SIGNA Architect AIR, Nuclear Magnetic Resonance Imaging… recallZ-0519-2025 · GE Medical Systems, LLC · initiated Oct 28, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Medical Systems, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning.

FDA root cause category: Device Design

What the company is doing

GE Healthcare issued an "URGENT MEDICAL DEVICE CORRECTION" notice to its consignees on 10/28/20224 delivered through a traceable means. The notice explained the issue, safety issue, and requested the following actions be taken for the continued safe usage of the affected devices: "Ensure hearing protection with a Noise Reduction Rating (NRR) of at least 33dB is used during exams on these systems. 200 pairs are provided with this letter. Please post this letter in your facility on or near the MR operator console. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records." Please complete and return the attached acknowledgement form to [email protected]. For questions or concerns, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.

Recall details

Recall number
Z-0519-2025
Status
Open, Classified
Initiated by the company
Oct 28, 2024
Posted by the FDA
Nov 22, 2024
Recalling firm
GE Medical Systems, LLC
Quantity
3 units
Product code
LNH
FDA submissions
K160618, K163331, K202966
Distribution
Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of: Canada, China, India, Japan, Korea, Latvia, Mexico, Poland, Russia, Saudi Arabia, Spain, and Taiwan.

Product description

SIGNA Architect AIR, Nuclear Magnetic Resonance Imaging System

Code information (models, lots, serial numbers)

Show the full code information (3k characters)

UDI/DI (01)00195278283481(11)231214(21)CNARC2300080TJ, System ID 82427110161; UDI/DI (01)00195278283481(11)230519(21)CNARC2300027TJ, System ID 82427090471; UDI/DI (01)00195278283481(11)231116(21)CNARC2300072TJ, System ID 82427120353; UDI/DI (01)00195278283481(11)230830(21)CNARC2300051TJ, System ID 82427140303; UDI/DI (01)00195278283481(11)230515(21)CNARC2300026TJ, System ID 82427200210; UDI/DI (01)00195278283481(11)230809(21)CNARC2300046TJ, System ID 82427090487; UDI/DI (01)00195278283481(11)230816(21)CNARC2300048TJ, System ID 82427010214; UDI/DI (01)00195278283481(11)230719(21)CNARC2300040TJ, System ID 82427070331; UDI/DI (01)00195278283481(11)230904(21)CNARC2300052TJ, System ID 82427090477; UDI/DI (01)00195278283481(11)231102(21)CNARC2300068TJ, System ID 82427160409; UDI/DI (01)00195278283481(11)230722(21)CNARC2300041TJ, System ID 82427040320; UDI/DI (01)00195278283481(11)231120(21)CNARC2300073TJ, System ID 82427090470; UDI/DI (01)00195278283481(11)231009(21)CNARC2300060TJ, System ID 82427030219; UDI/DI (01)00195278283481(11)240104(21)CNARC2400002TJ, System ID 82427090488; UDI/DI (01)00195278283481(11)231207(21)CNARC2300078TJ, System ID 82427090469; UDI/DI (01)00195278283481(11)230918(21)CNARC2300056TJ, System ID 82427200214; UDI/DI (01)00195278283481(11)240131(21)CNARC2400008TJ, System ID 82427030223; UDI/DI (01)00195278283481(11)240113(21)CNARC2400005TJ, System ID 82427060120; UDI/DI (01)00195278283481(11)231109(21)CNARC2300070TJ, System ID 82427130176; UDI/DI (01)00195278283481(11)230920(21)CNARC2300057TJ, System ID 82427100480; UDI/DI (01)00195278283481(11)231010(21)CNARC2300061TJ, System ID 82427100491; UDI/DI (01)00195278283481(11)230729(21)CNARC2300043TJ, System ID 82427250136; UDI/DI (01)00195278283481(11)240116(21)CNARC2400006TJ, System ID 82427250132; UDI/DI (01)00195278283481(11)240110(21)CNARC2400004TJ, System ID 82427110160; UDI/DI (01)00195278283481(11)231026(21)CNARC2300066TJ, System ID 82427100496; UDI/DI (01)00195278283481(11)231204(21)CNARC2300077TJ, System ID 82427100504; UDI/DI (01)00195278283481(11)230508(21)CNARC2300024TJ, System ID 82427160403; UDI/DI (01)00195278283481(11)231221(21)CNARC2300081TJ, System ID 82427030207; UDI/DI (01)00195278283481(11)240102(21)CNARC2400001TJ, System ID 82427120364; UDI/DI (01)00195278283481(11)231019(21)CNARC2300064TJ, System ID 82427120360; UDI/DI (01)00195278283481(11)240127(21)CNARC2400007TJ, System ID 82427200224; UDI/DI (01)00195278283481(11)231030(21)CNARC2300067TJ, System ID 82427200207; UDI/DI (01)00195278283481(11)231016(21)CNARC2300063TJ, System ID 82427140310; UDI/DI (01)00195278283481(11)230818(21)CNARC2300049TJ, System ID 82427140317; UDI/DI (01)00195278283481(11)230911(21)CNARC2300054TJ, System ID 82427140308; UDI/DI (01)00195278283481(11)230913(21)CNARC2300055TJ, System ID 82427190195

Other devices under the same FDA submission

10 devices cleared under K160618, K163331, K202966 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 95735).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page