Skip to content

MOSAIQ (UDI 07340201500026)Recalls & reported problems

Elekta Solutions AB · Class 2

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07340201500026), received between Nov 12, 2021 and Dec 13, 2024 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Computer Software Problem” (5 reports).

3 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 10, 2025Open, Classified

    Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice

    Z-1078-2026 · Root cause: Software change control · FDA record ↗

  2. Sep 17, 2025Open, Classified

    When appending a care plan that contains one or more wave medication orders, the occurrence and frequency of the appended orders may not match the intended schedule.

    Z-0503-2026 · Root cause: Software design · FDA record ↗

  3. Jan 15, 2025Open, Classified

    Image offsets calculated by oncology Information System software may be incorrect for Linear Accelerator (LINACs) not characterized at IEC61217 scale.

    Z-1162-2025 · Root cause: Software design · FDA record ↗

Most reported problems

Device problems

  • Computer Software Problem5
  • Application Network Problem1
  • Communication or Transmission Problem1

Patient problems

  • Insufficient Information5
  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 13, 2024Computer Software Problem; Computer Software Problem

    THE CUSTOMER REPORTED THE WRONG 2D MOSAIQ SHIFT FOR BEAMS AT COUCH ANGLES.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20924913 ↗
  • MalfunctionAug 28, 2024Computer Software Problem; Computer Software Problem

    THE CUSTOMER REPORTED THAT A HITACHI TREATMENT DOSE DID NOT RECORD IN MOSAIQ.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20084569 ↗
  • MalfunctionJun 26, 2024Computer Software Problem; Computer Software Problem

    THE CUSTOMER REPORTED THAT THE WRONG DOSE WAS DISPLAYED WHILE MAKING A PORTAL IMAGE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19613913 ↗
  • MalfunctionNov 12, 2021Computer Software Problem; Communication or Transmission Problem; Computer Software Problem; Communication or Transmission Problem

    THE CUSTOMER REPORTED THAT THE MACHINE WENT DOWN DURING THE TREATMENT OF A PATIENT. THE CUSTOMER TRIED TO CLOSE OUT THE PATIENT FROM THE MACHINE, BUT THEY WERE UNABLE TO, AND MANUALLY RECORDED THE DOSE TREATED. ONCE THE MACHINE WAS BACK UP AND RUNNING, THE PATIENT INFORMATION WAS STILL SHOWING, AND THE CUSTOMER HAD NOT ACTUALLY CLOSED OUT THE PATIENT AS THEY ORIGINALLY THOUGHT. WHEN THE CUSTOMER TREATED THE PATIENT, THE PARTIAL TREATMENT WAS RECORDED. THE CUSTOMER ATTEMPTED TO ZERO OUT THE MANUALLY RECORDED MU/DOSE WHICH CORRECTED THE DOSE RECORDING BUT DID NOT FIX THE MU LEFT TO BE TREATED. WHEN THE "TREAT" BUTTON IS CLICKED ON TO SEND THE FIELDS TO THE MACHINE, AN ERROR POPS-UP.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 12802299 ↗
  • MalfunctionNov 12, 2021Computer Software Problem; Application Network Problem; Computer Software Problem; Application Network Problem

    THE CUSTOMER REPORTED THAT THE RATM INTERFACE INITIALISATION FAILED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 12802178 ↗

Other Accelerator, Linear, Medical devices

All Accelerator, Linear, Medical devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.