EndoWrist 470249Recalls & reported problems

Intuitive Surgical, Inc. · DUAL BLADE RETRACTOR · Class 2

All models of EndoWrist (62)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
14
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00886874112403), received between Aug 20, 2018 and Jul 27, 2026 — classified as 14 malfunctions.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (7 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 3 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 19, 2024Class IIOpen, Classified

    Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.

    Z-0863-2025 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Split, Cut or Torn7
  • Material Fragmentation3
  • Material Frayed2
  • Break1
  • Detachment of Device or Device Component1
  • Unintended Movement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Insufficient Information6
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 27, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50001653 ↗
  • MalfunctionApr 11, 2026Material Frayed

    IT WAS REPORTED THAT THERE WAS A FRAYED CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24849001 ↗
  • MalfunctionMar 12, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24584933 ↗
  • MalfunctionDec 4, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ISSUE WITH AN INTUITIVE PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 23721075 ↗
  • MalfunctionJun 18, 2025Material Frayed; Material Frayed

    IT WAS REPORTED THAT THERE WAS A FRAYED CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 22288076 ↗
  • MalfunctionJun 11, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ISSUE WITH AN INTUITIVE PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 22190290 ↗
  • MalfunctionJun 11, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ISSUE WITH AN INTUITIVE PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 22190235 ↗
  • MalfunctionMay 30, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    REFER TO H11 FOR FOLLOW-UP INFORMATION.

    Patient problems: Insufficient Information

    MAUDE report 22119069 ↗
  • MalfunctionMay 7, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ISSUE WITH AN INTUITIVE PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21981190 ↗
  • MalfunctionJan 10, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED MITRAL VALVE REPAIR SURGICAL PROCEDURE, THE 8MM DUAL BLADE RETRACTOR INSTRUMENT HAD A LOOSE WIRE. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21121634 ↗
  • MalfunctionOct 30, 2024Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED PARAESOPHAGEAL HIATAL HERNIA SURGICAL PROCEDURE THE WIRE SNAPPED ON THE DUAL BLADE RETRACTOR INSTRUMENT. ON 10/21/2024, INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE CUSTOMER AND GATHERED THE FOLLOWING INFORMATION: THERE WAS A CABLE VISIBLY PROTRUDING FROM THE DISTAL END OF THE INSTRUMENT. THE CABLES WAS ONE THAT CONTROLS THE OPENING/CLOSING OF THE JAWS. NO FRAGMENT FELL INTO THE PATIENT. THE INSTRUMENT WAS INSPECTED PRIOR TO USE. NO PHOTOGRAPHIC IMAGES OF THE DEVICE OR RECORDING OF THE PROCEDURE IS AVAILABLE FOR ISI TO REVIEW.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20564361 ↗
  • MalfunctionAug 24, 2024Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED MITRAL VALVE REPAIR SURGICAL PROCEDURE, THE 8MM DUAL BLADE RETRACTOR INSTRUMENT HAD A BROKEN WIRE. THE USER COMPLETED THE PROCEDURE WITH NO FURTHER ISSUE REPORTED. NO FRAGMENT FELL INSIDE THE PATIENT. NO KNOWN IMPACT OR PATIENT CONSEQUENCE WAS REPORTED. ISI FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE INSTRUMENT WAS BEING USED TO EXPOSE THE VALVE WHEN THE ISSUE OCCURRED. THERE WAS NO INSTRUMENT COLLISION DURING THE PROCEDURE. THE ISSUE WAS RELATED TO THE OPENING/CLOSING OF THE GRIPS. THERE WAS 1 CABLE VISIBLY PROTRUDING FROM THE DISTAL END OF THE INSTRUMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20058159 ↗
  • MalfunctionMar 6, 2024Unintended Movement; Unintended Movement

    REFER TO H10/H11 FOR FOLLOW-UP INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18853209 ↗
  • MalfunctionAug 20, 2018Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED PROSTATECTOMY PROCEDURE, THE DUAL BLADE RETRACTOR INSTRUMENT HAD A BROKEN WIRE. THERE WAS NO REPORT OF FRAGMENT(S) FALLING INSIDE THE PATIENT, PATIENT HARM, ADVERSE OUTCOME OR INJURY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7797898 ↗

Other System,surgical,computer controlled instrument devices