Allura Xper FD Allura Xper FD20 (REF 722012)Recalls & reported problems
Philips Medical Systems Nederland B.V. · Monoplane Interventional X-Ray System. A ceiling or floor-mounted monoplane system with a 20-inch flat detector. · Class 2
All models of Allura Xper FD (19)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
22 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Philips Allura Xper FD20 is a ceiling or floor-mounted X-ray system for image-guided procedures. Many reports describe the X-ray not starting when activated, X-ray output failures, and positioning failures. The 3,612 reports include 2 deaths and 7 injuries. There are 23 recalls, most still open, covering hard drives, cooling hoses and table parts. Reports fell to 238 in the last 12 months from 1,149 in the previous 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 3,614
- Deaths reported
- 2
- Injuries reported
- 7
- Recalls naming it
- 23
Summary
The FDA's MAUDE database contains 3,614 adverse event reports that list this exact device (DI 00884838059054), received between Mar 23, 2020 and Sep 30, 2026 — classified as 2 deaths, 7 injuries, 3,605 malfunctions.
The most frequently reported specific device problem is “Activation Failure” (1,527 reports).
23 FDA recall records name this device's identifier, 22 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jun 17, 2026Class IIOpen, Classified
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
Z-2891-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Apr 30, 2026Class IIOpen, Classified
It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.
Z-2259-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Apr 28, 2026Class IIOpen, Classified
The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.
Z-2228-2026 · Root cause: Component design/selection · Recall details · FDA record ↗
- Mar 19, 2026Class IIOpen, Classified
Potential safety issue involving the Mean Aneurysm Flow Amplitude (MAFA) ratio of AneurysmFlow (Interventional Tool). Philips has determined that the MAFA ratio does not provide reliable prognostic information regarding aneurysm occlusion following Flow Diverter Stent treatment, however despite being stated in the IUF, it is being relied on for making clinical decisions.
Z-1805-2026 · Root cause: Labeling design · Recall details · FDA record ↗
- Mar 3, 2026Class IIOpen, Classified
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Z-1663-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Dec 15, 2025Class IIOpen, Classified
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Z-1070-2026 · Root cause: Other · Recall details · FDA record ↗
- Sep 3, 2025Class IIOpen, Classified
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
Z-2626-2025 · Root cause: Device Design · Recall details · FDA record ↗
- Sep 3, 2025Class IIOpen, Classified
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
Z-0005-2026 · Root cause: Device Design · Recall details · FDA record ↗
- Aug 1, 2025Class IIOpen, Classified
Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.
Z-2503-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Aug 1, 2025Class IIOpen, Classified
Potential that a component (air baffle) may be missing in some Nehalem FlexVision-2 PCs that are part of the Allura Xper series.
Z-2504-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Jan 13, 2025Class IOpen, Classified
Possibility of the patient falling from the table related to the mattress used on the patient table.
Z-1144-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Jan 13, 2025Class IIOpen, Classified
Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
Z-1081-2025 · Root cause: Device Design · Recall details · FDA record ↗
- Jan 13, 2025Class IIOpen, Classified
Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
Z-1080-2025 · Root cause: Device Design · Recall details · FDA record ↗
- Nov 11, 2024Class IIOpen, Classified
Potential safety issues in Allura Xper R7.6- R8.1 systems where the Lifetime Extension (LTE) kit is not installed in a timely manner. Root cause is under investigation, but the recall reason is listed as for systems over 10 years of life with LTE kits are experiencing propeller motor clamping bolt and x-ray tube locking bolt breaks. If this happens, it may cause erratic C-arm movements and potential collision with patients or bystanders, or detection of a false collisions and subsequent blocked movement of the c-arm. Movement may also cause image noise and black area on the x-ray image, and possible need to expose the patient to additional unnecessary radiation during re-examination.
Z-0676-2025 · Root cause: Equipment maintenance · Recall details · FDA record ↗
- Aug 22, 2024Class IIOpen, Classified
Potential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.
Z-3166-2024 · Root cause: Device Design · Recall details · FDA record ↗
- May 16, 2024Class IIOpen, Classified
When Software upgrade 8.1.100 is installed on an Allura Xper with an AD7NT (non tilt) table, the pivot brake does not engage. This issue may cause unexpected movements of the table.
Z-1895-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Jan 2, 2024Class IIOpen, Classified
Framegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be possible. Issue may result in delay of procedure.
Z-1144-2024 · Root cause: Process design · Recall details · FDA record ↗
- Jan 2, 2024Class IIOpen, Classified
Disk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.
Z-1151-2024 · Root cause: Process design · Recall details · FDA record ↗
- Jan 2, 2024Class IIOpen, Classified
Dual In-Line Memory Modules (DIMMs) may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.
