Neptune E-SEP 0703047000Recalls & reported problems
Stryker Corporation · Neptune E-SEP Smoke Evacuation Pencil, Coated, 70mm Blade, Rocker Switch · Class 2
All models of Neptune E-SEP (19)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Stryker Neptune smoke evacuation pencil has 131 reports, many describing the pencil turning on without a button being pressed or staying on, and some describing it catching fire. The reports mention 11 injuries, including some burns, and no deaths. There are 2 open recalls over unintended activation, which can cause localized burns from electrical currents.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 131
- Deaths reported
- 0
- Injuries reported
- 11
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 131 adverse event reports that list this exact device (DI 07613327117202), received between Jul 2, 2018 and Apr 15, 2026 — classified as 11 injuries, 120 malfunctions.
The most frequently reported specific device problem is “Unintended Power Up” (61 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Oct 4, 2024Open, Classified
There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
Z-0368-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Oct 4, 2024Open, Classified
There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
Z-0365-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Unintended Power Up61
- Device Remains Activated19
- Delivered as Unsterile Product18
- Excessive Heating10
- No Apparent Adverse Event9
- Overheating of Device8
- Fire5
- Output below Specifications5
Patient problems
- No Clinical Signs, Symptoms or Conditions87
- No Patient Involvement18
- Superficial (First Degree) Burn8
- Burn(s)5
- Insufficient Information4
- Laceration(s)4
- No Consequences Or Impact To Patient4
- Abrasion1
Recent adverse event reports
The 25 most recent of 131 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 15, 2026Fire
IT WAS REPORTED THAT DURING THE PROCEDURE AT THE USER FACILITY, THE ESEP PENCIL CAUGHT ON FIRE. IT WAS REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION AS A RESULT OF THIS EVENT. IT WAS ALSO REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24885179 ↗MalfunctionMar 4, 2026Unintended Power Up
IT WAS REPORTED THAT DURING INSPECTION THE ESEP PENCIL HAD FAULTY CONTINUITY. THERE WAS NO ASSOCIATED PROCEDURE; NO ADVERSE CONSEQUENCES OR INJURY WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24512862 ↗MalfunctionFeb 9, 2026Fire
IT WAS REPORTED THAT DURING THE PROCEDURE AT THE USER FACILITY, THE ESEP PENCIL CAUGHT ON FIRE. IT WAS ALSO REPORTED THAT A NON STRYKER ELECTRODE WAS USED. NO INFORMATION WAS AVAILABLE IF THE PROCEDURE WAS COMPLETED SUCCESSFULLY OR IF THERE WAS ANY ADVERSE CONSEQUENCES OR SURGICAL DELAY.
Patient problems: Insufficient Information
MAUDE report 24294009 ↗MalfunctionFeb 5, 2026Unintended Power Up; Fire
DURING A PROCEDURE, IT WAS REPORTED THAT THE ESEP PENCIL ACTIVATED WITHOUT THE SURGEON PRESSING ANY BUTTONS, AND THE PENCIL CAUGHT ON FIRE. IT WAS FURTHER REPORTED THAT A NON STRYKER ELECTRODE WAS USED AND THE HOLSTER WAS NOT USED WHEN THE DEVICE GOT CAUGHT ON FIRE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT SIGNIFICANT DELAY, AND NO ADVERSE CONSEQUENCES OR PATIENT INJURY WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24269375 ↗MalfunctionDec 17, 2025Unintended Power Up; Unintended Power Up
NO NEW INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23829308 ↗MalfunctionNov 4, 2025Unintended Power Up; Unintended Power Up
NO NEW INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23463660 ↗MalfunctionAug 19, 2025Fire; Fire
NO NEW INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22843026 ↗MalfunctionJul 31, 2025Output below Specifications; Output below Specifications
NO NEW INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22678044 ↗MalfunctionJul 31, 2025Output below Specifications; Output below Specifications
IT WAS REPORTED THAT DURING A PROCEDURE THE COAG FUNCTION ON THE E-SEP SAFEAIR PENCIL DID NOT WORK BUT THE CUT FUNCTION DID. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22678040 ↗MalfunctionJul 31, 2025Output below Specifications; Output below Specifications
IT WAS REPORTED THAT DURING A PROCEDURE THE COAG FUNCTION ON THE E-SEP SAFEAIR PENCIL DID NOT WORK BUT THE CUT FUNCTION DID. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22674945 ↗MalfunctionJul 30, 2025Output below Specifications; Output below Specifications
IT WAS REPORTED THAT DURING A PROCEDURE THE COAG FUNCTION ON THE E-SEP SAFEAIR PENCIL DID NOT WORK BUT THE CUT FUNCTION DID. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22664431 ↗MalfunctionJul 28, 2025Unintended Power Up; Unintended Power Up
