Microstream recallZ-1376-2025 · Oridion Medical 1987 Ltd. · initiated Feb 17, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Oridion Medical 1987 Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube for procedures (e.g. suctioning), may cause treatment delay and/or unintended extubation, which could lead to respiratory failure, hypoxia, aspiration, and/or low oxygen saturation.

FDA root cause category: Labeling design

What the company is doing

On 2/17/2025, correction notices were mailed to Risk Manager, Director of Respiratory Care, Director of Anesthesiology who were informed of the following: In September 2025, the firm will issue an update to the Instructions for Use (IFUs) for the The following IFU's: 01158, PT00156355, PT00156254, PT00156250 stating: Adverse events associated with attaching and detaching the airway adapter from the breathing circuit are listed in descending order of severity: unintended extubation, respiratory failure, hypoxia, low oxygen saturation, aspiration/inhalation and delay to treatment. Any serious incident related to device use that may occur should be reported immediately to the manufacturer, the local competent authority, and any other regulators as required. Pass on and post this notice for all those who need to be aware within your organization and to any organization where the affected product has been transferred or distributed. Complete and return the Customer Confirmation Form to [email protected] Adverse reactions or quality problems experienced with this product should be reported to the firm's Quality Assurance at 800-255-6774 option 1, then option 1. If you have questions contact the firm's Customer Service at 800-962-9888, Option 2. On 5/13/2025, additional correction notices identifying additional affected devices were mailed to additional customers.

Recall details

Recall number
Z-1376-2025
Status
Open, Classified
Initiated by the company
Feb 17, 2025
Posted by the FDA
Mar 17, 2025
Recalling firm
Oridion Medical 1987 Ltd.
Quantity
318292
Product code
CCK
FDA submissions
K181467
Distribution
Worldwide - US Nationwide distribution including in the states of TX, OH, FL, CO, IL, CA, MD, PA, VA, WI, IN, MI, NY, GA, OR, SC, NC, PR, AZ, KY, AR, TN, MO, DE, MA, MS, UT, DC, NE, IA, WV, NJ, KS, LA, WA, HI, MN, AL, ME, RI, OK, MT, ID, CT, NH, AK, NV, SD, WY, ND, NM and the countries of Algeria, Andorra, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, El Salvador, Estonia, Finland, France, French Guiana, French Polynesia, Germany, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Morocco, Namibia, Netherlands, New Caledonia, New Zealand, Northern Ireland, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Virgin Islands, U.S.

Product description

Microstream Instructions for Use and Part Number used with - Product Description, REF: FilterLine H Set & VitaLine H Set Infant/Neonatal, 006147 - MICROSTREAM FilterLine H Set Infant/Neonatal CO2 Sampling Line and Airway Adapter for Humid Environments, 006324; MICROSTREAM FilterLine H Set Long Infant/Neonatal CO2 Sampling Line and Airway Adapter for Humid Environments, 007738; MICROSTREAM VitaLine H Set Infant/Neonatal CO2 Sampling Line and Airway Adapter for Humid Environments, 010807.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Part Number and Revision - REF/UDI-DI(Lot): 006147 Rev N - 006324/20884521551692(C210417257, C210957017, C211102266, C211206691, C220108497, D191171644, D191171646, D191176150, D191278690, D200316307, D200538149, D200639546, D200742524, D200745256, D200950569, D200951852, D201055941, D201261296, D210101027, D210103472, D210310524, D210413725, D210726722, D220211210), 10884521551695(D191171644, D191171646, D191176150, D191278690, D200316307, D200745256, D201261296, D210103472); 007738/20884521216010(D191069375, D191171650, D191176156, D200314292, D200431907, D200537959, D200742194, D200745606, D200950373, D201260700, D210308577, D210728936, D211047680, D211155459); 010807/20884521551920(D200101284, D200426597, D200742219, D200745037, D201260710, D210310546, D210837166), 10884521551923(D200101284, D200745037).

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

15 devices cleared under K181467 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 96309).