EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-1TH190 (REF N4505841)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE (7)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus EVIS EXERA III gastroscope has 85 reports in the FDA database, with no deaths and no injuries recorded. Many reports describe the device failing to clean adequately, and others describe burned or melted plastic on the front end cover, with no patient involvement reported. There is 1 open recall, started in October 2024, over one disinfection method being removed from its instructions for use.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
85
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 85 adverse event reports that list this exact device (DI 04953170437014), received between Jul 18, 2023 and Oct 15, 2025 — classified as 85 malfunctions.

The most frequently reported specific device problem is “Failure to Clean Adequately” (59 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K232997); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 4 reportsDec 2023: 4 reportsJan 2024: 7 reportsFeb 2024: 13 reportsMar 2024: 4 reportsApr 2024: 11 reportsMay 2024: 6 reportsJun 2024: 4 reportsJul 2024: 6 reportsAug 2024: 4 reportsSep 2024: 5 reportsOct 2024: 6 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202613
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 31, 2024Class IIOpen, Classified

    High-level disinfection ("HLD") with Acecide-C disinfectant in the Olympus OER-Pro Automated Endoscope Reprocessor ("AER") has been removed as a compatible reprocessing method from the Instructions for Use when reprocessing the GIF-1THI90 EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE ( GIF-1TH190 ).

    Z-0620-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Clean Adequately59
  • Microbial Contamination of Device11
  • Thermal Decomposition of Device7
  • Device Reprocessing Problem2
  • Melted2
  • Contamination1
  • Contamination /Decontamination Problem1
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions85

Recent adverse event reports

The 25 most recent of 85 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 15, 2025Melted; Melted

    IT WAS OBSERVED DURING DEVICE EVALUATION THAT THE GASTROINTESTINAL VIDEOSCOPE HAD A BURNT CAMERA COVER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23303114 ↗
  • MalfunctionAug 4, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22701460 ↗
  • MalfunctionMar 19, 2025Device Reprocessing Problem; Device Reprocessing Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED PLASTIC DISTAL END COVER, BENDING SECTION COVER, AND CONNECTING TUBE HAD FOREIGN OBJECTS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21648602 ↗
  • MalfunctionJan 30, 2025Contamination; Contamination

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21277457 ↗
  • MalfunctionJan 27, 2025Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THE GASTROINTESTINAL VIDEOSCOPE PLASTIC DISTAL END COVER WAS BURNED. THE ISSUE OCCURRED DURING SERVICE / MAINTENANCE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21240141 ↗
  • MalfunctionDec 5, 2024Thermal Decomposition of Device; Thermal Decomposition of Device

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED MELTING OF THE PLASTIC DISTAL END COVER AT FORCEPS PORT. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20852913 ↗
  • MalfunctionNov 18, 2024Thermal Decomposition of Device; Thermal Decomposition of Device

    IT WAS REPORTED THE GASTROINTESTINAL VIDEOSCOPE'S PLASTIC DISTAL END COVER HAD A BURN. THE ISSUE OCCURRED DURING INSPECTION. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20706890 ↗
  • MalfunctionOct 25, 2024Thermal Decomposition of Device; Failure to Clean Adequately; Thermal Decomposition of Device; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED A BURN ON THE PLASTIC DISTAL END COVER AND FOREIGN OBJECT IN THE FOLLOWING COMPONENTS: NOZZLE, ADHESIVE OF THE LIGHT GUIDE LENS, ADHESIVE OF THE OBJECTIVE LENS, AND PLASTIC DISTAL END COVER. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20539779 ↗
  • MalfunctionOct 24, 2024Thermal Decomposition of Device; Melted; Thermal Decomposition of Device; Melted

    IT WAS REPORTED THE GASTROINTESTINAL VIDEOSCOPE PROBE COVER WAS BURNT. IT'S UNKNOWN WHEN THE ISSUE OCCURRED. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20520822 ↗
  • MalfunctionOct 16, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED FOREIGN MATERIAL IN THE DISTAL END COVER. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20461164 ↗
  • MalfunctionOct 15, 2024Thermal Decomposition of Device; Thermal Decomposition of Device

    IT WAS REPORTED THE GASTROINTESTINAL VIDEOSCOPE HAD A BURN ON THE PLASTIC DISTAL END COVER. THE ISSUE OCCURRED DURING AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20457981 ↗
  • MalfunctionOct 7, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING DEVICE EVALUATION THAT THE GASTROINTESTINAL VIDEOSCOPE HAD A WHITISH FOREIGN MATERIAL INSIDE THE AIR/WATER TUBE AND CYLINDER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20391567 ↗
  • MalfunctionOct 3, 2024Thermal Decomposition of Device; Thermal Decomposition of Device

