months from 39 in the prior 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 68
- Deaths reported
- 1
- Injuries reported
- 26
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 68 adverse event reports that list this exact device (DI 00858254006008), received between Apr 1, 2020 and Sep 29, 2026 — classified as 1 death, 26 injuries, 41 malfunctions.
The most frequently reported specific device problem is “Communication or Transmission Problem” (17 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K160754, K252316); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Sep 4, 2024Class IIOpen, Classified
The Venclose" RF Ablation Catheters may have been manufactured with internal wiring error. The red and yellow signal wires were soldered on swapped ports of the catheter PCB. This error in wire configuration would lead to power being applied through the proximal 3/4th of the heating coil, bypassing the thermocouple, which will then cause the generator to overdrive the catheter in an attempt to reach desired temperature set point. The wire attachment error causes the device to reach temperatures exceeding the levels intended for therapeutic efficacy, while the console is falsely indicating that temperature has not yet been reached, thus increasing the risk of harm due to excessive temperature.
Z-0371-2025 · Root cause: Employee error · Recall details · FDA record ↗
Most reported problems
Device problems
- Communication or Transmission Problem17
- Adverse Event Without Identified Device or Use Problem15
- Insufficient Heating12
- Thermal Decomposition of Device9
- Excessive Heating8
- Break6
- Contamination5
- Material Deformation5
Patient problems
- No Clinical Signs, Symptoms or Conditions44
- Burn(s)13
- Pain5
- Thrombosis/Thrombus4
- Ulcer3
- Electric Shock2
- Blister1
- Burn, Thermal1
Recent adverse event reports
The 25 most recent of 68 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Thermal Decomposition of Device; Smoking
On (b)(6) 2026, a patient underwent a radio frequency ablation procedure using the venclose catheter. During the procedure, the device functioned properly for several treatment cycles; however, during the final treatment cycle, it began smoking and emitted a burnt odor. Treatment was terminated early, and the coil was found unraveled upon removal from the patient. There was no reported patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50929314 ↗MalfunctionSep 16, 2026Break
On (b)(6)2026, a patient underwent a radio frequency ablation procedure using the venclose catheter. Prior to the procedure, the catheter tip broken. The procedure was completed using another catheter. There was no patient contact.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50785178 ↗MalfunctionSep 14, 2026Break; Material Deformation
On (b)(6) 2026, a patient underwent a radiofrequency ablation procedure using a Venclose catheter. During device insertion, a kink was observed in the catheter; however, the kinked portion remained outside the patient. It was further reported that the catheter material subsequently fractured completely. The procedure was completed successfully without any reported complications. There was no reported patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50749281 ↗MalfunctionSep 10, 2026Tear, Rip or Hole in Device Packaging; Communication or Transmission Problem
A patient underwent a radiofrequency ablation procedure using a Venclose catheter. During the procedure, it was found that the battery was depleted, the 60 cm catheter was not recognized by the generator when connected, and a hole was observed in the package. There was no reported patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50716238 ↗InjuryAug 10, 2026Contamination; Material Deformation; Excessive Heating
On (b)(6) 2026, a patient underwent a radiofrequency ablation procedure using the Venclose catheter. During the procedure, the catheter allegedly heated above the temperature displayed on the generator, and a thick clot formed on the tip of the catheter that could not be removed as expected. Additionally, Patient experienced pain and presented to the Emergency Department; ultrasound showed no EHITT or DVT. There was no reported patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions; Pain
MAUDE report 50190625 ↗InjuryJul 2, 2026Insufficient Heating; Nonstandard Device; Difficult to Remove
On (b)(6) 2026, a female patient underwent a radio frequency ablation procedure using the Venclose catheter. During the procedure, it was reported that upon initiation, the catheter failed to reach adequate temperature for effective treatment, with the generator displaying abnormal blue line patterns. During the final treatment cycle, the catheter was difficult to withdraw, with partial intimal tissue disruption observed. Patient experienced an atypical sensation during treatment; no electrical shock reported. Additionally, the generator displayed inconsistent cycle data, recording 160 seconds of treatment with only 5 cycles, indicating inaccurate cycle counting. The procedure was completed using the same catheter. The patient was reported to be stable, with no reported patient injury. The patient was scheduled for follow up with the hospital, including repeat imaging, to determine whether any additional treatment would be necessary.
