Skip to content

Proxima Drape labeled as UNIVERSAL DAVINCI ROBOTIC P.… recallZ-1369-2025 · MEDLINE INDUSTRIES, LP - Northfield · initiated Feb 13, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential breach in pouch packaging which could lead to loss of sterility.

FDA root cause category: Packaging process control

What the company is doing

Medline distributed an IMMEDIATE ACTION REQUIRED MEDICAL DEVICE RECALL notification, dated 2/13/25, to consignees who purchased individually packaged Proxima sterile surgical drapes and kits that contained Proxima sterile surgical drapes. Consignees that purchased kits containing recalled product are to immediately check stock for affected kits and place them in quarantine. Consignees are to complete the provided response form to receive over-labels to be applied to quarantined kits with additional instructions to remove the affected components prior to use of the kit. Consignees that purchased individual units of Proxima sterile surgical drapes are to immediately check stock for affected devices and immediately destroy identified product. Consignees are to complete the provided response form to obtain credit for the destroyed product. If a consignee is a distributor or have further distributed affected devices, they are instructed to forward the provided recall notification. Questions about this recall can be directed to Medline via phone at 866-359-1704 or by email to [email protected].

Recall details

Recall number
Z-1369-2025
Status
Open, Classified
Initiated by the company
Feb 13, 2025
Posted by the FDA
Mar 14, 2025
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Quantity
220 units
Product code
OHD
Distribution
US Nationwide distribution.

Product description

Proxima Drape labeled as UNIVERSAL DAVINCI ROBOTIC P. Sterile surgical drape.

Code information (models, lots, serial numbers)

Item No. DYNJ53968P; UDI (case): 40195327532759; UDI (each): 10195327532758; Lot No. (Shelf Life) 24KMC372 (3/31/2026), 24IMJ082 (4/30/2026), 24HMK377 (4/30/2026). .

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 96401).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page