EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE recallZ-0620-2025 · Olympus Corporation of the Americas · initiated Oct 31, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

High-level disinfection ("HLD") with Acecide-C disinfectant in the Olympus OER-Pro Automated Endoscope Reprocessor ("AER") has been removed as a compatible reprocessing method from the Instructions for Use when reprocessing the GIF-1THI90 EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE ( GIF-1TH190 ).

FDA root cause category: Under Investigation by firm

What the company is doing

Olympus notified consignees via letter on about 10/31/2024. Consignees were informed of the issue and of the upcoming revision of the labeling such that the GIF-1TH190 is no longer a compatible endoscope for reprocessing with Acecide-C in the OER-Pro. They were instructed that they would be contacted by their Olympus sales representative within the month to discuss potential OER-Elite and GIF-1TH190 transition programs. Additionally, they were instructed to carefully read the notification, ensure reprocessing staff and other users of the OER-Pro AER are aware of the corrective action, acknowledge receipt of the notification through the Olympus web portal, and to identify and notify recipients of the affected devices if they were further distributed.

Recall details

Recall number
Z-0620-2025
Status
Open, Classified
Initiated by the company
Oct 31, 2024
Posted by the FDA
Dec 5, 2024
Recalling firm
Olympus Corporation of the Americas
Quantity
1,770 units
Product code
FDS
FDA submissions
K232997
Distribution
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV.

Product description

EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190. Model Number: GIF-1TH190

Code information (models, lots, serial numbers)

Model Number: GIF-1TH190; UDI: 04953170343360, 04953170437014; Serial Number Range: Before 2446103 or After 2500453

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

1 device cleared under K232997 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates