Allura Xper FD recallZ-0676-2025 · Philips Medical Systems Nederland B.V. · initiated Nov 11, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips Medical Systems Nederland B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential safety issues in Allura Xper R7.6- R8.1 systems where the Lifetime Extension (LTE) kit is not installed in a timely manner. Root cause is under investigation, but the recall reason is listed as for systems over 10 years of life with LTE kits are experiencing propeller motor clamping bolt and x-ray tube locking bolt breaks. If this happens, it may cause erratic C-arm movements and potential collision with patients or bystanders, or detection of a false collisions and subsequent blocked movement of the c-arm. Movement may also cause image noise and black area on the x-ray image, and possible need to expose the patient to additional unnecessary radiation during re-examination.
FDA root cause category: Equipment maintenance
What the company is doing
Philips Medical Systems notified consignees on 11/11/2024 via letter. Consignees were instructed that affected systems may continue to be used in accordance with their Indications for Use and Instructions for Use (IFU), to keep the Urgent Medical Device Correction letter with the documentation of the system until Philips corrects the system, circulate this notice to all users of the system so that they are aware of the issue, and if you experience erratic movement of the C-arm and/or if you see a portion of black area on the X-ray image, please call your local Philips representative to report the event. Philips will contact all affected customers to arrange for a Field Service Engineer visit to install an updated Lifetime Extension Kit. Consignees were also instructed to notify customers if further distributed and inform Philips about the transfer through the email account: [email protected]. Philips requested consignees complete and return the provided response form. On 11/24/2025, Philips sent a follow up letter to consignees to provide an update to the Urgent Medical Device Correction (MDC) issued on 11/11/2024. The updated letter informed consignees that after further technical investigation on both the hydrogen embrittlement and fatigue conditions, Philips has confirmed that these failure mechanisms do not apply to the bolts in question. Therefore, replacement is not required, even after 10 years of use. As a result, the risks and actions outlined in the original MDC no longer apply to affected systems.
Recall details
- Recall number
- Z-0676-2025
- Status
- Open, Classified
- Initiated by the company
- Nov 11, 2024
- Posted by the FDA
- Dec 12, 2024
- Recalling firm
- Philips Medical Systems Nederland B.V.
- Quantity
- 2,782 units
- Product code
- OWB
- FDA submissions
- K161563
- Distribution
- Nationwide distribution. International distribution to Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mauritius, Mexico, Mongolia, Morocco, Nepal, Netherlands, New Zealand, Norway, Oman, Palestine, State of, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam, Yemen.
Product description
Philips Allura Xper Systems (R7.6-R8.1). Labeled as the following: 1. Allura Xper FD10, Model Number: 722010. 2. Allura Xper FD10/10, Model Number: 722011. 3. Allura Xper FD20, Model Number: 722012. 4. Allura Xper FD20 Biplane, Model Number: 722013. 5. Allura Xper FD10 OR Table, Model Number: 722022. 6. Allura Xper FD20 OR Table, Model Number: 722023. 7. Allura Xper FD20 Biplane OR Table, Model Number: 722025.
