Skip to content

Trilogy Evo EU2110X19Recalls & reported problems

Respironics, Inc. · Trilogy Evo, EU (non-BT) · Class 2

All models of Philips Trilogy Ventilator (90) · Trilogy Evo (20)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

3 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Oct 7, 2024Open, Classified

    Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.

    Z-0291-2025 · Root cause: Device Design · Recall details · FDA record ↗

  2. Jul 16, 2024Open, Classified

    Firm has released a mandatory software update Version 1.05.10.00 and a User Manual addendum for affected ventilators. These updates resolve safety issues identified in earlier recalls.

    Z-2944-2024 · Root cause: Device Design · Recall details · FDA record ↗

  3. Mar 19, 2024Open, Classified

    Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.

    Z-1504-2024 · Root cause: Device Design · Recall details · FDA record ↗

Other Continuous, ventilator, home use devices

All Continuous, ventilator, home use devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page