Z-1156-2024 · Root cause: Process design · Recall details · FDA record ↗
- Nov 3, 2023Class IIOpen, Classified
Ceiling mounted L-arm contains a rotation cover that may potentially be susceptible to falling if a collision between the L-arm and other hospital equipment (i.e., an operating light) were to occur result in injury or potential sterility issues due to the cover becoming loose
Z-0558-2024 · Root cause: Device Design · Recall details · FDA record ↗
- Jun 12, 2023Class IIOpen, Classified
The bolts supporting the FlexMove Carriage may become loose and/or break, and cracks may appear in the suspension rail of the FlexMove Carriage.
Z-2283-2023 · Root cause: Component design/selection · Recall details · FDA record ↗
- Feb 2, 2022Class IIOpen, Classified
Due to a software defect, the Philips StentBoost Live R2.0 might not process X-ray images of the current run and instead show an image processed in a previous run. The image shown could be from the same patient or from a different patient.
Z-0874-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Nov 19, 2021Class IITerminated
Set screws securing the two gliding rods of the Extra Monitors (called 7th and 8th monitor) mounted on the Monitor Ceiling Suspension (MCS) can come loose, the Extra monitors might disengage from the Mounting Ceiling Suspension structure and fall down, may lead to patient, user or service engineer harm that may require medical intervention.
Z-0416-2022 · Root cause: Other · Recall details · FDA record ↗
Most reported problems
Device problems
- Activation Failure1,527
- Radiation Output Failure1,214
- Positioning Failure452
- No Display/Image308
- Detachment of Device or Device Component29
- Display or Visual Feedback Problem21
- Failure to Power Up16
- Unintended System Motion14
Patient problems
- No Clinical Signs, Symptoms or Conditions3,586
- Insufficient Information28
- No Consequences Or Impact To Patient7
- Fall3
- Hair Loss3
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Cardiac Arrest1
- Headache1
Recent adverse event reports
The 25 most recent of 3,614 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Radiation Output Failure
It was reported to Philips that the fluoroscopy foot pedal was not functioning. The device was in clinical use at the time of the event. The patient¿s procedure was cancelled and rescheduled due to the loss of fluoroscopy functionality.Philips has started an investigation of this complaint.
Patient problems: Needle Stick/Puncture
MAUDE report 50943850 ↗MalfunctionSep 29, 2026Radiation Output Failure
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50929613 ↗MalfunctionSep 24, 2026Radiation Output Failure
It was reported to Philips that the system could not emit X-ray.The device was in clinical use at the time of the event. No harm to the patient or user was reported.Philips has started an investigation of this complaint.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50878025 ↗MalfunctionSep 23, 2026Radiation Output Failure
It was reported to Philips that the system could not emit X-ray. The system was in clinical use at the time of the event. No harm to the patient or user was reported to Philips. Philips has started an investigation of this complaint
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50866243 ↗MalfunctionMay 29, 2026No Display/Image
IT WAS REPORTED TO PHILIPS THAT THE IMAGES WERE UNCLEAR AND NOT VISIBLE. THE SYSTEM WAS IN CLINICAL USE AT THE TIME OF THE EVENT. NO HARM TO THE PATIENT OR USER WAS REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25328829 ↗MalfunctionMay 20, 2026Radiation Output Failure
IT WAS REPORTED TO PHILIPS THAT FOOTSWITCH CONNECTIVITY ISSUE PREVENTING IMAGING. THE SYSTEM WAS IN CLINICAL USE AT THE TIME OF THE EVENT. NO HARM TO THE PATIENT OR USER WAS REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25226775 ↗MalfunctionFeb 24, 2026Radiation Output Failure
IT WAS REPORTED TO PHILIPS THAT THE SYSTEM DID NOT HAVE ENOUGH PENETRATION POWER, WHICH CAN IMPACT IMAGING. NO HARM WAS REPORTED TO PHILIPS. WE ARE CONSERVATIVELY REPORTING THIS EVENT AS THE INVESTIGATION IS ONGOING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24424932 ↗MalfunctionFeb 18, 2026Activation Failure
IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT UP. NO HARM TO THE PATIENT OR USER WAS REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24377781 ↗MalfunctionFeb 18, 2026Detachment of Device or Device Component
IT WAS REPORTED TO PHILIPS THAT A JOINT SHELL OF THE SUSPENSION LEAD SCREEN FELL OFF. NO HARM TO THE PATIENT OR USER WAS REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24374965 ↗MalfunctionFeb 10, 2026Radiation Output Failure