Code information (models, lots, serial numbers)
Show the full code information (15k characters)
1. Model Number: 722010. UDI (Serial Numbers): (01)00884838054189(21)164 (164); (01)00884838059030(21)272 (272); (01)00884838059030(21)819 (819); (01)00884838059030(21)828 (828); No UDI (33, 39, 42, 43, 65, 67, 69, 79, 107, 119, 127, 158, 160, 169, 177, 190, 191, 209, 213, 214, 215, 218, 227, 229, 239, 240, 241, 252, 274, 280, 285, 286, 290, 296, 298, 306, 308, 316, 342, 343, 344, 360, 364, 365, 395, 411, 414, 422, 423, 453, 468, 477, 478, 480, 485, 488, 499, 500, 505, 521, 527, 531, 567, 572, 585, 592, 604, 605, 611, 624, 636, 637, 638, 641, 646, 648, 651, 656, 662, 668, 670, 671, 673, 677, 679, 683, 690, 694, 698, 714, 728, 730, 755, 763, 764, 780, 785, 787, 788, 791, 792, 794, 809, 818, 830, 833, 838, 839, 873, 878, 882, 890, 896, 902, 905, 914, 917, 920, 921, 931, 932, 937, 939, 946, 968, 970, 979, 986, 987, 992, 999, 1005, 1009, 1011, 1025, 1028, 1040, 1041, 1042, 1043, 1048, 1064, 1069, 1078, 1082, 1084, 1109, 1112, 1118, 1123, 1126, 1128, 1132, 1144, 1170, 1173, 1176, 1181, 1183, 1209, 1210, 1214, 1230, 1233, 1234, 1235, 1236, 1261, 1263, 1275, 1279, 1283, 1292, 1297, 1442, 601, 654, 1122, 1198, 332, 777, 666, 301, 608, 1081, 426, 924, 577, 265, 1047, 1187, 219, 959, 901, 552, 888, 1140, 64, 935, 1008, 1213, 1015, 1016, 1017, 1018, 374, 524, 1153, 1192, 1194, 841, 964, 996, 233, 1211, 721, 1190, 1188, 806, 907, 50, 418, 71, 400, 1168, 1116, 1093, 53, 370, 596, 45, 712, 1148, 323, 825, 402, 713, 195, 176, 508, 123, 1197, 91, 602, 245, 535, 230, 231, 232, 991, 248, 92, 187, 502, 1124, 537, 1199, 941, 553, 302, 289, 173, 134, 1137, 948, 509, 619, 1034, 447, 707, 708, 1285, 928, 1113, 802, 803, 192, 613, 614, 393, 506, 369, 1206, 264, 699, 759, 722, 1100, 350, 748, 1220, 657, 658, 659, 854, 565, 734, 735, 736, 737, 228, 642, 1054, 490, 391, 425, 513, 898, 424, 105, 993, 732, 580, 1256, 704, 695, 696, 238, 706, 1253, 1327, 1032, 319, 1252, 469, 541, 731, 1019, 1023, 1215, 247, 1191, 856, 51, 971, 603, 1003, 270, 267, 268, 1242, 47, 1089, 831, 530, 1091, 193, 578, 322, 1161, 126, 197, 884, 146, 62, 1095, 273, 29, 63, 412, 346, 622, 371, 740, 715, 635, 1134, 1135, 520, 936, 990, 1002, 911, 49, 978, 586, 1004, 1036, 653, 1086, 1180, 776, 288, 1245, 72, 625, 1221, 1087, 984, 253, 398, 1125, 1269, 773, 80, 380, 224, 222, 1010, 957, 282, 847, 1151, 1083, 1268, 711, 685, 686, 1104, 663, 664, 863, 799, 1224, 1225, 771, 631, 494, 1250, 652, 1106, 383, 52, 1278, 861, 906, 872, 382, 409, 969, 139, 1202, 28, 410, 1050, 915, 544, 299, 769, 540, 1007, 1305, 1306, 612, 1203, 1216, 620, 300, 1058, 600, 384, 90, 811, 