EndoWrist recallZ-0880-2025 · Intuitive Surgical, Inc. · initiated Dec 19, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Intuitive Surgical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.

FDA root cause category: Process change control

What the company is doing

On 12/19/2024, the firm send via email an "Urgent: Medical Device Correction" Letter informing customer that Intuitive Surgical has become aware of an increase in complaints regarding frayed or broken cables on some da Vinci X and Xi reusable instruments. Customer are instructed to: 1. Customers can continue using the products in accordance with the user manual. 2. As a reminder, when using da Vinci X and Xi reusable instruments, follow the Inspection Before Use and Warnings listed in the manual provided with your system to inspect for any broken cables. Refer to Figures C and D for examples. In addition, please refer to Appendix B for additional images for detection of grip cable failures. 3. If you observe any failed (frayed or broken) grip cables prior to use, during procedure, or during reprocessing, please stop use of instrument, remove from use and inform Intuitive via the standard complaint process. 4. Inform Intuitive of any Adverse Events*/Serious Incidents** or quality problems concerning the use of the subject device via the standard complaint process. For Question and assistance - " North and South America: (800) 876-1310, Option 3 (4 am to 5 pm PST) or mail: [email protected]. " Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 am to 6 pm CET) or [email protected] " South Korea: 02-3271-3200 (9 am to 6 pm KSTJ) or [email protected] " Japan: 0120-56-5635 or 03-5575-1362 (9 am to 6 pm JST) or [email protected] " India: +1-800-103-6952 (9 am to 6 pm IT) " Taiwan: +0800-86-8181 (9 am to 6 pm CT)

Recall details

Recall number
Z-0880-2025
Status
Open, Classified
Initiated by the company
Dec 19, 2024
Posted by the FDA
Jan 10, 2025
Recalling firm
Intuitive Surgical, Inc.
Quantity
88,840
Product code
NAY
FDA submissions
K131861, K214095
Distribution
U.S. - Nation-wide. Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming and Puerto Rico. O.U.S: Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Columbia, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, French Guiana, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Luxembourg, Malaysia, Malta, Martinique, Mexico, Monaco, Netherlands, New Zealand, Norway, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, and Vietnam

Product description

8MM,MEGA SUTURECUT ND,IS4000 REF 471309

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

Model Number: 471309 Part Number: 471309-15 UDI-DI code: 00886874119815 Batch Numbers: K10221113 K10221115 K10221212 K10221218 K10230103 K10230112 K10230122 K10230124 K10230129 K10230205 K10230212 K10230219 K10230720 K10230824 K10240621 K10240718 K11221113 K11221115 K11221212 K11230103 K11230112 K11230122 K11230202 K11230212 K11230217 K12221113 K12221212 K12230112 K12230122 K12230205 K12230217 K13221113 K13230217 K14230217 K15230217 K16240508 K16240531 U10230217 U10230223 Model Number: 471309 Part Number: 471309-16 UDI-DI code: 00886874119815 Batch Numbers: K10230302 K10230310 K10230316 K10230323 K10230330 K10230413 K10230420 K10230427 K10230504 K10230509 K10230511 K10230518 K10230525 K10230601 K10230608 K10230615 K10230622 K10230630 K10230706 K10230713 K10230727 K10230803 K10230810 K10230817 K10230831 K10230913 K10230914 K10230921 K10231005 K10231012 K10231019 K10231027 K10231102 K10231109 K10231123 K10231130 K10231207 K10231218 K10240104 K10240111 K10240118 K10240125 K10240215 K10240222 K10240229 K10240308 K10240314 K10240321 K10240322 K10240411 K10240418 K10240425 K10240502 K10240516 K10240523 K10240530 K10240606 K10240613 K10240614 K10240619 K10240711 K10240718 K10240822 K11230302 K11230310 K11230316 K11230323 K11230330 K11230413 K11230420 K11230511 K11230518 K11230601 K11230615 K11230706 K11230713 K11230727 K11230803 K11230810 K11230817 K11230913 K11230914 K11230921 K11231005 K11231012 K11231019 K11231027 K11231102 K11231109 K11231123 K11231130 K11231207 K11231218 K11240104 K11240111 K11240118 K11240125 K11240229 K11240308 K11240314 K11240321 K11240322 K11240411 K11240418 K11240502 K11240516 K11240523 K11240530 K11240606 K11240613 K11240711 K11240718 K11240822 K12230302 K12230310 K12230316 K12230323 K12230413 K12230420 K12230706 K12230810 K12230914 K12230921 K12231005 K12231012 K12231027 K12231109 K12231116 K12231123 K12231130 K12240104 K12240111 K12240118 K12240125 K12240321 K12240411 K12240418 K12240502 K12240516 K12240523 K12240530 K12240606 K12240613 K13230323 K13230914 K13231012 K13231027 K13231109 K13231116 K13240104 K13240111 K13240118 K13240125 K13240502 K13240523 K13240530 K13240606 K14240104 K14240111 K14240125 K14240523 U10230420 U10230426 U10230502 U10230522 U10230630 U10230707 U10230714 U10230824 U10230901 U10230914 U10231016 U10231103 U10231115 U10231120 U11230326 U11230418 U11230707 U11230824 U11230901 U11230914 U11231016 U11231020 U11231103

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 95938).