Microstream Instructions for Use and Part Number used with… recallZ-1381-2025 · Oridion Medical 1987 Ltd. · initiated Feb 17, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Oridion Medical 1987 Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube for procedures (e.g. suctioning), may cause treatment delay and/or unintended extubation, which could lead to respiratory failure, hypoxia, aspiration, and/or low oxygen saturation.
FDA root cause category: Labeling design
What the company is doing
On 2/17/2025, correction notices were mailed to Risk Manager, Director of Respiratory Care, Director of Anesthesiology who were informed of the following: In September 2025, the firm will issue an update to the Instructions for Use (IFUs) for the The following IFU's: 01158, PT00156355, PT00156254, PT00156250 stating: Adverse events associated with attaching and detaching the airway adapter from the breathing circuit are listed in descending order of severity: unintended extubation, respiratory failure, hypoxia, low oxygen saturation, aspiration/inhalation and delay to treatment. Any serious incident related to device use that may occur should be reported immediately to the manufacturer, the local competent authority, and any other regulators as required. Pass on and post this notice for all those who need to be aware within your organization and to any organization where the affected product has been transferred or distributed. Complete and return the Customer Confirmation Form to [email protected] Adverse reactions or quality problems experienced with this product should be reported to the firm's Quality Assurance at 800-255-6774 option 1, then option 1. If you have questions contact the firm's Customer Service at 800-962-9888, Option 2. On 5/13/2025, additional correction notices identifying additional affected devices were mailed to additional customers.
Recall details
- Recall number
- Z-1381-2025
- Status
- Open, Classified
- Initiated by the company
- Feb 17, 2025
- Posted by the FDA
- Mar 17, 2025
- Recalling firm
- Oridion Medical 1987 Ltd.
- Quantity
- 3800
- Product code
- CCK
- FDA submissions
- K181467
- Distribution
- Worldwide - US Nationwide distribution including in the states of TX, OH, FL, CO, IL, CA, MD, PA, VA, WI, IN, MI, NY, GA, OR, SC, NC, PR, AZ, KY, AR, TN, MO, DE, MA, MS, UT, DC, NE, IA, WV, NJ, KS, LA, WA, HI, MN, AL, ME, RI, OK, MT, ID, CT, NH, AK, NV, SD, WY, ND, NM and the countries of Algeria, Andorra, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, El Salvador, Estonia, Finland, France, French Guiana, French Polynesia, Germany, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Morocco, Namibia, Netherlands, New Caledonia, New Zealand, Northern Ireland, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Virgin Islands, U.S.
Product description
Microstream Instructions for Use and Part Number used with - Product Description, REF: Microstream Adv. Adult-Pediatric Intubated, PT00134333 - Microstream Advance Adult-Pediatric Intubated CO2 Filter Line 2m/6.5ft Short-term use: Procedural/Emergency, QMVAI; Microstream Advance Adult-Pediatric Intubated CO2 Filter Line 2m/6.5ft Extended Duration, QMVAIH
Code information (models, lots, serial numbers)
Part Number and Revision - REF/UDI-DI(Lot): PT00134333 Rev A - QMVAI/10884521762329(C201200644); QMVAIH/10884521762336(C210314314)
Other devices under the same FDA submission
15 devices cleared under K181467 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Microstream Advance MVIIHOridion Medical 1987 Ltd. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall27 reports
- Microstream Advance MVAIHOridion Medical 1987 Ltd. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall4 reports
- Microstream Advance MVIIHHOridion Medical 1987 Ltd. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall2 reports
- Microstream Advance MVAILOridion Medical 1987 Ltd. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall1 report
- Microstream Advance MVAIHHOridion Medical 1987 Ltd. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall1 report
- Microstream Advance MVAIOridion Medical 1987 Ltd. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall0 reports
- Microstream Advance MVAI100UOridion Medical 1987 Ltd. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall0 reports
- Microstream Advance MVAIH100UOridion Medical 1987 Ltd. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall0 reports
- Microstream Advance MVAIHLOridion Medical 1987 Ltd. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall0 reports
- Microstream Advance MVIIHLOridion Medical 1987 Ltd. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall0 reports
- Microstream Luer MLAIOridion Medical 1987 Ltd. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall0 reports
- Microstream Advance ZMVAIHCovidien LP · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 96309).
- Z-1376-2025 · Open, ClassifiedMicrostream Instructions for Use and Part Number used with - Product Description, REF: FilterLine H Set & VitaLine H Set Infant/Neonatal, 006147 - MICROSTREAM FilterLine H Set I…
- Z-1377-2025 · Open, ClassifiedMicrostream Instructions for Use and Part Number used with - Product Description, REF: FilterLine Sets & VitaLine Sets Adult/Pediatric, MQ04616 - MICROSTREAM FilterLine H Set L…
- Z-1378-2025 · Open, ClassifiedMicrostream Instructions for Use and Part Number used with - Product Description, REF: Filter Line Sets Adult/Pediatric & Infant/Neonatal, 10129497 - CapnoLine Neonatal-Infant …
- Z-1379-2025 · Open, ClassifiedMicrostream Instructions for Use and Part Number used with - Product Description, REF: Airway Adapter and CO2 Sampling Line, 01158 - Microstream Luer Adult-Pediatric Airway Ada…
- Z-1380-2025 · Open, ClassifiedMicrostream Instructions for Use and Part Number used with - Product Description, REF: OmniVentLine Set, 012808 - OmniVentLine Set Adult-Pediatric CO2 Sampling Line and Airway …
- Z-1382-2025 · Open, ClassifiedMicrostream Instructions for Use and Part Number used with - Product Description, REF: Microstream Adv. Infant Intubated, PT00134354 - Microstream Advance Neonatal-Infant Intub…
- Z-1383-2025 · Open, ClassifiedMicrostream Instructions for Use and Part Number used with - Product Description, REF: Microstream Luer Intubated Adult Pediatric, PT00156355 - Microstream Luer Adult-Pediatric…
- Z-1384-2025 · Open, ClassifiedMicrostream Instructions for Use and Part Number used with - Product Description, REF: Microstream Advance Intubated Adult-Pediatric, PT00156254 - Microstream Advance Adult-Ped…
- Z-1385-2025 · Open, ClassifiedMicrostream Instructions for Use and Part Number used with - Product Description, REF: Microstream Advance Intubated Infant, PT00156250 - Microstream Advance Neonatal-Infant In…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page