1138, 967, 795, 1154, 655, 610, 895, 136, 944, 255, 415, 866, 954, 720, 83, 1141, 86, 88, 246, 801, 498, 1039, 574, 495, 1265, 1177, 1178, 128, 1105, 1266, 97, 703, 526, 1237, 650, 98, 427, 496, 1286, 897, 649, 401, 963, 876, 275, 800, 1012, 102, 1238, 475, 1090, 261, 739, 356, 877, 34, 35, 249, 1196, 1276, 1294, 927, 375, 162, 333, 943, 293, 903, 1111, 797, 416, 955, 1145, 514, 100, 1273, 1157, 860, 942, 961, 287, 1021, 1033, 998, 733, 154, 353, 1150, 832, 474, 471, 962, 1037, 1053, 579, 211, 404, 428, 503, 493, 1120, 837, 406, 351, 528, 559, 217, 529, 58, 566, 588, 589, 590, 591, 593, 594, 891, 949, 1146, 1267, 518, 568, 842, 744, 106, 103, 104, 385, 956, 70, 850, 250, 1097, 904, 75, 199, 328, 182, 294, 77, 751, 810, 1280, 1300, 1077, 858, 362, 1207, 874, 131, 1218, 1130, 972, 710, 934, 587, 645, 855, 178, 790, 466, 94, 276, 952, 548, 115, 997, 116, 753, 606, 320, 114, 181, 1182, 1065, 562, 180, 789, 1067, 1129, 975, 626, 89, 523, 243, 223, 1617, 186, 555, 598, 1119, 702, 321, 338, 137, 1051, 660, 1088, 741, 742, 1029, 1136, 95, 844, 973, 1162, 1000, 394, 142, 561, 277, 269, 304, 558, 132, 804, 1044, 822, 852, 476, 862, 96, 396, 254, 373, 875, 312, 313, 900, 220, 563, 330, 367, 473, 515, 909, 112, 121, 762, 564, 1074, 1075, 20, 124, 357, 807, 834, 1066, 184, 697, 857, 634, 980, 782, 783, 125, 1096, 225, 812, 813, 1055, 913, 1185, 368, 349, 570, 781, 354, 386, 387, 525, 66, 388, 235, 236, 549, 251, 278, 868, 1006, 1060, 516, 517, 774, 60, 166, 501, 845, 846, 597, 381, 492, 908, 607, 507, 923, 775, 403, 719, 687, 78, 1030, 1031, 185, 1147, 689). 2. Model Number: 722011. No UDI. Serial Numbers: 157, 83, 308, 95, 107, 34, 59, 23, 319, 49, 170, 137, 173, 101, 84, 155, 133, 234, 345, 124, 332, 102, 270, 177, 132, 325, 15, 76, 79, 17, 87, 237, 255, 50, 171, 168, 22, 275, 344, 324, 140, 163, 172, 103, 82, 154, 215, 40, 69, 272, 286, 81, 224, 184, 266, 175, 146, 164, 334, 338, 165, 80, 231, 323, 327, 179, 145, 117, 6, 355, 151, 202, 21, 203, 214, 340, 18, 301, 302, 303, 304, 305, 306, 307, 257, 265, 127, 180, 259, 181, 211, 206, 217, 314, 98, 201, 341, 318, 347, 261, 178, 183, 271, 253, 230, 317, 198, 321, 204, 94, 104, 52, 290, 291, 244, 333, 287, 39, 282, 78, 216, 158, 329, 246, 295, 56, 337, 292, 125, 225, 300, 88, 309, 86, 222, 71, 156, 233, 31, 188, 77, 99, 238, 223, 51, 153, 68, 129, 186, 118, 119, 41, 55, 263, 278, 258, 32, 89, 108, 310, 311, 139, 141, 66, 274, 296, 243, 191, 227, 273, 161, 220, 242, 166, 189, 185, 100, 190, 299, 280, 199, 194, 74, 75, 228, 250, 251, 91, 213, 12, 212, 113, 152, 112, 111, 115, 114, 116, 43, 277, 313, 26, 27, 29. 