Trilogy Evo. Model Numbers: DS2110X11B, UDS2110X11B… recallZ-0291-2025 · Philips Respironics, Inc. · initiated Oct 7, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips Respironics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.
FDA root cause category: Device Design
What the company is doing
An URGENT Medical Device Correction notification was mailed to consignees on 10/07/2024. The notification provides instructions to users to prevent risks for patients when using in-line nebulizers and a set FiO2. Consignees are to distribute the provided notification to all members of their organization responsible for setting up and supervising patients who use these devices. Additionally, if affected devices were further distributed the notification needs to be provided to them. Philips Respironics is currently investigating the issue and plans to follow-up communications to customers to provide additional guidance and solutions at a later date. Consignees with any questions can contact their local Philips Respironics Customers Service at: For DME and Homecare Customers: 1-800-345-6443, option 4, option 5 For Hospital Customers: 1-800-722-9377, option 2. On 12/27/2024, Philips Respironics sent an updated letter to consignees. It included additional guidance to customers. Dependent upon the scenarios described, customers will be instructed to either contact one of the options provided in the Urgent Medical Device Correction letter for servicing of their device to replace the flow sensor or continue use of the device in accordance with the guidance provided in the Urgent Medical Device Correction letter. Execution of flow sensor replacement will take place through a Field Change Order (FCO). An UPDATED: URGENT FIELD SAFETY NOTICE dated 3/2/26 was mailed to consignees containing recall information for three different failure modes that can be experienced in all Trilogy Evo, Trilogy EVO O2, Trilogy Evo Universal, and Trilogy EV300 devices. The firm identified that using non-pneumatic nebulizers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume received by the patient. This can lead to under-delivery of therapy to the patient. Consignees are being instructed to update all devices with software ve
Recall details
- Recall number
- Z-0291-2025
- Status
- Open, Classified
- Initiated by the company
- Oct 7, 2024
- Posted by the FDA
- Nov 13, 2024
- Recalling firm
- Philips Respironics, Inc.
- Quantity
- 82,800 units
- Product code
- CBK
- FDA submissions
- K181166
- Distribution
- Worldwide distribution - US Nationwide and the countries of AE, AF, AR, AT, AU, BD, BE, BG, BH, BO, BR, BS, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GE, GF, GI, GR, GT, HK, HR, HU, ID, IE, IL, IN, IT, JE, JO, JP, KE, KR, KW, LB, LT, LY, MA, ME, MK, MM, MQ, MT, MX, MY, NG, NL, NO, NP, NZ, OM, PA, PE, PF, PH, PK, PL, PT, PY, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW, TZ, UA, UY, VN, YT, ZA, ZW.
Product description
Trilogy Evo. Model Numbers: DS2110X11B, UDS2110X11B, RDS2110X11B, CA2110X12B, JP2110X16B, IN2110X15B, DE2110X13B, AU2110X15B, BL2110X15B, EU2110X15B, BR2110X18B, LA2110X15B, EE2110X15B, FR2110X14B, GB2110X15B, ND2110X15B, ES2110X15B, IA2110X15B, IT2110X21B, KR2110X15B, TR2110X15B, EU2110X19, FP2110X10, PP2110X10, RBRBR2110X18B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Model Number (UDI-DI Number): AU2110X15B (606959051980); BL2110X15B (606959055520); BR2110X18B (606959055476); CA2110X12B (606959052307); DE2110X13B (606959054110); DS2110X11B (606959051942); EE2110X15B (606959055544); ES2110X15B (606959054080); EU2110X15B (606959054981); EU2110X19 (606959055537); FP2110X10 (606959062092); FR2110X14B (606959055506); GB2110X15B (606959054127); IA2110X15B (606959055490); IN2110X15B (606959051959); IT2110X21B (606959055421); JP2110X16B (606959051973); KR2110X15B (606959055483); LA2110X15B (606959055452); ND2110X15B (606959055513); PP2110X10 (606959054035); RBRBR2110X18B (606959055476); RDS2110X11B (606959060517); TR2110X15B (606959055551); UDS2110X11B (606959060500). All Serial Numbers.
Devices named in this recall
18 devices in the FDA device registry whose identifier appears in the recall's code information.
- Trilogy Evo Trilogy Evo, USARespironics, Inc. · Ventilator, continuous, facility use6 recalls50 deaths4,393 reports
- Trilogy Evo, Japan JP2110X16BRespironics, Inc. · Continuous, ventilator, home use5 recalls27 deaths1,112 reports
- Trilogy Trilogy Evo Ventilator, IntlRespironics, Inc. · CONTINUOUS, VENTILATOR, HOME USE4 recalls3 deaths648 reports
- Trilogy Evo KR2110X15BRespironics, Inc. · Ventilator, continuous, facility use6 recalls3 deaths466 reports
- Trilogy Evo GB2110X15BRespironics, Inc. · Continuous, ventilator, home use5 recalls1 death141 reports
- Trilogy Evo EE2110X15BRespironics, Inc. · Continuous, ventilator, home use5 recalls1 death85 reports
- Trilogy Evo FR2110X14BRespironics, Inc. · Ventilator, continuous, facility use5 recalls77 reports
- Trilogy Trilogy Evo Ventilator, IberiaRespironics, Inc. · Oximeter4 recalls94 reports
- Trilogy IT2110X21BRespironics, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE5 recalls30 reports
- Trilogy Trilogy Evo Ventilator, GermanyRespironics, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE5 recalls26 reports
- Trilogy Trilogy Evo Ventilator, EURespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE4 recalls69 reports
- Trilogy Evo ND2110X15BRespironics, Inc. · Continuous, ventilator, home use4 recalls37 reports
- Trilogy Evo BL2110X15BRespironics, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE4 recalls28 reports
- Trilogy Evo CA2110X12BRespironics, Inc. · Ventilator, continuous, facility use4 recalls23 reports
- Trilogy Evo AU2110X15BRespironics, Inc. · Ventilator, continuous, facility use3 recalls3 deaths54 reports
- Trilogy Evo TR2110X15BRespironics, Inc. · Ventilator, continuous, facility use3 recalls9 reports
- Trilogy Evo EU2110X19Respironics, Inc. · Continuous, ventilator, home use3 recalls0 reports
- Trilogy Evo Trilogy Evo, PostponementRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 95486).
- Z-0290-2025 · Open, ClassifiedTrilogy Evo O2. Model Numbers: FP2100X10, DS2100X11B, IN2100X15B, DE2100X13B, IA2100X15B, EU2100X15B, LA2100X15B, EE2100X15B, FX2100X15B, ES2100X15B, ND2100X15B, FR2100X14B, IT2100…
- Z-0292-2025 · Open, ClassifiedTrilogy Evo Universal Ventilator. Model Number DS2000X11B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in insti…
- Z-0293-2025 · Open, ClassifiedTrilogy EV300. Model Numbers: DS2200X11B, IN2200X15B, BL2200X15B, BR2200X18B, CA2200X12B, DE2200X13B, GB2200X15B, ND2200X15B, ES2200X15B, FR2200X14B, FX2200X15B, EU2200X15B, IA2200…
- Z-0294-2025 · Open, ClassifiedAeris EVO. Model Number: VT2110X24B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and n…
- Z-0295-2025 · Open, ClassifiedGarbin Evo. Model Number: LD2110X23B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and …
- Z-0296-2025 · Open, ClassifiedLifeVent EVO2. Model Number: SP2100X26B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, an…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page