Discovery MR750w 3.0T, whole body magnetic resonance scanner recallZ-0517-2025 · GE Medical Systems, LLC · initiated Oct 28, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Medical Systems, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning.
FDA root cause category: Device Design
What the company is doing
GE Healthcare issued an "URGENT MEDICAL DEVICE CORRECTION" notice to its consignees on 10/28/20224 delivered through a traceable means. The notice explained the issue, safety issue, and requested the following actions be taken for the continued safe usage of the affected devices: "Ensure hearing protection with a Noise Reduction Rating (NRR) of at least 33dB is used during exams on these systems. 200 pairs are provided with this letter. Please post this letter in your facility on or near the MR operator console. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records." Please complete and return the attached acknowledgement form to [email protected]. For questions or concerns, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Recall details
- Recall number
- Z-0517-2025
- Status
- Open, Classified
- Initiated by the company
- Oct 28, 2024
- Posted by the FDA
- Nov 22, 2024
- Recalling firm
- GE Medical Systems, LLC
- Quantity
- 8 units
- Product code
- LNH
- FDA submissions
- K160618, K163331, K142085
- Distribution
- Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of: Canada, China, India, Japan, Korea, Latvia, Mexico, Poland, Russia, Saudi Arabia, Spain, and Taiwan.
Product description
Discovery MR750w 3.0T, whole body magnetic resonance scanner
Code information (models, lots, serial numbers)
System ID 760724DVMR1, UDI/DI To be provided; System ID 214820TS750WMR, UDI/DI To be provided; System ID 281333MR2, UDI/DI To be provided; System ID 82427240081, UDI/DI (01)00840682103817(11)181200(21)PG75W1800047SC; System ID 82427120136, UDI/DI Not Applicable; System ID 82427200092, UDI/DI To be provided; System ID 82427250061, UDI/DI To be provided; System ID EM0232, UDI/DI Not Applicable
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- NA Discovery MR750w 3.0TGE Medical Systems, L.L.C. · System, nuclear magnetic resonance imaging13 recalls0 reports
Other devices under the same FDA submission
18 devices cleared under K160618, K163331, K142085 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Not Applicable Optima MR450w 1.5TGE Medical Systems, L.L.C. · System, nuclear magnetic resonance imaging11 recalls0 reports
- Not Applicable Discovery MR750 3.0TGE Medical Systems, L.L.C. · System, nuclear magnetic resonance imaging10 recalls1 report
- NA SIGNA ArtistGE Medical Systems, L.L.C. · System, nuclear magnetic resonance imaging5 recalls1 report
- NA SIGNA ArchitectGE Medical Systems, L.L.C. · System, nuclear magnetic resonance imaging4 recalls0 reports
- NA 5265300-22GE Medical Systems, L.L.C. · System, nuclear magnetic resonance imaging4 recalls0 reports
- NA SIGNA ArchitectGE Medical Systems, L.L.C. · System, nuclear magnetic resonance imaging4 recalls0 reports
- NA SIGNA ArtistGe Healthcare(Tianjin) Company Limited · System, nuclear magnetic resonance imaging3 recalls0 reports
- NA 5265300-20GE Medical Systems, L.L.C. · System, nuclear magnetic resonance imaging3 recalls0 reports
- NA 5767030GE Medical Systems, L.L.C. · System, nuclear magnetic resonance imaging3 recalls0 reports
- NA SIGNA ArtistGe Healthcare(Tianjin) Company Limited · System, nuclear magnetic resonance imaging2 recalls0 reports
- NA SIGNA ArtistGe Healthcare(Tianjin) Company Limited · System, nuclear magnetic resonance imaging2 recalls0 reports
- NA 5265300-800GE Medical Systems, L.L.C. · System, nuclear magnetic resonance imaging2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 95735).
- Z-0518-2025 · Open, ClassifiedSIGNA Architect, Nuclear Magnetic Resonance Imaging System
- Z-0519-2025 · Open, ClassifiedSIGNA Architect AIR, Nuclear Magnetic Resonance Imaging System
- Z-0520-2025 · Open, ClassifiedTomographic Imager Combining Emission Computed Tomography With Nuclear Magnetic Resonance: 1) SIGNA PET/MR, 2) DISCOVERY PET/MR,
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page