MOSAIQ recallZ-1162-2025 · Elekta, Inc. · initiated Jan 15, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Elekta, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Image offsets calculated by oncology Information System software may be incorrect for Linear Accelerator (LINACs) not characterized at IEC61217 scale.

FDA root cause category: Software design

What the company is doing

On January 15, 2025 Elekta issued a recall notification to affected consignees via Email. Elekta asked consignees to take the following actions. 1. Customers included in the scope of the Field Safety Notice, should not calculate shifts from 2D verification images taken at a non-zero patient support angle (couch angle). 2. Please post this notice in a place accessible to all users, e.g. Instructions for Use, until this action is closed. 3. Advise the appropriate personnel, working with this product, on the content of this letter. 4. You are required to either acknowledge receipt of this notification via the Elekta Care" Community or complete this form and return it to Elekta immediately upon receipt, but no later than within 30 days.

Recall details

Recall number
Z-1162-2025
Status
Open, Classified
Initiated by the company
Jan 15, 2025
Posted by the FDA
Feb 12, 2025
Recalling firm
Elekta, Inc.
Quantity
64 units
Product code
IYE
FDA submissions
K223229
Distribution
US: PA, AL, FL, IA UT OUS: Austria Belgium Bolivia Bosnia and Herzegowina Brazil Canada France Germany Hong Kong Italy Japan Korea, Republic of Malta Netherlands Panama South Africa Spain Taiwan Thailand Turkey Venezuela

Product description

MOSAIQ Oncology Information System

Code information (models, lots, serial numbers)

Software versions /UDI: 3.1.2.0 (01)07340201500026 3.1.3.0 (01)07340201500026; 3.1.3.1 (01)07340201500026; 3.1.3.2 (01)07340201500026; 3.1.3.3 (01)07340201500026; 3.1.3.4 (01)07340201500026; 3.2.1.0 (01)07340201500071; 3.2.1.1 (01)07340201500071; 3.2.1.2 (01)07340201500071; 3.2.1.3 (01)07340201500071; 3.2.1.4 (01)07340201500071