3. Model Number: 722012. UDI (Serial Numbers): (01)00884838059054(21)1356 (1356); (01)00884838059054(21)191 (191); (01)00884838059054(21)351 (351); (01)00884838059054(21)583 (583); (01)00884838059054(21)655 (655); (01)00884838059054(21)796 (796); (01)00884838059054(21)950 (950); No UDI (1117, 1001, 1673, 1712, 585, 517, 1545, 70, 1715, 504, 757, 935, 1250, 1613, 2006, 1733, 1734, 2201, 1922, 741, 92, 1034, 1359, 1848, 865, 1908, 1166, 1113, 1659, 494, 335, 1804, 1473, 1408, 1401, 1588, 576, 595, 590, 2302, 1279, 2151, 481, 348, 543, 893, 1187, 1455, 2085, 2202, 423, 525, 1798, 1656, 399, 990, 1871, 951, 1652, 894, 1590, 1597, 596, 1054, 1189, 1099, 823, 1037, 566, 72, 448, 1606, 924, 1610, 686, 1108, 1294, 568, 500, 459, 469, 1422, 1077, 1537, 1035, 285, 243, 84, 1317, 1649, 1895, 1900, 1723, 878, 919, 1581, 2221, 2222, 1515, 1803, 1731, 1017, 1020, 380, 1811, 2171, 1497, 782, 418, 1053, 1558, 1336, 601, 1014, 684, 1265, 331, 389, 732, 751, 1324, 2276, 2086, 1285, 2219, 298, 814, 815, 1120, 278, 297, 1713, 1506, 580, 1098, 1852, 1307, 1858, 377, 758, 1598, 172, 2054, 856, 1690, 1407, 480, 690, 579, 1327, 733, 1091, 1254, 797, 213, 735, 1280, 2138, 828, 458, 1118, 740, 1249, 542, 1033, 2265, 490, 1075, 2203, 1321, 1396, 605, 606, 444, 66, 2092, 1269, 67, 244, 1002, 1022, 1023, 2264, 2234, 1805, 1192, 1823, 483, 484, 615, 2150, 892, 1478, 1262, 1688, 1961, 1732, 1062, 649, 650, 767, 1104, 1641, 1277, 1975, 276, 316, 1335, 1461, 159, 489, 766, 639, 669, 671, 1583, 451, 625, 651, 1047, 107, 1447, 99, 670, 124, 1616, 1660, 1083, 1807, 1924, 511, 1714, 122, 101, 75, 1406, 352, 492, 472, 1105, 642, 643, 1270, 577, 578, 1306, 855, 1456, 2269, 71, 1078, 2285, 1793, 1797, 1216, 948, 387, 498, 1662, 104, 1868, 1743, 369, 1239, 1561, 464, 2293, 1960, 616, 1508, 1084, 1038, 1058, 318, 1990, 1796, 1026, 1337, 685, 656, 1560, 1441, 2926, 1854, 2288, 473, 2068, 624, 1586, 1617, 1904, 1354, 1918, 1906, 160, 383, 560, 68, 114, 1948, 1195, 518, 888, 1329, 1992, 323, 379, 792, 1003, 1219, 89, 115, 2003, 2004, 135, 1832, 2133, 2166, 562, 1941, 760, 859, 1593, 1608, 1947, 1950, 731, 520, 1328, 848, 2292, 1946, 274, 1063, 611, 1587, 1226, 613, 1271, 1842, 205, 2147, 1799, 1665, 187, 1513, 2015, 2207, 319, 2161, 1116, 1993, 208, 943, 150, 1281, 1945, 1919, 110, 337, 1735, 1326, 788, 488, 2002, 132, 1218, 1519, 1533, 336, 694, 1334, 2013, 1405, 1141, 923, 925, 533, 406, 1293, 513, 1446, 1625, 1675, 1683, 1148, 1181, 207, 1305, 790, 499, 1716, 934, 1726, 85, 476, 1657, 1045, 1531, 1630, 851, 1827, 277, 1923, 1516, 404, 453, 321, 334, 197, 1869, 2215, 574, 1611, 1559, 2924, 1686, 1635, 858, 1224, 1345, 86, 2057, 1430, 1255, 1599, 302, 610, 1209, 920, 666, 1989, 2014, 2022, 1126, 