Allura Xper FD Allura Xper FD20/15Recalls & reported problems

Philips Medical Systems Nederland B.V. · Biplane Interventional X-Ray System with floor-mounted frontal stand (20-inch flat detector) and ceiling-mounted lateral stand ( 15-inch flat detector). · Class 2

All models of Allura Xper FD (19)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

20 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Philips Allura Xper X-ray system has 780 FDA reports, including 773 malfunctions, 5 injuries and 2 deaths. Many reports describe the system failing to start, X-ray output failing and the stand not moving. There are 20 open recalls, covering issues such as aging hard drives, patient table finger pinch and loss of imaging. Reports fell from 287 in the previous 12 months to 58 in the most recent 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
776
Deaths reported
2
Injuries reported
5
Recalls naming it
20

Summary

The FDA's MAUDE database contains 776 adverse event reports that list this exact device (DI 00884838059115), received between Jun 15, 2017 and Sep 23, 2026 — classified as 2 deaths, 5 injuries, 769 malfunctions.

The most frequently reported specific device problem is “Activation Failure” (398 reports).

20 FDA recall records name this device's identifier, 20 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 12 reportsDec 2023: 24 reportsJan 2024: 13 reportsFeb 2024: 29 reportsMar 2024: 10 reportsApr 2024: 16 reportsMay 2024: 7 reportsJun 2024: 19 reportsJul 2024: 19 reportsAug 2024: 20 reportsSep 2024: 10 reportsOct 2024: 14 reportsNov 2024: 14 reportsDec 2024: 4 reportsJan 2025: 11 reportsFeb 2025: 26 reportsMar 2025: 8 reportsApr 2025: 18 reportsMay 2025: 24 reportsJun 2025: 39 reportsJul 2025: 27 reportsAug 2025: 17 reportsSep 2025: 52 reportsOct 2025: 47 reportsNov 2025: 26 reportsDec 2025: 7 reportsJan 2026: 3 reportsFeb 2026: 4 reportsMar 2026: 6 reportsApr 2026: 1 reportsMay 2026: 6 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports20242025202652
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 29, 2026Class IIOpen, Classified

    Firmware issue potentially leading to loss of geometry movements.

    Z-2928-2026 · Root cause: Software design · Recall details · FDA record ↗

  2. Jun 17, 2026Class IIOpen, Classified

    Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.

    Z-2891-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  3. Apr 30, 2026Class IIOpen, Classified

    It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.

    Z-2259-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  4. Apr 28, 2026Class IIOpen, Classified

    The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.

    Z-2228-2026 · Root cause: Component design/selection · Recall details · FDA record ↗

  5. Mar 19, 2026Class IIOpen, Classified

    Potential safety issue involving the Mean Aneurysm Flow Amplitude (MAFA) ratio of AneurysmFlow (Interventional Tool). Philips has determined that the MAFA ratio does not provide reliable prognostic information regarding aneurysm occlusion following Flow Diverter Stent treatment, however despite being stated in the IUF, it is being relied on for making clinical decisions.

    Z-1805-2026 · Root cause: Labeling design · Recall details · FDA record ↗

  6. Mar 3, 2026Class IIOpen, Classified

    Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

    Z-1671-2026 · Root cause: Device Design · Recall details · FDA record ↗

  7. Sep 3, 2025Class IIOpen, Classified

    Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.

    Z-2633-2025 · Root cause: Device Design · Recall details · FDA record ↗

  8. Sep 3, 2025Class IIOpen, Classified

    The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.

    Z-0010-2026 · Root cause: Device Design · Recall details · FDA record ↗

  9. Aug 1, 2025Class IIOpen, Classified

    Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.

    Z-2503-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  10. Aug 1, 2025Class IIOpen, Classified

    Potential that a component (air baffle) may be missing in some Nehalem FlexVision-2 PCs that are part of the Allura Xper series.

    Z-2504-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  11. Jun 30, 2025Class IIOpen, Classified

    Potential for temporary loss of imaging (X-ray) functionality due to software issue.

    Z-2189-2025 · Root cause: Software design · Recall details · FDA record ↗

  12. Jan 13, 2025Class IIOpen, Classified

    Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel

    Z-1081-2025 · Root cause: Device Design · Recall details · FDA record ↗

  13. Jan 13, 2025Class IIOpen, Classified

    Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel

    Z-1080-2025 · Root cause: Device Design · Recall details · FDA record ↗

  14. Jan 13, 2025Class IOpen, Classified

    Possibility of the patient falling from the table related to the mattress used on the patient table.

    Z-1149-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  15. Jan 13, 2025Class IIOpen, Classified

    Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel

    Z-1086-2025 · Root cause: Device Design · Recall details · FDA record ↗

  16. Aug 22, 2024Class IIOpen, Classified

    Potential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.

    Z-3166-2024 · Root cause: Device Design · Recall details · FDA record ↗

  17. Jan 2, 2024Class IIOpen, Classified

    Disk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.

    Z-1151-2024 · Root cause: Process design · Recall details · FDA record ↗

  18. Jan 2, 2024Class IIOpen, Classified

    Framegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be possible. Issue may result in delay of procedure.

    Z-1144-2024 · Root cause: Process design · Recall details · FDA record ↗

  19. Jan 2, 2024Class IIOpen, Classified

    Dual In-Line Memory Modules (DIMMs) may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.

    Z-1156-2024 · Root cause: Process design · Recall details · FDA record ↗

  20. Nov 3, 2023Class IIOpen, Classified

    Ceiling mounted L-arm contains a rotation cover that may potentially be susceptible to falling if a collision between the L-arm and other hospital equipment (i.e., an operating light) were to occur result in injury or potential sterility issues due to the cover becoming loose

    Z-0558-2024 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Activation Failure398
  • Radiation Output Failure236
  • No Display/Image57
  • Positioning Failure57
  • Detachment of Device or Device Component6
  • Display or Visual Feedback Problem5
  • Adverse Event Without Identified Device or Use Problem4
  • Output Problem4

Patient problems

  • No Clinical Signs, Symptoms or Conditions759
  • Insufficient Information12
  • Fall3
  • Needle Stick/Puncture2
  • No Consequences Or Impact To Patient2
  • Abrasion1
  • Bone Fracture(s)1
  • Embolism/Embolus1

Recent adverse event reports

The 25 most recent of 776 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 23, 2026Activation Failure

    It was reported to Philips that the system was unable to boot up.The system was not in clinical use at the time of the event. No harm to the patient or user was reported.Philips has started an investigation of this complaint.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50864359 ↗
  • MalfunctionMay 29, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM'S FLEXVISION COMPUTER WAS UNABLE TO BOOT. THE SYSTEM WAS NOT IN CLINICAL USE AT THE TIME OF THE EVENT. NO HARM TO THE PATIENT OR USER WAS REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25328686 ↗
  • MalfunctionMay 26, 2026Positioning Failure

    IT WAS REPORTED TO PHILIPS THAT C-ARM WAS STUCK IN A ROTATED POSITION AND WILL NOT MOVE. PROCEDURE WAS COMPLETED BY MOVING THE PATIENT TO ANOTHER ROOM. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25276394 ↗
  • MalfunctionMay 20, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT UP SUCCESSFULLY AFTER A WARM RESTART, RESULTING IN A FAILURE TO BOOT. THE SYSTEM WAS OUTSIDE OF CLINICAL USE AT THE TIME OF THE EVENT. NO HARM TO THE PATIENT OR USER WAS REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25226774 ↗
  • MalfunctionMay 19, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM KEPT RESTARTING, WHICH COULD IMPACT BOOTING.THE SYSTEM WAS NOT IN CLINICAL USE AT THE TIME OF THE EVENT. NO HARM WAS REPORTED.PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25223603 ↗
  • MalfunctionMay 11, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM'S HOST COMPUTER WAS UNABLE TO BOOT, PREVENTING A FULL SYSTEM BOOT. THE SYSTEM WAS NOT IN CLINICAL USE AT THE TIME OF THE EVENT. NO HARM TO THE PATIENT OR USER WAS REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25134678 ↗
  • MalfunctionMay 7, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM FROZE DURING A PROCEDURE AND WAS UNABLE TO START. THE SYSTEM WAS IN CLINICAL USE AT THE TIME OF THE EVENT. NO HARM TO THE PATIENT OR USER WAS REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25112148 ↗
  • MalfunctionApr 17, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT UP. THE DEVICE WAS OUTSIDE OF CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24912252 ↗
  • MalfunctionMar 30, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT POWER ERRORS WERE OBSERVED IN THE DEVICE LOGS, WHICH CAN INDICATE AN ISSUE WITH BOOTING. NO HARM TO THE PATIENT OR USER WAS REPORTED. WE ARE CONSERVATIVELY REPORTING THIS EVENT AS THE INVESTIGATION IS ONGOING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24714251 ↗
  • MalfunctionMar 20, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT SYSTEM WAS UNABLE TO BOOT UP. THE DEVICE WAS OUTSIDE OF CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24647256 ↗
  • MalfunctionMar 12, 2026Unexpected/Unintended Radiation Output