Patient problems: Electric Shock; No Clinical Signs, Symptoms or Conditions
MAUDE report 25690477 ↗MalfunctionJun 9, 2026Material Separation
ON (B)(6) 2026, A PATIENT UNDERWENT A RADIO FREQUENCY ABLATION PROCEDURE USING THE VENCLOSE CATHETER. DURING TO THE PROCEDURE, THE SEPARATION OF CATHETER OUTER COMPONENTS RESULTING IN EXPOSURE OF INTERNAL COMPONENTS. THE PROCEDURE WAS COMPLETED USING ANOTHER CATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25440063 ↗MalfunctionJun 4, 2026Self-Activation or Keying
ON (B)(6) 2026, A PATIENT UNDERWENT A RADIO FREQUENCY ABLATION PROCEDURE USING THE VENCLOSE CATHETER. DURING TO THE PROCEDURE, THE CATHETER INITIATED AUTOMATIC TREATMENT WITHOUT ACTIVATION AFTER SETUP; DEVICE WAS PROMPTLY DISCONNECTED FROM THE GENERATOR. THE PROCEDURE WAS COMPLETED USING ANOTHER CATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25387305 ↗InjuryMay 20, 2026Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2026, A PATIENT UNDERWENT A RADIO FREQUENCY ABLATION PROCEDURE USING THE VENCLOSE CATHETER. IT WAS REPORTED THAT DURING THE PROCEDURE, THE PATIENT ALLEGEDLY EXPERIENCED BURN. THE PATIENT STATUS IS UNKNOWN.
Patient problems: Burn(s)
MAUDE report 25226663 ↗InjuryMay 7, 2026Insufficient Heating
ON (B)(6) 2026, A PATIENT UNDERWENT A RADIO FREQUENCY ABLATION PROCEDURE USING THE VENCLOSE CATHETER. DURING THE PROCEDURE, THE CATHETER TEMPERATURE FLUCTUATED BETWEEN ONE HUNDRED TWENTY DEGREES AND ONE HUNDRED THIRTY DEGREES CELSIUS, THEN DROPPED BACK TO HUNDRED DEGREES CELSIUS. BLUE SQUIGGLY LINES APPEARED AFTER EACH BURN. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Burn(s); No Clinical Signs, Symptoms or Conditions
MAUDE report 25110825 ↗InjuryOct 30, 2025Insufficient Heating; Insufficient Heating
ON (B)(6) 2025, A PATIENT UNDERWENT A RADIO FREQUENCY ABLATION PROCEDURE USING THE VENCLOSE EVSRF CATHETER. DURING THE PROCEDURE, THE TUMESCENT ADMINISTRATION, THE TEMPERATURE DID NOT DECREASE BELOW 30-DEGREE CELSIUS AND THE PATIENT REPORTED EXPERIENCING PAIN. REPORTEDLY, THE CATHETER WAS REMOVED AND INSERTED A EVLT INSTEAD AND COMPLETED THE CASE SUCCESSFULLY WITHOUT ISSUE. THE CURRENT STATUS OF THE PATIENT WAS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions; Pain
MAUDE report 23422341 ↗MalfunctionOct 3, 2025Communication or Transmission Problem; Communication or Transmission Problem
ON (B)(6) 2025, A PATIENT UNDERWENT A RADIO FREQUENCY ABLATION PROCEDURE USING THE VENCLOSE CATHETER. DURING THE PROCEDURE, THE GENERATOR SCREEN DISPLAYED A RED X WHEN THE CATHETER WAS PLUGGED IN. THE PROCEDURE WAS COMPLETED USING ANOTHER CATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23213330 ↗MalfunctionOct 3, 2025Communication or Transmission Problem; Communication or Transmission Problem
ON (B)(6) 2025, A PATIENT UNDERWENT A RADIO FREQUENCY ABLATION PROCEDURE USING THE VENCLOSE CATHETER. DURING THE PROCEDURE, THE GENERATOR SCREEN DISPLAYED A RED X WHEN THE CATHETER WAS PLUGGED IN. THE PROCEDURE WAS COMPLETED USING ANOTHER CATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23213317 ↗MalfunctionOct 3, 2025Communication or Transmission Problem; Communication or Transmission Problem
ON (B)(6) 2025, A PATIENT UNDERWENT A RADIO FREQUENCY ABLATION PROCEDURE USING THE VENCLOSE CATHETER. DURING THE PROCEDURE, THE GENERATOR SCREEN DISPLAYED A RED X WHEN THE CATHETER WAS PLUGGED IN. THE PROCEDURE WAS COMPLETED USING ANOTHER CATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23213315 ↗MalfunctionOct 3, 2025Communication or Transmission Problem; Communication or Transmission Problem
ON (B)(6) 2025, A PATIENT UNDERWENT A RADIO FREQUENCY ABLATION PROCEDURE USING THE VENCLOSE CATHETER. DURING THE PROCEDURE, THE GENERATOR SCREEN DISPLAYED A RED X WHEN THE CATHETER WAS PLUGGED IN. THE PROCEDURE WAS COMPLETED USING ANOTHER CATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23213273 ↗MalfunctionOct 3, 2025Communication or Transmission Problem; Communication or Transmission Problem