1943, 1188, 1073, 1404, 1664, 103, 1272, 2063, 1467, 1596, 326, 926, 675, 756, 1027, 424, 1894, 2145, 2146, 2148, 2065, 1527, 1477, 2227, 1177, 1645, 93, 95.1167, 1595, 88, 1672, 909, 1210, 1512, 1101, 787, 1399, 123, 126, 136, 952, 953, 954, 1005, 1661, 2156, 320, 186, 455, 570, 1622, 1115, 1316, 1212, 975, 976, 1032, 1746, 986, 971, 1594, 233, 239, 640, 446, 69, 430, 1542, 76, 1607, 1211, 1004, 2046, 974, 1044, 505, 310, 192, 314, 852, 224, 1016, 1097, 315, 1532, 1964, 2135, 1357, 589, 1092, 248, 846, 963, 964, 1658, 1344, 1902, 328, 781, 1909, 1132, 1361, 1504, 1433, 229, 142, 962, 1260, 1094, 837, 333, 1681, 1815, 922, 1333, 774, 1492, 2047, 1655, 1855, 2098, 930, 1052, 1421, 627, 1428, 1994, 1802, 1956, 931, 1817, 309, 2168, 1168, 1562, 967, 167, 910, 632, 863, 957, 289, 137, 1040, 1444, 409, 993, 995, 57, 1072, 516, 2042, 1332, 809, 795, 1460, 170, 914, 1156, 347, 1291, 1971, 664, 180, 1376, 2107, 139, 350, 628, 1865, 550, 1872, 759, 1464, 1363, 1457, 2020, 553, 288, 1980, 1580, 1708, 1495, 1244, 556, 546, 806, 1289, 1284, 1144, 1145, 223, 1081, 1436, 726, 755, 648, 1391, 1972, 215, 871, 629, 1631, 125, 432, 966, 1157, 1623, 1702, 916, 2112, 2123, 2118, 98, 1297, 2103, 211, 1147, 1303, 2300, 2120, 199, 1833, 1874, 631, 1887, 891, 1382, 1471, 1818, 2270, 1878, 1462, 1384, 798, 165, 210, 1203, 1966, 1178, 1999, 793, 794, 667, 2196, 1669, 502, 1021, 1142, 1879, 1703, 2031, 659, 1572, 1076, 2291, 1093, 479, 1043, 1626, 2056, 608, 1624, 216, 2204, 1138, 839, 805, 381, 752, 555, 1548, 1814, 1808, 979, 390, 261, 269, 2188, 1056, 866, 1699, 619, 1267, 311, 477, 1174, 1485, 59, 541, 1415, 764, 388, 876, 1150, 1701, 1693, 2080, 1986, 772, 1241, 1819, 1898, 911, 978, 2093, 116, 56, 2181, 599, 2034, 1071, 622, 941, 1088, 1927, 1412, 181, 548, 582, 588, 1567, 1367, 1143, 1423, 1268, 279, 1592, 1122, 2028, 1375, 1059, 400, 166, 746, 771, 770, 2177, 280, 1202, 753, 2274, 2198, 1028, 581, 1309, 810, 1086, 524, 1390, 1721, 374, 738, 360, 992, 219, 1825, 1839, 652, 2102, 2127, 1834, 829, 1434, 665, 742, 737, 1360, 1349, 391, 415, 2095, 730, 1967, 1996, 763, 332, 867, 902, 917, 1366, 2096, 1201, 587, 1061, 2072, 1691, 1861, 407, 1741, 833, 554, 214, 779, 799, 2101, 1228, 2117, 534, 1348, 813, 1740, 1530, 998, 1266, 680, 1997, 441, 60, 803, 2195, 895, 509, 1736, 355, 1381, 112, 478, 227, 785, 291, 185, 1450, 540, 1976, 2016, 1591, 443, 106, 127, 382, 661, 727, 1137, 1795, 1128, 1377, 392, 1134, 1130, 296, 266, 1074, 376, 2067, 152, 1978, 663, 1678, 569, 1481, 2349, 436, 1929, 1654, 287, 944, 2176, 1413, 188, 2140, 1998, 635, 1089, 1668, 1256, 2272, 660, 1671, 743, 58, 128, 1287, 1079, 61, 982, 693, 1388, 184, 1479, 