IT WAS REPORTED TO PHILIPS THAT THE SYSTEM GAVE AN ALERT INDICATING THE IMAGE PROCESSING COMPUTER WAS UNABLE TO BOOT UP, PREVENTING CINE. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24302796 ↗MalfunctionFeb 3, 2026Radiation Output Failure
IT WAS REPORTED TO PHILIPS THAT THE SYSTEM'S WIRELESS FOOTSWITCH CHARGER WAS NOT WORKING, WHICH CAN IMPACT IMAGING. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24243246 ↗MalfunctionJan 12, 2026Activation Failure; Radiation Output Failure
IT WAS REPORTED TO PHILIPS THAT THE SYSTEM'S IMAGE PROCESSING COMPUTER WAS UNABLE TO BOOT, PREVENTING CINE. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24035073 ↗MalfunctionJan 9, 2026Radiation Output Failure
IT WAS REPORTED TO PHILIPS THAT THE SYSTEM FROZE DURING A PROCEDURE, WHICH CAN IMPACT IMAGING. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24023109 ↗MalfunctionJan 8, 2026Radiation Output Failure
IT HAS BEEN REPORTED TO PHILIPS THAT THE SYSTEM TUBE ANODE WAS DEFECTIVE , WHICH CAN IMPACT IMAGING. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24004029 ↗MalfunctionJan 6, 2026Failure to Power Up
IT WAS REPORTED TO PHILIPS THAT INTERMITTENTLY THE SYSTEM COULD NOT START UP. THE DEVICE WAS OUTSIDE OF CLINICAL USE AT THE TIME OF THE EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23971931 ↗MalfunctionJan 2, 2026Positioning Failure
IT WAS REPORTED TO PHILIPS THAT THE SYSTEM'S C-ARM WAS UNABLE TO PERFORM GEOMETRY MOVEMENTS THE DEVICE WAS IN CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23943216 ↗MalfunctionDec 31, 2025Radiation Output Failure; Radiation Output Failure
IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO PERFORM EXPOSURE. THE USER PERFORMED A RESTART, BUT THE ISSUE PERSISTED. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF THE EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23931845 ↗MalfunctionDec 23, 2025Radiation Output Failure; Radiation Output Failure
IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO PERFORM EXPOSURE, IMPACTING IMAGING. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23880138 ↗MalfunctionDec 12, 2025Activation Failure; Activation Failure
IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT UP. THE DEVICE WAS NOT IN CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23791077 ↗MalfunctionDec 9, 2025Positioning Failure; Positioning Failure
IT WAS REPORTED TO PHILIPS THAT THE SYSTEM'S C-ARM WAS UNABLE TO PERFORM GEOMETRY MOVEMENTS THE DEVICE WAS IN CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23759473 ↗MalfunctionDec 5, 2025Activation Failure; Activation Failure
IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT UP. THE DEVICE WAS NOT IN CLINICAL USE AT THE TIME OF THE EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23727546 ↗MalfunctionDec 2, 2025Radiation Output Failure; Radiation Output Failure
IT WAS REPORTED TO PHILIPS THAT THE IMAGE DISK WAS FULL, PREVENTING FLUOROSCOPY AND CINE. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF THE EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23690215 ↗MalfunctionDec 1, 2025Activation Failure; Activation Failure
IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT UP. THE DEVICE WAS OUTSIDE OF CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23682271 ↗MalfunctionNov 25, 2025Activation Failure; Activation Failure
IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT UP. THE DEVICE WAS OUTSIDE OF CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23644431 ↗MalfunctionNov 25, 2025Positioning Failure; Positioning Failure
IT WAS REPORTED TO PHILIPS THAT C-ARM MOVEMENT ISSUE OCCURRED DUE TO GEO IPC COMMUNICATION TIMEOUT ON AN ALLURA XPER FD20. THE CLINICAL USE OF THE SYSTEM WAS IN USE. THERE WAS NO REPORTED PATIENT OR USER HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23643717 ↗
Other Interventional Fluoroscopic X-Ray System devices
- Allura Xper FD Allura Xper FD20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System25 recalls2 deaths3,984 reports
- Azurion Azurion 7 M20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System32 recalls2 deaths2,308 reports
- O-ARM® BI70002000Medtronic Navigation, Inc. · Interventional fluoroscopic x-ray system2 recalls1 death3,667 reports
- Azurion Azurion 7 M20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System28 recalls1 death2,176 reports
- Allura Xper FD Allura Xper FD10Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System23 recalls2 deaths2,015 reports
- Allura Xper FD Allura Xper FD10Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System20 recalls2 deaths1,837 reports
- Azurion Azurion 7 M12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System24 recalls2 deaths1,087 reports
- Azurion Azurion 7 M12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System24 recalls2 deaths835 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page