IT WAS REPORTED THAT DURING A PROCEDURE THE GENERATOR ALARM WENT OFF AND IT WAS NOTED THAT THE E-SEP SAFEAIR PENCIL ACTIVATED WITHOUT PRESSING ANY BUTTONS AND COULD NOT BE TURNED OFF. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR INJURY WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22630154 ↗MalfunctionJul 23, 2025Output below Specifications; Output below Specifications
IT WAS REPORTED THAT DURING A PROCEDURE THE COAG FUNCTION ON THE E-SEP SAFEAIR PENCIL DID NOT WORK BUT THE CUT FUNCTION DID. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22577653 ↗MalfunctionMay 14, 2025Device Remains Activated; Device Remains Activated
IT WAS REPORTED THAT DURING A PROCEDURE, THE CUT FUNCTION OF THE BOVIE PENCIL REMAINED ACTIVATED EVEN AFTER THE SURGEON HAD SWITCHED IT OFF. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT SIGNIFICANT DELAY, AND NO ADVERSE CONSEQUENCES OR INJURIES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22019008 ↗MalfunctionApr 24, 2025Unintended Power Up; Unintended Power Up
IT WAS REPORTED THAT DURING PREPARATION, PRIOR TO A PROCEDURE, THE BOVIE TRIGGERED AN ERROR ON THE GENERATOR, WHICH WAS NOT IMMEDIATELY NOTED. THE NURSE BELIEVED THE BOVIE MAY HAVE ACTIVATED ON ITS OWN. THE DEVICE WAS QUICKLY DISCONNECTED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY. NO ADVERSE CONSEQUENCES OR INJURY WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21899771 ↗InjuryMar 24, 2025Unintended Power Up; Unintended Power Up
ACCORDING ADDITIONAL INFORMATION PROVIDED, THE E-SEP PENCIL WAS USED PARALLEL TO AND IN CLOSE PROXIMITY WITH TWO OTHER ELECTRICAL DEVICES. IT WAS FURTHER REPORTED THAT THE POWER SETTING WAS INCREASED FROM COAG 30/40, CUT 30 TO COAG 60¿80 DURING THE PROCEDURE.
Patient problems: Burn(s)
MAUDE report 21679250 ↗MalfunctionJan 6, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
NO NEW INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21078691 ↗MalfunctionOct 25, 2024Overheating of Device; Excessive Heating; Overheating of Device; Excessive Heating
IT WAS REPORTED THAT DURING USE IN A PROCEDURE AT THE USER FACILITY, A CAUTERY TIP BURNED HOLES IN DRAPES. IT WAS REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY, WITHOUT A SIGNIFICANT DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20532116 ↗MalfunctionOct 21, 2024Overheating of Device; Excessive Heating; Overheating of Device; Excessive Heating
IT WAS REPORTED THAT DURING USE IN A PROCEDURE AT THE USER FACILITY, A CAUTERY TIP BURNED HOLES IN DRAPES. IT WAS REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY, WITHOUT SIGNIFICANT DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20496355 ↗InjuryOct 18, 2024Improper or Incorrect Procedure or Method; Excessive Heating; Improper or Incorrect Procedure or Method; Excessive Heating
IT WAS REPORTED THAT DURING USE IN A PROCEDURE AT THE USER FACILITY, A BURN TO THE PATIENT WHILE USING SMOKE EVAC PENCIL. IT WAS FURTHER REPORTED THAT THE BURN REQUIRED MEDICAL INTERVENTION TO SUTURE THE BURN, AS IT WAS 1.5CM SKIN LACERATION. IT WAS ALSO REPORTED THAT THE EVENT WILL RESULT IN A SCAR TO THE PATIENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY, WITHOUT DELAY.
Patient problems: Burn(s)
MAUDE report 20480853 ↗MalfunctionOct 11, 2024Overheating of Device; Overheating of Device
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE DEVICE REMAINS ¿HOT¿ AFTER USE AND BURNS HOLES IN THE DRAPES. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20427191 ↗InjuryOct 10, 2024Improper or Incorrect Procedure or Method; Excessive Heating; Improper or Incorrect Procedure or Method; Excessive Heating
IT WAS REPORTED THAT DURING USE IN AN ANTERIOR CRUCIATE LIGAMENT(ACL) RECONSTRUCTION PROCEDURE AT THE USER FACILITY, THE DEVICE WHEN PLACED ON THE DRAPES CAUSED A BURN TO THE PATIENT. IT WAS FURTHER REPORTED THAT THE BURN REQUIRED MEDICAL INTERVENTION TO CUT AWAY CHARRED TISSUE AND SUTURE THE WOUND. IT WAS ALSO REPORTED THAT THE EVENT WILL RESULT IN A SCAR TO THE PATIENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: Burn(s)
MAUDE report 20417718 ↗MalfunctionJul 2, 2024Melted; Melted
IT WAS REPORTED THAT DURING A PROCEDURE THE CHAR WAS BUILDING UP ON THE TIP AND IT LOOKS AS IF IT MELTED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19655102 ↗MalfunctionJun 7, 2024Delivered as Unsterile Product; Device Contaminated During Manufacture or Shipping; Delivered as Unsterile Product; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT PRIOR TO A PROCEDURE IT WAS OBSERVED THAT THERE WAS A HAIR INSIDE THE STERILE PACKAGING FOR THE E-SEP PENCIL. THERE WAS NO PATIENT IMPACT, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19484396 ↗MalfunctionMay 10, 2024Unintended Power Up; Unintended Power Up
IT WAS REPORTED THAT DURING A PROCEDURE, THE E-SEP PENCIL PLUGGED IN AND IMMEDIATELY STARTED 'ACTIVATING' CUT ON CAUTERY MACHINE. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19288052 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page