1801, 2074, 1342, 405, 1535, 1882, 1127, 980, 2058, 1705, 260, 539, 2278, 748, 1435, 528, 506, 1885, 1926, 2073, 572, 688, 903, 410, 901, 869, 728, 91, 218, 419, 275, 830, 2094, 1411, 1222, 1397, 272, 2153, 1711, 1965, 1480, 1937, 1298, 414, 338, 1237, 1151, 1152, 1347, 1977, 1836, 1131, 1242, 573, 1484, 117, 2082, 1695, 817, 1536, 630, 593, 773, 1214, 936, 679, 1816, 544, 2392, 1862, 1364, 2033, 1296, 2200, 2017, 2208, 1175, 1233, 1653, 1463, 955, 2087, 1698, 1064, 1958, 271, 1135, 1385, 870, 507, 2182, 1223, 1288, 2008, 398, 1234, 965, 662, 868, 1483, 1489, 647, 1232, 2175, 1984, 222, 840, 175, 176, 1261, 437, 557, 1082, 1633, 426, 672, 1379, 872, 1973, 1822, 144, 692, 1820, 1821, 1312, 2077, 161, 140, 1696, 678, 780, 73, 74, 1292, 1466, 34, 119, 2043, 155, 1837, 118, 1451, 598, 1158, 1153, 1663, 1546, 835, 177, 1720, 749, 1400, 1897, 1246, 1245, 1939, 897, 2137, 111, 2206, 1585, 880, 1845, 777, 403, 26, 523, 1951, 2144, 1494, 2921, 385, 1930, 609, 1748, 1425, 1988, 2000, 148, 592, 1870, 674, 1049, 510, 2294, 2316, 1829, 1838, 1915, 1172, 687, 1710, 80, 1987, 428, 602, 623, 2220, 775, 988, 1582, 1979, 928, 807, 1322, 529, 1891, 465, 1719, 808, 1048, 1024, 824, 745, 617, 985, 1259, 1136, 2927, 1534, 1350, 1452, 1620, 802, 913, 1180, 354, 1813, 522, 983, 1121, 421, 1867, 1198, 416, 1840, 1812, 2066, 212, 933, 1676, 1634, 2059, 1541, 1338, 1520, 1955, 178, 249, 1418, 1251, 1176, 1417, 1899, 1454, 1055, 1962, 1015, 1124, 1547, 2266, 1230, 1543, 1609, 1744, 2110, 1553, 2922, 940, 1639.657, 2141, 1619, 143, 1368, 1149, 292, 887, 113, 105, 1742, 2280, 1949, 475, 1850, 1901, 845, 2100, 273, 1319, 2036, 468, 1341, 545, 526, 769, 1554, 1651, 932, 1060, 1647, 1220, 1200, 1932, 1931, 1935, 491, 514, 294, 1227, 236, 408, 1847, 2089, 873, 874, 1046, 994, 1745, 173, 1700, 2097, 341, 342, 1648, 853, 744, 786, 1355, 1925, 1313, 1991, 373, 1171, 1551, 1794, 2091, 1615, 1439, 1911, 149, 1080, 1, 618, 471, 1844, 1697, 972, 353, 2114, 1283, 1448, 267, 1123, 1206, 497, 425, 1330, 1888, 1050, 2253, 1273, 1974, 1907, 818, 621, 890, 1374, 1747, 162, 1371, 1125, 1031, 1085, 1864, 1565, 179, 946, 1372, 945, 1449, 1517, 1186, 2078, 1318, 1264, 1351, 1640, 1667, 1248, 1637, 1576, 1575, 1493, 646, 1577, 1579, 1574, 329, 2069, 1914, 1173, 343, 304, 306, 307, 308, 1684, 433, 413, 1159, 429, 2010, 1490, 1670, 921, 882, 1051, 2347, 567, 422, 1304, 849, 1853, 1057, 1295, 2055, 603, 1090, 463, 1841, 1129, 1959, 604, 850, 454, 460, 2053, 812, 1468, 1921, 1896, 1982, 1503, 295, 1969, 2130, 411, 183, 375, 969, 1628, 634, 1555, 847, 1666, 240, 1320, 1706, 1528, 241, 242, 157, 156, 804, 1432, 1008, 1068, 658). 