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED THE DISTAL END COVER WAS BURNT. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20363598 ↗
  • MalfunctionSep 17, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED THE NOZZLE BLOCKED BY FOREIGN MATERIAL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20238339 ↗
  • MalfunctionSep 17, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED AIRWATER-TUBE AND AIRWATER-CYLINDER HAD WHITE FOREIGN MATERIAL INSIDE THE TUBES. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20238117 ↗
  • MalfunctionSep 17, 2024Thermal Decomposition of Device; Failure to Clean Adequately; Thermal Decomposition of Device; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE INSULATION RESISTANCE VALUE AT DISTAL END DID NOT MEET THE STANDARD VALUE DUE TO BURN ON PLASTIC DISTAL END COVER. ADDITIONALLY, AIR/WATER CYLINDER, AIR/WATER TUBE, ADHESIVE ON BENDING SECTION COVER, AND JET TUBE HAD FOREIGN OBJECTS. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20237159 ↗
  • MalfunctionSep 13, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THE AIR/WATER CHANNEL AND JET TUBE OF THE GASTROINTESTINAL VIDEOSCOPE HAD WHITE FOREIGN MATERIAL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20212182 ↗
  • MalfunctionSep 3, 2024Microbial Contamination of Device; Microbial Contamination of Device

    IT WAS REPORTED THAT THE GASTROINTESTINAL VIDEOSCOPE TESTED POSITIVE FOR AN UNSPECIFIED NUMBER OF STREPTOCOCCUS PARASANGUINIS, STREPTOCOCCUS SANGUINIS, AND STREPTOCOCCUS VESTIBULARIS. THE ISSUE WAS FOUND DURING A ROUTINE CULTURE OF THE SCOPE. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20124936 ↗
  • MalfunctionAug 22, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED FOREIGN MATERIAL IN THE NOZZLE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20039109 ↗
  • MalfunctionAug 22, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED, DURING THE DEVICE INSPECTION. THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED FOREIGN MATERIAL INSIDE THE JET TUBE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20033014 ↗
  • MalfunctionAug 9, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED, THAT DURING THE DEVICE EVALUATION. THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED FOREIGN MATERIAL IN THE CHANNEL TUBE. ADHESIVE ON BENDING SECTION COVER AND BETWEEN PLASTIC DISTAL END COVER AND DISTAL END. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19956452 ↗
  • MalfunctionAug 6, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED WHITE FOREIGN MATERIAL INSIDE THE AIR/WATER CHANNEL. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19924495 ↗
  • MalfunctionJul 23, 2024Microbial Contamination of Device; Microbial Contamination of Device

    IT WAS REPORTED THAT, DURING REPROCESSING, THE GASTROINTESTINAL VIDEOSCOPE TESTED POSITIVE FOR AN UNKNOWN COLONY FORMING UNITS (CFUS) OF AN UNKNOWN MICROBES. ALL CHANNELS WERE SAMPLED. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19813194 ↗
  • MalfunctionJul 19, 2024Microbial Contamination of Device; Microbial Contamination of Device

    IT WAS REPORTED THAT, DURING REPROCESSING, THE GASTROINTESTINAL VIDEOSCOPE TESTED POSITIVE FOR 14 COLONY FORMING UNITS (CFUS) OF KLEBSIELLA GROUPE PNEUMONIAE AND 46 CFUS OF ENTEROBACTER CLOACAE COMPLEX. THE USER DID NOT REPORT ANY CONTAMINATION OR ANY OTHER SERIOUS DETERIORATION IN STATE OF HEALTH OF ANY PERSON, TO WHICH THE SCOPE COULD HAVE BEEN A CONTRIBUTORY CAUSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19780740 ↗
  • MalfunctionJul 16, 2024Failure to Clean Adequately; Failure to Clean Adequately

    IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE GASTROINTESTINAL VIDEOSCOPE EXHIBITED FOREIGN MATERIAL INSIDE THE AIR/WATER CYLINDER, ON THE DISTAL END COVER, ON THE BENDING SECTION COVER, AND ON THE BENDING SECTION COVER ADHESIVE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19757705 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K232997. One submission often covers many models, so check the product and lot details in each record.

  • Z-0498-2025 · Sep 18, 2024 · Open, Classified
    Brand Name: EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190 Product Name: Olympus GIF-1TH190 Model/Catalog Number: GIF-1TH190 Software Version: N/A Product Description: This

Other Gastroscope And Accessories, Flexible/Rigid devices

All Gastroscope And Accessories, Flexible/Rigid devices →