Allura Xper FD recallZ-1149-2025 · Philips Medical Systems Nederland B.V. · initiated Jan 13, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips Medical Systems Nederland B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Possibility of the patient falling from the table related to the mattress used on the patient table.
FDA root cause category: Under Investigation by firm
What the company is doing
On January 14, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. 4. Actions that should be taken by the customer / user in order to prevent risks for patients or users " Circulate this Urgent Medical Device Correction letter to all users of the system so that they are aware of the issue. " Follow the additional instructions on the use of the mattress included in Appendix B. These instructions include information for the transfer of the patient and the correct use and positioning of the mattress. The information of this Appendix is being included in an Addendum to the Instructions for Use. Additionally, Philips is creating a Quick Reference Card. The Addendum and the Quick Reference Card will be available as from end of March 2025. Please refer to Appendix C on how to download the Addendum to the Instructions for Use and the Quick Reference Card. " In case that the affected system has been transferred to another organization, please send a copy of this Urgent Medical Device Correction letter to that organization and inform Philips about this transfer through the email account: [email protected] " Please complete and return the Response form attached to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction letter and understanding of the issue and required actions to be taken. " Should you experience the issue reported in this letter, please report the event to Philips through your local Philips representative (1- 800-722-9377). 5. Actions planned by Philips Image Guided Therapy Systems to correct the problem Philips is informing customers of this issue through this Urgent Medical Device Correction. Additionally, Philips is releasing an Addendum to the Instructions for Use and a Quick Reference Card that will be available for customers at the end of March 2025. This Urgent Medical Device Correction letter has been reported to the appropriate Regulatory Agencies.
Recall details
- Recall number
- Z-1149-2025
- Status
- Open, Classified
- Initiated by the company
- Jan 13, 2025
- Posted by the FDA
- Feb 20, 2025
- Recalling firm
- Philips Medical Systems Nederland B.V.
- Quantity
- 302 units (69 US, 233 OUS)
- Product code
- OWB
- FDA submissions
- K130842, K162859, K033737
- Distribution
- Worldiwde distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Fiji, Finland, France, French Polynesia, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guatemala, Guinea, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, JAMAICA, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Viet Nam, Yemen.
Product description
Allura Xper FD20/15; Catalog numbers: 722058
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
Catalog number: 722058; UDI-DI: 00884838059115; Serial Numbers: 172, 337, 178, 249, 22, 202, 279, 254, 285, 71, 186, 82, 85, 162, 138, 24, 365, 203, 167, 146, 168, 304, 200, 123, 129, 149, 259, 4, 197, 134, 38, 290, 171, 56, 264, 238, 199, 61, 156, 99, 65, 59, 109, 100, 305, 319, 215, 278, 124, 196, 405, 325, 265, 378, 277, 274, 94, 182, 340, 366, 190, 219, 373, 239, 23, 205, 257, 212, 111, 110, 35, 108, 32, 161, 6, 49, 89, 18, 19, 336, 392, 416, 398, 399, 412, 379, 333, 418, 385, 331, 413, 419, 400, 421, 395, 394, 408, 380, 409, 393, 369, 360, 389, 356, 357, 382, 415, 358, 411, 387, 335, 381, 370, 422, 407, 423, 397, 367, 396, 404, 420, 66, 163, 211, 214, 55, 40, 133, 114, 72, 96, 15, 9, 39, 150, 25, 81, 84, 119, 177, 268, 3, 60, 131, 377, 164, 7, 206, 101, 246, 384, 220, 261, 244, 348, 233, 106, 229, 223, 145, 136, 137, 288, 76, 86, 179, 112, 351, 130, 166, 276, 237, 45, 210, 181, 33, 251, 144, 64, 79, 29, 113, 21, 102, 36, 266, 295, 245, 243, 16, 41, 242, 187, 10, 345, 218, 361, 90, 193, 70, 26, 194, 53, 153, 234, 34, 235, 247, 272, 195, 87, 43, 31, 221, 75, 20, 330, 208, 189, 303, 63, 37, 117, 73, 230, 116, 148, 284, 236, 98, 95, 240, 350, 169, 376, 313, 343, 241, 192, 185, 115, 375, 154, 327, 92, 334, 232, 151, 201, 269, 332, 292, 142, 256, 391, 128, 344, 226, 141, 78, 183, 125, 126, 320, 248, 77, 368, 315, 68, 139, 311, 67, 347, 222, 93, 213, 354, 355, 58, 143, 301, 170, 159, 30, 103, 88, 386, 349, 252, 388, 11, 372, 281, 204, 127, 258, 289, 291, 293, 27, 97, 48, 188, 273.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Allura Xper FD Allura Xper FD20/15Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System20 recalls2 deaths776 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 96132).
- Z-1142-2025 · Open, ClassifiedAllura Xper FD10 - Catalog numbers: (1) 722003, (2) 722010, (3) 722026,
- Z-1143-2025 · Open, ClassifiedAllura Xper FD10/10; Catalog numbers: (1) 722005, (2) 722011, (3) 722027.
- Z-1144-2025 · Open, ClassifiedAllura Xper FD20; Catalog numbers: (1) 722006, (2) 722012, (3) 722028.
- Z-1145-2025 · Open, ClassifiedAllura Xper FD20 Biplane; Catalog numbers: (1) 722008, (2) 722013.
- Z-1146-2025 · Open, ClassifiedAllura Xper FD20 Biplane OR Table; Catalog numbers: (1) 722020, (2) 722025.
- Z-1147-2025 · Open, ClassifiedAllura Xper FD20 OR Table; Catalog numbers: (1) 722023, (2) 722035.
- Z-1148-2025 · Open, ClassifiedAllura Xper FD20/10; Catalog numbers: 722029
- Z-1150-2025 · Open, ClassifiedAllura Xper FD20/20; Catalog numbers: 722038
- Z-1153-2025 · Open, ClassifiedAzurion 3 M12; Catalog numbers: (1) 722063, (2) 722221.
- Z-1154-2025 · Open, ClassifiedAzurion 3 M15; Catalog numbers: (1) 722064, (2) 722222 (3) 722280(OUS ONLY).
- Z-1155-2025 · Open, ClassifiedAzurion 5 M12; Catalog numbers: (1) 722227, (2) 722231 (OUS ONLY).
- Z-1156-2025 · Open, ClassifiedAzurion 5 M20; Catalog numbers: (1) 722228, (2) 722232 (OUS ONLY), (3) 722281 (OUS ONLY).
- Z-1157-2025 · Open, ClassifiedAzurion 7 B12; Catalog numbers: (1) 722067, (2) 722225, (3) 722235 (OUS ONLY).
- Z-1158-2025 · Open, ClassifiedAzurion 7 B20; Catalog numbers: (1) 722068, (2) 722226, (3) 722236.
- Z-1159-2025 · Open, ClassifiedAzurion 7 M12; Catalog numbers: (1) 722078, (2) 722223, (3) 722233 (OUS ONLY).
- Z-1160-2025 · Open, ClassifiedAzurion 7 M20; Catalog numbers: (1) 722079, (2) 722224, (3) 722234, (4) 722282 (OUS ONLY);
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page