4. Model Number: 722013. No UDI. Serial Numbers: 505, 217, 313, 492, 34, 153, 410, 360, 290, 74, 83, 580, 160, 404, 126, 244, 106, 346, 461, 307, 1, 442, 435, 293, 297, 46, 289, 638, 342, 69, 267, 214, 402, 113, 86, 603, 82, 149, 424, 140, 254, 255, 266, 475, 518, 519, 303, 194, 115, 150, 496, 599, 441, 491, 124, 151, 179, 550, 377, 575, 562, 178, 309, 601, 327, 73, 77, 264, 78, 308, 141, 357, 85, 40, 216, 637, 422, 174, 392, 109, 218, 228, 511, 389, 152, 479, 262, 61, 401, 497, 84, 125, 241, 564, 573, 426, 556, 42, 100, 495, 500, 96, 515, 105, 428, 427, 531, 375, 316, 278, 347, 205, 246, 43, 639, 35, 480, 304, 147, 15, 52, 283, 168, 171, 265, 272, 332, 509, 444, 400, 292, 117, 598, 415, 454, 222, 215, 335, 416, 47, 384, 559, 361, 358, 549, 476, 133, 36, 315, 31, 490, 417, 48, 276, 247, 618, 390, 157, 53, 238, 525, 17, 528, 224, 76, 585, 271, 172, 50, 544, 570, 359, 296, 530, 232, 478, 457, 547, 183, 184, 185, 186, 187, 188, 189, 190, 413, 437, 107, 138, 227, 318, 341, 225, 29, 508, 103, 434, 473, 135, 551, 557, 541, 294, 128, 388, 207, 258, 429, 95, 411, 353, 542, 533, 523, 182, 574, 139, 137, 250, 563, 502, 450, 436, 275, 524, 354, 510, 395, 458, 261, 595, 586, 535, 569, 425, 334, 344, 355, 356, 465, 345, 501, 522, 64, 396, 208, 408, 555, 394, 57, 520, 493, 201, 366, 527, 65, 449, 167, 70, 312, 321, 161, 324, 311, 176, 181, 554, 104, 459, 130, 325, 470, 440, 144, 28, 385, 403, 81, 134, 487, 256, 374, 405, 380, 367, 108, 352, 71, 455, 468, 60, 301, 299, 191, 300, 302, 239, 243, 110, 111, 158, 378, 223, 343, 165, 453, 291, 532, 364, 370, 22, 193, 336, 132, 236, 163, 2266, 368, 177, 98, 156, 443, 259, 540. 5. Model Number: 722022. No UDI. Serial Number: 6. 6. Model Number: 722023. No UDI. Serial Numbers: 27, 116, 12, 175, 98, 35, 72, 142, 46, 64, 110, 49, 124, 66, 117, 109, 42, 138, 82, 83, 127, 75, 118, 41, 33, 146, 121, 119, 70, 112, 47, 137, 25, 114, 105, 90, 107, 74, 5, 9, 4, 14, 13, 16, 132, 88, 73, 99, 38, 141, 32, 84, 68, 134, 79, 52, 56, 69, 37, 21, 53, 123, 157, 102, 126, 145, 144, 15, 135, 152, 149, 60, 150, 108, 100, 95, 34, 54, 85, 87, 94, 57, 30, 130, 58, 86, 125, 96, 3, 103, 10, 31, 106, 67. 7. Model Number: 722025. No UDI. Serial Numbers: 3, 2, 6.
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- Allura Xper FD Allura Xper FD20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System23 recalls2 deaths3,614 reports
- Allura Xper FD Allura Xper FD10Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System23 recalls2 deaths2,015 reports
- Allura Xper FD Allura Xper FD10Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System20 recalls2 deaths1,837 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.