ON (B)(6) 2025, A PATIENT UNDERWENT A RADIO FREQUENCY ABLATION PROCEDURE USING THE VENCLOSE CATHETER. DURING THE PROCEDURE, THE GENERATOR SCREEN DISPLAYED A RED X WHEN THE CATHETER WAS PLUGGED IN. THE PROCEDURE WAS COMPLETED USING ANOTHER CATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23213272 ↗MalfunctionOct 3, 2025Communication or Transmission Problem; Communication or Transmission Problem
ON (B)(6) 2025, A PATIENT UNDERWENT A RADIO FREQUENCY ABLATION PROCEDURE USING THE VENCLOSE CATHETER. PRIOR TO THE PROCEDURE, THE GENERATOR SCREEN DISPLAYED A RED X WHEN THE CATHETER WAS PLUGGED IN. THE PROCEDURE WAS COMPLETED USING ANOTHER CATHETER. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23213271 ↗MalfunctionOct 3, 2025Communication or Transmission Problem; Communication or Transmission Problem
ON (B)(6) 2025, A PATIENT UNDERWENT A RADIO FREQUENCY ABLATION PROCEDURE USING THE VENCLOSE CATHETER. DURING THE PROCEDURE, THE GENERATOR SCREEN DISPLAYED A RED X WHEN THE CATHETER WAS PLUGGED IN. THE PROCEDURE WAS COMPLETED USING ANOTHER CATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23213270 ↗MalfunctionOct 3, 2025Communication or Transmission Problem; Communication or Transmission Problem
ON (B)(6) 2025, A PATIENT UNDERWENT A RADIO FREQUENCY ABLATION PROCEDURE USING THE VENCLOSE CATHETER. PRIOR TO THE PROCEDURE, THE GENERATOR SCREEN DISPLAYED A RED X WHEN THE CATHETER WAS PLUGGED IN. THE PROCEDURE WAS COMPLETED USING ANOTHER CATHETER. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23213269 ↗MalfunctionOct 3, 2025Communication or Transmission Problem; Communication or Transmission Problem
A PATIENT UNDERWENT A RADIO FREQUENCY ABLATION PROCEDURE USING THE VENCLOSE CATHETER. DURING THE PROCEDURE, THE GENERATOR SCREEN DISPLAYED A RED X WHEN THE CATHETER WAS PLUGGED IN. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23213267 ↗MalfunctionOct 3, 2025Communication or Transmission Problem; Communication or Transmission Problem
ON (B)(6) 2025, A PATIENT UNDERWENT A RADIO FREQUENCY ABLATION PROCEDURE USING THE VENCLOSE CATHETER. DURING THE PROCEDURE, THE GENERATOR SCREEN DISPLAYED A RED X WHEN THE CATHETER WAS PLUGGED IN. THE PROCEDURE WAS COMPLETED USING ANOTHER CATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23213266 ↗MalfunctionOct 3, 2025Communication or Transmission Problem; Communication or Transmission Problem
ON (B)(6) 2025, A PATIENT UNDERWENT A RADIO FREQUENCY ABLATION PROCEDURE USING THE VENCLOSE CATHETER. PRIOR TO THE PROCEDURE, THE GENERATOR SCREEN DISPLAYED A RED X WHEN THE CATHETER WAS PLUGGED IN. THE PROCEDURE WAS COMPLETED USING ANOTHER CATHETER. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23213264 ↗MalfunctionOct 3, 2025Communication or Transmission Problem; Communication or Transmission Problem
A PATIENT UNDERWENT A RADIO FREQUENCY ABLATION PROCEDURE USING THE VENCLOSE CATHETER. DURING THE PROCEDURE, THE GENERATOR SCREEN DISPLAYED A RED X WHEN THE CATHETER WAS PLUGGED IN. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23213262 ↗MalfunctionOct 3, 2025Communication or Transmission Problem; Communication or Transmission Problem
ON (B)(6) 2025, A PATIENT UNDERWENT A RADIO FREQUENCY ABLATION PROCEDURE USING THE VENCLOSE CATHETER. DURING THE PROCEDURE, THE GENERATOR SCREEN DISPLAYED A RED X WHEN THE CATHETER WAS PLUGGED IN. THE PROCEDURE WAS COMPLETED USING ANOTHER CATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23213172 ↗MalfunctionOct 3, 2025Communication or Transmission Problem; Communication or Transmission Problem
ON (B)(6) 2025, A PATIENT UNDERWENT A RADIO FREQUENCY ABLATION PROCEDURE USING THE VENCLOSE CATHETER. DURING THE PROCEDURE, THE GENERATOR SCREEN DISPLAYED A RED X WHEN THE CATHETER WAS PLUGGED IN. THE PROCEDURE WAS COMPLETED USING ANOTHER CATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23213171 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K160754, K252316. One submission often covers many models, so check the product and lot details in each record.
- Z-2638-2025 · Aug 21, 2025 · Open, ClassifiedVenclose digiRF Generators, Model VCRFG1, with software 3.35 designed to identify internal wiring anomalies in the Venclose EVSRF Ablation Catheter prior to use. The Venclose digiR
Other Electrosurgical, Cutting & Coagulation & Accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page