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM'S WIRED FOOTSWITCH WAS STICKING, AND THE FLUOROSCOPY REMAINED ON FOR APPROXIMATELY 30 SECONDS, WHICH COULD ADVERSELY AFFECT THE PATIENT. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF THE EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24577296 ↗
  • MalfunctionMar 9, 2026Radiation Output Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM'S FOOTSWITCH BRACKET HAD LOOSED, WHICH CAN IMPACT IMAGING. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24542607 ↗
  • MalfunctionMar 4, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE ALLURA DEVICE WOULD NOT BOOT UP. IT DISPLAYED 0:00 AND WOULD NOT PROGRESS FURTHER THROUGH THE BOOT UP SEQUENCE. THE DEVICE WAS OUTSIDE OF CLINICAL USE AT THE TIME OF THE EVENT. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24508353 ↗
  • MalfunctionMar 3, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE ALLURA DEVICE WOULD NOT BOOT UP. IT DISPLAYED 0:00 AND WOULD NOT PROGRESS FURTHER THROUGH THE BOOT UP SEQUENCE. THE DEVICE WAS OUTSIDE OF CLINICAL USE AT THE TIME OF THE EVENT. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24499856 ↗
  • MalfunctionFeb 27, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT UP. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24469179 ↗
  • MalfunctionFeb 17, 2026Radiation Output Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM COULD NOT EMIT X-RAY. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF THE EVENT. NO HARM TO THE PATIENT OR USER WAS REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24362325 ↗
  • MalfunctionFeb 9, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE THE SYSTEM'S FLEXVISION COMPUTER WAS UNABLE TO BOOT, PREVENTING A FULL SYSTEM BOOT. THE DEVICE WAS NOT IN CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24293826 ↗
  • MalfunctionFeb 2, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE HOST COMPUTER'S MEMORY HAD PROBLEM, WHICH CAN IMPACT A FULL SYSTEM BOOT. THE DEVICE WAS OUTSIDE OF CLINICAL USE AT THE TIME OF THE EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24233424 ↗
  • MalfunctionJan 20, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT UP. THE DEVICE WAS OUTSIDE OF CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24116803 ↗
  • MalfunctionJan 12, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM'S GEOMETRY MODULE BUTTON WAS DAMAGED. UPON REBOOTING, THE SYSTEM WAS UNABLE TO BOOT, STALLING AT 00:00. THE DEVICE WAS IN CLINICAL USE AT THE TIME OF THE EVENT. NO HARM TO THE PATIENT OR USER WAS REPORTED. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24034908 ↗
  • MalfunctionJan 8, 2026Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM'S HOST COMPUTER WAS UNABLE TO BOOT PROPERLY. THE DEVICE WAS OUTSIDE OF CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24007406 ↗
  • MalfunctionDec 31, 2025Activation Failure; Activation Failure

    IT HAS BEEN REPORTED TO PHILIPS THAT THE SYSTEM DID NOT BOOT UP SUCCESSFULLY. THE DEVICE WAS OUTSIDE OF CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23933343 ↗
  • MalfunctionDec 24, 2025Activation Failure; Activation Failure

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM WAS UNABLE TO BOOT, STALLING AT 00:00. THE DEVICE WAS OUTSIDE OF CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23886188 ↗
  • MalfunctionDec 22, 2025No Display/Image; No Display/Image

    IT WAS REPORTED TO PHILIPS THAT THE SYSTEM LOST THE LIVE IMAGE. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23866628 ↗
  • MalfunctionDec 17, 2025Activation Failure; Activation Failure

    IT HAS BEEN REPORTED TO PHILIPS THAT THE SYSTEM DID NOT BOOT UP SUCCESSFULLY. THE DEVICE WAS OUTSIDE CLINICAL USE AT THE TIME OF REPORTED EVENT. NO HARM WAS REPORTED TO PHILIPS. PHILIPS HAS STARTED AN INVESTIGATION OF THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23834260 ↗

Other Interventional Fluoroscopic X-Ray System devices