Proxima Drape labeled as MINOR PACK and LAPAROSCOPY.… recallZ-1366-2025 · MEDLINE INDUSTRIES, LP - Northfield · initiated Feb 13, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential breach in pouch packaging which could lead to loss of sterility.
FDA root cause category: Packaging process control
What the company is doing
Medline distributed an IMMEDIATE ACTION REQUIRED MEDICAL DEVICE RECALL notification, dated 2/13/25, to consignees who purchased individually packaged Proxima sterile surgical drapes and kits that contained Proxima sterile surgical drapes. Consignees that purchased kits containing recalled product are to immediately check stock for affected kits and place them in quarantine. Consignees are to complete the provided response form to receive over-labels to be applied to quarantined kits with additional instructions to remove the affected components prior to use of the kit. Consignees that purchased individual units of Proxima sterile surgical drapes are to immediately check stock for affected devices and immediately destroy identified product. Consignees are to complete the provided response form to obtain credit for the destroyed product. If a consignee is a distributor or have further distributed affected devices, they are instructed to forward the provided recall notification. Questions about this recall can be directed to Medline via phone at 866-359-1704 or by email to [email protected].
Recall details
- Recall number
- Z-1366-2025
- Status
- Open, Classified
- Initiated by the company
- Feb 13, 2025
- Posted by the FDA
- Mar 14, 2025
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Quantity
- 506 units
- Product code
- FDE
- Distribution
- US Nationwide distribution.
Product description
Proxima Drape labeled as MINOR PACK and LAPAROSCOPY. Sterile surgical drape.
Code information (models, lots, serial numbers)
Item No. DYNJ34603F & DYNJ903916I; UDI (case): 40198459143596 & 40195327580118; UDI (each): 10198459143595 & 10195327580117; Lot No. (Shelf Life) 25AMD464(11/30/2029), 24JMC011(04/30/2029), 25ABA362(09/30/2025), 24KBT718(02/28/2026), 24JBE997(09/30/2025), 24IBM203(02/28/2026).
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJ34603FMedline Industries, Inc. · laparoscopy kit3 recalls0 reports
- Medline Industries, Inc. DYNJ903916IMedline Industries, Inc. · laparoscopy kit2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 96401).
- Z-1363-2025 · Open, ClassifiedProxima Drape labeled as DRAPE C-SECTION W/ WINDOW; SHEET, DRAPE, SPLIT, ST,10/CS; DRAPE,LAPROSCOPY,CHOLE,W/TROUGH,ST,8/CS; DRAPE,LAPROSCOPY,CHOLE,W/TROUGH,ST,8/CS; DRAPE,ABDOMINAL…
- Z-1364-2025 · Open, ClassifiedProxima Drape labeled as PACK,BASIC,SIRUS,V,5/CS; PACK,BASIC IV,SIRUS,5/CS; PACK,SET UP,NO DRAPE,12/CS; PACK,SET UP,NO DRAPE,12/CS; PACK,SET UP,NO DRAPE,12/CS; PACK,SET UP,NO DRAPE…
- Z-1365-2025 · Open, ClassifiedProxima Drape labeled as ENT-PLASTICS PACK. Sterile surgical drape.
- Z-1367-2025 · Open, ClassifiedProxima Drape labeled as FACE LIFT. Sterile surgical drape.
- Z-1368-2025 · Open, ClassifiedProxima Drape labeled as HEAD AND NECK OSIS, WILLIS SHOULDER PACK, EXTREMITY PACK-LF, TOTAL HIP PACK-LF, SHOULDER W/POUCH-LF, VEIN PACK. Sterile surgical drape.
- Z-1369-2025 · Open, ClassifiedProxima Drape labeled as UNIVERSAL DAVINCI ROBOTIC P. Sterile surgical drape.
- Z-1370-2025 · Open, ClassifiedProxima Drape labeled as ORTHO TRAUMA CDS, TOTAL KNEE AUX PACK, TOTAL HIP PACK, TOTAL HIP, TOTAL KNEE, SPINE. Sterile surgical drape.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page